Nutradip GrowBoss User manual

OWNERS MANUALOWNERS MANUAL

Contents
Meet the GrowBoss ......................................................... 3
Before you Begin ............................................................. 4
Unpacking your new Meter ........................................... 4
First Time Use .................................................................... 5
Attaching the pH Probe ................................................. 5
Attaching the Nutrient/Temperature probe ............... 5
Attaching the Power adaptor ...................................... 5
Preparing the pH Probe ................................................. 5
Calibration ....................................................................... 6
pH Probe Calibration ..................................................... 6
What you will need ......................................................... 6
What to do ....................................................................... 6
Nutrient Probe Calibration ............................................. 7
What to do ....................................................................... 7
Care & Cleaning .............................................................. 8
Caring for pH Probes ....................................................... 8
Cleaning pH Probes ........................................................ 8
Every Cleaning ................................................................ 8
Every Third Cleaning .............................................................. 8
Storing the pH Probe........................................................ 9
If the probe has been allowed to dry out .................... 9
Caring for Calibration Solutions ..................................... 9
Storage ............................................................................. 9
Replacement .................................................................. 9
Specifications .................................................................. 10
Abilities ............................................................................. 10
Operational Environment ............................................... 10
Troubleshooting Guide ................................................... 11
Problem - No Display ...................................................... 11
Problem pH Probe Inaccurate ...................................... 11
Problem Temperature Inaccurate ................................ 11
Problem Nutrient reading is inaccurate ....................... 11
Problem Won't Hold Calibration ................................... 12
Warranty Coverage........................................................ 13
Nutradip 12 Month Warranty Coverage ..................... 13
Nutradip Accidental Damage Coverage .................. 13
What is not Covered ................................................... 14
Making a Warranty Claim.............................................. 15

3
Meet the GrowBoss
Push Button
Nutrient Scale
PPM | 442 | EC
Temperature
F|C
Min | Max
Total Recall
Min | Max
Indicators
On | Off
Power up
External Connections
Nutrient Scale
pH BNC | USB Nutrient | Power Supply
Display Panel
Plant Friendly
High vis LED
pH 7
Calibration
trim pot
pH 4 | 10
Min | Max
Nutrient
Calibration trim pot
Nutrient
Scale indicator
.5 | .67 | EC
On | Off
Power up

Before You Begin
Your Nutradip GrowBoss is engineered to provide years of
dependable service, however, as with any monitoring
equipment, it does require periodic calibration and
maintenance.
In order to optimize meter longevity and performance, we
strongly recommend that you thoroughly read this manual prior
to using your new meter for the first time. Store this manual in
a safe place. Should you require a new copy of this manual,
please visit www.futureharvest.com
Unpacking Your New Meter
Carefully unpack your meter and additional contents.
Your carton should contain the following pieces:
1. Nutradip Growboss
2. PPM/Temperature probe
3. pH probe with storage bottle cap
4. Power adaptor
5. 125mL pH 4 calibration solution
6. 125mL pH 7calibration solution
7. 125mL 1000 PPM calibration solution
8. Wall mounting bracket
9. Mounting bracket screws (2)
10. Calibration screw driver
Contact your dealer immediately if there is damage to any of the
above contents, or if a component is missing.
4
Before You Begin
Your Nutradip GrowBoss is engineered to provide years of
dependable service, however, as with any monitoring
equipment, it does require periodic calibration and
maintenance.
In order to optimize meter longevity and performance, we
strongly recommend that you thoroughly read this manual prior
to using your new meter for the first time. Store this manual in
a safe place. Should you require a new copy of this manual,
please visit www.futureharvest.com
Unpacking Your New Meter
Carefully unpack your meter and additional contents.
Your carton should contain the following pieces:
1. Nutradip Growboss
2. PPM/Temperature probe
3. pH probe with storage bottle cap
4. Power adaptor
5. 125mL pH 4 calibration solution
6. 125mL pH 7calibration solution
7. 125mL 1000 PPM calibration solution
8. Wall mounting bracket
9. Mounting bracket screws (2)
10. Calibration screw driver
Contact your dealer immediately if there is damage to any of the
above contents, or if a component is missing.
4

