On-X On-X Mitral Heart Valve with Standard Sewing... User manual

0459
Prosthetic Heart Valve
Instructions for Use
English
On-X®Aortic Heart Valve with Standard Sewing Ring - REF ONXA
On-X®Mitral Heart Valve with Standard Sewing Ring - REF ONXM
On-X®Aortic Heart Valve with Conform-X®Sewing Ring - REF ONXAC
On-X®Mitral Heart Valve with Conform-X®Sewing Ring - REF ONXMC
On-X®Aortic Heart Valve with Anatomic Sewing Ring - REF ONXAN
On-X®Aortic Heart Valve and Extended Holder - REF ONXAE
On-X®Aortic Heart Valve with Conform-X®Sewing Ring and Extended Holder - REF ONXACE
On-X®Aortic Heart Valve with Anatomic Sewing Ring and Extended Holder - REF ONXANE
On-X®

English
On-X®Prosthetic Heart Valve | Instructions for Use
2
TABLE OF CONTENTS
ON-X®PROSTHETIC HEART VALVE........................................................ 2
INSTRUCTION FOR USE ......................................................................... 4
1. DEVICE DESCRIPTION ...................................................................... 4
2. INDICATIONS FOR USE .................................................................... 4
3. CONTRAINDICATIONS...................................................................... 4
4. WARNINGS AND PRECAUTIONS ...................................................... 5
4.1 WARNINGS ...................................................................................... 5
4.2 PRECAUTIONS ................................................................................ 5
5. POTENTIAL ADVERSE EVENTS ......................................................... 5
6. INDIVIDUALIZATION OF TREATMENT .............................................. 5
6.1 SPECIFIC PATIENT POPULATION.................................................... 5
7. PATIENT COUNSELING...................................................................... 5
8. HOW SUPPLIED ................................................................................ 6
8.1 AVAILABLE MODELS AND SIZES.................................................... 6
8.2 PACKAGING .................................................................................... 6
8.3 STORAGE ........................................................................................ 6
8.4 ACCESSORIES ................................................................................. 7
8.5 ACCESSORY CLEANING AND STERILIZATION.............................. 8
9. DIRECTIONS FOR USE....................................................................... 8
9.1 PHYSICIAN TRAINING ..................................................................... 8
9.2 STERILIZATION AND RESTERILIZATION ........................................ 8
9.3 HANDLING AND PREPARATION INSTRUCTIONS ........................ 8
9.4 DEVICE IMPLANTATION ...............................................................10
9.5 SUTURING TECHNIQUES...............................................................10
9.6 LEAFLET MOTION ASSESSMENT AND VALVE ROTATION...........11
9.7 VALVE ORIENTATION .....................................................................12
10. POSTOPERATIVE INFORMATION.................................................. 12
10.1 MAGNETIC RESONANCE IMAGING (MRI) COMPATIBILITY .......12
10.2 RETURNED GOODS .....................................................................12
11. PATIENT INFORMATION................................................................ 12
11.1 PATIENT REGISTRATION ..............................................................12
11.2 PATIENT RECORD CARD..............................................................13
11.3 PATIENT INFORMATION BOOKLET.............................................13
12. DISCLAIMER OF WARRANTIES ..................................................... 13
APPENDIX A ........................................................................................ 13
1. ADVERSE EVENTS ........................................................................... 13
1.1 OBSERVED ADVERSE EVENTS .......................................................13
2. CLINICAL STUDIES ......................................................................... 13
ON-X®PROSTHETIC
HEART VALVE
INSTRUCTIONS FOR USE
On-X®Aortic Heart Valve with
Standard Sewing Ring
On-X®Mitral Heart Valve with
Standard Sewing Ring
On-X®Aortic Heart Valve with
Conform-X®Sewing Ring
On-X®Mitral Heart Valve with
Conform-X® Sewing Ring
On-X®Aortic Heart Valve with
Anatomic Sewing Ring
On-X®Aortic Heart Valve and
Extended Holder
On-X®Aortic Heart Valve with
Conform-X®Sewing Ring and
Extended Holder
On-X®Aortic Heart Valve with
Anatomic Sewing Ring and
Extended Holder
For more information visit
www.onxlti.com/ifu

English
On-X®Prosthetic Heart Valve | Instructions for Use
3
LIST OF FIGURES
Figure 1: Aortic and Mitral Profiles ............................................................... 4
Figure 2: Standard or Extended Aortic Valve Holders ................................ 6
Figure 3a: Sizer and Replica Sizer ................................................................. 7
Figure 3b: Sizer ............................................................................................ 7
Figure 4: Instrument Handle ........................................................................ 8
Figure 5: Rotator ........................................................................................... 8
Figure 6: Leaflet Probe ................................................................................. 8
Figure 7a. Twist-off outer lid ........................................................................ 9
Figure 7b. Remove by pull tab... .................................................................. 9
Figure 7c. ...or invert on sterile field.............................................................. 9
Figure 8a. Peel-off outer lid .......................................................................... 9
Figure 8b. Remove by pull tab...................................................................... 9
Figure 8c. ...or invert on sterile field ............................................................. 9
Figure 9. Opening the inner container.........................................................10
Figure 10. Inserting the instrument handle..................................................10
Figure 11. Aortic replica sizers verify the aortic valve..................................10
Figure 12. Supra-annular valve positioning .................................................11
Figure 13. Sewing ring cross-sections..........................................................11
Figure 14. Removing the valve holder .........................................................11
Figure 15. Insert valve rotator .....................................................................12
Figure 16. Pivot axis of the mitral valve positioned anti-anatomically.........12
LIST OF CHARTS
Chart 1: Patient Follow-up Over Time .........................................................15
LIST OF TABLES
Table 1: On-X Valve Specifications (millimeters)........................................... 6
Table 2: Sizer Selections - Regardless of Aortic Holder Type....................... 7
Table 3: Aortic Replacement Observed Adverse Event Rates ....................15
Table 4: Mitral Replacement Observed Adverse Event Rates.....................16
Table 5: Preoperative Patient Demographics..............................................16
Table 6: Operative Patient Demographics ..................................................17
Table 7: Number Implanted and Years by Valve Size...................................18
Table 8: Valve Effectiveness Outcomes.......................................................18
Table 9: Effectiveness Outcomes, Hemodynamic Results...........................19

English
On-X®Prosthetic Heart Valve | Instructions for Use
4
orifice internal diameter (D)
orifice internal diameter (D)
tissue annulus (mounting) diameter (A)
external sewing ring diameter (S)
profile height open (H)profile height closed (h)
orifice internal diameter (D)
tissue annulus (mounting) diameter (A)
external sewing ring diameter (S)
profile height open (H)
profile height closed (h)
orifice internal diameter (D)
external sewing ring diameter (S)
orifice internal diameter (D)
tissue annulus (mounting) diameter (A)
external sewing ring diameter (S)
tissue annulus (mounting) diameter (A)
external sewing ring diameter (S)
profile height open (H)
profile height closed (h)
tissue annulus (mounting) diameter (A)
profile height open (H)
profile height closed (h)
profile height open (H)
profile height closed (h)
INSTRUCTION FOR USE
CAUTION: Federal law (USA) restricts this device to
sale by or on the order of a physician.
1. DEVICE DESCRIPTION
The On-X®Prosthetic Heart Valve (Figure 1) is a bileaflet
mechanical heart valve, which consists of an orifice
housing and two leaflets. The orifice inflow area has a
flared inlet designed to reduce flow turbulence, and
the outflow rim consists of leaflet guards designed to
protect the leaflets while in the closed position. The
leaflets rotate around tabs located within the inner
circumference of the orifice ring. In the closed position,
each leaflet forms a nominal angle of 40º relative to the
plane of the orifice. In the open position, the plane of
each leaflet forms a nominal angle of 90º relative to the
plane of the orifice. The leaflets have a travel arc of 50º
to the closed position.
The orifice is composed of graphite substrate coated
with On-X®Carbon, a pure unalloyed form of pyrolytic
carbon. The leaflets consist of On-X®Carbon deposited
on a graphite substrate, which is impregnated with 10
weight% tungsten to provide radiopacity.
The sewing ring is constructed of
polytetrafluoroethylene (PTFE) fabric mounted on the
orifice using titanium retaining rings and 5-0 suture
material. This form of sewing ring attachment to the
orifice allows for rotation of the sewing ring in situ during
implantation. Orientation reference marks are provided
on the sewing ring for valve orientation.
The On-X®Prosthetic Heart Valve is available in 3
aortic and 2 mitral sewing ring configurations. All
aortic configurations are available in sizes 19, 21, 23,
25, and 27/29 mm. The standard mitral sewing ring is
available in sizes 23, 25, 27/29 and 31/33, while the mitral
Conform-X®sewing ring is available in size 25/33 only.
Aortic valves, size 19 mm through 25 mm, are designed
for intrasupra-annular sewing ring position, while the
valve size 27/29 mm is designed for intra-annular sewing
ring position. All mitral valve sizes are designed for the
supra-annular sewing ring position.
2. INDICATIONS FOR USE
The On-X Prosthetic Heart Valve is indicated for the
replacement of diseased, damaged, or malfunctioning
native or prosthetic heart valves in the aortic and mitral
positions.
3. CONTRAINDICATIONS
The On-X Prosthetic Heart Valve is contraindicated for
patients unable to tolerate anticoagulation therapy.
Figure 1: Aortic and Mitral Profiles
(See Table 1 for corresponding dimensions)
(C) AORTIC ANATOMIC (ONXAN, ONXANE)
(B) AORTIC CONFORM-X®(ONXAC, ONXACE)
(A) AORTIC (ONXA, ONXAE)
(D) MITRAL (ONXM)
(E) MITRAL CONFORM-X®(ONXMC)

