Metal Samples MS4500E-HC User manual

MS4500E-HC
High Resolution ER Data Logger
Operator’s Manual
Metal
Samples Company
ADivisionof Alabama SpecialtyProducts,
Inc.
152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358-4202 Fax: (256)
358-4515

Table of
Contents
I.Introduction............................................................................................................................................1
A.GeneralDescription.................................................................................................................1
B.PrinciplesofOperation............................................................................................................1
C.TechnicalSpecifications...........................................................................................................3
II.StartUpandOperation..........................................................................................................................4
A.ReceivingtheMS4500E‐HCDataLogger.................................................................................4
B.StartUp....................................................................................................................................4
1.PortDetailsandProbeConnection............................................................................4
2.PowerUp....................................................................................................................5
a.SafeAreaUsage.............................................................................................5
b.HazardousAreaUsage....................................................................................6
c.BatteryReplacement.......................................................................................8
3.SetupandOperation.....................................................................................................9
a.SelectProbe...................................................................................................9
i.NewProbe................................................................................................10
ii.ExistingProbe...........................................................................................10
iii.DeleteExistingProbe................................................................................11
b.MakeMeasurement.....................................................................................12
c.RecallReadings..............................................................................................13
d.DeleteReadings............................................................................................14
e.SetTime&Date..........................................................................................15
f.SetMeasurementTimer................................................................................16
g.Communications...........................................................................................17
h.Language......................................................................................................23
i.PowerOff.......................................................................................................23
j.BacklightOn/Off.............................................................................................23
C.Maintenance..........................................................................................................................24
1.TestingtheMS4500E‐HCHighResolutionDataLoggerwiththeMeterProver......24
D.Troubleshooting.....................................................................................................................25
III.ServiceandWarrantyInformation.....................................................................................................26
A.Warranty................................................................................................................................26
B.ObtainingServiceandReturningtheInstrumentforRepair.................................................26
C.InstrumentRepairForm.........................................................................................................27
AppendixA–RevisionHistory.................................................................................................................28
AppendixB–Hazardousareacertificationdetails..................................................................................29
AppendixC‐Drawings.............................................................................................................................31
ControlDrawing(HazardousAreaWiringDiagram)...................................................................32

1
I. Introduction
A.GeneralDescription
TheMS4500E‐HCisahand‐held,batterypowered,corrosionmetercapableofmeasuringandstoringdata
fromalltypesofelectricalresistance(ER)corrosionprobes.Theinstrumentislightweight,
microprocessor‐based,andfeaturesasimple,menu‐driveninterfaceusingakeypadandabacklit
graphicalLCDdisplay.
Corrosionratemeasurementsaremadeusingtheelectricalresistancemethod.Essentially,theinstrument
measurestheresistanceoftheprobeelementwhichchangesovertime,asmetallossoccurs.Therateof
changeisdirectlyproportionaltocorrosionrate.Thismethodfindsawidevarietyofapplicationssinceit
canbeusedinconductiveandnonconductiveenvironmentssuchaspetroleum,chemical,water,soil,or
evenatmosphere.Thenewhigh‐resolutionmeasurementoftheMS4500E‐HCdetectssmallerincrements
ofmetalloss,providingfasterresponsethantraditionalERinstruments.
Aftertakingareading,theinstrumentdisplaysmetallossinmilsandcorrosionrateinmilsperyear(mpy).
Thereadingcanthenbestoredtomemoryordiscarded.Allstoredreadingsareautomaticallytimeand
datestamped.Readingsarestoredtonon‐volatileFlashmemorywhichretainsdatawithouttheneedfor
abatterybackup.
TheMS4500E‐HCHighResolutionERDataLoggercanstore16,000readingsperprobeonupto250
differentprobes(4milliontotal).StoreddatacanbedownloadeddirectlytoaUSBFlash(“jump”)drivein
non‐hazardousareas.DatacanbeopenedandchartedusingtheprovidedCDMSsoftware,orcanbe
importedintoanystandarddataanalysis(spreadsheet)programsuchasMicrosoftExcel.Datacanalsobe
reviewedandchartedontheinstrument’sLCDdisplayforquickreference.
B.PrinciplesofOperation
TheMS4500E‐HCHighResolutionERDataLoggeroperatesontheElectricalResistance(ER)technique
andisusedinconjunctionwithanERprobe.TheERprobeutilizesaresistivesensingelement
manufacturedfromthematerialofinterest(oracloseapproximation)whichisexposedtoacorroding
environment.ThisiscalledtheExposedorCorrodingElement.TheresistanceoftheExposedElement
isdirectlyrelatedtoitsthickness,soastheelementcorrodestheresultinglossofmetalcausesa
proportional
increasein
theelement’sresistance.TheprobealsocontainsaninternalReference
ElementwhichisusedtocompensatefortheinfluencesoftemperatureontheExposedElement.
TheMS4500E‐HCHighResolutionERDataLogger isdesignedtoworkwithanystandardERprobe,
butitisrecommendedthatCylindricalandLargeFlushtypeprobesbeusedtoensureoptimum
performance.TheirphysicaldesignplacestheReferenceElementincloserproximitytotheExposed
Elementcomparedtootherprobetypes,providingmoreeffectivetemperaturecompensationand
thusreducingtheeffectsofthermalnoise.

