Zeiss LSM800-02 Operator's manual

1
Zeiss LSM800-02 + Airyscan confocal microscope 20/11/2019
System description:
I. Upright confocal microscope ZEN blue 2.6 HotFix-8
II. 2 GaAsP +1 AiryScan detector (GaAsP)
III. Motorized stage for multiple position imaging
IV. Heating chamber
V. NO CO2 mixer
VI. Multi-slide holder, dish holder, single slide holder
Objectives:
-Position1: Objective Plan-Apochromat 10x / NA 0.45 Air (WD=2.1mm)
-Position2: Objective Plan-Apochromat 20x / NA 1.0 W DIC (WD=1.8mm)
-Position3: Objective Plan-Apochromat 20x / NA 0.8 Air
-Position4: Objective Plan-Apochromat 40x/1.4 Oil DIC (WD=0.13 mm)
-Objective Plan-Apochromat 63x / NA 1.4 Oil (not mounted)
-Objective Plan-Apochromat 40x / NA 1.2 Water (not mounted)
Laser lines: 405, 488, 561, 640 nm
Detectors:
Two very sensitive Gallium arsenide phosphide photomultiplier tube (GaAsP PMTs) detectors
with free choice of the spectral range. (Uses a tunable short pass and band pass filter in front
of the PMTs)
Be very careful with using it to prevent instrument damage! If oversaturated, detectors will
switch off automatically.
1 GaASP PMT for the Airyscan detector that can function as a normal detector in confocal
mode (detection range adjustable by a tunable short pass and a tunable long pass filter, no
bandpass filters)
An ESID detector for transmitted light.
Simultaneous detection of 3 dyes possible.
VII. Starting up:
1. Switch on the multiple socket (1) “Hanging on the right side of the rack”. Provides power
to the computer and microscope controllers

2
2. Switch on SYSTEM (2).
3. Switch on COMPONENTS (3).
4. Make sure that lasers are on. Laser key should be in position “I”. (4)
5. Switch on the computer.
6. Turn on Cool LED (by pressing Big blue on/off button), if you need it for visual
examination and don’t forget to close the shutter when you don’t need it or at the end
of your session.
7. Login: LSMuser (password: Useme!11)
8. Open ZEN Blue Software and press “ZEN System” -> Restart the ZEN if it is open!
(preventing bugs and accidental objective crash)S
9. “Image Processing” is for offline analysis of your images, please do not use it here
VIII. Visual examination of the sample
1. Put mechanical switch on the right hand side of the microscope to the ”Eye” position.
2. Press “Locate” in the main microscope menu.
3. Since this is an upright system, changing objectives can only be manually done by
rotating the objective revolver. Rotation should work very easily when objectives are in
the “up” position.
4. For imaging or visual inspection, the objectives must be lowered entirely with the black
wheel in front of the microscope stand.
5. Please make sure that the chosen objective corresponds to the one shown in the
software. If this is not the case, please restart the system. (Otherwise, your pinhole
and pixel sizes, as well as the sectioning thickness are all incorrect.)
6. If you wish to view your sample in Brightfield mode, press the “TL” button
7. For visual examination of fluorescence, press the buttons DAPI, GFP or RFP
8. Before switching to acquisition mode, close transmitted and reflected light shutters by
pressing ‘off’ and turn off the Cool LED by pressing ON/Off button
IX. Acquiring confocal images

3
•Select the ‘Acquisition’ tab in the main menu of Zen blue
•Go to “Smart Setup” choose your dyes (set the LUT) and configuration, you can choose
between simultaneous (fastest), Best compromise line scanning mode (Smart) and sequential
(best signal) scan, press ”Apply”. (Avoid Line mode as it blocks the changing of filters and
sliders, therefore any change will affect all the track!)
•Make sure that Red and FarRed dyes are not set to the 1st detector (due to the Variable
dichroic arrangement)
•If you are using the detector for transmitted light (ESID): activate (ESID) represented in the
light path scheme. Adjust intensity in the ESID Gain, then adjust digital offset and digital gain
of the ESID detector in order to “stretch” the histogram and use the whole dynamic range of
the detector.
•
Make sure Köhler illumination is set properly if you intend to use the ESID detector
(for bright field images)!
•Press the “AF” (Autofocus) button to find your focus, alternatively go to “Live” acquisition
modus and find it manually

4
•Set pinhole to Airy Unit (AU) 1 in the channel dialog to achieve confocality. Note, that activating
the ‘1AU’ button by clicking on it once (gets light blue background color) fixes the pinhole to
1AU. In this case, the match pinhole function in the Z-stack setup will not change pinhole sizes.
•In the ‘Channels’ dialog you can adjust the pinhole, gain and laser power manually
•Press the “Auto exposure” button and the software will adjust the gain of the PMTs but not the
laser power, according to the brightness of your dyes.
•What Auto exposure does, is to increase the PMT voltages until about 1% of all pixels are
overexposed. If the resulting gain is below 400 V, decrease the laser power. If the gain is above
900 V, increase the laser power.
•To zoom in your image, go to ‘Acquisition mode’ and adjust the slider accordingly. With “Crop”,
you can select the desired area of your image, then press “Live” or “Snap” to apply. “Reset all”
will remove the zoom again. Depending on the scan speed, the lowest zoom is 0.5 and not 1.0.
•Before taking an image, press ‘optimal’in the ‘acquisition mode’tab to adjust optimal number
of pixels and the pixel size according to Nyquist sampling criteria.
Note: More pixel than suggested by the software as ‘optimal’ rarely make sense. On the other
hand, feel free to decrease the pixel number to increase acquisition speed at the cost of
resolution.
•To capture single images, click on ‘Snap’ (Snapped images are overwritten if you press Live
before saving them)
1. Setting up Z-stack:
Enable Z-Stack and the Z-Stack dialog will appear
Press “Live” and set your first and last slice
The software will suggest optimal slice number and interval which you can change manually
Activate the ‘Optimize sectioning and step’ checkbox to see slice overlapping. Here you can
again set an optimal interval to ensure ~50% overlap between sections.
If you wish to use the match pinhole function, make sure that the ‘1AU’ button in the
‘Channels’ menu is deactivated on all tracks (no blue background of button). This function will
match pinhole sizes to the optimal pinhole size of the lowest wavelength of your imaging setup
and ensure that all sections have the same thickness.
Press “Start experiment”
2. Setting up time series:
Enable “time series”
Time series dialog will open
Set cycle number and interval time
Press “start experiment”
3. Tile Scan (in this mode you have: Tiles, Multipositions):
Enable “tile scan”

