manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Omega
  6. •
  7. Measuring Instrument
  8. •
  9. Omega DP25B-E User manual

Omega DP25B-E User manual

5 6
Scaling Without Known Loads Continued)
3. Press /TARE again. The unit displays the
last setting f r IN!1 .
(The first digit flashes.)
4. Change IN!1 as necessary:
•
Press

/MAX t set r change the digit's current value. C ntinue t
press

/MAX until the meter displays the
desired value f r the
flashing digit. Values can range fr m 0 t 9.
F r the first digit, y u can
als enter a minus sign (–) r –1.
• Press

/TARE t scr ll t the digit(s) y u want t change.
5. Press MENU t st re IN!1 . The unit displays:
6. Press /TARE. The unit displays the last
setting f r RD!1 . (The first digit flashes.)
7. Change RD!1 as necessary:
• Press

/MAX
until the meter displays the desired value f r the
flashing digit. Values can range fr m 0 t 9. F r the first digit, y u
can als enter a minus sign (–) r –1.
• Press

/TARE t scr ll t the digit(s) y u want t change.
8. Press MENU t st re the value sh wn f r RD!1 .
The unit displays:
T define the maximum l ad (IN!2 and RD!2 ):
1. Repeat steps 3–8 above, entering the values or IN!2
and RD!2 .
2. Once y u've c mpleted all steps, the unit displays:
T begin perati n:
Reinitialize the unit (press RESET twice r
press MENU until RST flashes n the
display). When a numeric reading appears,
the unit is perati nal.
Determining Reading Offset
The run m de reading f r meters scaled with ut kn wn l ads may
reflect an ffset. F r example, say y u set
RD!1
t 0 and
RD!2
t
100, but when the minimum l ad is applied, a negative value f –1.5
displays n the fr nt panel.
T c rrect the reading ffset:
1. With zer l ad applied, n te the reading n the display.
2. Subtract that am unt fr m the RD!1 and RD!2 values y u
riginally entered.
In the example, the ffset w uld be –1.5. If RD!1 is t read 0 in
Run M de, it must be reentered as 1.5. RD!2 must likewise be
reentered as 101.5 if the meter is t read 100 when the
maximum l ad is applied.
3.
Repeat the steps f r "Scaling With ut Kn wn L ads," but when
the values f r IN!1 and IN!2 display, d n t change them.
Instead, press MENU t m ve t the pr mpts f r RD!1 and RD!2
and make the necessary changes.
4. Reinitialize the unit and resume perati n.
Scaling With Known Loads Continued)
6. Press MENU t st re IN!1 . The unit displays:
7. Press /TARE. The unit displays the last setting f r RD!1 .
8. Change RD!1 as necessary:
• Press

/TARE t scr ll t the digit(s) y u want t change
(it flashes n the display).
• Press