2
Becausetheyaredesignedtocorrode,ERprobesaresacrificialinnature.EachERprobewillhavea
finitelifethatisbasedontheelementthickness.ERprobesareavailableinanumberofgeometries
andthicknessesdesignedtosuitawidevarietyofapplications.Table1liststhecommonERelement
optionsavailablefromMetalSamplesandtheeffectivelifeofeach.
ElementTypeCompatibilityThicknessProbeLife
(Span)
TubularLoopCompatible
42
84
WireLoopCompatible
4010
8020
Flush(Small)
Compatible
42
84
2010
Cylindrical
Preferred
105
2010
5025
Flush(Large)
Preferred
52.5
105
2010
4020
Table1.StandardERProbeElements
TheMS4500E‐HCHighResolutionERDataLoggermeasuresanERprobeutilizingahigh‐resolution,16‐
bitmeasurement.Thisallowstheunittodetectmuchsmalleramountsofmetalloss,thus
respondingfastertocorrosioneventsandupsets. At16‐bitresolutiontheMS4500E‐HCERData
Loggercanmeasuremetallossamountsassmallas0.004%oftheprobelife. Forhighlysensitive
probessuchasthe5‐millargeflushshownabove,thatresultsinadetectionlimitoflessthanone
angstrom.

3
C.Technic alSpecifications
ModelMS45000E‐HCHighResolutionERDataLogger
Physical Data
InstrumentWeight(w/boot):2.21lb.(1.00kg)
TotalWeightw/Case&Accessories:7.46lb.(3.38kg)
InstrumentDimensions(w/boot):8.75"Lx4.54"Wx2.20"D(21.77cmx11.53cmx5.59cm)
InstrumentDimensions(w/oboot):7.50"Lx4.00"Wx2.00"D(19.05cmx10.16cmx5.08cm)
CarryingCaseDimensions:14.50"Lx11.38"Wx5.88"D(36.83cmx28.89cmx14.92cm)
OperatingTemperature: ‐4°to158°F(‐20°to70°C)
StorageTemperature: ‐40°to158°F(‐40°to70°C)
Performance Data
MeasurementType:ERmeasurementusinganystandardERProbeTypes
PreferredProbeTypes:Cylindrical,Flush(Large)
Compatibleprobetypes:TubeLoop,WireLoop,Flush(Small)
Range:0‐25,000probelifeunits(displayedas0.00to1000.00PLU’sin
0.04increments)
Resolution:0.004%ofProbeLife
Repeatability:±0.1%ofFullScale
Electrical Data
PowerRequirements:Two'C'size3.6VLithiumBatteries(XenoEnergyXL‐145For
TadiranTL4920)
MaximumProbeCableDistance:200ft.(61m)
DownloadMethod:Non‐hazardousarea:DirectlytoUSBFlashdriveortoPCthrough
certifiedUSBBarrierHazardousarea:DirectlytocertifiedUSB
DataTransferUnit
Hazardous Location Certifications –Intrinsic Safety
USA/CanadaForuseinClassIZone0AExia[ia]IICT4Ga
Exia[ia]IICT4Ga
ClassI,Division1,GroupsA,B,C&D,T4
ProvidesoutputstoClassI,Division1[Exia]
‐40°C≤Ta≤+70°C
EuropeandWorldwideII1GExia[ia]IICT4Ga
(ATEXandIECEx)‐40°C≤Ta≤+70°C
ATEXCertificateNo:ITS18ATEX203161X
IECExCertificateNo:IECExETL18.0007X
X–seeSpecialConditions(page6)
Included AccessoryItems
Carryingcase,ProbeCable(1’coiled‐6’extended),MeterProver,OperationManual,CorrosionData
ManagementSoftware,ProtectiveBoot.
OptionalAccessoryItems
CertifiedDataTransferUnit,CertifiedUSBBarrier