First Time Use
Working from left to right, attach the probes and power adaptor
to the Nutradip GrowBoss.
Attaching the pH Probe
The pH probe uses a twist-lock connector. Ensure the pins on
GrowBoss are aligned with the slot on the probe, and push firmly.
Twist the locking nut clockwise to tighten.
Attaching the Nutrient/Temperature Probe
The nutrient and temperature probe attaches in only
one direction. Ensure that the side of the plug that reads
“1394” faces up, and push firmly into place.
Attaching the Power Adaptor
Push the power adaptor barrel plug firmly into the GrowBoss
Preparing the pH Probe
1. Remove the protective liquid filled cap from the pH probe. Do
not touch the pH probe bulb. Put cap in a safe place. The pH
buffer solution is required for storage of the pH probe.
2. Gently shake the probe downward to remove any air bubbles
which may be trapped within the glass bulb.
Your Nutradip Growboss has been personally calibrated by a
trained service technician before leaving our assembly facility in
Kelowna British Columbia, Canada and arrives to you in ready-to-
use condition. However, we also recommend that calibration be
performed prior to initial use in the environment that the device will
be used most often to ensure the greatest possible degree of
accuracy from your new meter.
5
5

Calibration
Calibration should be done on your Nutradip GrowBoss at
least once a month; whenever a probe is cleaned; when a
new cycle is about to begin, or when the GrowBoss has been
taken out of storage.
Over time, all probes naturally degrade due to environmental
stresses, and calibration ensures that the natural changes in
the integrity of the probe are compensated for over the life
of the probes. To ensure the longest possible life for probes,
please refer to the CARE & CLEANING section of this manual.
pH Probe Calibration
For calibrating pH, we recommend using a two point calibration
between 4.0 and 7.0 to ensure that the meter is accurate across
the scale, and calibrated within the most desirable pH target
areas.
What you will need:
for 10 seconds. Let stand for 60 seconds.
12. Repeat steps 6 to 11 as needed. This may take several
N utricycles,ent Pr obeespeciallyCali brwithati onolder or
worn probes.
What you will need:
•ph 4 Buffer Solution (Future Harvest Part Number 1100002)
•ph 7 Buffer Solution (Future Harvest Part Number 1100001)
•Calibration screwdriver
•Three small, clean plastic cups
•30 minutes
•Clear tap water
What to do:
1. Place all supplies on your calibration table,and let sit. All
supplies MUST be the same temperature before starting to
achieve an accurate calibration. It is recommended to
float the calibration solution bottle in your reservoir for five to
ten minutes to reach the reservoir temperature. This will
allow for a more accurate calibration.
2. Place the GrowBoss on a level surface and connect all
probes and power supply. Power up the GrowBoss.
3. Pour pH7.0 solution into one of the small cups (enough to
cover pH bulb)
4. Pour pH 4.0 solution into one of the small cups (enough to
cover pH bulb)
5. Pour clear tap water in the third cup for rinsing the pH probe.
6. Place the probe in the pH 7.0 solution, and agitate for 10
seconds. Let stand for 60 seconds.
7. Rotate the 7.0 calibration screw counter clockwise very
slowly and gently until the display increases to 7.0.
8. Place the probe in the clear tap water cup, and agitate for
10 seconds. Let stand for 60 seconds.
6
Calibration
Calibration should be done on your Nutradip GrowBoss at
least once a month; whenever a probe is cleaned; when a
new cycle is about to begin, or when the GrowBoss has been
taken out of storage.
Over time, all probes naturally degrade due to environmental
stresses, and calibration ensures that the natural changes in
the integrity of the probe are compensated for over the life
of the probes. To ensure the longest possible life for probes,
please refer to the CARE & CLEANING section of this manual.
pH Probe Calibration
For calibrating pH, we recommend using a two point calibration
between 4.0 and 7.0 to ensure that the meter is accurate across
the scale, and calibrated within the most desirable pH target
areas.
What you will need:
for 10 seconds. Let stand for 60 seconds.
12. Repeat steps 6 to 11 as needed. This may take several
N utricycles,ent Pr obeespeciallyCali brwithati onolder or
worn probes.
What you will need:
•ph 4 Buffer Solution (Future Harvest Part Number 1100002)
•ph 7 Buffer Solution (Future Harvest Part Number 1100001)
•Calibration screwdriver
•Three small, clean plastic cups
•30 minutes
•Clear tap water
What to do:
1. Place all supplies on your calibration table,and let sit. All
supplies MUST be the same temperature before starting to
achieve an accurate calibration. It is recommended to
float the calibration solution bottle in your reservoir for five to
ten minutes to reach the reservoir temperature. This will
allow for a more accurate calibration.
2. Place the GrowBoss on a level surface and connect all
probes and power supply. Power up the GrowBoss.
3. Pour pH7.0 solution into one of the small cups (enough to
cover pH bulb)
4. Pour pH 4.0 solution into one of the small cups (enough to
cover pH bulb)
5. Pour clear tap water in the third cup for rinsing the pH probe.
6. Place the probe in the pH 7.0 solution, and agitate for 10
seconds. Let stand for 60 seconds.
7. Rotate the 7.0 calibration screw counter clockwise very
slowly and gently until the display increases to 7.0.
8. Place the probe in the clear tap water cup, and agitate for
10 seconds. Let stand for 60 seconds.
6

