2Installation
2.1 Installation of the binocular viewing head
1) Loosen the thumb screw (11) on the top of the body (12) and
remove the plastic cover on the top.
2) Remove the cap on the circular dovetail of the binocular viewing
head (10)
3) Insert the dovetail into the socket on the top of body (12);
ensure that it is completely seated into the socket; tighten the thumb screw (11).
Caution: Do not release the head from your hand grip until you are sure the head is
installed securely.
2.2 Installation of the eyepieces
1) Remove the protective caps from the eyepiece tubes (3).
2) Insert the eyepieces (1) into the eyepiece tubes (3).
2.3 Installation of the objectives
1) Adjust the coarse focus knob (14) until the mechanical stage (7) is at its lowest
position.
2) Install the 4x objective into the nosepiece (4). Then in a clock-wise direction, rotate
the nosepiece and install each succeeding higher magnification objective.
Note: Inspect the objectives frequently for dirt or oil; clean if necessary. Use the 10x
objective to initially focus the image of your specimen. When changing the objective
magnification, rotate the objective nosepiece until you hear a “click” sound. This
ensures the objective is centered in the optical light path.
2.4 Installation of the filter
1) Swing out the filter holder under the condenser (21).
2) Put the filter onto the holder, swing the holder back in.
2.5 Installing (or changing) the halogen bulb
1) Turn the power off and disconnect the power cord.
2) Allow some time to cool down the lamp.
3) Turn over the microscope on its side; find the bulb compartment at the bottom.
4) Open the cover of the bulb compartment by loosening the thumb screw. Take out
the dead bulb and insert the new bulb. Be sure the pins on the bulb
are completely inserted into the lamp socket. You may also loosen
the two screws on the cover to adjust the position of the bulb to get
centered and even brightness. Screw the cover on.
Caution: Before you turn over the microscope, be sure to take the
eyepieces off and be certain that the head is securely locked by the
thumb screw.
2.6 Replacing the fuse
1) Turn off the power and disconnect the power cord.
2) Turn over the microscope on its side; find the fuse holder at the bottom of the base.
3) Turn the fuse holder counter-clockwise to take it out, replace the fuse, and then turn