4
II.Start‐upandOperation
A.ReceivingtheMS4500E‐HC‐HCHighResolutionERDataLogger
ChecktheMS4500E‐HCHighResolutionDataLoggerforanyshippingdamagewhenitisfirstreceived.
Whentheunitisunpacked,verifythatthefollowingitemsareincluded:
CarryingCase
DataLogger
MeterProver
User’sManual
ProbeCable
Intheeventofshippingdamage,quantityshortage,ormissingitems,itisrecommendedthatthe
eventisdocumentedimmediatelyandthatdigitalphotographsaretaken.Anyshortagesormissing
itemsshouldbereportedtoMetalSamplesimmediately.Intheeventofshippingdamage,aclaim
shouldbeopenedwiththeresponsiblecarrier.
B.Start‐Up
CAUTION:
Usingthisproductinanywayotherthanthatspecifiedwithinthismanualmayimpairthe
intrinsicsafetyprotection.
ATTENTION:
l'utilisationdeceproduitdetouteautremanièrequecellespécifiéedansleprésentmanuel
peutaltérerlaprotectiondesécuritéintrinsèque.
Start‐upoftheMS4500E‐HCHighResolutionDataLoggerinvolvesthefollowingsteps:
1. PortDetailsandProbeConnection
2. Power‐Up
3. SetupandOperation
1.PortDetailsandProbeConnection
Attention: Utilisez Seulement
dans des endroits Sécurisés

5
TheMS4500E‐HCHighResolutionDataLoggerissuppliedwithaprobecableanditcanbe
connectedtotheprobewhosemetallossvalueneedstobemeasured,noadditionalhardwareis
necessary.Connecttheprobecable‘Lemo’connectorendtothedataloggerprobeconnector‘J9’
availableontopofthedatalogger.Connecttheotherendoftheprobecablei.e.probeconnector
(MSConnector)totheprobe.
2.Power‐Up
Theunitissuppliedwithtwo(2)Ctypebatteriesinstalled.Theinstrumentcanbepoweredupby
pressingthe‘ON’buttonontheKeypad.TheStart‐upscreenwiththeMetalSampleslogowillappearon
thescreen;thestart‐upscreenconfirmsthesoftwareversioncurrentlyrunningandtheSDcardstatus.
PowerOnButtonDisplayWindow
1. SELECT PROBE
2. MAKE MEASUREMENT
3. RECALL READINGS
4. DELETE READINGS
5. SET TIME & DATE
6. SET MEASUREMENT TIMER
7. COMMUNICATIONS
8. LANGUAGE / IDIOMA / LINGUAGEM
9. POWER OFF
10. BACKLIGHT ON/OFF
Version 1.7

