manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Omron
  6. •
  7. Blood Pressure Monitor
  8. •
  9. Omron HEM-9200T User manual

Omron HEM-9200T User manual

Introduction
Intended Use
This device is a digital monitor intended for use in measuring blood pressure
and pulse rate in adult patient population. The device detects the
appearance of irregular heartbeats during measurement and gives a
warning signal with the measurement result.
Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your
own blood pressure, CONSULT YOUR PHYSICIAN.
Important Safety Information
Warning: Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
General Usage
DO NOT adjust medication based on measurement results from this
blood pressure monitor. Take medication as prescribed by your physician.
Only a physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
Consult your physician before using the device for any of the following
conditions: common arrhythmias such as atrial or ventricular premature
beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age,
pregnancy, pre-eclampsia, renal diseases.
Note that PATIENT motion, trembling, shivering may affect the
measurement reading.
Do not use the device on the injured arm or the arm under medical
treatment.
Do not apply the arm cuff on the arm while being on an intravenous drip
or blood transfusion.
Consult your physician before using the device on the arm with an
arterio-venous (A-V) shunt.
Do not use the device with other medical electrical (ME) equipment
simultaneously.
Do not use the device in the area the HF surgical equipment, MRI, or CT
scanner exists, or in the oxygen rich environment.
The air tube or the AC adapter cable may cause accidental strangulation
in infants.
Contains small parts that may cause a choking hazard if swallowed by
infants.
Keep the battery out of reach of children.
Data Transmission
Do not use this product on aircraft or in hospitals. Please remove the
battery from the unit. This product emits radio frequencies (RF) in the
2.4 GHz band, use of this product in locations where RF is restricted is
not recommended.
The use of RF in this product is licensed for use by the FCC, for further
information on any potential restrictions refer to documentation on
Bluetooth®usage by the FCC.
AC Adapter (optional accessory)
Do not use the AC adapter if the device or the power cord is damaged.
Turn off the power and unplug the power cord immediately.
Plug the AC adapter into the appropriate voltage outlet. Do not use in a
multi-outlet plug.
Never plug in or unplug the power cord from the electric outlet with wet
hands.
Caution: Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury to the user
or patient or damage to the equipment or other property.
General Usage
Always consult your physician. Self-diagnosis of measurement results
and self-treatment are dangerous.
Consult your physician before using the device for any of the following
conditions:
•If you have had a mastectomy.
•People with severe blood flow problems or blood disorders as cuff
inflation can cause bruising.
Do not take measurements more than necessary. It may cause bruising
due to blood flow interference.
Remove the arm cuff if it does not start deflating during the measurement.
Do not use this device on infants or persons who cannot express their
intentions.
Do not use the device for any purpose other than measuring blood
pressure.
Use only the approved arm cuff for this device. Use of other arm cuffs
may result in incorrect measurement results.
Do not use a mobile phone or other devices that emit electromagnetic
fields, near the device. This may result in incorrect operation of the
device.
Do not disassemble the monitor or arm cuff. This may cause an
inaccurate reading.
Do not use in a location with moisture, or a location where water may
splash on the device. This may damage the device.
Do not use the device in a moving vehicle (car, airplane).
Read “What to do if your systolic pressure is more than 210 mmHg” of this
instruction manual, if your systolic pressure is known to be more than 210
mmHg. Inflating to a higher pressure than necessary may result in
bruising where the arm cuff is applied.
AC Adapter (optional accessory)
Fully insert the power plug into the outlet.
When disconnecting the power plug from the outlet, do not pull the power
cord. Be sure to pull from the power plug safely.
When handling the power cord, take care not to do the following:
Wipe the dust off from the power plug.
Unplug monitor when not in use.
Disconnect the power plug before cleaning.
Use only an OMRON AC adapter designed for this device. Use of
