Omega OM-CP-HITEMP140-FP User manual

OM-CP-HITEMP140-FP
High Temperature Data Logger
with Flexible RTD Probe
For latest product manuals:
www.omegamanual.info
Shop online at
omega.com
User’s Guide
TM

Product Overview
The OM-CP-HITEMP140-FP is a durable, user friendly high temperature data logger featuring a long, flexible
RTD probe with a narrow diameter, making it ideal for use in lyophilization processes.
Commonly used for mapping, validation and monitoring of high temperature surfaces and environments, this
stainless steel data logger is available in two models, the OM-CP-HITEMP140-FP-36 and the
OM-CP-HITEMP140-FP-72, which feature either 36” or 72” flexible probe lengths, respectively. The flexible
probe is coated with PFA insulation and can withstand temperatures up to 260°C with an accuracy of ±0.1°C.
The OM-CP-HITEMP140-FP probe design is narrow and lightweight making it ideal for placement within small
vials, tubing, test tube and other small diameter or delicate applications. Because of the flexible probe, the
risks of breakage (both vial and probe) generally associated with stainless steel probe loggers are diminished
and the location and placement of the probe is easy to manipulate. The device records and stores up to
32,700 time stamped readings and is equipped with non-volatile solid state memory that will continue
recording even if battery becomes discharged.
The OM-CP-HITEMP140-FP can be configured for delayed start and is capable of reading rates as often as four
readings per second, up to once every 24 hours.
Submergibility
The OM-CP-HITEMP140-FP is rated IP68 and is fully submersible. It can be placed in environments up to 230’
(70m) of water.
Bend Radius
The probe should not be bent within 1” where the probe meets the logger or less than 1” from the probe tip.
O-Rings
O-ring maintenance is a key factor when properly caring for the OM-CP-HITEMP140-FP. The O-rings ensure a
tight seal and prevent liquid from entering the inside of the device.
Trigger Settings
The device can be programmed to only record based o user configured trigger settings.
1. In the Connected devices panel, select the intended device to change the settings.
2. On the Device tab, in the Information group, click Properties. Users can also right-click on the device and
select Properties in the context menu.
3. Click Trigger and configure the Trigger settings. Trigger formats are available in Window and Two Point
(bi-level) mode. Window mode allows for one range of temperature monitoring and two point mode
allows for two ranges.
Note: This product is rated for use up to 140°C. Please heed the battery warning. Additionally, this product is not
ideal for steam sterilization application above 121°C /1.1 bar.
OM-CP-HITEMP140-FP
2

Troubleshooting Tips
Why is the data logger not appearing in the software?
If the OM-CP-HITEMP140-FP doesn’t appear in the Connected Devices panel, or an error message is received
while using the OM-CP-HITEMP140-FP, try the following:
• Check that the OM-CP-IFC400 is properly connected. For more information, see Troubleshooting Interface
Cable problems (below).
• Ensure that the battery is not discharged. For best voltage accuracy, use a voltage meter connected to the
battery of the device. If possible, try switching the battery with a new OM-CP-BAT110.
• Ensure that no other Omega software is running in the background.
• Ensure that Omega Software is being used.
• Ensure that the Connected Devices panel is large enough to display devices. This can be verified by
positioning the cursor on the edge of the Connected Devices panel until the resize cursor appears, then
dragging the edge of the panel to resize it. The screen layout may also be reset in the options menu by
selecting File, Options, and scrolling to the bottom.
Troubleshooting Interface Cable problems
Check that the software properly recognizes the connected OM-CP-IFC400.
If the data logger is not appearing in the Connected Devices list, it may be that the OM-CP-IFC400 is not
properly connected.
1. In the software, click the File button, then click Options.
2. In the Options window, click Communications.
3. The Detected Interfaces box will list all of the available communication interfaces. If the OM-CP-IFC400 is
listed here, then the software has correctly recognized and is ready to use it.
Check that Windows recognizes the connected OM-CP-IFC400.
If the software does not recognize the OM-CP-IFC400, there may be a problem with Windows or the USB
drivers.
1. In Windows, click Start, right-click Computer and choose Properties or press Windows+Break as a
keyboard shortcut.
2. Click Device Manager in the left hand column.
3. Double click Universal Serial Bus Controllers.
4. Look for an entry for Data logger Interface.
5. If the entry is present, and there are no warning messages or icons, then windows has correctly
recognized the connected OM-CP-IFC400.
6. If the entry is not present, or has an exclamation point icon next to it, the USB drivers may need to be
installed. These are available on the software flash drive included with the OM-CP-IFC400.
Ensure that the USB end of the OM-CP-IFC400 is securely connected to the computer.
1. Locate the USB-A plug of the OM-CP-IFC400.
2. If the interface cable is connected to the PC, unplug it. Wait ten seconds.
3. Reconnect the cable to the PC.
4. Check to make sure that the red LED is lit, indicating a successful connection.
3
Quick Start Manual