9. Place the probe in the pH 4.0 solution, and
agitate for 10 seconds. Let stand for 60 seconds.
10. Rotate the 4.0 calibration screw counter clockwise
very slowly and gently until the display increases to
4.0.
11. Place the probe in the clear tap water cup, and
agitate for 10 seconds. Let stand for 60 seconds.
12. Repeat steps 6 to 11 as needed. This may take
several cycles, especially with worn or older
probes.
Nutrient Probe Calibration
What you will need:
•1000ppm Calibration
Solution
•Calibration Screw Driver
•Two small, clean plastic cups
•30 minutes
•Clear Tap Water
What to do:
1. Place all supplies on your calibration table, and let sit. All
supplies MUST be the same temperature before starting to
achieve an accurate calibration. It is recommended to float
the calibration solution bottle in your reservoir for five to ten
minutes to reach the reservoir temperature. This will allow for
a more accurate calibration.
2. Place the GrowBoss on a level surface and connect all
probes and power supply. Power on the GrowBoss.
3. Pour some calibration solution into the cup. Use enough
solution to completely cover the probe pins.
4. Pour clear tap water into the second cup for rinsing the ppm
probe.
5. Place the probe in the EC/PPM solution and agitate for 10
seconds. Let stand for two minutes.
6. Set the scale factor on the meter to .5 via the push button
selector. This is the NaCl scale. Rotate the PPM calibration
screw counter clockwise very slowly until the display reads
1000 ppm.
7. Due to the sensitivity of the ppm probe, this process can take
10 + minutes during initial setup, where fine tuning of the
calibration trimpot may be necessary. Turn the screw
counterclockwise to increase the value.
8. If you wish to calibrate to a different scale then choose your
preferred setting and follow the above procedure. Refer to
the conversion graph on page 10. 7
9. Place the probe in the pH 4.0 solution, and
agitate for 10 seconds. Let stand for 60 seconds.
10. Rotate the 4.0 calibration screw counter clockwise
very slowly and gently until the display increases to
4.0.
11. Place the probe in the clear tap water cup, and
agitate for 10 seconds. Let stand for 60 seconds.
12. Repeat steps 6 to 11 as needed. This may take
several cycles, especially with worn or older
probes.
Nutrient Probe Calibration
What you will need:
•1000ppm Calibration
Solution
•Calibration Screw Driver
•Two small, clean plastic cups
•30 minutes
•Clear Tap Water
What to do:
1. Place all supplies on your calibration table, and let sit. All
supplies MUST be the same temperature before starting to
achieve an accurate calibration. It is recommended to float
the calibration solution bottle in your reservoir for five to ten
minutes to reach the reservoir temperature. This will allow for
a more accurate calibration.
2. Place the GrowBoss on a level surface and connect all
probes and power supply. Power on the GrowBoss.
3. Pour some calibration solution into the cup. Use enough
solution to completely cover the probe pins.
4. Pour clear tap water into the second cup for rinsing the ppm
probe.
5. Place the probe in the EC/PPM solution and agitate for 10
seconds. Let stand for two minutes.
6. Set the scale factor on the meter to .5 via the push button
selector. This is the NaCl scale. Rotate the PPM calibration
screw counter clockwise very slowly until the display reads
1000 ppm.
7. Due to the sensitivity of the ppm probe, this process can take
10 + minutes during initial setup, where fine tuning of the
calibration trimpot may be necessary. Turn the screw
counterclockwise to increase the value.
8. If you wish to calibrate to a different scale then choose your
preferred setting and follow the above procedure. Refer to
the conversion graph on page 10. 7
the conversion graph on page 10.