5
Tile scan dialog will open
Set desired number of horizontal and vertical tiles
You can choose different regions to scan. Choose regions and press button “+” to add it.
Press “start experiment”
Advanced Set up allows you to do preview scan of region of interest.
You can also add multipositions in this tab:
Important: to make sure that the software images the positions in the order you added them,
make sure that you DISABLE the feature “tile regions/positions” in the options menu of the
Tiles bar (see picture on the next page)
4. Data Saving:
Save your data to D: data folder. Note that this computer is not backed up. Please move your
data as soon as to the file server, all the user data will be deleted automatically after one
month.
Disable this option to keep
the order of positions during
imaging!

6
X. Switching off:
1. DO NOT save any “Experiment” or overwrite the default “Experiment” (marked with a home
logo) as this setting will be used in the next startup of the system and might cause potential
damage to GaAsP PMTs.
2. Close the ZEN Software
3. Shut down the computer.
4. Turn off the cool LED
5. Switch off COMPONENTS (3).
6. Switch off SYSTEM (2).
7. Switch off multiple sockets (1).

7
Laser Safety Instructions
During operation of class 3B and class 4 lasers red warning lights have to be switched on manually
Red warning light at the door of the room, containing laser-based equipment, prohibits the
entrance
Optical path of the laser beam at all setups has to stay intact and should never be disassembled
by a user. Users are never permitted to disconnect optical connections (pipes, fibers etc), remove
protective coveringsor disassemble any parts of the setups, especially those parts that are labeled
with laser-warning signs.
User has to make sure, that objectives mounts are blocked by objectives or light- blocking plugs,
before switching the system on or starting the work
Any cleaning activities (objectives, stage cleanings) as well as changing of objectives or filters have
to be performed only after blocking of the laser light is ensured. This can be ensured by closing
the scanhead shutter or switching off the laser.
Laser class-specific warnings at each setup have to be observed and considered
Eye contact with direct beam of Class 3B laser, or eye contact with mirror reflection from class 3B
laser, should be avoided at all times
Eye or skin contact with direct or diffuse light of Class 4 laser, should be avoided at all times
Laser safety goggles are situated at all workspaces and should be used in any situation where
potential contact of eyes with Laser light of the classes 3B or 4 is possible, according to the
previous two points
Laser safety goggles have to be worn at all times of operation of Laser Class 4
Laser safety goggles have to be worn at all times of operation laser Class 3 at the optogenetic
setups
Laser safety goggles are assigned to each setup and matched to the corresponding laser
wavelengths. Matching laser safety goggles should be used at all times, and should not be carried
over between the setups
Only one person is allowed to be in the corresponding compartment during Laser Class 4
operation and optogenetic setup operation
Users are not allowed to wear any reflective objects (rings, watches etc) during laser operation
Using of the equipment is only allowed after the introduction from a laser safety officer of IST
Austria. Introduction has to be done individually for each setup.
Changing experimental conditions, that involves changes in the laser application, have to be
reported to the laser safety officer prior to the start of experiment
Users have to understand that any violations against the instructed rules and also withholding information
leading to safety hazards will ultimately result in denial of admission to all laser equipped instruments at
the IST Austria.
Table of contents
Other Zeiss Microscope manuals

Zeiss
Zeiss LSM 710 User manual

Zeiss
Zeiss Standard Pol User manual

Zeiss
Zeiss ?IGMA VP-FE-SEM User manual

Zeiss
Zeiss SV8 User manual

Zeiss
Zeiss Axiocam MR Installation and operating manual

Zeiss
Zeiss LSM 700 User manual

Zeiss
Zeiss LSM 710 User manual

Zeiss
Zeiss HBO 100 User manual

Zeiss
Zeiss LSM 700 User manual

Zeiss
Zeiss Axio Examiner User manual

Zeiss
Zeiss OPMI PROergo S7 User manual

Zeiss
Zeiss LSM 880 User manual

Zeiss
Zeiss OPMI PRO magis User manual

Zeiss
Zeiss Axiolab A User manual

Zeiss
Zeiss SV8 User manual

Zeiss
Zeiss SL 800 Operator's manual

Zeiss
Zeiss LSM 880 User manual

Zeiss
Zeiss Stemi 305 cam User manual

Zeiss
Zeiss VISU 200 User manual

Zeiss
Zeiss Stemi 508 User manual