6
a.SafeAreaUsage
TheMS4500E‐HCHighResolutionDataLoggerisapprovedforuseinhazardousareas,butcan
beusedinnon‐hazardousareasaswell.
ThedatatransferfromtheunittoPCshouldbedoneinasafearea.AcertifiedUSBisolator/
BarriershouldbeusedbetweentheunitandPCfordatatransfer.
b.HazardousAreaUsage
CAUTION:Thissectionprovidesgeneralguidelinesforhazardousareausage.However,
regardlessofanythingstatedhere,theMS4500E‐HCHighResolutionDataLoggermustbeused
infullcompliancewiththecontroldrawinglocatedinAppendixAandallofthelocalarea
requirements.Theentityparametersoftheunitaregivenbelow.
ATTENTION:Cettesectionfountdesdirectivesgénéralespourl’utilisationenzonedangereuse
Toutefois,etindépendammentdetoutedéclarationfaiteici,L’enregistreurdedonnéesde
hauterésolutionMS4500E‐HCdoitêtreutiliseenpleineconformitéavecleschémadecontrôle
donneenpage20etavectouteslesexigencesdulocalezonedangereuse.
J9-USB Port J8 – Mini USB Port Entity Parameters):
Uo= 4.94V Uo= 4.94V
Io= 0.332A Io= 0.215A
Po= 0.410 W Po= 0.322 W
Co= 1.0 uF
Lo= 100 uH
SPECIALCONDITIONS
ConnectionoftheequipmenttotheET1867USBbarrier(oranyothercertifiedbarrierusedwiththe
equipment)mayonlybemadewhilstboththebarrierandtheequipmentarelocatedinthenon‐
hazardousarea.
Externalnon‐metallicmaterialsutilizeaconductivecoatingtopreventtheriskofelectrostaticcharging.
Theequipmentshallberemovedfromserviceifdamagetothiscoatingisobserved.Refertothe
manufacturer’sinstructionmanualforfurtherinformationonthedurabilityandanychemicalvulnerability
ofthiscoating.
CONDITIONSPARTICULIERES
•Laconnexiondel'équipementàlabarrièreUSBET1867(outouteautrebarrièrecertifiée
utiliséeavecl'équipement)nedoitêtreeffectuéequelorsquelabarrièreetl'équipementsont
situésdansunezonenondangereuse.
•Lesmatériauxexternesnonmétalliquesutilisentunrevêtementconducteurpouréviterle
risquedechargeélectrostatique.L'équipementdoitêtreretiréduservicesidesdommagesàce
revêtementsontobservés.Reportez‐vousaumanueld'instructionsdufabricantpourplus
d'informationssurladurabilitéetlavulnérabilitéchimiquedecerevêtement.

7
CAUTION
1.MiniUSBporttotransferthedatafromtheDataLoggertoPCallowedonlyinsafearea
throughcertifiedUSBbarrier.
2.ConnectionoftheequipmenttotheMetalSamples#ET1867USBbarrier(oranyother
certifiedbarrierusedwiththeequipment)mayonlybemadewhileboththebarrierandthe
equipmentarelocatedinthenon‐hazardousarea.
3.OnlyCertifiedUSBflashdrivesareallowedtouseinHazardouslocations.
4.Conductivecoatingutilizedontheequipmentispermanentlystaticdissipativeandhumidity
independent,nofurthermaintenanceisnecessary.Theunitneedstobereturnedto
manufacturerifanydamages/scratchesnoticedontheoutersurfaceoftheequipment.
*DONOTUSECHEMICALSOLVENTSTOCLEANTHEOUTERSUFRACEOFTHEEQUIPMENTORBOOT.
ATTENTION:
1. MiniportUSBpourtransférerlesdonnéesdel'enregistreurdedonnéesauPCautorisé
uniquementdansunezonesûregrâceàunebarrièreUSBcertifiée.
2. Connexiondel'équipementauxéchantillonsmétalliques#ET1867LabarrièreUSB(outoute
autrebarrièrecertifiéeutiliséeavecl'équipement)nepeutêtrefabriquéequesilabarrière
etl'équipementsontsituésdansunezonenondangereuse.
3. SeulsleslecteursflashUSBcertifiéssontautorisésàutiliserdansdesendroitsdangereux.
4. Lerevêtementconducteurutilisésurl'équipementestenpermanenceantistatiqueet
indépendantdel'humidité,aucunentretiensupplémentairen'estnécessaire.L'unitédoit
êtreretournéeaufabricantsidesdommages/rayuresontétéconstatéssurlasurface
extérieuredel'équipement.
*N'UTILISEZPASDESOLVANTSCHIMIQUESPOURNETTOYERLESUFRACEEXTERNEDEL'ÉQUIPEMENT
OUDELABOTTE.