Care & Cleaning
Regular care and cleaning of the Nutradip GrowBoss and its
probes is the best way of ensuring that your device has a long, and
trouble free life.
Caring for pH Probes
The pH Probe is the only component of the GrowBoss that
requires periodic replacement. pH probes age through normal use,
and over time accuracy decreases. Eventually, the probe fails
completely. The life span of a pH probe depends on the level of
care it receives, with good care, they will give a long service life.
•DO NOT let the pH probe dry out. If allowed to completely dry
out, the probe will no longer function. Return the probe to the
buffer cap when not in use.
DO NOT knock the probe. It contains a sensitive thin walled
glass tube. If that tube breaks, the probe will no longer
function.
•DO NOT allow the connector of the probe to become wet.
Corrosion on the connector will prevent the signal from
reaching the GrowBoss
•DO NOT rapidly move the probe from cold to hot liquids. The
thermal stress will fracture the internal glass and the probe will
no longer function
•DO NOT put in greasy or oily substances
•DO NOT bend the probe. It is made of glass and it will break
•DO NOT lengthen the cord either with extension pieces or by
modifying the cord. The GrowBoss expects every pH cord to
be the same length, and changing this length will confuse the
GrowBoss resulting in inaccurate readings.
Cleaning pH Probes
As water evaporates from the solution that the pH probe is in,
residue can be left behind on the pH Probe.
Every Cleaning
The probe bulb should be carefully cleaned with a soft
toothbrush and pH 4.0 calibration solution. Do not touch the bulb
with your fingers.
Every Third Cleaning
The probe bulb should be carefully cleaned with a mixture
of 250mL of water and two-three drops of mild dish
washing liquid using a soft toothbrush, then rinsed with tap
water to remove all traces of detergent, and left overnight in
pH 4 storage solution.
8
Care & Cleaning
Regular care and cleaning of the Nutradip GrowBoss and its
probes is the best way of ensuring that your device has a long, and
trouble free life.
Caring for pH Probes
The pH Probe is the only component of the GrowBoss that
requires periodic replacement. pH probes age through normal use,
and over time accuracy decreases. Eventually, the probe fails
completely. The life span of a pH probe depends on the level of
care it receives, with good care, they will give a long service life.
•DO NOT let the pH probe dry out. If allowed to completely dry
out, the probe will no longer function. Return the probe to the
buffer cap when not in use.
DO NOT knock the probe. It contains a sensitive thin walled
glass tube. If that tube breaks, the probe will no longer
function.
•DO NOT allow the connector of the probe to become wet.
Corrosion on the connector will prevent the signal from
reaching the GrowBoss
•DO NOT rapidly move the probe from cold to hot liquids. The
thermal stress will fracture the internal glass and the probe will
no longer function
•DO NOT put in greasy or oily substances
•DO NOT bend the probe. It is made of glass and it will break
•DO NOT lengthen the cord either with extension pieces or by
modifying the cord. The GrowBoss expects every pH cord to
be the same length, and changing this length will confuse the
GrowBoss resulting in inaccurate readings.
Cleaning pH Probes
As water evaporates from the solution that the pH probe is in,
residue can be left behind on the pH Probe.
Every Cleaning
The probe bulb should be carefully cleaned with a soft
toothbrush and pH 4.0 calibration solution. Do not touch the bulb
with your fingers.
Every Third Cleaning
The probe bulb should be carefully cleaned with a mixture
of 250mL of water and two-three drops of mild dish
washing liquid using a soft toothbrush, then rinsed with tap
water to remove all traces of detergent, and left overnight in
pH 4 storage solution.
8