English
On-X®Prosthetic Heart Valve | Instructions for Use
5
4. WARNINGS AND PRECAUTIONS
4.1 Warnings
FOR SINGLE USE ONLY.
DO NOT use the On-X Prosthetic Heart Valve if:
• the prosthesis has been dropped, damaged, or
mishandled in any way;
• the expiration date has elapsed;
• the tamper evident seal is broken;
• the serial number tag does not match the serial
number on the container label.
DO NOT pass a catheter, surgical instrument, or
transvenous pacing lead through the prosthesis as
this may cause valvular insufficiency, leaflet damage,
leaflet dislodgment, and/or catheter/instrument/lead
entrapment.
DO NOT resterilize the On-X Prosthetic Heart Valve.
4.2 Precautions
Handle the prosthesis with only On-X Life
Technologies, Inc. (On-XLTI) On-X Prosthetic Heart
Valve Instruments. Only On-XLTI On-X Prosthetic
Heart Valve sizers should be used during the
selection of the valve size; other sizers may result in
improper valve selection.
Avoid contacting the carbon surfaces of the valve
with gloved fingers or any metallic or abrasive
instruments as they may cause damage to the valve
surface not seen with the unaided eye that may lead
to accelerated valve structural dysfunction, leaflet
escape, or serve as a nidus for thrombus formation.
Avoid damaging the prosthesis through the
application of excessive force to the valve orifice or
leaflets.
5. POTENTIAL ADVERSE EVENTS
Adverse events potentially associated with the use of
prosthetic heart valves (in alphabetical order) include,
but are not limited to:
• angina
• cardiac arrhythmia
• endocarditis
• heart failure
• hemolysis
• hemolytic anemia
• hemorrhage
• myocardial infarction
• prosthesis leaflet entrapment (impingement)
• prosthesis nonstructural dysfunction
• prosthesis pannus
• prosthesis perivalvular leak
• prosthesis regurgitation
• prosthesis structural dysfunction
• prosthesis thrombosis
• stroke
• thromboembolism
It is possible that these complications could lead to:
• reoperation
• explantation
• permanent disability
• death
Mechanical prosthetic heart valves produce audible
sounds as a normal function of their operation. In some
patients, these sounds may be objectionable.
Risk of Re-Use Statement
In accordance with the EU Medical Device Directive,
93/42/EEC, Annex I, Section 13.6h, the device
manufacturer must provide information on risks
associated with re-use of a single use device. Therefore,
the following statement is provided:
The implanted On-X prosthetic heart valve is designed
for single use only. Do not re-use the device. In addition
to the risks listed in Section 5, re-use may cause
procedural complications including device damage,
compromised device biocompatibility, and device
contamination. Re-use may result in infection, serious
injury, or patient death.
6. INDIVIDUALIZATION OF TREATMENT
Adequate anticoagulant or anticoagulant/antiplatelet
therapy should be administered. Selection of an
anticoagulant or anticoagulant/antiplatelet regimen is
based on the particular needs of the patient and the
clinical situation.
6.1 Specific Patient Population
The safety and effectiveness of the On-X Prosthetic
Heart Valve has not been established for the following
specific populations because it has not been studied in
these populations:
• patients who are pregnant;
• nursing mothers;
• patients with chronic endocarditis;
• patients requiring pulmonary or tricuspid
replacement.
7. PATIENT COUNSELING
• Prophylactic antibiotic treatment must be provided
to all patients with prosthetic valves undergoing
dental procedures or other potentially bacteremic
procedures.
• Patients require anticoagulation or anticoagulant/
antiplatelet therapy.
• Patients should be encouraged to complete the
Patient ID card provided with the valve and carry it
with them at all times.

English
On-X®Prosthetic Heart Valve | Instructions for Use
6
8. HOW SUPPLIED
8.1 Available Models and Sizes
The On-X Prosthetic Heart Valve is available in 3 aortic
and 2 mitral sewing ring configurations. All aortic
configurations are available in sizes 19, 21, 23, 25, and
27/29 mm. The standard mitral sewing ring is available
in sizes 23, 25, 27/29 and 31/33 mm, while the Mitral
Conform-X sewing ring is available in size 25/33 only.
Aortic valves, size 19 mm through 25 mm, are designed
for intrasupra-annular sewing ring position, while the
valve size 27/29 mm is designed for intra-annular sewing
ring position. All mitral valve sizes are designed for the
supra-annular sewing ring position.
Aortic valves are available with either standard or
extended valve holders (Figure 2).
The dimensional and model specifications for all
available sizes of the On-X Prosthetic Heart Valve are
shown in Table 1 and Figure 1. The symbol SZ mm on
the box, container labels, and implant registration card
refers to the tissue annulus diameter of the valve in
millimeters.
8.2 Packaging
The On-X Prosthetic Heart Valve is provided sterile,
mounted on a holder, in a double-sealed plastic
container. The package consists of the following items:
• Outer box
• Patient record card
• Plastic valve container
• Implant registration card
• Plastic valve holder
• Valve serial number tag
• Instructions for use
Instruments for implantation of the On-X Prosthetic
Heart Valve are supplied separately, NON-STERILE, and
must be cleaned and sterilized prior to use as outlined in
section 8.5.
8.3 Storage
The On-X Prosthetic Heart Valve has been qualified
for a maximum storage life of 5 years from the date of
manufacture for the twist-off lid design, or 6 years for
the peel-off lid design. The sterility expiration date of
the On-X Prosthetic Heart Valve is recorded on the outer
package label. Appropriate inventory control should be
maintained so that prostheses with earlier expiration
dates are preferentially implanted and expiration is
avoided. To protect the valve, it should be stored in its
outer box until used. The storage environment should
be clean, cool, and dry.
Figure 2: Standard or Extended Aortic Valve Holders
Extended Holder Standard Holder
Model Designator
Size /Type
Tissue Annulus
(mounting)
Diameter
(A)
Orifice Internal
Diameter
(D)
External
Sewing Ring
Diameter
(S)
Profile Height
(closed)
(h)
Profile Height
(open)
(H)
Internal Orifice
Area
(mm²)Extended Holder Standard Holder
ONXAE-19* ONX A-19 19 Aortic 19 17.4 23 10.8 13.3 228
ONXAE-21* ONXA-21 21 Aortic 21 19.4 26 11.9 14.7 284
ONXAE-23* ONXA-23 23 Aortic 23 21.4 29 13.1 16.1 344
ONXAE-25* ONXA-25 25 Aortic 25 23.4 32 14.2 17.8 411
ONXAE-27/29* ONX A-27/29 27/29 Aortic 27-29 23.4 34 14.2 17. 8 411
ONXACE-19* ONXAC-19* 19 Aortic Conform-X 19 17.4 27 10.8 13.3 228
ONXACE-21* ONXAC-21* 21 Aortic Conform-X 21 19.4 30 11.9 14.7 284
ONXACE-23* ONXAC-23* 23 Aortic Conform-X 23 21.4 33 13.1 16 .1 344
ONXACE-25* ONXAC-25* 25 Aortic Conform-X 25 23.4 34 14.2 17. 8 411
ONXACE-27/29* ONX AC-27/29* 27/29 Aortic Conform-X 27-29 23.4 36 14.2 17. 8 411
ONXANE-19* ONX AN-19* 19 Aortic Anatomic 19 17.4 27 10.8 13.3 228
ONXANE-21* ONX AN-21* 21 Aortic Anatomic 21 19.4 30 11.9 14.7 284
ONXANE-23* ONXAN-23* 23 Aortic Anatomic 23 21.4 33 13.1 16.1 344
ONXANE-25* ONXAN-25* 25 Aortic Anatomic 25 23.4 34 14.2 17. 8 411
ONXANE-27/29* ONXAN-27/29* 27/29 Aortic Anatomic 27/2 9 23.4 36 14.2 17.8 411
ONXM-23** 23 Mitral 23 21.4 31 13.1 16.1 344
ONXM-25 25 Mitral 25 23.4 33 14.2 17.8 411
ONXM -27/2 9 27/29 Mitral 27-29 23.4 34 14.2 17. 8 411
ONXM-31/33 31/33 Mitral 31-33 23.4 36 14.2 17. 8 411
ONXMC-25/33 Mitral Conform-X 25-33 23.4 39 14.2 17.8 411
Table 1: On-X Valve Specifications (millimeters)
* Not available in all markets ** Not available in the USA Refer to Figure 1 for location of measured dimensions. Values given are nominal within the tolerance band.