8
c.BatteryReplacement
Thebatterycompartmentislocatedonthebackoftheinstrument.Youmustfirstremoveittoaccess
thebatterycompartment.
Toreplacethebatteries,firstloosenandremovethecoverscrew,thenremovethebatterycoverto
exposethebatteries.Removetheoldbatteriesandreplacethemwithtwonew3.6V'C'sizebatteries
forMS4500E‐HC,beingsuretofollowthepolarity(+/‐)symbolsinthebatterycompartment.Replace
thebatterycoverandcoverscrew.
WARNING:
Donotmixoldandnewbatteries.
Donotmixbatteriesofdifferenttypes(suchasalkalineandlithium.)
MS4500E‐HC
XenoXL‐145F/TadiranTL4920batteryisrequiredforintrinsicsafety.
Thecoverscrewisusedtopreventunauthorizedtamperingofthebatteriesandisrequiredtoensure
intrinsicsafety.
Donotover‐tightenthecoverscrew,asthismaydamagetheinstrumentcase.
ATTENTION:
Avertissement ! Ne mélangez pas des piles neuves et usées.
Avertissement ! Ne mélangez pas des piles de types différents (telles que des piles alcalines etdes
piles lithium.)
Avertissement ! MS4500E-HC
La pile Xeno XL-145F / Tadiran TL4920 est requise pour la
sécurité intrinsèque
Avertissement ! La vis du couvercle est utilisé pour éviter toute manipulation non autorisée des
batteries, et est nécessaire pour assurer la sécurité intrinsèque.
Avertissement ! Ne pas trop serrer les vis du couvercle, pour ne pas endommager le boîtier de
l'instrument

9
3.SetupandOperation
MAINMENU
TheMS4500E‐HCallowsyoutomeasuremetallossandcorrosionrate.Reviewthissectionwhich
describesthefunctionsthatdisplayontheMainMenu:
SelectthefunctionsbyusingtheupordownarrowandpressingENTERordirectlyselectingthenumber
usingthekeypadandpressENTER.Pressing‘EXIT’willreturntheselectiontothefirstitemoftheMain
Menu.
a. SelectProbe
Thedisplaywillshowasbelowifyouselect‘SELECTPROBE’menufromthemainmenu.Youcanentera
newprobe,deleteaprobe,orselectanexistingprobefromthissection.
1. SELECT PROBE
2. MAKE MEASUREMENT
3. RECALL READINGS
4. DELETE READINGS
5. SET TIME & DATE
6. SET MEASUREMENT TIMER
7. COMMUNICATIONS
8. LANGUAGE / IDIOMA / LINGUAGEM
9. POWER OFF
10. BACKLIGHT ON/OFF
V
e
r
s
i
o
n 1.7
ENTER NEW PROBE
DELETE EXISTING PROBE
1000: WIRE LOOP PL:10.0
2500: CYLINDRICAL PL:10.0
2500: CYLINDRICAL PL:10.0
0434: TUBE LOOP PL:10.0
PRESS ENTER TO SELECT

10
ENTERNEWPROBE‐ToentertheIDofanewprobe,select‘ENTERNEWPROBE’;thenextscreen
allowsyoutoentertheIDnumberandthetype.
1.EntertheProbeIDusingthenumberkeys.(UsetheUp/Downarrowstoscrollbetween
alphanumericcharacters.UsetheRight/Leftarrowstoadvancetothenextcharacterorgobackto
thepreviouscharacter.)Whena4‐digitProbeIDisenteredthecursorautomaticallyadvancesto
theProbeTypefield.
2.UsetheUp/DownarrowstoselecttheProbeType.UsetheRightarrowtoadvancetotheProbe
Lifefield.
3.EntertheProbeLifeusingthenumberkeys.
4.PressENTER.
SELECTEXISTINGPROBE
ToassistyouwithselectingaprobeID,thescreendisplaysalistofallprobeIDscurrentlyinmemory.
Selectfromthelistofprobesinmemorytheoneforwhichyouwishtorunafunction(Make
Measurement,RecallReadingsandDeleteReadings).
oPressthearrowkeytoselectaprobeID.
oPressENTER.
ENTER NEW PROBE
DELETE EXISTING PROBE
1000: WIRE LOOP PL:10.0
2500: CYLINDRICAL PL:10.0
2500: CYLINDRICAL PL:10.0
0434: TUBE LOOP PL:10.0
PRESS ENTER TO SELECT
PROBE ID: 5385
PROBE TYPE: WIRE 40
PROBE LIFE: 10.0
PRESS UP / DOWN ARROW TO CHANGE
PRESS ENTER TO SET