/MAX t change the value f the flashing digit. Values
can range fr m 0 t 9. F r the first digit, y u can als enter a
minus sign (–) r –1.
9. Press MENU t st re the value sh wn f r RD!1 .
The unit displays:
T identify the maximum kn wn l ad (IN!2 and RD!2 ):
1. Apply the maximum kn wn l ad (100%).
2. Repeat steps 4–9 above, or IN!2 and RD!2 .
Once y u've c mpleted all steps, the unit displays:
T begin perati n:
Reinitialize the unit (press RESET twice r
press MENU until RST flashes n the display).
When a numeric reading appears, the unit is
perati nal.
Scaling Without Known Loads
F r 4-20 mA sens rs, the values f r the minimum and maximum
input l ads are always as f ll ws:
• Minimum l ad (IN!1 )— 2000
• Maximum l ad (IN!2 )— 9999.
If y ur installati n uses a different sens r type, y u must calculate
the values f r IN!1 and IN!2 bef re pr ceeding with the steps
bel w. Use the f rmula pr vided in the Operat r's Manual.
T define the minimum l ad
(IN!1 and RD!1 ):
1. If it's n t already sh wn, press MENU until
the unit displays:
2. Press /TARE. The unit displays:
This device is marked with the internati nal cauti n symb l. It is imp rtant t read the
Setup Guide bef re installing r c mmissi ning this device, as the guide c ntains
imp rtant inf rmati n relating t safety and EMC.
WARNING: These pr ducts are n t designed f r use in, and sh uld n t be used f r, patient-
c nnected applicati ns.
It is the p licy f OMEGA t c mply with all w rldwide safety and EMC/EMI regulati ns that apply.
OMEGA is c nstantly pursuing certificati n f its pr ducts t the Eur pean New Appr ach Directives.
OMEGA will add the CE mark t every appr priate device up n certificati n.
The inf rmati n c ntained in this d cument is believed t be c rrect, but OMEGA Engineering, Inc.
accepts n liability f r any err rs it c ntains, and reserves the right t alter specificati ns with ut
n tice.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects
in materials and workmanship for a period of 61 months from
date of purchase. OMEGA’s WARRANTY adds an additional one (1)
month grace period to the normal five (5) year product warranty
to cover handling and shipping time. This ensures that OMEGA’s
customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evalua-
tion. OMEGA’s Customer Service Department will issue an Authorized
Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it
will be repaired or replaced at no charge. OMEGA’s WARRANTY does
not apply to defects resulting from any action of the purchaser, includ-
ing but not limited to mishandling, improper interfacing, operation
outside of design limits, improper repair, or unauthorized modifica-
tion. This WARRANTY is VOID if the unit shows evidence of having
been tampered with or shows evidence of having been damaged as a
result of excessive corrosion; or current, heat, moisture or vibration;
improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear
is not warranted, include but are not limited to contact points, fuses,
and triacs.
OMEGA is pleased to offer suggestions on the use of its vari-
ous products. However, OMEGA neither assumes responsibil-
ity for any omissions or errors nor assumes liability for any
damages that result from the use if its products in accordance
with information provided by OMEGA, either verbal or writ-
ten. OMEGA warrants only that the parts manufactured by
the company will be as specified and free of defects. OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT
THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING
ANY WARRANTY OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set
forth herein are exclusive, and the total liability of OMEGA
with respect to this order, whether based on contract, warran-
ty, negligence, indemnification, strict liability or otherwise,
shall not exceed the purchase price of the component upon
which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used,
nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC),
used in or with any nuclear installation or activity; or (2) in medical appli-
cations or used on humans. Should any Product(s) be used in or with
any nuclear installation or activity, medical application, used on humans,
or misused in any way, OMEGA assumes no responsibility as set forth
in our basic WARRANTY/DISCLAIMER language, and, additionally,
purchaser will indemnify OMEGA and hold OMEGA harmless from any
liability or damage whatsoever arising out of the use of the Product(s)
in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA
Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER
SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be marked on the
outside of the return package and on any correspondence.
FOR WARRANTY RETURNS,
please have the following
information available BEFORE
contacting OMEGA:
1. Purchase Order number
under which the product
was PURCHASED,
2.
3.
Model and serial number of the
product under warranty, and
Repair instructions and/or
specific problems relative
to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair
charges. Have the following
information available BEFORE
contacting OMEGA:
1. Purchase Order number to cover
the COST of the repair or
calibration,
2.
3.
Model and serial number of the
product, and
Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes,
whenever an improvement is possible. This affords our customers
the latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC.
© Copyright 2018 OMEGA ENGINEERING, INC. All rights reserved.
This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable
form, in whole or in part, without the prior written consent of OMEGA
ENGINEERING, INC.
MQS3597/0918
DP25B-E
Process Panel Meter
omega.com [email protected]
The information contained in this document is believed to be correct, but
OMEGA accepts no liability for any errors it contains, and reserves the right to
alter specifications without notice.
Servicing North America:
U.S.A. Omega Engineering, Inc.
Headquarters: Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only)
Engineering Service: 1-800-872-9436 (USA & Canada only)
Tel: (203) 359-1660 Fax: (203) 359-7700
e-mail: [email protected]
For Other Locations Visit omega.com/worldwide
For complete product manual:
www.omega.com/manuals/manualpdf/M3597.pdf
TM
Using This Quick Start Manual
Use this Quick Start Manual t set up y ur Pr cess Meter and begin
perati n. Inf rmati n is pr vided n h w t :
• C nnect ac p wer • Set basic pti ns f r perati n
C nnect the sens r • Scale the meter.
Features with are f r the “B” versi n which has three-c l r
pr grammable “Big” LED display - All segment characters sh wn are f r
the “B” versi n.
IMPORTANT: F r c mplete inf rmati n n all setup pti ns, please
refer t the Operat r’s Manual.
This Quick Start Manual includes speci ic con iguration
parameters or transducers with an output range o 4–20 mA
and 24 V excitation. Other sensor types may require di erent
parameters or additional ones. When this is the case, we re er
you to the Operator's Manual or detailed instructions.
Safety Consideration
This device is marked with the international Caution symbol.
The instrument is a panel m unt device pr tected in acc rdance with
2014/35/EU, electrical safety requirements f r electrical equipment f r
measurement, c ntr l and lab rat ry. Remember that the unit has n
p wer- n switch. Building installati n sh uld include a switch r circuit-
breaker that must be c mpliant t IEC 947-1 and 947-3.
SAFETY:
• D n t exceed v ltage rating n the label l cated n the t p f the instrument
h using.
• Always disc nnect p wer bef re changing signal and p wer c nnecti ns.
• D n t use this instrument n a w rk bench with ut its case f r safety
reas ns.
• D n t perate this instrument in flammable r expl sive atm spheres.
• D n t exp se this instrument t rain r m isture.
EMC:
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and p wer wires in the same c nduit.
• Use signal wire c nnecti ns with twisted-pair cables.
• Install Ferrite Bead(s) n signal wire cl se t the instrument if EMC pr blems
persist.
Mount the Unit
1. Cut a panel pening using the dimen-
si ns sh wn t the right.
2. P siti n the unit in the pening, making
sure the fr nt bezel is flush with the
panel.
3. Install retaining clips n b th sides f
the meter and tighten against the panel.
Wiring
Warning: Do not connect AC power to your device until you
have completed all input and output connections. This device
must only be installed by a specially trained electrician with
corresponding quali ications. Failure to ollow all instructions
and warnings may result in injury!
1. Rem ve the panel at the back f the unit.
2. L cate the TB1 c nnect r.
3. Insert the c rrect wire in each terminal as sh wn in the f ll wing figure
and tighten the l ckd wn screws.
4. Tug gently n the wires t verify the c nnecti ns.
External Fuse Required:
Time-delay, UL 248-14 listed Time-lag, IEC 127-3 rec gnized
175 mA (115 Vac line) 125 mA (115 Vac line)
80 mA (230 Vac line) 63 mA (230 Vac line)
AC Powered Unit Connections
START HERE
Wiring continued)
When using DC p wer, d n t use internal excitati n r Is lated