English
On-X®Prosthetic Heart Valve | Instructions for Use
7
8.4 Accessories
The On-X Prosthetic Heart Valve is designed to be used
only with On-XLTI On-X instruments. The instruments,
supplied separately, are provided as a set, which
includes sizers, rotators, an instrument handle, and a
leaflet probe. The instruments are reusable.
CAUTION: Sizers and instrument handles have
metallic regions that are bendable. Repeated
bending of these metallic regions can lead to fatigue
and fracture. To avoid instrument fracture during
use, the stem should be inspected for surface
cracks before and after each time it is bent. If metal
fatigue surface cracks are present, the sizer and/
or instrument handle should be discarded and
replaced. Contact On-XLTI Customer Service to order
replacements.
CAUTION: Leaflet probes and rotators are flexible,
but are not intended to be bent to a permanently
deformed state.
Sizer
The sizer is used to gauge the resulting tissue annulus
diameter after the annulus is prepared for implant. The
sizer has a bendable stem on each end. The sizers are
cylindrical for size 19 mm through 25 mm valves and
conical for size 27/29 mm and 31/33 mm valves (Figure 3a
and 3b). To facilitate sizer selection, refer to Table 2.
Sizer Choice Position of sewing ring
Size Valve Type Sizer Type Use Replicate Sizer
19 Aortic Cylindrical YES Intrasupra-annular
21 Aortic Cylindrical YES Intrasupra-annular
23 Aortic Cylindrical YES Intrasupra-annular
25 Aortic Cylindrical YES Intrasupra-annular
27/29 Aortic Conical YES Intra-annular
19* Aortic Conform-X Cylindrical YES Intrasupra-annular
21* Aortic Conform-X Cylindrical YES Intrasupra-annular
23* Aortic Conform-X Cylindrical YES Intrasupra-annular
25* Aortic Conform-X Cylindrical YES Intrasupra-annular
27/29* Aortic Conform-X Conical YES Intra-annular
19* Aortic Anatomic Cylindrical YES Intrasupra-annular
21* Aortic Anatomic Cylindrical YES Intrasupra-annular
23* Aortic Anatomic Cylindrical YES Intrasupra-annular
25* Aortic Anatomic Cylindrical YES Intrasupra-annular
27/29* Aortic Anatomic Conical YES Intra-annular
23* Mitral Cylindrical NO Supra-annular
25 Mitral Cylindrical NO Supra-annular
27/29 Mitral Conical NO Supra-annular
31/33 Mitral Conical NO Supra-annular
25/33 Mitral Conform-X Cylindrical or Conical NO Supra-annular
Replica Sizers
Aortic replica sizers are provided for all aortic valve
sizes (Figure 3a). They model the On-X standard aortic
valve profile. They are used after sizing for standard,
Conform-X, and Anatomic sewing ring configurations
to assure fit of the aortic valve without obstruction of
the coronary arteries. The size 19 through 25 aortic
replica sizers shape is intended to model intrasupra-
annular positioning. The size 27/29 aortic replica sizer is
intended to model intra-annular positioning.
Table 2: Sizer Selections - Regardless of Aortic Holder Type
Figure 3a: Sizer and Replica Sizer
Figure 3b: Sizer
Cylindrical
Aortic Replica
Conical
* Not available in all markets

English
On-X®Prosthetic Heart Valve | Instructions for Use
8
Instrument Handle
The instrument handle (Figure 4) facilitates holding the
valve or the rotator during surgery. The instrument handle
is comprised of a grip, a bendable stem, and a tip.
Rotator
The valve rotator (Figure 5) is used for reorienting an in
situ valve and may be used to verify leaflet mobility. The
rotator consists of a plastic head with a centrally located
leaflet probe and an attached handle.
The rotator may be used with or without the instrument
handle attached. To attach the rotator to the instrument
handle, insert the instrument handle tip directly into the
slot on the end of the rotator handle. The rotator snaps
into place after the application of a light insertion force.
Leaet Probe
The leaflet probe (Figure 6) is a flexible rod with tapered
ends. The leaflet probe may be used to gently move the
leaflets to verify that they open and close freely.
WARNING: DO NOT sterilize instruments with any
method of sterilization other than steam. Damage
to some items could result from use of other
sterilization methods.
WARNING: The rotator must be removed from
the handle after use and prior to cleaning. A force
greater than the insertion force is required to
remove the rotator from the instrument handle.
9. DIRECTIONS FOR USE
WARNING: DO NOT use the On-X Prosthetic Heart
Valve if:
• the prosthesis has been dropped, damaged, or
mishandled in any way;
• the expiration date has elapsed;
• the tamper evident seal is broken;
• the serial number tag does not match the serial
number on the container label.
9.1 Physician Training
No special training is required to implant the On-X
Prosthetic Heart Valve. The techniques for implanting
this prosthesis are similar to those used for any
mechanical heart valve prosthesis.
9.2 Sterilization and Resterilization
The On-X Prosthetic Heart Valve is provided sterile.
If the sterility expiration date has passed or if upon
removal from the outer box, the valve container is
damaged or the sterility barrier is broken, do not use
the valve. Call On-XLTI Customer Service and arrange to
return the valve and receive a replacement.
WARNING: If during surgery the valve is removed
from its container but not used, it must not be
repackaged or resterilized. In this situation, the valve
must be returned to On-XLTI. Call Customer Service for
information before any return is made.
WARNING: Do not resterilize the On-X Prosthetic
Heart Valve.
9.3 Handling and Preparation Instructions
CAUTION: Handle the prosthesis with only On-XLTI
On-X Prosthetic Heart Valve Instruments. Only On-
XLTI On-X Prosthetic Heart Valve sizers should be
used during the selection of the valve size; other
sizers may result in improper valve selection.
CAUTION: Avoid contacting the carbon surfaces
of the valve with gloved fingers or any metallic or
abrasive instruments as they may cause damage to
the valve surface not seen with the unaided eye that
may lead to accelerated valve structural dysfunction,
leaflet escape, or serve as a nidus for thrombus
formation.
CAUTION: Avoid damaging the prosthesis through
the application of excessive force to the valve orifice
or leaflets.
Figure 4: Instrument Handle
Figure 5: Rotator
Figure 6: Leaflet Probe
8.5 Accessory Cleaning and Sterilization
Instruments for implantation of the On-X Prosthetic
Heart Valve are supplied separately, NON-STERILE, and
must be cleaned and sterilized prior to use. Standard
hospital surgical instrument cleaning procedures
must be used. Note: the metallic instruments are
made of titanium. The plastic instruments are made of
polyphenylsulfone. Materials used in these instruments
can withstand standard steam and flash steam
sterilization.
WARNING: These instruments are NOT provided
sterile. They must be properly cleaned and sterilized
prior to each use.