11
DELETEEXISTINGPROBE
ItisoccasionallynecessarytoremovetheIDofaprobefrommemory,forexample:
oWhenthememorybankisfull
oWhenaprobeisnolongerused
TodeletetheIDofaprobeinmemory,selectDELETEEXISTINGPROBE.
1. EntertheprobeID‐usethenumberkeys(presstheDOWNarrowifyoumakeanerror).
2. PresstheUParrowwhenyou’veenteredthecorrectIDnumber.
TheinstrumentwillconfirmyourchoiceofprobeIDbeforedeletingitfrommemory.
DELETE PROBE
PROBE ID: 1000
USE UP / DOWN ARROW TO CHANGE
PRESS ENTER TO DELETE

12
b.MakeMeasurement
Selectaprobebeforeyourunthefunction
Inorderforyoutorun‘MAKEMEASUREMENT’‘RECALLREADINGS’&‘DELETEREADINGS’,
youmusthavealreadyselectedaprobefrommemory.See‘SELECTPROBE’fromtheMain
menus(page8).
Totakemeasurementfromprobesinmemory,select‘MAKE MEASUREMENT’menufrom
theMainMenu:
ThenextscreendisplaystheprobeIDthatyouselectedearlierfromtheMainMenuitem
‘SELECT PROBE’.
Press‘ENT’buttontostartmeasuringtheprobevalue.Duringmeasurement‘MEASURING..‘
messagedisplaysonthescreen.Oncompletionofmeasurementthedisplayshowsthe
probereading.
Press‘ENT’buttontosavethereading.Press‘Up’Arrowtore‐measuretheprobeagain.
1. SELECT PROBE
2. MAKE MEASUREMENT
3. RECALL READINGS
4. DELETE READINGS
5. SET TIME & DATE
6. SET MEASUREMENT TIMER
7. COMMUNICATIONS
8. LANGUAGE / IDIOMA / LINGUAGEM
9. POWER OFF
10. BACKLIGHT ON/OFF
PROBE ID:1234
TUBE
PRESS ENTER TO START
PROBE ID:1234
PREVIOUS READING : 860.61
PROBE CHECK : 808.69
READING: 860.60
MEASURING…..
PRESS UP ARROW TO REMEASURE
PRESS ENTER TO SAVE

13
c. RecallReadings
Select‘RECALL READINGS’tolookatdatafromdifferentdates.
Selecttheprobeifitisnotalreadyselectedtolookintothereadingsbyusingup/down
arrowsandpress‘ENT’button.Thereadingsoftheselectedprobewillbedisplayedinthe
nextscreen.
Press‘ENT’buttononthekeypadtoplotthegraphoftheselectedprobe.Thegraphwillbe
displayedonthenextscreenwithmetallossvalueandcorrosionrate.
ENTER NEW PROBE
DELETE EXISTING PROBE
1000: WIRE LOOP PL:10.0
2500: CYLINDRICAL PL:10.0
2500: CYLINDRICAL PL:10.0
0434: TUBE LOOP PL:10.0
PRESS ENTER TO SELECT
PROBE ID:1000
RDG# READING DATE(M/D/Y) TIME CHECK
8 859.60 04/24/15 18:45 809
7 759.59 04/24/15 18:45 704
6 841.60 04/24/15 18:45 705
5 459.00 04/24/15 18:45 809
4 118.00 04/24/15 18:45 704
3 434.00 04/24/15 18:45 705
2 118.00 04/24/15 18:45 809
1 456.54 04/24/15 18:45 809
PRESS ENTER TO GRAPH
PROBE ID:1000
P 1000 L
R 04/01/15
O RDG : 860.04
B ML : 8.60
E R
04/27/15
RDG : 860.32
500 ML : 8.60
COR RATE:
U 0.05 MPY
N
I
T
S 0 124 248 372 496
<ENT = UPDATE> # OF READINGS