unsupported adapters may damage and/or may be hazardous to the
device.
Battery Usage
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 “AA” alkaline or manganese batteries with this device. Do not
use other types of batteries. Do not use new and used batteries together.
Remove the batteries if the device will not be used for three months or
more.
General Precautions
• Do not forcibly crease the arm cuff or the air tube excessively.
• Do not press the air tube while taking a measurement.
• To unplug the air plug, pull on the air plug at the connection with the
monitor, not the tube itself.
• Do not drop the monitor or subject device to strong shocks or vibrations.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not use the device outside the specified environment. It may cause an
inaccurate reading.
• Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
1. Know Your Device
Contents:
Monitor, arm cuff, 4 “AA” batteries, instruction manual, quick start
guide
Monitor:
Arm cuff: Arm circumference 9" - 17" (22 - 42 cm)
Display:
1.1 Display symbols
Irregular Heartbeat Symbol ( )
When the monitor detects an
irregular rhythm two or more times
during the measurement, the
irregular heartbeat symbol will
appear on the display with the
measurement values.
An irregular heartbeat rhythm is
defined as a rhythm that is 25%
less or 25% more than the average
rhythm detected while the monitor
is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol displays with your measurement
results, we recommend you consult your physician. Follow the
directions of your physician.
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during
the measurement. Please remove the arm cuff, and wait 2 - 3 minutes.
Take another measurement, remain still during measurement.
SYNC Symbol ( )
The SYNC symbol ( ) is displayed if device is not connected to
Telehealth service receiver or when data is not transmitted
successfully. Please see “Connecting failure. Data is not being
transmitted.” in section 4.2.
1.2 Wireless Function
Before use, thoroughly read the instruction manual included with the
Telehealth service receiver (in some cases, receiver maybe your smart
phone) being used with this blood pressure monitor for instruction
about your monitor to Telehealth service receiver, and receiver
transmission range. It will be necessary for the blood pressure monitor
to be within the receiver’s transmission range to successfully transfer
data. This Omron blood pressure monitor is designed to connect to
specific Bluetooth®Smart receivers, and is not guaranteed to connect
to all Bluetooth®Smart compatible devices and Telehealth service
receiver. “Omron Healthcare, Inc. shall not be liable for loss of use
or any other special, incidental, consequential or indirect costs,
expenses or damages due to improper usage of this monitor.”
1.3 Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and
eating for 30 minutes before taking a measurement. Rest for at
least 5 minutes before taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during
stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your arm.
2. Preparation
Battery Installation
Below procedures are ONLY done at inserting batteries.
1. Remove the battery cover.
2. Insert 4 “AA” batteries as
indicated in the battery
compartment.
3. Put back the battery cover.
Notes:
• When the depleted battery symbol ( ) appears on the display, turn
the monitor off, then replace all batteries at the same time. Long life
alkaline batteries are recommended.
• The measurement values continue to be stored in memory even after
the batteries are replaced.
• The supplied batteries may have a shorter life.
• Dispose of the device, components and optional accessories
according to applicable local regulations. Unlawful disposal may
cause environmental pollution.
4. Connect the device to the Telehealth service receiver.
As soon as insert batteries, it will automatically start the connection
process, as below.
When the display shown above is not appeared, see “Connecting
failure.” in section 4.2 first.
To retry connecting the Telehealth service receiver, remove
batteries and press START/STOP button for 2-3 times. Then start
with step 2 again.
Note: If your Telehealth service receiver asks for PIN code, enter 6
digits of PIN code on the rating label at the bottom of the device
to complete.
5. Confirm if the device is successfully connected.
If the device is connected successfully to the Telehealth service
receiver, OK symbol will appear on the display, as below.
When “Err” appears, see “Connecting failure.” in section 4.2 for
more detail.
Notes:
• We recommend to keep batteries all the time, even if you chose to
use AC adapter.
• If only AC adapter is used without keeping batteries, above device
connection process (step 4 and 5) is necessary each time you unplug
and plug back the AC adapter.
3. Using the Device
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your left
upper arm.
Don’t place the arm cuff over thick clothes.
1. Connect cuff plug to the unit.
2. Wrap the arm cuff firmly
in place around your left
upper arm.
The bottom edge of the arm
cuff should be 1/2 inch (1 to 2
cm) above the elbow. Air tube
is on the inside of your arm
and aligned with your middle
finger.
3. Secure closed with
the fabric fastener.
Notes:
• When you take a measurement on the
right arm, the air tube will be at the
side of your elbow. Be careful not to
rest your arm on the air tube.
• The blood pressure can differ between the right arm and the left arm,
and the measured blood pressure values can be different. OMRON
recommends to always use the same arm for measurement. If the
values between both arms differ substantially, please check with your
physician which arm to use for your measurements.
3.2 How to Sit Correctly
To take a measurement, you need
to be relaxed and comfortably
seated, under comfortable room
temperature.
• Sit on a chair with your legs uncrossed
and your feet flat on the floor.
• Sit upright with your back straight.
• Sit with your back and arm being
supported.
• The arm cuff should be placed on your
arm at the same level as your heart.
3.3 Taking a Measurement
Notes:
• To stop the measurement, press the START/STOP button once to
deflate the arm cuff.
• Remain still and do not talk while taking a measurement.
• The memory cannot view on the display.
1. Press the START/STOP button.
The arm cuff will start to inflate automatically.
2. Transferring measured data.
As soon as the measurement is completed, the data transfer will start
automatically, as below.
When the connection symbol ( ) is not appeared, see
“Data is not being transmitted.” in section 4.2 for more detail.
3.
After data transfer is completed, press the START/STOP
button to turn the monitor off.
It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes before taking another measurement. Waiting
between measurements allows the arteries to return to the
condition prior to taking a measurement.
4. Remove the arm cuff.
DO NOT adjust medication based on measurement results from this
blood pressure monitor. Take medication as prescribed by your physician.
Only a physician is qualified to diagnose and treat High Blood Pressure.
This device is not intended to be a diagnostic device.
Always consult your physician. Self-diagnosis of measurement results
and self-treatment are dangerous.
Inflating to a higher pressure than necessary may result in bruising where
the arm cuff is applied.
5130713-0A
Blood Pressure Monitor
Model HEM-9200T
Instruction Manual
Do not damage. Do not break it.
Do not tamper with it. Do not forcibly bend or pull.
Do not twist. Do not bundle during use.
Do not pinch. Do not place under heavy objects.
A. Display
B. START/STOP button
C. Battery compartment
D. Air jack
E. AC adapter jack
(for optional AC adapter)
C
D
E
A
B
F
H
G
F. Ar m c u f f
G. Air plug
H. Air tube
I. SYNC symbol
J. Connection symbol
K. OK symbol
L. Systolic blood pressure
M. Diastolic blood pressure
N. Battery symbol (low/depleted)
O. Heartbeat symbol
(Flashes during measurement.)
P. Movement error symbol
Q. Irregular heartbeat symbol
R. Deflation symbol
S. Pulse display
L
M
N
LO
K
Q
P
R
S
I
J
Irregular Heartbeat
Normal Heartbeat
Pulse
Blood pressure
Blood pressure
Short Long
Pulse
1/2 inch
(1 - 2 cm)
START INFLATE DEFLATE COMPLETE
What to do if your systolic pressure is more
than 210 mmHg
After the arm cuff starts to inflate, press and hold the
START/STOP button until the monitor inflates 30 to 40 mmHg
higher than your expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
Inflating to a higher pressure than necessary may result in
bruising where the arm cuff is applied.
HEM-9200T-Z_A_M10_141211.pdf
EN
HEM-9200T-Z_A_M.fm Page 1 Thursday, December 11, 2014 3:33 PM
4. Error Messages and Troubleshooting
4.1 Error Messages
4.2 Troubleshooting
5. Maintenance and Storage
5.1 Maintenance
To protect your device from damage, please follow the
directions below:
• Store the device and the components in a clean, safe
location.
• Do not use any abrasive or volatile cleaners.
• Do not wash the device and any components by
immersing them in water.
• Do not use gasoline, thinners or similar solvents to
clean the device.
• Use a soft dry cloth, or a soft cloth moistened with