English
On-X®Prosthetic Heart Valve | Instructions for Use
9
Circulating Nurse
1. Check the expiration date on the outer box.
WARNING: DO NOT use the On-X Prosthetic Heart
Valve if the expiration date has elapsed. If a valve
is unused, its plastic container is undamaged, and
the sterility expiration date has passed, the valve
should be returned to On-XLTI.
2. Remove the valve container and package inserts from
the outer box. Inspect the container for damage.
WARNING: DO NOT use the On-X Prosthetic
Heart Valve if the prosthesis has been dropped,
damaged, or mishandled in any way. If any
damage is found, use another valve and arrange
for a return through On-XLTI Customer Service.
3. Fill out the implant registration card as completely
as local law allows and return to On-XLTI as soon as
possible. This allows the patient to be entered into
the tracking database, which could be important for
future notices regarding the valve. Give the patient
record card to the patient or place it in the patient’s
records.
Figure 7a. Twist-off outer lid Figure 7b. Remove by pull tab... Figure 7c. ...or invert on sterile field
4. Open the outer container
Twist-off outer lid package design: Rotate the lid
counter-clockwise until it stops, then lift the lid off off
of the container (Figure 7a).
Peel-off Tyvek®lid package design: Grasp the peel
tab corner of the lid and pull back towards the center
of the container (Figure 8a). Continue peeling until
the lid is completely removed.
5. The scrub nurse may remove the sterile inner
container from the outer container by gently lifting
the pull tab attached to the top of the inner container
(Figure 7b or Figure 8b). The inner container is then
placed onto the instrument tray. Alternately, the inner
container can be placed on the sterile field by gently
inverting the outer container slightly above the sterile
field (Figure 7c or Figure 8c) and allowing the inner
container to slip out onto the sterile field.
Twist-off Lid Design
Figure 8a. Peel-off outer lid Figure 8b. Remove by pull tab... Figure 8c. ...or invert on sterile field
Peel-off Lid Design
On-X®Prosthetic Heart Valve | Instructions for Use

English
On-X®Prosthetic Heart Valve | Instructions for Use
10
Scrub Nurse/Surgeon:
1. Check the tamper evident seal of the inner container.
WARNING: DO NOT use the On-X Prosthetic Heart
Valve if the tamper evident seal has been broken.
If the tamper evident seal has been broken, use
another valve and arrange for return through On-
XLTI Customer Service.
2. Open the inner container by gently twisting the lid
to break the tamper-proof seals (Figure 9) and then
lifting the lid off the base.
3. Press the instrument handle tip into the slot on the
valve holder until it snaps firmly into position (Figure
10). Gently lift the valve out of the container and slide
the holder plate off the holder.
Carefully grasp the sewing ring with a gloved hand
using a light grip and gently turn the instrument
handle in either direction. The valve should easily
rotate within the sewing ring. Stop rotation testing
with an orientation mark aligned with the pivot axis.
WARNING: DO NOT use the On-X Prosthetic
Heart Valve if the valve does not rotate easily. Use
another valve and arrange for return through On-
XLTI Customer Service.
6. The valve is now ready for implantation. To ease
positioning during implantation, the instrument
handle stem can be bent by grasping the ends of the
handle and the stem, then bending. Avoid grasping
the valve.
WARNING: DO NOT use the valve for leverage in
bending the instrument handle. This could damage
the valve and lead to mechanical failure.
9.4 Device Implantation
WARNING: All accessory instruments must be
cleaned and sterilized prior to use according to the
instrument instructions.
Sizing
Use only On-X Prosthetic Heart Valve sizers when sizing
the annulus. Sizers contain cylindrical, conical, and aortic
replica ends. Refer to Table 2 to facilitate sizer selection.
Cylindrical sizers correspond to the valve sizes 19 mm
through 25 mm. Conical sizers correspond to the valve
size 27/29 mm and 31/33 mm. These types of sizers may
be used for both aortic and mitral valves.
The correct valve size is determined by obtaining a
comfortable, not tight, fit of the sizer within the annulus.
When a comfortable fit is found, the corresponding
valve size is signified by the identification on the sizer.
On-X Mitral Conform-X Prosthetic Heart Valves may be
used when a comfortable fit is at or between size 25 and
size 33.
Aortic replica sizers are provided for all aortic valve sizes.
For size 19 mm through 25 mm aortic valves, the aortic
replica sizers are used to verify that the aortic valve can
be properly seated in the annulus and that the coronary
arteries remain unobstructed. Size 19 mm through 25
mm aortic valves of standard, Conform-X, and Anatomic
sewing ring configurations are designed to fit within the
annulus at implant such that the exposed carbon flare
rests in the annulus and the sewing ring is intrasupra-
annular (Figure 11).
Figure 11. Aortic replica sizers verify the aortic valve
Figure 10. Inserting the
instrument handle
Figure 9. Opening the inner container
4. Check the serial number tag against the label on the
outer container.
WARNING: DO NOT use the On-X Prosthetic Heart
Valve if the serial number tag does not match the
container label. Use another valve and arrange for
a return through On-XLTI Customer Service.
5. Remove the serial number tag by cutting the suture
that holds it on the valve. If desired, the tag can
be used to check for sterility by standard culture
techniques immediately after it is removed.

English
On-X®Prosthetic Heart Valve | Instructions for Use
11
Figure 13. Sewing ring cross-sections
(A)
Aortic
(B)
Mitral
(D)
Aortic
Conform-X
and Aortic
Anatomic
(C)
Mitral
Conform-X
Housing
Titanium
Rings
Layered
Cloth
Housing
Titanium
Rings
Layered
Cloth
Figure 14. Removing the valve holder
WARNING: DO NOT size the sewing ring of the size
19 mm through 25 mm aortic valve to fit within the
annulus.
Size 27/29 mm aortic valves of standard, Conform-X, and
Anatomic sewing ring configurations are designed to
be placed in an intra-annular position and have a replica
sizer to mimic this placement.
All mitral valves, including the On-X Mitral Conform-X
Prosthetic Heart Valve, are designed to be placed in a
supra-annular position (Figure 12).
CAUTION: Avoid oversizing the valve, as this could
lead to interference with valve function.
Figure 12. Supra-annular valve positioning
9.5 Suturing Techniques
The suturing techniques vary according to the
preferences of the implanting surgeon and patient
condition. The aortic valve is designed to have the tissue
annulus abut the orifice flare. The general consensus
among surgeons is that the non-everting interrupted
mattress suture technique, with or without pledgets,
provides the best conformation of the valve annulus to
the outer surface of the flare.
The mitral valves have generally been implanted using a
pledgetted or non-pledgetted everting mattress suture
technique, although non-everting and continuous suture
techniques have also been used with success.
CAUTION: When seating the valve, ensure that no
suture material or anatomic structures interfere with
leaflet motion. The valve’s rotation capability may be
helpful in avoiding abnormal residual pathology that
could interfere with leaflet motion.
The sutures should be passed through the mid-point of
the sewing ring. This allows the sewing ring to remain
flexible and conform to the annulus. It also prevents the
suture needle from contacting the titanium rings that lie
within the sewing ring (Figure 13). The orientation marks on
the sewing ring may be used to aid in suture placement.
CAUTION: For the Anatomic sewing ring, the sutures
at the three valve commissures must correspond to
the three orientation marks on the sewing ring.
When all the sutures are in place, the valve is advanced
into the annulus and the sutures are tied down. For
aortic valves, it is suggested that the first 3 knots be tied
equidistant to one another and midway between the
commissures to stabilize the valve in the annulus. The
holder is removed from the valve by carefully cutting the
retaining suture as shown in Figure 14, then gently lifting
the valve holder with handle out of the valve.
WARNING: Do NOT attempt to reinsert the valve
holder into the valve once it has been removed.
CAUTION: Suture ties should be cut short to avoid
any potential interference with leaflet motion.