When Heavily Contaminated
Heat a diluted Potassium Chloride (KCl) solution to 60-80 degrees
Celsius. Place the tip of the probe into the heated solution and let
sit for an hour. Remove the probe from the solution and let it cool to
room temperature. Then clean with a toothbrush. Rinse with pH 4
solution.
Storing the pH Probe
The pH probe ships with a cap that contains pH 4.0 buffer
solution. This ensures that the tip of the probe is always wet. The
probe should always be stored in pH 4.0 solution. Storing the
probe in distilled, reverse osmosis, or deionized water will cause
premature wear to the probe. Rinse the probe before putting it
into the storage bottle solution to prevent contamination of the
solution in the bottle. Storage solution that becomes contaminated
should be promptly replaced.
If the probe has been allowed to dry out
Place the probe in pH 4.0 Calibration solution for 24 hours,
then attempt to complete a calibration to see if the probe
has suffered permanent damage.
Cleaning the Nutrient Probe
The nutrient probe should be carefully cleaned with a
soft toothbrush to clean off any tds residue. Ensure the probe pins
are dry.
Caring for Calibration Solutions
Your calibration is only as good as the solution used
to perform the calibration. Nutradip manufactures its
own calibration solution at our Kelowna British Columbia
laboratory to ensure that our solution is of the highest quality.
Storage
Store your calibration solutions in a cool dark place with the lids
firmly secured.
Ensure caps are always replaced. Calibration solution that has
been stored uncovered or with the seal compromised should be
discarded.
ALWAYS discard calibration solution after use. NEVER add solution
to the storage bottle.
Replacement
Replace your calibration solution within 2 months of opening
for continued accuracy.
9
When Heavily Contaminated
Heat a diluted Potassium Chloride (KCl) solution to 60-80 degrees
Celsius. Place the tip of the probe into the heated solution and let
sit for an hour. Remove the probe from the solution and let it cool to
room temperature. Then clean with a toothbrush. Rinse with pH 4
solution.
Storing the pH Probe
The pH probe ships with a cap that contains pH 4.0 buffer
solution. This ensures that the tip of the probe is always wet. The
probe should always be stored in pH 4.0 solution. Storing the
probe in distilled, reverse osmosis, or deionized water will cause
premature wear to the probe. Rinse the probe before putting it
into the storage bottle solution to prevent contamination of the
solution in the bottle. Storage solution that becomes contaminated
should be promptly replaced.
If the probe has been allowed to dry out
Place the probe in pH 4.0 Calibration solution for 24 hours,
then attempt to complete a calibration to see if the probe
has suffered permanent damage.
Cleaning the Nutrient Probe
The nutrient probe should be carefully cleaned with a
soft toothbrush to clean off any tds residue. Ensure the probe pins
are dry.
Caring for Calibration Solutions
Your calibration is only as good as the solution used
to perform the calibration. Nutradip manufactures its
own calibration solution at our Kelowna British Columbia
laboratory to ensure that our solution is of the highest quality.
Storage
Store your calibration solutions in a cool dark place with the lids
firmly secured.
Ensure caps are always replaced. Calibration solution that has
been stored uncovered or with the seal compromised should be
discarded.
ALWAYS discard calibration solution after use. NEVER add solution
to the storage bottle.
Replacement
Replace your calibration solution within 2 months of opening
for continued accuracy.
9