Installation Guide
Installing the Interface cable
- OM-CP-IFC400 or OM-CP-IFC406
Refer to the “Quick Start Guide” included in the package.
Installing the software
Insert the Omega Software Flash Drive in an open USB port. If the autorun does not appear, locate the drive
on the computer and double click on Autorun.exe. Follow the instructions provided in the Installation
Wizard.
Device Operation
Connecting and Starting the data logger
1. Once the software is installed and running, plug the interface cable into the data logger.
2. Connect the USB end of the interface cable into an open USB port on the computer. Place the data logger
into the docking station.
3. The data logger will automatically appear under Connected Devices within the software.
4. For most applications, select “Custom Start” from the menu bar and choose the desired start method,
reading rate and other parameters appropriate for the data logging application and click “Start”. (“Quick
Start” applies the most recent custom start options, “Batch Start” is used for managing multiple loggers at
once, “Real Time Start” stores the dataset as it records while connected to the logger.)
5. The status of the device will change to “Running”, “ Waiting to Start” or “Waiting to Manual Start”,
depending upon your start method.
6. Disconnect the data logger from the docking station and place it in the environment to measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this
point the device cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
1. Connect the logger to the interface cable.
2. Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
3. Once the data logger is stopped, with the logger highlighted, click “Download”. You will be prompted to
name your report.
4. Downloading will ooad and save all the recorded data to the PC.
OM-CP-HITEMP140-FP
4

Product Maintenance
Battery Replacement
Materials: OM-CP-BAT110
1. Unscrew the bottom of the logger and remove the battery.
2. Place the new battery into the logger. Note the polarity of the battery.
3. Screw the cover back onto the logger.
Recalibration
The OM-CP-HITEMP140-FP standard calibrations are two points at 30°C and 140°C.
Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in
the software when the device is due.
Notice: Steam Sterilization Applications
The pervasive nature of pressurized steam creates a very dicult environment for electronics.
Please refer to the following preventative maintenance procedure when using this device in steam sterilization
applications.
Additionally, this device is not ideal for steam sterilization applications above 121 °C/1.1 bar.
PREVENTATIVE MAINTENANCE
After every 3 hours of steam exposure:
1. Remove the endcap and battery from the device (ref. battery change procedure on Product User Guide)
2. Place open logger (minus battery) in an oven at 120°C (250°F) for a minimum of 30 minutes
3. Remove logger from oven and allow to cool to room temp
4. Re-assemble the logger with the battery (note polarity) and endcap
Quick Start Manual
5

OM-CP-HITEMP140-FP
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH,
PENETRATE OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 80 °C (176 °F).
Specifications subject to change.
See Omega’s terms and conditions at www.omega.com
Temperature Sensor Flexible RTD Probe
Probe Range -60°C to +260°C (-76°F to +500°F)
Temperature Resolution 0.01°C (0.02°F)
Calibrated Accuracy ±0.1°C (0.18°F)
Memory 32,767 readings
Number of Readings in
Trigger Settings Mode 10,922 readings
Reading Rate 1 reading every 0.25 seconds up to 1 reading every 24 hours
Required Interface Package OM-CP-IFC400 or OM-CP-IFC406 USB docking station required
Baud Rate 125,000 baud
Typical Battery Life 1 year typical (1 minute reading rate at 25°C/ 77°F)
Operating System
Compatibility XP SP3/Vista/Windows 7/Windows 8
OM-CP Software
Compatibility
• OM-CP Standard Software version 4.2.1.1
• OM-CP Secure Software version 4.2.0.1 or later
Operating Environment -40°C to +140°C (-40°F to +284°F) 0%RH to 100%RH
Material • Body: 316 Stainless Steel
• Probe: PFA Insulated Cable
Dimensions (Body) 2.95” x 0.97” x 0.97” (75mm x 24.6mm x 24.6mm)
Dimensions (Probe) • OM-CP-HITEMP140-FP-36: 36” x 0.10” (914mm x 2.5mm)
• OM-CP-HITEMP140-FP-72: 72” x 0.10” (1829mm x 2.5mm)
Weight 85g (3oz)
Submersible Yes (IP68)
Approvals CE
OM-CP-HITEMP140-FP General Specifications
6

WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months
from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product
warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an
Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to
be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action
of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper
repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows
evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specifica-
tion; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warrant-
ed, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes respon-
sibility for any omissions or errors nor assumes liability for any damages that result from the use of its products
in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts
manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES
OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are
exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negli-
gence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which
liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10
CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will
indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in
such a manner.
FOR WARRANTY RETURNS, please have the following infor-
mation available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was
PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the
product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current
repair charges. Have the following information available BEFORE
contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the
product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and
engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2015 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER
SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the
outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
Servicing North America:
U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047
Stamford, CT 06907-0047 USA
Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only)
Engineering Service: 1-800-872-9436 (USA & Canada only)
Tel: (203) 359-1660 Fax: (203) 359-7700
For Other Locations Visit omega.com/worldwide
omega.com [email protected]
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the
right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
MQS5090/1118
TM
Table of contents
Other Omega Data Logger manuals

Omega
Omega OM-CP-RFPULSE2000A User manual

Omega
Omega OM-DLAC User manual

Omega
Omega OM-EL-WiFi-TP User manual

Omega
Omega HHAQ-104 User manual

Omega
Omega OM-USB-5203 User manual

Omega
Omega OM-EL-WIN-USB User manual

Omega
Omega OM-DAQLINK-TEMP User manual

Omega
Omega OM-NET-1608 User manual

Omega
Omega OM-CP-RHTEMP101A User manual

Omega
Omega OM-60-TE User manual

Omega
Omega OM-VIB-101 User manual

Omega
Omega OM-DLAV User manual

Omega
Omega OM-CP-PRHTEMP2000 User manual

Omega
Omega OMB-DAQ-2408 Series User manual

Omega
Omega OMB-DAQ-TC-RACK User manual

Omega
Omega OM-CP-PROCESS101A User manual

Omega
Omega HH520 User manual

Omega
Omega OMP-MODL User manual

Omega
Omega OM-CP-QUADVOLT User manual

Omega
Omega OM-21A User manual