
Manual | NEP-9500-PLUS series turbidity probe
Status: Final | Not confidential
7 Installation
There are aspects to consider when preparing to install the Analite NEP-9500 series probes in the field:
1. Installation of the probe properly into the environment where measurements are to take place
2. Physical cabling considerations
3. Electrical cabling considerations
7.1 Installing the probe for the first time
The probe is normally installed with the optics pointing downwards (but not vertical) or in a horizontal
alignment. In a simple application the probe is immersed into the water to the desired depth, but within the
depth rating of the probe. Please note the depth rating is based on stagnant water. Allowances must be
made for the effect of flowing water to ensure the stagnant depth rating is not exceeded. If the probe is to
be installed downwards, then it is recommended to install it a few degrees away from the vertical to allow
bubbles to escape away from the optic face.
Probes with integrated wiping should be installed such that they can be easily retrieved from time to time
to replace the wiper arm assembly. It is important that the optic end of the probe is kept clear of
obstruction such as the river bed. The minimum distance between the optic head and any object should
be 50mm (2”).
The standard casing of the Analite NEP-9500 series probes is made of 6 stainless steel but the optic face
is made of plastic materials and so should be protected from accidental scratching or abrasion. The wiper
arm assembly should be replaced periodically to avoid abrasive material build-up in the pad that may
eventually abrade the optic face. The optic face is partially protected from damage by the protruding
castellation in the probe casing.
To avoid crevice corrosion on the 6 stainless steel casing in salt and acidic water deployments, it is
strongly recommended that the probe be thoroughly washed in clean water after usage and prior to
storage. Failure to do this will cause crevice corrosion and this is not covered under warranty. A composite
cased version is available for corrosive and sea water environments but its pressure rating is limited to
30m.
If the probe is to be installed in a glanded fitting (for insertion into a pipe etc.) then care must be taken to
ensure the sealing surface pressures offered by the gland fitting are not excessive so as to not cause
distortion of the probe casing and force leakage. The Analite NEP-9500 series probes are thin wall
instruments and so glanding pressure must be minimal and spread over the largest possible area. Do not
cut or damage the outer sheath of the cable. Water may enter the probe through holes or cuts in the cable
sheath. Where damage may occur due to river rocks striking or rolling over the probe body, a protective
shroud should be used (plastic or stainless steel, whichever the situation warrants. A shroud is available
for the NEP-9500 as an accessory under the part number NEP-9SHRD. Such a shroud not only protects
the probe but also assists in maintaining a minimum distance between the probe optics and any local
obstructions.