9
OM-CP-ULTRASHOCK-A & OM-CP-SHOCK300
OM-CP-ULTRASHOCK-A Specifications
Battery Warning
FIRE, EXPLOSION AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CRUSH, PENETRATE, INCINERATE OR DISASSEMBLE. AVOID
TEMPERATURES ABOVE THE MAXIMUM OPERATING TEMPERATURE OF THE PRODUCT. DISPOSE OF PROPERLY. CHARGE ONLY WITH THE
PROVIDED OMEGA CHARGER, OR FROM OTHER USB POWER SOURCE VIA THE OMEGA PROVIDED CABLE.
Specifications subject to change.
See Omega’s terms and conditions at www.omega.com
TEMPERATURE
Sensor Semiconductor
Range -20 ºC to +60 ºC
Resolution 0.1 ºC
Accuracy ±0.5 °C (+5 °C to +60 °C)
HUMIDITY
Sensor Capacitive Polymer
Range 0 %RH to 95 %RH
Resolution 0.1 %RH
Calibrated
Accuracy
±3 %RH (±2 %RH typical at 25 ºC)
Specified
Accuracy Range
+20 °C to +40 °C; 25 %RH to 75 %RH
PRESSURE
Sensor Semiconductor Strain Gage
Range 100 mbar to 1300 mbar
Resolution 0.05 mbar
Calibrated
Accuracy
±1.5 mbar at 25 ºC; at 750 mbar
SHOCK
Accelerometer
Type
MEMS Semiconductor
Acceleration
Range (g)
±15 g ±100 g ±300 g
Acceleration
Resolution (g)
0.02 g 0.05 g 0.2 g
Calibrated
Accuracy (g)
±0.3 g ±2.0 g ±6.0 g
Sampling Rate 1000 Hz
Accelerometer
Freq. Resp.
>1000 Hz (15 g)
>500 Hz (100 g, 300 g)
Reading Rate 1024 Hz to 5 minutes for shock,
selectable in software. Temperature,
pressure and humidity sampled
approx. every 2 seconds at intervals
shorter than 2 seconds. Otherwise,
sampled at the reading rate.
GENERAL
Memory 3,593,358 readings (598,893 per
channel, all channels in use)
Start Modes Software programmable immediate
start or delay start, up to 6 months in
advance
Real Time
Recording
May be used with PC to monitor and
record data in real time at a 1 second
sample rate
Password
Protection
An optional password may be
programmed into the device to restrict
access to configuration options. Data
may be read out with the password.
Calibration Digital calibration through software
Calibration Date Automatically recorded within device
Battery Type Internal Lithium Ion pack, charger
included
Battery Life 90 days typical at 64Hz rate
Data Format Date and time stamped gravities (g
and mg), temperature (°C, °F, K, °R),
humidity (%RH, mg/ml water vapor
concentration), pressure (PSIA, inHg,
mmHg, bar, atm, Torr, Pa, kPa, MPa)
IP Rating IP64
Time Accuracy 10 seconds/month (at 0 ºC to 50 ºC)
Computer
Interface
USB-C cable required (included);
1MBaud
Supported
Software
Windows XP SP3/7/8/10
Software
Compatibility
Standard Software version 4.2.15.0 or
later
Secure Software version 4.2.14.0 or
later
Operating
Environment
-20 ºC to +60 ºC,
0 %RH to 95 %RH non-condensing
Dimensions 3.4 in x 1.7 in x 1.3 in
(86 mm x 43 mm x 33 mm)
Weight 8 oz (227 g)
Enclosure Anodized aluminum
Approvals CE compliant
EMC Directive 2014/30/EU
RoHS Directive 2011/65/EU