Specifications
Abilities
10
Operational Environment
-20 to 80C / -4 to 176 F

Troubleshooting Guide
Problem - No Display
Follow the steps below until a solution is reached:
1. Check that the Growboss is connected to the 5dcv/0.5 amp
power supply
2. Check that the Power supply is connected to the wall outlet
3. Verify that the outlet has power by connecting another
device such as a lamp or fan
4. Unplug all cords and probes, wait 15 minutes, reconnect
and attempt to power on
5. Replace the Power Supply
Problem – pH Probe Inaccurate
Follow the steps below until a solution is reached:
1. Ensure the pH probe is connected to the GrowBoss and
locked into place
2. Ensure probe and solution are the same temperature, this
can take up to 15 minutes
3. Ensure the probe to meter connection is not soiled. Clean
with compressed air
4. Ensure the probe is clean. See cleaning instructions
5. Calibrate your Growboss following the calibration guide
6. Rehydrate your pH Probe following the rehydration guide
7. Replace the pH Probe
Problem - Temperature Inaccurate
Follow the steps below until a solution is reached:
1. Ensure device is set to either Celsius or Fahrenheit,
depending on your preference
2. Ensure probe and solution are the same temperature, this
can take up to 15 minutes
3. Ensure the PPM probe is fully plugged into the meter
Problem - Nutrient reading is inaccurate
Follow the steps below until a solution is reached:
1. Ensure the device is set to display Nutrient in your preferred
method of display
2. Ensure probe and solution are the same temperature, this can
take up to 15 minutes
3. Ensure the probe to meter connection is not soiled. Clean with
compressed air
4. Ensure the probe pins are clean. See cleaning instructions
5. Calibrate your GrowBoss following the calibration guide
6. Ensure the PPM probe is fully plugged into the meter
7. Replace your Nutrient probe
11
Troubleshooting Guide
Problem - No Display
Follow the steps below until a solution is reached:
1. Check that the Growboss is connected to the 5dcv/0.5 amp
power supply
2. Check that the Power supply is connected to the wall outlet
3. Verify that the outlet has power by connecting another
device such as a lamp or fan
4. Unplug all cords and probes, wait 15 minutes, reconnect
and attempt to power on
5. Replace the Power Supply
Problem – pH Probe Inaccurate
Follow the steps below until a solution is reached:
1. Ensure the pH probe is connected to the GrowBoss and
locked into place
2. Ensure probe and solution are the same temperature, this
can take up to 15 minutes
3. Ensure the probe to meter connection is not soiled. Clean
with compressed air
4. Ensure the probe is clean. See cleaning instructions
5. Calibrate your Growboss following the calibration guide
6. Rehydrate your pH Probe following the rehydration guide
7. Replace the pH Probe
Problem - Temperature Inaccurate
Follow the steps below until a solution is reached:
1. Ensure device is set to either Celsius or Fahrenheit,
depending on your preference
2. Ensure probe and solution are the same temperature, this
can take up to 15 minutes
3. Ensure the PPM probe is fully plugged into the meter
Problem - Nutrient reading is inaccurate
Follow the steps below until a solution is reached:
1. Ensure the device is set to display Nutrient in your preferred
method of display
2. Ensure probe and solution are the same temperature, this can
take up to 15 minutes
3. Ensure the probe to meter connection is not soiled. Clean with
compressed air
4. Ensure the probe pins are clean. See cleaning instructions
5. Calibrate your GrowBoss following the calibration guide
6. Ensure the PPM probe is fully plugged into the meter
7. Replace your Nutrient probe
11

Problem Won't Hold Calibration
Follow the steps below until a solution is reached:
1. Power off the unit. Unplug probe(s) and carefully plug back in,
ensuring connections are tight.
2. Replace calibration solution and recalibrate as per
procedures
3. Leave probe in solution for at least 10 minutes
4. Replace probe(s)
12
Problem Won't Hold Calibration
Follow the steps below until a solution is reached:
1. Power off the unit. Unplug probe(s) and carefully plug back in,
ensuring connections are tight.
2. Replace calibration solution and recalibrate as per
procedures
3. Leave probe in solution for at least 10 minutes
4. Replace probe(s)
12