neutral soap to clean on the monitor and the arm cuff.
• Changes or modification not approved by the
manufacturer will void the user warranty. Do not
disassemble or attempt to repair the device or
components.
5.2 Storage
1. Unplug the air plug from the air jack.
2. Gently fold the air tube into the
arm cuff.
Note: Do not bend or crease the air tube
excessively.
Do not store the device in the following situations:
•If the device is wet.
•Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapors such as bleach.
•Locations exposed to vibrations, shocks or where it will be at an
angle.
5.3 Optional Accessories
Using the Optional AC Adapter
1. Insert the AC adapter
plug into the AC adapter
jack on the rear side of
the monitor.
2. Plug the AC adapter into
an electrical outlet.
To disconnect the AC adapter,
unplug the AC adapter from the electrical outlet first, and then remove
the AC adapter plug from the monitor.
6. Specifications
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, the 5th phase was used on 85 subjects
for determination of diastolic blood pressure.
• This device has not been validated for use on pregnant patients.
• IP classification is degrees of protection provided by IEC 60529.
7. FCC/IC Statement and Trademarks
FCC CAUTION
Changes or modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate the equipment.
Note:
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
This transmitter must not be co-located or operated in conjunction with any
other antenna or transmitter.
This equipment complies with FCC/IC radiation exposure limits set forth for
an uncontrolled environment and meets the FCC radio frequency (RF)
Exposure Guidelines and RSS-102 of the IC radio frequency (RF) Exposure
rules. This equipment has very low levels of RF energy that are deemed to
comply without testing of specific absorption ratio (SAR).
This Product operates in the unlicensed ISM band at
2.4GHz. In case this Product is used around the other
wireless devices including microwave and wireless LAN,
which operate same frequency band of this Product,
there is a possibility that interference occurs between
this Product and such other devices. If such interference
occurs, please stop the operation of other devices or
relocate this Product before using this Product or do not
use this Product around the other wireless devices.
The Bluetooth®Smart word mark and logos are
registered trademarks owned by Bluetooth SIG, Inc. and
any use of such marks by OMRON HEALTHCARE
Co., Ltd. is under license. Other trademarks and trade
names are those of their respective owners.
8. Limited Warranty
Your HEM-9200T Automatic Blood Pressure Monitor, excluding batteries, is
warranted to be free from defects in materials and workmanship appearing
within 2 years from the date of purchase, when used in accordance with the
instructions provided with the monitor. The above warranty extends only to
the original retail purchaser.
We will, at our option, replace without charge any monitor or arm cuff
covered by the above warranty. Replacement is our only responsibility and
your only remedy under the above warranty.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY
DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND
OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED
IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other
rights which vary from state to state. Some states do not allow limitations on
how long an implied warranty lasts, so the above limitation may not apply to
you. Some states do not allow the exclusion or limitation of incidental or
consequential damages, so the above limitation or exclusion may not apply
to you.
9. Guidance and Manufacturer’s Declaration
OMRON Blood Pressure Monitor (BPM) including AC-adapter
Information for accompanying documents in the scope of
IEC60601-1-2:2007
Error Display Cause Solution
Irregular heartbeats are
detected.
Remove the arm cuff. Wait 2
- 3 minutes and then take
another measurement.
Repeat the steps in section
3.3. If this error continues to
appear, contact your
physician.
Movement during
measurement.
Carefully read and repeat
the steps in section 3.3.
Connecting failure.
Data is not being
transmitted.
Refer “Connecting failure.
Data is not being
transmitted.” in section 4.2.
The batteries are low.
Recommend to replace the
batteries with new ones
ahead of time.
Refer to section 2.
The batteries are depleted.
Replace 4 batteries with new
ones at once.
Refer to section 2.
Air plug disconnected. Insert the plug securely.
Refer to section 3.1.
Arm cuff is applied too
loosely.
Apply the arm cuff tighter.