English
On-X®Prosthetic Heart Valve | Instructions for Use
12
10. POSTOPERATIVE INFORMATION
10.1 Magnetic Resonance Imaging (MRI)
Compatibility
MR Conditional:
The On-X Prosthetic Heart Valve, Mitral Conform-X
Heart Valve Prosthesis, Size 25-33*, was determined
to be MR-conditional according to the terminology
specified in the American Society for Testing and
Materials (ASTM) International, Designation: F2503-
08. Standard Practice for Marking Medical Devices
and Other Items for Safety in the Magnetic Resonance
Environment. ASTM International, 100 Barr Harbor Drive,
PO Box C700, West Conshohocken, Pennsylvania.
Non-clinical testing demonstrated that the On-X Mitral
Conform-X Heart Valve Prosthesis, Size 25-33, is MR
Conditional. A patient with this device can be scanned
safely immediately after placement under the following
conditions:
Static Magnetic Field:
• Static magnetic field of 3-Tesla or less
• Maximum spatial gradient magnetic field of
720-Gauss/cm or less
9.6 Leaflet Motion Assessment and Valve Rotation
Leaet Motion Testing
Once the valve is in place, free motion of the leaflets
must be tested. To test leaflet mobility, use the rotator
probe or the leaflet probe to gently move the leaflets to
verify that they open and close freely.
WARNING: Test the leaflet mobility only with the
On-XLTI On-X leaflet probe or the leaflet probe on
the end of the rotator.
Rotation
If the leaflets do not move freely, gently rotate the valve
in either direction until it reaches a position where leaflet
interference is not encountered.
CAUTION: Do not attempt to rotate the valve if any
significant resistance to rotation is encountered.
The torque required to rotate the valve in situ
should be about the same as that required when
testing rotation before implantation. If noticeably
greater torque is required to rotate, stop attempting
rotation. If rotation is necessary and cannot be
performed, remove the valve.
The rotator may be used with or without the instrument
handle attached. As needed, attach the instrument
handle to the rotator by inserting the instrument handle
tip into the slot on the end of the rotator handle until it
snaps firmly into position.
WARNING: Use only the On-XLTI On-X rotator to
rotate the valve in situ. Use only the correspondingly
sized rotator. Use of the wrong size rotator could
damage the valve.
With the rotator leaflet probe between the leaflets and
the cross-bar aligned with the leaflet pivot axis of the
valve, carefully insert the valve rotator into the valve until
it seats easily in place (Figure 15).
CAUTION: No resistance should be experienced
when inserting the rotator. If resistance is
encountered, stop, remove, and realign the rotator
before attempting to insert the rotator again.
Retest leaflet motion after rotation. If free leaflet motion
cannot be achieved, remove the valve.
9.7 Valve Orientation
Aortic:
Based on clinical studies, there is no preferred
orientation for the Aortic On-X Prosthetic Heart Valves
with the standard, Conform-X, or Anatomic sewing ring
configurations.
CAUTION: Once the valve is implanted, visually
confirm that the coronary ostia are free from
potential interference.
Mitral:
Literature suggests that the pivot axis of the mitral
valve should be positioned anti-anatomically. Refer to
Figure 16.
Figure 16. Pivot axis of the mitral valve positioned anti-
anatomically
Mitral Standard and Conform-X
Figure 15. Insert valve rotator
Cross-bar

English
On-X®Prosthetic Heart Valve | Instructions for Use
13
MRI-Related Heating:
In non-clinical testing, the On-X Prosthetic Heart Valve,
Mitral Conform-X Heart Valve Prosthesis, Size 25-
33, produced the following temperature rise during
MRI performed for 15-min of scanning (i.e., per pulse
sequence) in the 3-Tesla (3-Tesla/128-MHz, Excite,
HDx, Software 14X.M5, General Electric Healthcare,
Milwaukee, WI) MR system:
Highest temperature change +1.6˚C
Therefore, the MRI-related heating experiments for the
On-X Mitral Conform-X Heart Valve Prosthesis, Size 25-
33, at 3-Tesla, using a transmit/receive RF body coil at an
MR system reported whole body averaged SAR of 2.9-
W/kg (i.e., associated with a calorimetry measured whole
body averaged value of 2.7-W/kg), indicated that the
greatest amount of heating that occurred in association
with these specific conditions was equal to or less than
+1.6˚C .
Artifact Information:
MR image quality may be compromised if the area of
interest is in the exact same area or relatively close to
the position of the On-X Mitral Conform-X Heart Valve
Prosthesis, Size 25-33. Therefore, optimization of MR
imaging parameters to compensate for the presence of
this device may be necessary.
11. PATIENT INFORMATION
11.1 Patient Registration
In each valve package, there is a Patient Record Card
and an Implant Registration Card. On-XLTI requests that
the Implant Registration Card be filled out immediately
and that the mailing copy be returned to On-XLTI
Customer Service. For multiple valve implants, please
fill out a card for each valve. On-XLTI will use these data
for notification purposes and to help with inventory
restocking in the hospital. All patient information
remains strictly confidential, and the release of patient-
identifying information can be refused if allowed by law.
11.2 Patient Record Card
A Patient Record Card is provided with the prosthesis.
Patients should be encouraged to complete the card
and carry it with them at all times.
11.3 Patient Information Booklet
On-XLTI has made available a patient information
booklet that the physician may choose to provide to
the patient prior to discharge. Copies of this booklet
are available on request from your On-XLTI sales
representative.
12. DISCLAIMER OF WARRANTIES
Because of the complications listed previously that may
occur with the use of any heart valve prosthesis and the
possibilities of damage, also noted previously, before,
during or after implantation, On-XLTI warrants only
that the product shall conform to On-XLTI’s standard
specifications. No other warranty is made by On-XLTI
concerning the function of the product in use, and
On-XLTI assumes no risk whatsoever as to the results
of the use of this product. The entire risk with use of
the product is that of the buyer. On-XLTI disclaims all
other warranties, respecting the product, expressed or
implied, including but not limited to those related to
the product’s merchantability or fitness for a particular
purpose. On-XLTI shall not be liable for any direct,
special, consequential or incidental loss, damage or
expense related to the use of the product. No person
has any authority to alter any of these conditions or to
bind On-XLTI to any additional responsibility or warranty
in connection with the use of the product.
*The MRI findings apply to this specific largest heart
valve prosthesis and all other smaller sizes which are
made from similar materials.
10.2 Returned Goods
Prior authorization from On-XLTI Customer Service is
required for the return of any product. For any questions
regarding the valve or for return authorization, please
contact Customer Service.
Licensed under U.S. Patent Nos. 5,308,361; 5,137,532;
5,545,216; 5,772,694; 5,641,324; 5,908,452; 5,284,676;
5,305,554; 5,328,713, 5,332,337; 5,336,259; 5,514,410;
5,677,061; 6,096,075; Serial No. 09/010,449 allowed; Serial
No. 09/224,816 allowed; and other permits and patents
pending.
Pulse Sequence Signal Void Size
(mm2)
Plane Orientation
T1-SE 1,090 Parallel
T1-SE 686 Perpendicular
GRE 1,478 Parallel
GRE 1,014 Perpendicular