14
d. DeleteReadings
Select‘DELETE READINGS’fromMainMenutoeraseareadingfrommemory.
Periodicallydeletereadingstocreatemorefreestorageareainmemory.
Selecttheprobefromwhichthereadingsneedtobedeletedbyusingup/downarrowsand
press‘ENT’button.Thereadingsoftheselectedprobewillbedisplayedinthenextscreen.
Selecttheprobereading(RDG#)numberwhichneedstobedeletedbyusingup/downarrows
andpress‘ENT’button.Theselectedprobereadingwillbedeletedpermanentlyfromthe
instrumentmemory.
ENTER NEW PROBE
DELETE EXISTING PROBE
1000: WIRE LOOP PL:10.0
2500: CYLINDRICAL PL:10.0
2500: CYLINDRICAL PL:10.0
0434: TUBE LOOP PL:10.0
PRESS ENTER TO SELECT
PROBE ID:1000
RDG# READING DATE(M/D/Y) TIME CHECK
8 859.60 04/24/15 18:45 809
7 759.59 04/24/15 18:45 704
6 841.60 04/24/15 18:45 705
5 459.00 04/24/15 18:45 809
4 118.00 04/24/15 18:45 704
3 434.00 04/24/15 18:45 705
2 118.00 04/24/15 18:45 809
1 456.54 04/24/15 18:45 809
PRESS ENTER TO DELETE

15
e. SetTime&Date
Select‘SET TIME & DATE’fromMainMenutosettheinternalclockoftheMS4500E‐
HCDataLogger.Inorderforinformationinmemorytobeusefulandaccurate,besure
thetimeanddateareaccurate.
CAUTION:TheTimeandDatearesetatthefactory,butmayneedtobeadjustedtoyour
timezone.Also,ifyoutrytochangetheDateorTimeanditconflictswithstored
information,theMS4500E‐HCwilldisplayN/A.
ATTENTION:l'heureetladatesontdéfiniesàl'usine,maisilspeuventêtreajustéesàvotre
fuseauhoraire.Aussi,sivousessayezdechangerladateoul'heureetquececiconflitavec
l'informationmémorisée,leMS4500E‐HC‐HCafficheraN/A.
Usethenumbersonthekeypadandthearrowkeystomakechangestothetimeanddate.
Press‘ENT’buttononthekeypadtoconfirmyourchangesandreturntotheMainMenu,or
Press‘EXIT’buttontodiscardthechanges.
DATE (MM / DD / YY):
04 / 27 / 15
TIME (HH:MM):
03: 42 PM
USE LEFT / RIGHT ARROW TO SELECT
USE UP / DOWN ARROW TO CHANGE
PRESS ENTER TO SAVE

16
f. SetMeasurementTimer
Select‘SET MEASUREMENT TIMER’fromMainMenutosetautomaticprobe
readingatsetintervals.TheMS4500E‐HCDataLoggerperiodicallywakesupandtakes
themeasurementandstoresthereading.
Selectaprobebeforeyourunthefunction
InorderforyoutorunMAKEMEASUREMENT,youmusthavealreadyselectedaprobe
frommemory.SeeSelectProbeIDfromtheMainmenus(seepage8).
Ifprobeisnotselectedearlier,selecttheprobefromthedisplay.UseUp/Downarrow
keystoselecttheprobeandpress‘ENT’button.
UseRight/Leftarrowkeystoturnon/offthedatalogging.UseUp/Downarrowto
changedatalogginginterval.Minimumcanbesetas1Hoursandmaximumcanbeset
as99Days.Theunittakesthereadingonceperthesetinterval.
ENTER NEW PROBE
DELETE EXISTING PROBE
1000: WIRE LOOP PL:10.0
2500: CYLINDRICAL PL:10.0
2500: CYLINDRICAL PL:10.0
0434: TUBE LOOP PL:10.0
PRESS ENTER TO SELECT
DATA LOGGING INTERVAL SETUP
PROBE ID : 2500
SELECT DATA LOGGING INTERVAL:
ONCE EVERY 1 HOURS
LOGGING OFF
RIGHT / LEFT ARROW TO TURN ON/OFF
UP/DOWN ARROW TO CHANGE INTERVAL
PRESS ENTER TO SAVE