Refer to section 3.1.
Air is leaking from the arm
cuff.
Replace the arm cuff with a
new one.
Refer to section 5.3.
Movement during
measurement and the arm
cuff has not been inflated
sufficiently.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the arm cuff manually
until it is 30 to 40 mmHg
above your previous
measurement result.
Refer to section 3.3.
The arm cuff was inflated
exceeding the maximum
allowable pressure, and then
deflated automatically when
inflating the arm cuff
manually.
Do not touch the arm cuff
and/or bend the air tube
while taking a measurement.
Do not inflate the arm cuff
more than necessary.
Refer to section 3.3.
Movement during
measurement.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Clothing is interfering with
the arm cuff.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Device error. Contact your Telehealth
service provider.
Problem Cause and Solution
No power.
No display appears on the monitor.
Replace all batteries with new ones.
Check the battery installation for
proper placement of the battery
polarities.
Refer to section 2.
Measurement values appear too
high or too low.
Blood pressure varies constantly.
Many factors including stress, time of
day, and how you wrap the cuff, may
affect your blood pressure. Review
the section 1.3 and section 3.3.
Connecting failure.
Data is not being transmitted.
The blood pressure monitor might
not be properly placed within the
receiver’s transmission range and is
too far from the receiver. If there are
no causes of data transmission
interference found near the blood
pressure monitor, move the blood
pressure monitor within 16 ft. (5 m)
of the receiver and try again.
The Bluetooth®feature on the
receiver is turned off. Turn the
Bluetooth®feature on the receiver
ON, and try to transmit the readings
again.
The blood pressure monitor did not
pair successfully to the receiver. Try
to pair the devices again. Refer to
section 2.
The application on the receiver or
destination device is not ready.
Check the application then try to
transmit the readings again. Refer to
section 2. If the ERR symbol is on
the screen after checking the
application, contact your Telehealth
service provider.
Problem Cause and Solution
Arm cuff AC Adapter
Arm circumference
9" - 17" (22 - 42 cm)
Arm circumference
7" - 9" (17 - 22 cm)
CD-WR17
(Model: HEM-RML31)
• Same as the arm cuff
provided with the
product.
CD-CS9
(Model: HEM-CS24)
HEM-ADPTW5
Model HEM-9200T
Display LCD digital display
Measurement range Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats / min.
Accuracy Pressure: ±3 mmHg or 2% of reading
Pulse: ±5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Measurement
method Oscillometric method
IP classification IP 20
Power source
4 “AA” batteries 1.5V or optional AC adapter (INPUT AC100-240V
50/60Hz 0.12A)
Battery life Approximately 1000 measurements (using new alkaline
batteries)
Operating
temperature /
humidity
50°F to 104°F (10°C to 40°C) / 15 to 90% RH
Storage temperature /
humidity / air
pressure
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH /
700 to 1060 hPa
Weight Monitor : Approximately 10 oz. (290 g)
not including batteries
Arm cuff : Approximately 6 oz. (170 g)
Dimensions Monitor : Approximately 4 1/4" (w) ×3 1/8" (h) ×5 1/2" (l)
(107 mm ×79 mm ×141 mm)
Arm cuff : Approximately 5 3/4" ×23 1/2" (air tube: 29 1/2")
(145 mm
×
594 mm (air tube: 750 mm))
Cuff circumference 9" to 17" (220 to 420 mm)
Contents Monitor, arm cuff, 4 “AA” batteries, instruction manual,
quick start guide
Applied part
Protection against
electric shock
Internally powered ME equipment (When using only the
batteries)
= Type BF
= Class II ME equipment (Optional AC adapter)
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic
interference from other devices. Electromagnetic interference may result in incorrect
operation of the medical device and create a potentially unsafe situation. Medical
devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the
aim to prevent unsafe product situations, the IEC60601-1-2 standard has been
implemented. This standard defines the levels of immunity to electromagnetic
interferences as well as maximum levels of electromagnetic emissions for medical
devices.
Medical devices manufactured by OMRON Healthcare conform to this
IEC60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• The use of accessories and cables other than those specified by OMRON, with the
exception of cables sold by OMRON as replacement parts for internal components,
may result in increased emission or decreased immunity of the device.
• The medical devices should not be used adjacent to or stacked with other