English
On-X®Prosthetic Heart Valve | Instructions for Use
14
APPENDIX A
Clinical information as required by FDA (USA)
1. ADVERSE EVENTS
A total of 184 aortic On-X Prosthetic Heart Valves were
implanted in 184 patients at 11 centers. The mean
follow-up was 2.2 years (range of 0 to 4.0 years) with a
total of 411.8 patient-years. In the mitral position 229
valves were implanted in 229 patients at 16 centers.
Mean mitral follow-up was 1.8 years (range of 0 to 4.5
years) with a total of 417.9 patient-years.
In aortic patients, a total of 7 deaths occurred during the
study and 2 of these were characterized as valve-related.
The causes of the aortic valve-related deaths were early
thromboembolism (1 patient) and sudden, unexplained
death (1 patient). In mitral patients, a total of 18
deaths occurred during the study and 3 of these were
characterized as valve-related. The causes of the mitral
valve-related deaths were early, uncontrolled bleeding (1
patient) and sudden, unexplained death (2 patients).
1.1 Observed Adverse Events
Adverse events were reported in the clinical study as
shown in Tables 3 and 4
2. CLINICAL STUDIES
The On-X Prosthetic Heart Valve clinical trials were
designed to study the safety and effectiveness of the
valve in aortic and mitral valve replacement. Patients
requiring isolated aortic heart valve replacement
were enrolled from 1996 to 2000 at 11 centers in an
international multicenter, prospective, non-randomized
study with retrospective controls. Patients requiring
isolated mitral heart valve replacement were enrolled
from 1996 to 2001 at 16 centers in an international
multicenter, prospective, non-randomized study with
retrospective controls.
The aortic cohort included 184 patients (121 men, 63
women), aged from 20 to 80 years (mean of 60.2 years).
The cumulative follow-up was 411.8 patient-years with a
mean follow-up of 2.2 years (SD = 0.8 years, range = 0
to 4.0 years). The mitral cohort included 229 patients (86
men, 143 women), aged from 21 to 78 years (mean of 59.2
years). The cumulative follow-up was 417.9 patient-years
with a mean follow-up of 1.8 years (SD = 1.3 years, range
= 0 to 4.5 years). Tables 3 and 4 present preoperative
and operative patient demographics. Figure 14 shows
the number of patients implanted versus duration of
follow-up. Tables 5 presents implant information by valve
size, including the number of patients implanted and the
number of patient-years.
Safety endpoints captured in the studies were
complications; blood analyses were used to confirm the
absence or presence of certain complications. Safety
results are provided in Tables 1 and 2. Effectiveness
endpoints were New York Heart Association (NYHA)
classification and echocardiographic assessments.
NYHA and blood data were obtained pre-operatively,
intra-operatively, and post-operatively at 3 to 6 months,
at one year, and annually thereafter. Hemodynamic data
were obtained at discharge and at one year. Tables 6
and 7 present these effectiveness results.

English
On-X®Prosthetic Heart Valve | Instructions for Use
15
Complication Early Events Late Events2Freedom from Event3, % [SE]
n % (n/N)4n %/pt-yr 1 Year Postoperative (n=138) 3 Year Postoperative (n=37)
Mortality (all) 42.2% 30.7% 97. 8% [1.1] 96.0% [1.5]
Mortality (valve-related) 10.5% 10.2% 99.4% [0.5] 98.8% [0.9]
Endocarditis 00.0% 20.5% 99.4% [0.6] 98.9% [0.8]
Explant 10.5% 20.5% 98.4% [0.9] 97. 8 % [1.1]
Hemolysis500.0% 00.0% 100.0% [0] 100.0%[0]
Hemorrhage6 (all) 10.5% 30.7% 99.4% [0.5] 97. 3% [1.4 ]
Hemorrhage (major) 10.5% 10.2% 100.0% [0] 99.1% [0.9]
Perivalvular Leak (all) 42.2% 30.7% 96.7% [1.3] 96.7% [1.3]
Perivalvular Leak (major) 10.5% 00.0% 100.0% [0] 100.0% [0]
Nonstructural Valve Dysfunction 00.0% 00.0% 100.0% [0] 100.0% [0]
Reoperation (valve-related) 21.1% 30.7% 97.8 % [1.1] 97.2% [1.2]
Structural Valve Dysfunction 0 0.0% 00.0% 100.0% [0] 100.0% [0]
Thromboembolism 10.5% 71.7% 97. 8% [1.1] 93.9% [2.5]
Thrombosis 00.0% 00.0% 100.0% [0] 100.0% [0]
Notes:
1. Data does not include results from double valve replacement.
2. Late events calculated as linearized rates based on total patient-years.
3. Freedom from event was calculated based on the method of Kaplan-Meier. SE = Standard Error.
4. n = number of patients in each category; N = total number of study patients.
5. Blood studies conducted at a core laboratory established that the valve creates a low level of fully compensated hemolysis typified by an increase in SLDH
with a mean within normal range, a decrease in haptoglobin to below normal in 69% Aortic Valve Replacement (AVR) and 65% Mitral Valve Replacement (MVR)
patients at 1-year, and all other analytes within normal range.
6. The anticoagulant agents used were reported. The target International Normalized Ratio was 2.5 to 3.5 in AVR and 3.0 to 4.5 in MVR.
Table 3: Aortic Replacement Observed Adverse Event Rates1
All patients implanted, N = 184, Cumulative follow-up = 411.8 patient-years
Patients
Followed, Nf
Discharge 1 Year
Postoperative
2 Year
Postoperative
3 Year
Postoperative
Aortic 184 138 66 37
Mitral 216 134 74 44
Chart 1: Patient Follow-up Over Time
Aortic patients implanted N = 184,
Cumulative follow-up = 411.8 patient-years
Mitral patients implanted N = 229,
Cumulative follow-up = 417.9 patient-years

English
On-X®Prosthetic Heart Valve | Instructions for Use
16
Mitral Preoperative Patient Demographics
All patients implanted, N = 229,
Cumulative follow-up = 417.9 patient-years
Patient Characteristic N % (n/N)1
Age at implant in years 59.2 ± 10.6
Gender: • Male 86 37.6%
• Female 143 62.4%
NYHA Clas-
sification:
• I 5 2.2%
• II 68 29.7%
• III 134 58.5%
• IV 18 7.9 %
• Unknown 41.7%
Valve Lesion: • Stenosis 29 12.7%
• Insufficiency 111 48.5%
• Mixed 87 38.0%
• Other 2 0.9%
Aortic Preoperative Patient Demographics
All patients implanted, N = 184,
Cumulative follow-up = 411.8 patient-years
Patient Characteristic N % (n/N)1
Age at implant in years 60.2 ± 8.4
Gender: • Male 121 65.8%
• Female 63 34.2%
NYHA Clas-
sification:
• I 9 4.9%
• II 91 49.5%
• III 79 42.9%
• IV 5 2.7%
• Unknown 00.0%
Valve Lesion: • Stenosis 86 46.7%
• Insufficiency 39 21.2%
• Mixed 59 32.1%
• Other 0 0%
Table 5: Preoperative Patient Demographics
Notes: 1. n = number of patients in each category; N = total number
of study patients.
Notes: 1. n = number of patients in each category; N = total number of study
patients.
Complication Early Events Late Events2Freedom from Event3, % [SE]
n % (n/N)4n %/pt-yr 1 Year Postoperative (n=134) 3 Year Postoperative (n=44)
Mortality (all) 93.9% 92.2% 95.4% [1.4] 89.2% [2.7]
Mortality (valve-related) 10.4% 20.5% 99.5% [0.5] 97.2% [1.7]
Endocarditis 00.0% 3 0.7% 99.0% [0.7] 99.0% [0.7]
Explant 10.4% 3 0.7% 98.0% [1.0] 98.0% [1.0]
Hemolysis500.0% 00.0% 100.0% [0] 100.0%[0]
Hemorrhage6(all) 41.8% 61.4% 96.4% [1.3] 94.4% [2.0]
Hemorrhage (major) 41.8% 20.5% 97.0% [1.2] 97.0% [1.2]
Perivalvular Leak (all) 20.9% 3 0.7% 98.0% [1.0] 97.1% [1.2]
Perivalvular Leak (major) 10.4% 10.2% 99.4% [0.6] 99.4% [0.6]
Nonstructural Valve Dysfunction 00.0% 10.2% 100.0% [0] 99.1% [0.9]
Reoperation (valve-related) 31.3% 51.2% 97.0% [1.2] 97.0% [1.2]
Structural Valve Dysfunction 0 0.0% 00.0% 100.0% [0] 100.0% [0]
Thromboembolism 20.9% 71.7% 97.0% [1.2] 96.3% [1.4]
Thrombosis 00.0% 00.0% 100.0% [0] 100.0% [0]
Notes:
1. Data does not include results from double valve replacement.
2. Late events calculated as linearized rates based on total patient-years.
3. Freedom from event was calculated based on the method of Kaplan-Meier. SE = Standard Error.
4. n = number of patients in each category; N = total number of study patients.
5. Blood studies conducted at a core laboratory established that the valve creates a low level of fully compensated hemolysis typified by an increase in SLDH
with a mean within normal range, a decrease in haptoglobin to below normal in 69% AVR and 65% MVR patients at 1-year, and all other analytes within normal
range.
6. The anticoagulant agents used were reported. The target International Normalized Ratio was 2.5 to 3.5 in AVR and 3.0 to 4.5 in MVR.
Table 4: Mitral Replacement Observed Adverse Event Rates1
All patients implanted, N = 229, Cumulative follow-up = 417.9 patient-years