Anal g Output f r high c l r brightness. F r l w r medium
brightness, internal excitati n is limited t 24 V @ 25 mA; 5 V, 10
V, 12 V @ 35 mA.
In rder t maintain the same degree f pr tecti n as the AC units,
always use a Safety Agency Appr val DC s urce with the same
Overv ltage Categ ry and P lluti n Degree.
DC Powered Unit Connections
Connect the Sensor
1. L cate the TB2 c nnect r n the rear f the unit.
2. Attach the sens r wires and tighten the l ckd wn screws.
The diagram bel w sh ws the wiring f r 4–20 mA sens rs with internal
excitati n.
Refer t the Operat r’s Manual f r setup requirements f r ther
sens r types.
Current Input Connections (4–20 mA)
with Internal Excitation
3. Tug gently n the wires t verify the c nnecti ns.
4. Replace the panel at the back f the unit.
Using the Configuration Menu
T c nfigure the meter, y u use the butt ns n the fr nt panel.
To: Take This Action:
Display the Press the MENU butt n. The first functi n
C nfigurati n Menu n the menu, INPT, displays.
Select a submenu 1. Press MENU until the functi n y u
functi n want is sh wn.
2. Press /TARE.
The inf rmati n y u can change flashes.
Select a value 1. Press /MAX t display the pti n
f r that submenu y u want.
functi n 2. Press MENU t st re it.
STRD quickly flashes, indicating that
the selecti n has been st red in mem ry.
Then the next menu functi n displays.
G back t previ us Press RESET nce.
menu functi n
Exit the Press RESET twice. The unit displays
C nfigurati n RST as it reinitializes. When a numeric
Menu value displays, the unit is in run m de.
(Opti nally, y u can press MENU t
m ve thr ugh all the menu functi ns
until the unit reinitializes.)
2 3 4
To Set the Decimal Point continued)
3. Press /MAX t m ve the decimal p int t the
desired l cati n. The ch ices are FFFF,
F.FFF, FF.FF, and FFF.F.
4. Press MENU t select the decimal p int p siti n sh wn. The
unit displays:
To Scale the Meter
Y u can scale the meter in ne f tw ways:
1. With a kn wn l ad — This meth d uses input (l ad) inf rmati n
sent fr m an ther device such as a scale r a simulat r f r
v ltage r current.
2. With ut a kn wn l ad — This inv lves calculating the l ad
based n transducer specificati ns and manually entering it t
the meter.
F r b th meth ds, y u must first identify the minimum input l ad
IN!1 and the c rresp nding display reading y u want RD!1 . Then
y u identify the maximum input l ad IN!2 and its c rresp nding
display reading RD!2 .
The decimal p int is f r display purp ses nly — y u set it
where y u want it t display f r y ur applicati n.
When entering IN!1 and IN!2 values, ign re any decimal
p int n the display. H wever, y u must enter RD!1 and
RD!2 values with the decimal p int in the desired p siti n.
Scaling With Known Loads
T identify the minimum kn wn l ad (IN!1 and RD!1 ):
1. If it's n t already sh wn, press MENU
until the unit displays:
2. Apply the minimum kn wn l ad (0%).
3. Press /TARE. The unit displays:
4. Press /TARE again. The unit displays the
last setting f r
IN!1
.
5. Press /TARE again. The unit displays the
actual reading being received fr m the
sending device.
MENU SUBMENU /TARE DESCRIPTION
INPT 100M, ±50M, !10V , !± 5 V , 0-20 *Input
DEC.P FFFF*, F.FFF, FF.FF, FFF.F Decimal P int
RD.S.O IN!1, RD!1 , IN!2 , RD!2 Scale and Offset
RD.CF R.1=T*, R.1=N
Reading C nfigurati n
R.2=0, R.2=1, R.2=2, R.2=3, R.2=4
*
R.3=F*, R.3=U
COLR GRN, RED, AMBR Display C l r
S1.CF S.1=A*, S.1=B
Setp int 1 C nfigurati n
S.2=U*, S.2=L
S2.CF S.1=A*, S.1=B
Setp int 2 C nfigurati n
S.2=U*, S.2=L
S1.DB 0003*
Setp int 1, Deadband
S2.DB 0003*
Setp int 2, Deadband
OT.CF O.1=E *, 0.1=D Anal g Output
O.2=C *, 0.2=V
C nfigurati n
O.3=A *, 0.3=P
O.4=D , 0.4=R
O.5=F , 0.5=H
P.BND 0000 sh wn if 0.3 = P Pr p rti nal Band
M.RST 0000 sh wn if 0.3 = P Manual Reset
OT.S.O RD!1, OUT1, RD!2 , OUT2
Output Scale & Offset
L .CF RS=E *, RS=D
L ck ut C nfigurati n
SP=E *, SP=D
L3=0*, L3=1
BRIT M.BrT, .BrT, H.BrT
Display Brightness
* Fact ry Default Settings
To Set the Input Type
1. Press MENU until the unit displays:
2. Press /TARE. The unit displays:
3. F r this applicati n y u want 0-20. If 0-20 is n t displayed,
press /MAX until it appears. Other ch ices are 100M, 50M,
10V and 5V.
Re er to the Operator's Manual or more in ormation on
changing ranges.
4. Press MENU t select the sens r sh wn. The meter displays the
next menu item. If y u changed input type, the meter displays:
To Set the Decimal Point
1. If it's n t already sh wn, press MENU until the
unit displays:
2. Press /TARE. The unit displays:
Using the Configuration Menu continued)
DEVICE