17
g. Communications
TheMS4500E‐HCDataLoggerhastheabilitytostorereadingsastheyaretaken.These
readingscanlaterbetransferredtoyourPCviaUSBport.TotransferdatatoaPCitis
necessarytoinstalltheCorrosionDataManagementSoftware(included).
Todownloaddatafromaninstrument:
•
ConnecttheinstrumenttoanavailablePCUSBport.
•
TurntheMS4500E‐HCon.
•
RuntheCorrosionDataManagementSoftware
•
OpentheInstrumentDownloadCenter
•
Selecttheserialportandinstrument
•
ClicktheDownloadbutton(ensurethatPartStatustogglesto“On”).
•
Select‘COMMUNICATIONS’menufromtheMainMenuoftheinstrument.
•
Select’TRANSFER ALL’orSelecttheProbedatayouwishtodownload.
DatawillbegintotransferfromtheinstrumenttothePC,andwillappearinthedatawindowofthe
InstrumentDownloadCenterasshownbelow.Ifthedataappearsgarbled,thewronginstrument
typehasbeenselected.Selectanotherinstrumenttypeandtryagain.
TRANSFER ALL
1001: WIRE LOOP
PRESS ENTER TO TRANSFER

18
Ifdatadoesnotappearinthedatawindowatall,verifythat:
TheinstrumentisconnectedtoavalidserialorUSBport
Allcablesaresecurelyconnected
Themessage“PortStatus:On”appearsinthestatusbar
Therearenoerrorsontheinstrument
IfusingaUSBadapter,ensurethatthedevicedrivershavebeeninstalled.
SelectingaSerialPort
WhentheMS4500E‐HCisconnectedtotheUSBportitwillappearasavirtualCOMport.Use
theSerialPortselectionboxtoselectthevirtualCOMportnumber.IftheCOMportnumberis
unknown,itcanbefoundinWindowsDeviceManager
under“Ports(COM&LPT)”.
SelectinganInstrument
UsetheInstrumentoptionboxtoselectthemodelofMetalSamplesinstrumentbeingdownloaded.
Thissetstheappropriatecommunicationparameters,whichwillbedisplayedinthestatusbaratthe
bottomofthewindow(“9600,N,8,1”foranMS1500,“2400,N,8,1”foranMS3500).
DownloadingData
Totoggletheselectedportonandoff,clicktheDownloadbutton.Togglingtheportonandoff
willalsoclearthedatawindow.
Onceavalidserialportandinstrumenthavebeenselected,clicktheDownloadbuttontoturnthe
portonandenablethecomputertoreceivedata.Ifavalidserialporthasbeenselected,the
statusbaratthebottomofthewindowwilldisplaythemessage“PortStatus:On”.
Ifaninvalidserialporthasbeenselected,anerrormessagewillappear,andthestatusbarwilldisplay
themessage“PortStatus:Off”.Ifthishappens,anotherserialportshouldbeselected.
SavingData
Tosavethedatainthedatawindow,clicktheSavebutton.Thedataiscomma‐delimitedASCIItext.
Itcanbesavedtoastandardtext(.txt)file,oritcanbesavedtoacomma‐separatedvalues(.csv)file,
whichgreatlysimplifiestheprocessofimportingthedataintosomespread‐sheetprogramssuchas
Excel®orQuattroPro®.
PrintingData
Toprintthecontentsofthedatawindow,clickthePrintbutton.Aprintdialogboxwillbedisplayedto
allowprinterselectionandsetup.
ChartingData
Tochartthecontentsofthedatawindow,clicktheChartDatabutton.Ifthedatahasnotyetbeen
saved,theuserwillbepromptedtodosobeforethechartingprocessbegins.Formoreinformation
onchartingdata,seetheDataAnalysissection.
Table of contents
Other Metal Samples Data Logger manuals
Popular Data Logger manuals by other brands

Elitech
Elitech LogET 6 user manual

cloudworks
cloudworks CDS552 user manual

MadgeTech
MadgeTech HiTemp150 Product Information Card

Huace Navigation Technology
Huace Navigation Technology LT500XYZ Operational description

National Instruments
National Instruments Isolated Analog Input Device NI PXI-4224 user manual

Ebro
Ebro EBI 20 Series manual