English
On-X®Prosthetic Heart Valve | Instructions for Use
17
Table 6: Operative Patient Demographics
Notes:
1. Ordered by frequency of occurrence, except for valve size.
2. n = number of patients in each category; N = total number of study patients.
3. May be more than one per patient.
Operative Mitral Patient Demographics
All patients implanted, N = 229,
Cumulative follow-up = 417.9 patient-years
Variable Category1N % (n/N)2
Etiology3Rheumatic 86 37. 6%
Degenerative 62 27.1%
Calcific 36 15.7%
Endocarditis 16 7.0 %
Prosthetic Valve Dysfunction 62.6%
Congenital 41.8%
Other 38 16.6%
Concomitant
Procedures3
None 130 56.8%
Coronary Artery Bypass Graft 44 19.2%
Tricuspid Repair 22 9.6%
Closure of Atrial Appendage 12 5.2%
Mitral Repair 12 5.2%
Maze Procedure 12 5.2%
Septal Defect Closure 8 3.5%
Ventricular Aneurysm Repair 31.3%
Muscularization 20.9%
Tricuspid Replacement 10.4%
Explant of Annuloplasty Ring 10.4%
Pre-existing
Conditions3
Atrial Arrhythmias 137 59.3%
Pulmonary Hypertension 108 46.8%
Systemic Hypertension 88 38.1%
Hyperlipidemia 88 38.1%
Congestive Heart Failure 80 34.6%
Other 77 33.3%
Coronary Artery Disease 67 29.0%
Cigarette Smoker 64 27.7 %
Left Ventricular Dysfunction 47 20.4%
Cerebrovascular Accident 43 18.6%
Diabetes Mellitus 40 17.3%
Angina 38 16.4%
Myocardial Infarction 30 13.0%
Hyperthyriodism 27 11.7%
Chronic Obstructive Pulmonary Disease 25 10.8%
Endocarditis 18 7.8%
Gastroinestinal Ulcer 18 7.8%
Chronic Kidney Failure 13 5.6%
Carotid Artery Disease 12 5.2%
Coronary Artery Bypass Graft 10 4.4%
Cancer 10 4.4%
Previous Mitral Valve Replacement 93.9%
Cardiomyopathy 83.5%
Pacemaker Implant 62.6%
Valve Size 25 mm 33 14.4%
27/2 9 mm 131 5 7. 2%
31/33 mm 65 28.4%
Operative Aortic Patient Demographics
All patients implanted, N = 184,
Cumulative follow-up = 411.8 patient-years
Variable Category1n % (n/N)2
Etiology3Calcific 92 50.0%
Degenerative 51 27.7 %
Rheumatic 24 13.0%
Congenital 18 9.8%
Endocarditis 84.4%
Prosthetic Valve Dysfunction 00.0%
Other 63.3%
Concomitant
Procedures3None 141 76.7%
Coronary Artery Bypass Graft 21 11.4%
Myotomy 10 5.4%
Mitral Repair 52.7%
Aorta Repair or Replacement 42.2%
Tricuspid Repair 10.5%
Muscle Bridge 10.5%
Tricuspid Replacement 00.0%
Explant of Annuloplasty Ring 00.0%
Maze Procedure 00.0%
Closure of Atrial Appendage 00.0%
Ventricular Aneurysm Repair 00.0%
Other 00.0%
Pre-existing
Conditions3Systemic Hypertension 90 48.9%
Hyperlipidemia 83 45.1%
Angina 42 22.8%
Coronary Artery Disease 42 22.8%
Diabetes Mellitus 33 17.9 %
Atrial Arrhythmias 25 13.6%
Left Ventricular Dysfunction 23 12.5%
Congestive Heart Failure 22 12.0%
Myocardial Infarction 12 6.5%
Cerebrovascular Accident 10 5.4%
Carotid Artery Disease 73.8%
Endocarditis 42.2%
Cardiomyopathy 31.6%
Pacemaker Implant 21.1%
Coronary Artery Bypass Graft 1 0.5%
Previous Aortic Valve Replacement 10.5%
Previous Mitral Valve Replacement 00.0%
Other 27 14.8%
Valve Size 19 mm 17 9.2%
21 mm 35 19.0%
23 mm 70 38.0%
25 mm 38 20.6%
27/2 9 mm 24 13.0%

English
On-X®Prosthetic Heart Valve | Instructions for Use
18
Number of Aortic Patients Implanted and Number of Patient-
years by Valve Size
All patients implanted, N = 184,
Cumulative follow-up = 411.8 patient-years
Numbers by Valve size
19
mm
21
mm
23
mm
25
mm
27/29
mm
Total
Number of Patients Implanted 17 35 70 38 24 184
Number of Patient-years 36.9 82.2 151.5 85.9 55.3 411.8
Number of Mitral Patients Implanted and Number of Patient-
years by Valve Size
All patients implanted, N = 229,
Cumulative follow-up = 417.9 patient-years
Numbers by Valve size
25
mm
27/29
mm 31/33 mm Total
Number of Patients Implanted 33 131 65 229
Number of Patient-years 60.2 239.1 118.6 417.9
Aortic Effectiveness Outcomes, Functional New York Heart
(NYHA) Classification1
All patients implanted, N = 184,
Cumulative follow-up = 411.8 patient-years
NYHA Class Preoperative
Assessment
(Nd = 184)
Postoperative Assessments
1 Year
(10-14 Months)
(Nf = 138,
Nd = 129)2
2 Year
(22-26 Months)
(Nf = 66,
Nd = 66)
3 Year
(34-38 Months)
(Nf = 37,
Nd = 36)
N3% (n/
Nd)
n % (n/
Nd)
n % (n/
Nd)
n % (n/
Nd)
I94.9 83 64.3 48 72.7 20 55.6
II 91 49.5 35 27.1 12 18.2 10 2 7.8
III 79 42.9 43 .1 69.1 411.1
IV 52.7 000000
Undetermined40075.4 0025.6
Missing50N/A 9N/A 0N/A 1N/A
Notes:
1. Data does not include results from double valve replacement.
2. Nf = number of patients followed (reproduced from Figure 2); Nd = number
of patients for which NYHA data were collected (not including missing).
3. n = number of patients in each category.
4. Undetermined means data were collected but Class could not be
determined during exam
5. Missing refers to the difference between the number of patients followed,
Nf, and the number of patients for which NYHA data were collected, Nd.
Mitral Effectiveness Outcomes, Functional New York Heart
(NYHA) Classification1
All patients implanted, N = 229,
Cumulative follow-up = 417.9 patient-years
NYHA Class Preoperative
Assessment
(Nd = 229)
Postoperative Assessments
1 Year
(10-14 Months)
(Nf = 134,
Nd = 127)2
2 Year
(22-26 Months)
(Nf = 74,
Nd = 69)
3 Year
(34-38 Months)
(Nf = 44,
Nd = 42)
n3% (n/
Nd)
N % (n/
Nd)
n % (n/
Nd)
N % (n/
Nd)
I52.2 85 66.9 35 50.7 14 33.3
II 68 29.7 29 22.8 24 34.8 22 52.4
III 134 58.5 53.9 57. 2 614.3
IV 18 7.9 0011.4 0 0
Undetermined441.7 86.3 45.8 0 0
Missing50N/A 7N/A 5N/A 2N/A
Notes:
1. Data does not include results from double valve replacement.
2. Nf = number of patients followed (reproduced from Figure 2); Nd = number
of patients for which NYHA data were collected (not including missing).
3. n = number of patients in each category.
4. Undetermined means data were collected but Class could not be
determined during exam
5. Missing refers to the difference between the number of patients followed,
Nf, and the number of patients for which NYHA data were collected, Nd.
Table 7: Number Implanted and Years by Valve Size Table 8: Valve Effectiveness Outcomes

