Ohm-Labs, Inc. 611 E. Carson St. Pittsburgh, PA 15203-1021 Tel. 412-431-0640 www.ohm-labs.com
100-Series Resistance Standards
Instruction Manual
3) Use
On receipt, inspect the standard for physical damage. If damaged, please immediately contact the
carrier. We will assist with any damage claims and/or necessary repair.
Review the Report of Calibration accompanying the standard. The reported value is at 23 °C.
At 10 Kand below, these standards must be used as four-terminal resistors to realize their stated
accuracy. Make current and potential connections via the binding posts on the top of the standard. The
standard is screened to identify these terminals.
Connection may be made with bare copper wire, silver plated solid copper wire, gold plated copper
spade lugs or gold plated 4mm banana plugs. Wire may be passed through the hole in the binding post or
wrapped around the post. Do not over tighten the binding posts; a snug finger tight pressure is adequate.
The terminals are marked for Current (C) and Potential (P). On 4-wire ohmmeters, connect C to Input and P
to Sense. Resistors below 10 K should always be measured in 4-wire mode. A ground terminal is provided
for the case.
Allow 24 hours for the standard to acclimatize at ambient temperature (23 °C nominal).
For best measurement accuracy, do not exceed the current ratings of the standard. Although
momentary application of current above the rated value will not damage these standards, self-heating will
change the resistance from the reported value.
Caution: Do not exceed maximum rated current. Damage may occur.
4) Measured Value and Temperature Coefficients of Resistance
Each standard’s Report of Calibration includes its measured value at 23 °C and its alpha and beta
temperature coefficients of resistance. Barring damage, the temperature coefficients of resistance will not
change over the life of the standard and do not need to be re-determined.
The alpha () coefficient is the change in resistance with temperature at 23 °C; the beta () coefficient is
the curvature of this change. Within a temperature range around ambient (18-30 °C), the resistance of a
standard may be accurately expressed as:
Rt= R23[1+(t-23)+(t-23)2]
Where: Rt= Resistance at temperature ‘t’
R23 = Resistance at 23°C
t = Temperature of resistor
Each resistor is supplied with a table of resistance versus temperature. As a visual aid, this data is also
presented in a graph.