13
Warranty Coverage
The Nutradip GrowBoss is covered by three separate
warranty programs.
The 12 month warranty coverage protects the product from
material/workmanship defects.
The Accidental Damage Coverage plan protects product from
damage caused by user error or negligence.
The theft coverage plan allows a replacement Nutradip
GrowBoss to obtained at a discounted rate in the event of loss or
theft. (Please visit http://www.futureharvest.com/warranty/nutradip/loss-
theft-coverage/ for more information)
Registration of the product within 30 days of purchase is required in
order warranty coverage to be obtained. PLEASE SAVE YOUR
RECEIPT AS A COPY MAY BE REQUIRED FOR ALL WARRANTY SERVICES.
Nutradip 12 Month Warranty Coverage
Nutradip Accidental Damage Coverage
1. The product must be shipped back to Future Harvest
2. Only a photo proving the item is defective must be sent
Do not return to the store of purchase. Warranty claims are
handled directly by Nutradip. All warranty claims must be
submitted online at: www.futureharvest.com/warranty
13
Warranty Coverage
The Nutradip GrowBoss is covered by three separate
warranty programs.
The 12 month warranty coverage protects the product from
material/workmanship defects.
The Accidental Damage Coverage plan protects product from
damage caused by user error or negligence.
The theft coverage plan allows a replacement Nutradip
GrowBoss to obtained at a discounted rate in the event of loss or
theft. (Please visit http://www.futureharvest.com/warranty/nutradip/loss-
theft-coverage/ for more information)
Registration of the product within 30 days of purchase is required in
order warranty coverage to be obtained. PLEASE SAVE YOUR
RECEIPT AS A COPY MAY BE REQUIRED FOR ALL WARRANTY SERVICES.
Nutradip 12 Month Warranty Coverage
Nutradip Accidental Damage Coverage
1. The product must be shipped back to Future Harvest
2. Only a photo proving the item is defective must be sent
Do not return to the store of purchase. Warranty claims are
handled directly by Nutradip. All warranty claims must be
submitted online at: www.futureharvest.com/warranty

What is Not Covered:
1. Use of parts not assembled or installed in accordance with
the instructions of Future Harvest. The product guarantee
becomes effective from the date of purchase by the first
purchaser. Coverage terminates if you sell or otherwise
transfer the product.
2. The repair of your product under guarantee will not extend
the period of the guarantee.
3. Purchase of the product from a third party site that is not an
authorized distributor or retailer.
Provided you supply proof of purchase via a store-printed receipt,
we will repair or replace your product if your product is found,
within the guarantee period, to be defective due to defective
materials or workmanship existing at the time of purchase. If any
part is no longer available or out of maufacture, Future Harvest will
replace it with a functionally-equivalent replacement part.
Future Harvest shall not be liable for costs of repair or replacement
of a product incurred as a result of:
1. Normal wear and tear.
2. Accidental damage, faults caused by negligent use or care,
neglect, careless operation or handling of the product which
is not in accordance with the Nutradip GrowBoss Instruction
Manuals. See accidental warranty coverage.
3. Use of parts or accessories other than those produced or
recommended by Future Harvest.
4. External sources such as transit damage or weather.
5. Repairs or alterations carried out by parties other than Future
Harvest or its authorized agents.
6. Serial numbers defaced or missing.

15
Making a Warranty Claim
All warranty plans are subject to change without notice. Visit
www.futureharvest.com/warranty for more information. Shipping
and handling charges may apply.
TO THE MAXIMUM EXTENT PERMITTED BY LAW,THIS GUARANTEE AND THE REMEDIES SET
FORTH ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, GUARANTEES AND
REMEDIES, WHETHER ORAL OR WRITTEN, EXPRESS OR IMPLIED. ANY IMPLIED WARRANTIES
THAT MAY BE IMPOSED BY LAW (INCLUDING, WITHUT LIMITATION, MERCHANTABILITY) ARE
LIMITED IN DURATION TO THE PERIOD OF THIS LIMITED GUARANTEE.
TO THE MAXIMUM EXTENT PERMITTED BY LAW, FUTURE HARVEST IS NOT RESPONSIBLE FOR,
AND CONSUMER RELEASES FUTURE HARVEST FROM SPECIAL, INCIDENTAL OR
CONSEQUENTIAL DAMAGES RESULTING FROM ANY BREACH OF WARRANTY, OR RELATED IN
ANY WAY TO THE PRODUCT, OR UNDER ANY OTHER LEGAL OR EQUITABLE THEORY.