Other manuals for DP25B-E

1

Other Omega Measuring Instrument manuals

Omega FMG470 Series User manual

Omega

Omega FMG470 Series User manual

Omega DPG2001B User manual

Omega

Omega DPG2001B User manual

Omega DP18-P7 User manual

Omega

Omega DP18-P7 User manual

Omega Vortex FV100 Series User manual

Omega

Omega Vortex FV100 Series User manual

Omega DPF20 Series User manual

Omega

Omega DPF20 Series User manual

Omega HFH81 User manual

Omega

Omega HFH81 User manual

Omega PRESSURE GAUGES DPG5000 User manual

Omega

Omega PRESSURE GAUGES DPG5000 User manual

Omega AOK 20SS Ex User manual

Omega

Omega AOK 20SS Ex User manual

Omega DP18-SC User manual

Omega

Omega DP18-SC User manual

Omega DP18-P8 User manual

Omega

Omega DP18-P8 User manual

Omega TCL-3M165E2 Series User manual

Omega

Omega TCL-3M165E2 Series User manual

Omega HHLM112SD User manual

Omega

Omega HHLM112SD User manual

Omega HHG1394 User manual

Omega

Omega HHG1394 User manual

Omega FPDM1000 User manual

Omega

Omega FPDM1000 User manual

Omega DP6060 User manual

Omega

Omega DP6060 User manual

Omega HX71 Series User manual

Omega

Omega HX71 Series User manual

Omega ACC320 User manual

Omega

Omega ACC320 User manual

Omega SP-016 User manual

Omega

Omega SP-016 User manual

Omega FTB-1000 Series User manual

Omega

Omega FTB-1000 Series User manual

Omega DP41-B Series User manual

Omega

Omega DP41-B Series User manual

Omega Volume VII-System Description User manual

Omega

Omega Volume VII-System Description User manual

Omega PHH444 User manual

Omega

Omega PHH444 User manual

Omega FMG3000 SERIES User manual

Omega

Omega FMG3000 SERIES User manual

Omega FPD1000-HP Series User manual

Omega

Omega FPD1000-HP Series User manual

Popular Measuring Instrument manuals by other brands

Rainforest Automation EMU-2 quick start guide

Rainforest Automation

Rainforest Automation EMU-2 quick start guide

HART Proline Promag 55 manual

HART

HART Proline Promag 55 manual

Amprobe THWD-3 user manual

Amprobe

Amprobe THWD-3 user manual

VOLTCRAFT NTU-600 operating instructions

VOLTCRAFT

VOLTCRAFT NTU-600 operating instructions

Endress+Hauser PROline promag 53 Documentation

Endress+Hauser

Endress+Hauser PROline promag 53 Documentation

ATAGO PAL-36S instruction manual

ATAGO

ATAGO PAL-36S instruction manual

ERMA SSI 9005 instruction manual

ERMA

ERMA SSI 9005 instruction manual

Onick LSP320 user manual

Onick

Onick LSP320 user manual

Chroma 66203 user manual

Chroma

Chroma 66203 user manual

Geco G-913-P01 Service & operating manual

Geco

Geco G-913-P01 Service & operating manual

Johnson LDM150 quick start guide

Johnson

Johnson LDM150 quick start guide

Bender ISOMETER iso685W-S-B manual

Bender

Bender ISOMETER iso685W-S-B manual

HP 8560A Service manual

HP

HP 8560A Service manual

CST/BERGER Lasermark LD500 quick guide

CST/BERGER

CST/BERGER Lasermark LD500 quick guide

ROOTECH Accura 3550S Quick setup guide

ROOTECH

ROOTECH Accura 3550S Quick setup guide

TELEVAC MX2A instruction manual

TELEVAC

TELEVAC MX2A instruction manual

AdValue Photonics AP-ML1 user guide

AdValue Photonics

AdValue Photonics AP-ML1 user guide

Dinel CLM-36 instruction manual

Dinel

Dinel CLM-36 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.