English
On-X®Prosthetic Heart Valve | Instructions for Use
19
Effectiveness Outcomes, Aortic Hemodynamic Results1
All patients implanted, N = 184,
Cumulative follow-up = 411.8 patient-years
Hemodynamic
Parameter
Results by Valve Size
19 mm 21 mm 23 mm 25 mm 27/29 mm
Early Postoperation (< 30 days), Nf
2= 184
Mean Gradient3Nd
4= 20 Nd= 31 Nd= 58 Nd= 33 Nd= 20
•Mean ± SD 11.6 ± 4.5 9.4 ± 3.6 8.4 ± 4.3 7.5 ± 3.8 6.1 ± 2.9
•Min, max 5.6, 21.5 4.0, 18.4 2.0, 26.4 2.1, 18.6 1.0, 11.5
EOA5Nd= 19 Nd= 31 Nd= 57 Nd= 33 Nd= 20
•Mean ± SD 1.4 ± 0.2 1.8 ± 0.3 2.1 ± 0.5 2.5 ± 0.8 2.8 ± 0.4
•Min, max 1.1, 1.9 1.3, 2.4 1.0, 3.6 0.9, 4.3 1.9, 3.5
Regurgitation6Nd= 22 Nd= 40 Nd= 72 Nd= 38 Nd= 24
n7%
(n/Nd)
n %
(n/Nd)
n %
(n/Nd)
N %
(n/Nd)
n %
(n/Nd)
•0 940.9% 14 35.0% 31 43.1% 19 50.0% 937.5%
•1-2+ 12 54.6% 25 62.5% 37 51.4% 19 50.0% 13 54.2%
•3+ 0 10.0% 0 0.0% 22.8% 00.0% 00.0%
•4+ 00.0% 00.0% 00.0% 00.0% 00.0%
•Not available 14.6% 12.5% 22.8% 00.0% 28.3%
1 Year Postoperation, Nf= 138
Mean Gradient Nd= 13 Nd= 22 Nd= 55 Nd= 24 Nd= 16
•Mean ± SD 9.7 ± 2.6 7.7 ± 2.8 6.6 ± 3.0 3.7 ± 2.2 5.6 ± 2.9
•Min, max 5.7, 14.3 3.1, 15.2 2.0, 16.0 0.5, 11.3 1.0, 10.8
EOA Nd= 13 Nd= 22 Nd= 54 Nd= 25 Nd= 16
•Mean ± SD 1.4 ± 0.3 1.9 ± 0.4 2.3 ± 0.6 2.8 ± 0.8 2.8 ± 0.6
•Min, max 0.9, 1.8 1.2, 2.9 1.0, 4.1 0.8, 4.2 2.0, 4.1
Regurgitation Nd= 16 Nd= 28 Nd= 60 Nd= 30 Nd= 21
n % (n/
Nd)
n % (n/
Nd)
n % (n/
Nd)
N % (n/
Nd)
n % (n/
Nd)
•0 425.0% 621.4% 24 40.0% 12 40.0% 523.8%
•1-2+ 11 68.8% 21 75.0% 33 55.0% 16 53.3% 15 71.4%
•3+ 00.0% 00.0% 23.3% 26.7% 14.8%
•4+ 00.0% 00.0% 00.0% 00.0% 00.0%
•Not available 16.2% 13.6% 11.7% 00.0% 00.0%
> 1 Year Postoperation, Nf= 103 (total of 2 yr (66) and 3 yr (37) follow-up)
Mean Gradient Nd= 17 Nd= 29 Nd= 61 Nd= 30 Nd= 18
•Mean ± SD 9.0 ± 3.2 8.1 ± 3.2 6.6 ± 3.1 4.2 ± 2.5 5.5 ± 3.0
•Min, max 2.2, 14.3 3.5, 16.6 2.0, 14.1 0.8, 12.8 1.0, 10.8
EOA Nd= 17 Nd= 29 Nd= 60 Nd= 31 Nd= 18
•Mean ± SD 1.5 ± 0.2 1.8 ± 0.5 2.3 ± 0.7 2.7 ± 0.8 2.9 ± 0.8
•Min, max 0.9, 1.9 0 .7, 2.9 1.4, 4.7 0.8, 4.2 2.0, 4.3
Regurgitation Nd= 20 Nd= 37 Nd= 68 Nd= 36 Nd= 25
n % (n/
Nd)
n % (n/
Nd)
n % (n/
Nd)
N % (n/
Nd)
n % (n/
Nd)
•0 525.0% 924.3% 27 39.7% 17 47. 2% 728.0%
•1-2+ 12 60.0% 25 67. 6% 37 54.4% 16 44.4% 17 68.0%
•3+ 2 10.0% 0 0.0% 34.4% 25.6% 14.0%
•4+ 00.0% 00.0% 00.0% 12.8% 00.0%
•Not available
15.0% 38.1% 11.5% 00.0% 00.0%
Effectiveness Outcomes, Mitral Hemodynamic Results1
All patients implanted, N = 229,
Cumulative follow-up = 417.9 patient-years
Hemodynamic
Parameter
Results by Valve Size
25 mm 27/29 mm 31/33 mm
Early Postoperation (< 30 days), Nf
2= 216
Mean Gradient3Nd= 31 Nd= 117 Nd= 59
•Mean ± SD 4.3 ± 1.3 4.3 ± 1.6 4.5 ± 2.2
•Min, max 1.7, 7.5 1.2, 10.0 1.0, 11.7
EOA5 Nd= 25 Nd= 97 Nd= 53
•Mean ± SD 2.4 ± 0.8 2.2 ± 0.6 2.2 ± 0.8
•Min, max 0.9, 4.2 1.0, 4.3 0.8, 4.4
Regurgitation6Nd= 28 Nd= 104 Nd= 56
n % (n/
Nd)
N % (n/
Nd)
N % (n/
Nd)
•0 20 71.4% 73 70.2% 40 71.4%
•1-2+ 414.3% 25 24.0% 16 28.6%
•3+ 00.0% 00.0% 00.0%
•4+ 00.0% 00.0% 00.0%
•Not available 414.3% 65.8% 00.0%
1 Year Postoperation, Nf= 134
Mean Gradient Nd= 18 Nd= 79 Nd= 30
•Mean ± SD 3.7 ± 2.0 4.4 ± 1.8 4.0 ± 1.5
•Min, max 1.7, 7.5 1.7, 10.0 2.0, 7.1
EOA Nd= 15 Nd= 70 Nd= 28
•Mean ± SD 2.1 ± 0.6 2.1 ± 0.6 2.1 ± 0.6
•Min, max 1. 2, 3.1 0.9, 4.0 1.4, 4.3
Regurgitation Nd= 15 Nd= 66 Nd= 29
n % (n/
Nd)
n % (n/
Nd)
N % (n/
Nd)
•0 11 73.3% 53 80.3% 23 79.3%
•1-2+ 320.0% 11 16.7% 620.7%
•3+ 16.7% 11.5% 00.0%
•4+ 00.0% 00.0% 00.0%
•Not available 00.0% 11.5% 00.0%
Table 9: Effectiveness Outcomes, Hemodynamic Results1
Notes:
1. Hemodynamic evaluations were performed using transthoracic echocardiography (TTE) and in some cases, transesophageal echocardiography (TEE). Data
does include results from double valve replacement.
2. Nf= number of patients followed (reproduced from Figure 2).
3. Mean gradient represents the pressure drop measured across the valve in mmHg.
4. Nd= number of patients for which hemodynamic data were collected.
5. EOA = effective orifice area measured in cm2.
6. Regurgitation represents the valvular backflow of blood due to normal leakage and perivalvular leakage; 0 = none, 1+ = mild, 2+ = moderate,
3+ = moderate/severe, 4+ = severe.
7. n = number of patients in each category.

On-X Life Technologies, Inc.
1300 East Anderson Lane, Bldg. B Austin, Texas 78752 U.S.A.
TEL: (512) 339-8000 FAX: (512) 339-3636
CryoLife Europa, Ltd.
Bramley House, The Guildway, Old Portsmouth Road
Guildford, Surrey, GU3 1LR, United Kingdom
010122 103 A (06/2017)
Moist Heat (Steam)
This manual suits for next models
7
Table of contents
Other On-X Medical Equipment manuals
Popular Medical Equipment manuals by other brands

DEHACO
DEHACO NANO user manual

Pari
Pari Trio Instructions for use

Orliman
Orliman TP-6101A INSTRUCTIONS FOR USE AND PRESERVATION

MTS Systems
MTS Systems 111 Series Product information

Shenzhen Mindray Bio-Medical Electronics
Shenzhen Mindray Bio-Medical Electronics BeneHeart D6 Operator's manual

Ferno
Ferno Kwik Klip 031-4002 quick start guide