equipment.
In case adjacent or stacked use is necessary, the medical device should be
observed to verify normal operation in the configuration in which it will be used.
• Refer to further guidance below regarding the EMC environment in which the device
should be used.
• The MEDICAL ELECTRICAL EQUIPMENT BPM including AC-adapter needs
special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this documentations.
• The Essential Performance of the BPM including AC-adapter is to measure a blood
pressure and a pulse rate and using the memory function.
The BPM including AC-adapter may be interfered with by other equipment, even if that
other equipment complies with CISPR EMISSION requirements.
Guidance and manufacturer’s declaration - electromagnetic emissions
OMRON BPM including AC-adapter is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BPM including AC-adapter should
assure that it is used in such environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11 Group 1
The OMRON BPM including AC-adapter uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11 Class B
The OMRON BPM including AC-adapter is suitable
for use in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BPM including AC-adapter is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BPM including AC-adapter should
assure that it is used in such environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be
that of a typical commercial
and/or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial
and/or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply
inputlines
IEC 61000-4-11
<5 % UT(>95 %
dip in UT)
for 0.5 cycle
<5 % UT(>95 %
dip in UT)
for 0.5 cycle
Mains power quality should be
that of a typical commercial
and/or hospital environment. If
the user of the OMRON BPM
including AC-adapter requires
continued operation during
power mains interruption, it is
recommended that the
OMRON BPM including
AC-adapter be powered from
an uninterruptible power
supply.
40 % UT(60 %
dip in UT)
for 5 cycles
40 % UT(60 %
dip in UT)
for 5 cycles
70 % UT(30 %
dip in UT)
for 25 cycles
70 % UT(30 %
dip in UT)
for 25 cycles
<5 % UT(>95 %
dip in UT)
for 5 sec.
<5 % UT(>95 %
dip in UT)
for 5 sec.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: UTis the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BPM including AC-adapter is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BPM including AC-adapter should
assure that it is used in such environment.
Immunity
test
IEC 60601
test level
Compl
iance
level
Electromagnetic environment - guidance
Conducted
RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 V rms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 V
rms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the OMRON
BPM including AC-adapter and cables, than the
recommended separation distance calculated from
the equation appropriate to the frequency of the
transmitter.
Recommend separation distance
d= 1.2
d= 1.2
80 MHz to 800 MHz
d= 2.3
800 MHz to 2.5 GHz
where Pis the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and dis the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/ cordless)
telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the OMRON BPM including AC-adapter is used
exceeds the applicable RF compliance level above, the OMRON BPM including
AC-adapter should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
OMRON BPM including AC-adapter.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distance between portable and mobile RF communications
equipment
and the OMRON BPM including AC-adapter
OMRON BPM including AC-adapter is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of this OMRON
BPM including AC-adapter can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the OMRON BPM including AC-adapter as recommended below,
according to the maximum output power of the communications equipment.
Output Power of Transmitter in
Watt
Separation distance according to frequency of
transmitter in meter
150 kHz to 80
MHz
d= 1.2
80 MHz to 800
MHz
d= 1.2
800 MHz to
2.5GHz
d= 2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.21.22.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance din meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where Pis the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Manufactured by: OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002
JAPAN
Distributed by: OMRON HEALTHCARE, INC.
1925 West Field Court, Lake Forest, IL 60045 U.S.A.
1-800-634-4350
www.omronhealthcare.com
© 2015 OMRON HEALTHCARE, INC.
Made in Japan
P
P
P
PPP
HEM-9200T-Z_A_M.fm Page 2 Thursday, December 11, 2014 3:33 PM

Other manuals for HEM-9200T

2

Other Omron Blood Pressure Monitor manuals

Omron HEM-7320T-CA User manual

Omron

Omron HEM-7320T-CA User manual

Omron HBP-1300 User manual

Omron

Omron HBP-1300 User manual

Omron HEM-670ITCAN User manual

Omron

Omron HEM-670ITCAN User manual

Omron INTELLI-SENSE HEM-780 Setup guide

Omron

Omron INTELLI-SENSE HEM-780 Setup guide

Omron M3W User manual

Omron

Omron M3W User manual

Omron HEM-711AC User manual

Omron

Omron HEM-711AC User manual

Omron M6 Comfort User manual

Omron

Omron M6 Comfort User manual

Omron 7 Series User manual

Omron

Omron 7 Series User manual

Omron HEM-6131 User manual

Omron

Omron HEM-6131 User manual

Omron M7 Intelli IT User manual

Omron

Omron M7 Intelli IT User manual

Omron M7 Intelli IT User manual

Omron

Omron M7 Intelli IT User manual

Omron M1 Compact User manual

Omron

Omron M1 Compact User manual

Omron HEM-7203 User manual

Omron

Omron HEM-7203 User manual

Omron BP7350 User manual

Omron

Omron BP7350 User manual

Omron JPN2 User manual

Omron

Omron JPN2 User manual

Omron BP786CAN User manual

Omron

Omron BP786CAN User manual

Omron HEM-7155T-ESL User manual

Omron

Omron HEM-7155T-ESL User manual

Omron M2 Compact User manual

Omron

Omron M2 Compact User manual

Omron HEM-FL31 User manual

Omron

Omron HEM-FL31 User manual

Omron IntelliSense HEM-650 User manual

Omron

Omron IntelliSense HEM-650 User manual

Omron M6 Comfort User manual

Omron

Omron M6 Comfort User manual

Omron HEM-7130 User manual

Omron

Omron HEM-7130 User manual

Omron M10-IT User manual

Omron

Omron M10-IT User manual

Omron M3 Intellisense User manual

Omron

Omron M3 Intellisense User manual

Popular Blood Pressure Monitor manuals by other brands

ForaCare TD-3124 owner's manual

ForaCare

ForaCare TD-3124 owner's manual

Gima 32902 user manual

Gima

Gima 32902 user manual

uebe Visomat double comfort Instructions for use

uebe

uebe Visomat double comfort Instructions for use

Microlife BP3NA1-1X Instruction booklet

Microlife

Microlife BP3NA1-1X Instruction booklet

BIOS Diagnostics PRECISION 4.0+ instruction manual

BIOS Diagnostics

BIOS Diagnostics PRECISION 4.0+ instruction manual

Braun iCheck 7 manual

Braun

Braun iCheck 7 manual

dyras BPSS-4128 user manual

dyras

dyras BPSS-4128 user manual

TrueLife Pulse BT Instructions for use

TrueLife

TrueLife Pulse BT Instructions for use

AND UM-102 instruction manual

AND

AND UM-102 instruction manual

Microlife CARDIO+ instruction manual

Microlife

Microlife CARDIO+ instruction manual

Henry Schein 900-9733 Instructions for use

Henry Schein

Henry Schein 900-9733 Instructions for use

VISIOMED bewell connect MyTensio BW-BA1 user manual

VISIOMED

VISIOMED bewell connect MyTensio BW-BA1 user manual

zewa UAM-950 user manual

zewa

zewa UAM-950 user manual

Livongo HT900 user guide

Livongo

Livongo HT900 user guide

ADC 6015 instruction manual

ADC

ADC 6015 instruction manual

Sunbeam 7682 instruction manual

Sunbeam

Sunbeam 7682 instruction manual

MABIS SmartRead Plus Guide book

MABIS

MABIS SmartRead Plus Guide book

IBP A40 instruction manual

IBP

IBP A40 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.