- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
Equipment Type: Digital Stereo Microscope
Model: VM-500
Electrical Requirements: 110/220 V
Frequency: 50/60 Hz
Manual Revision Date: June 2021
Noise Levels: >40 dB Idle
>50 dB Working

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
Contents
PAGE
Warranty i
1.0 Safety Guidelines 1
2.0 Product Description 4
3.0 Shipping, Unpacking, and Installation 6
4.0 Steps for Observation 7
5.0 Adjusting Features 8
6.0 Image Software Operation 10
7.0 Trouble shooting 14

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
WARRANTY
Terms and Conditions applying to all PACE Technologies Products
1. LIMITED WARRANTY AND DISCLAIMER:
PACE Technologies microscopes and hardness testers are warranted for one year from the purchase
date to be free from defects in material and workmanship under correct use, normal operating
conditions, and proper application. PACE Technologies obligation under this warranty shall be limited
to the repair or exchange, at PACE Technologies option, of any PACE Technologies Product or part
which proves to be defective as provided herein. PACE Technologies reserves the right to either
inspect the product at Buyer’s location or require it to be returned to the factory for inspection. Buyer is
responsible for freight to and from factory on all warranty claims. The above warranty does not extend
to goods damaged or subjected to accident, abuse or misuse after release from PACE Technologies
warehouse, nor goods altered or repaired by anyone other than specifically authorized PACE
Technologies representatives. PACE Technologies shall not in any way be responsible for the
consequences of any alteration, modification or misuse unless previously approved in writing by an
officer of PACE Technologies. Note: Corrosion is considered a maintenance issue and not a warranty
issue.
PACE TECHNOLOGIES MAKES NO EXPRESS WARRANTIES OTHER THAN THOSE WHICH
ARE SPECIFICALLY DESCRIBED HEREIN. Any description of the goods sold hereunder, including
any reference to Buyer’s specifications and any description in catalogs, circulars and other written
material published by PACE Technologies, is the sole purpose of identifying such goods and shall
not create an express warranty that the goods shall conform to such description.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED. THERE ARE NO IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR
PARTICULAR PURPOSE. THIS WARRANTY STATES PACE TECHNOLOGIES ENTIRE AND
EXCLUSIVE LIABILITY AND BUYER’S EXCLUSIVE REMEDY FOR ANY CLAIM FOR DAMAGES
IN CONNECTIONS WITH PACE TECHNOLOGIES PRODUCTS. PACE TECHNOLOGIES WILL IN
NO EVENT BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES WHATSOEVER,
NOR FOR ANY SUM IN EXCESS OF THE PURCHASE PRICE.
2. LIABILITY CAP:
PACE Technologies maximum aggregate liability for loss and damage arising under, resulting from or in
connection with the supply or use of the Equipment and Consumables provided under this purchase, or from
the performance or breach of any obligation (s) imposed hereunder, whether such liability arises from any
one or more claims or actions for breach of contract, tort, (including negligence), delayed completion,
warranty, indemnity, strict liability or otherwise, unless otherwise limited by the terms hereof, shall be
limited to one hundred percent (100%) of the purchase price.
3. DELIVERY:
Customer assumes and shall bear the risk of all loss or damage to the Products from every cause whatsoever,
whether or not insured, and title to such Products shall pass to Customer upon PACE Technologies delivery
of the Products to the common carrier of Pace Technologies choice, or the carrier specified in writing by
Customer, for shipment to Customer. Any claims for breakage, loss, delay, or damage shall be made to the
carrier by the Customer and Pace Technologies will render customer reasonable assistance in prosecuting
such claims.
4. ACCEPTANCE:
Customer shall inspect the Products promptly upon receipt of delivery. Unless customer objects in writing
within thirty (30) business days thereafter, customer shall be deemed to have accepted the Products. All
i

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
5. PAYMENT:
Customer agrees to provide timely payment for the Products in accordance with the terms of payment set
forth on the reverse side hereof or in any proposal submitted herewith. If any payment is not paid on or
before its due date, Customer shall pay interest on such late payment from the due date until paid at the
lesser of 12% per annum or the maximum rate allowed by law.
6. DEFAULT:
If Buyer is in default (including, but not limited to, the failure by Buyer to pay all amounts due and payable
to Seller) under the work or purchase order or any other agreement between Buyer and Seller, Buyer’s rights
under the warranty shall be suspended during any period of such default and the original warranty period
will not be extended beyond its original expiration date despite such suspension of warranty rights.
7. MISCELLANEOUS PROVISIONS:
This agreement has been made in and shall be governed by the laws of the State of Arizona. All disputes
arising under or relating to the purchase of the equipment shall be brought and resolved solely and
exclusively in the State of Arizona, Pima County . These terms and conditions and the description of the
Products on the reverse side hereof or in any proposal submitted herewith constitute the entire agreement
and understanding of the parties with respect to this sale and supersede all prior and contemporaneous
agreements or understandings, inducements or representations, expressed or implied, written or oral,
between the parties with respect hereto. Any term or provision of this Agreement may be amended, and
any observance of any term of this Agreement may be waived, only by a writing signed by the party to be
bounds. The waiver by a party of any breach shall not be deemed to constitute a waiver of any other
breach. Should suit be brought on this Agreement, the prevailing party shall be entitled to recover its
reasonable attorneys’ fees and other costs of suit including costs and attorneys’ fees incurred on appeal or
in collection of any judgment., errors, or shortage in Products delivered shall be made by Customer in
writing within such five (5) business day period. Failure to make any claim timely shall constitute
acceptance of the Products.
8. RESTOCKING FEE:
All Returns are subject to a restocking charge equal to 15% (fifteen percent) of the Invoice, unless the
Goods are proved to be non-conformed by PACE Technologies.
ii

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
1
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
1.0 Safety Guidelines
1.1 Warning Sign
This sign points to special safety features on the machine.
!
1.2 Safety Precautions
!Careful attention to this instruction manual and the recommended safety guidelines is
essential for the safe operation of the VM-500.
!
Proper operator training is required for the operation of the VM-500. Any unauthorized
mechanical or electrical modifications made to the VM-500, as well as improper
operation, voids all warranty claims. All service issues need to be reported to the
manufacturer or supplier.
!Operate unit as specified in this manual.
!Disconnect from power before opening unit.
!Ensure that any air slots on the machine remain unobstructed.
1.3 Emergency Statement
Do not attempt to service the machine without expert help. Contact your PACE representative
before attempting repairs. Failure to do so could result in breaking the warranty provided with
the machine.
The VM-500 is a digital stereo microscope designed for capturing images of samples under
light and magnification. Be mindful of surroundings and follow all operational guidelines. In the
event of an injury, report to appropriate medical service.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
2
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
2.0 Product Description
Housing
Frame
Illumination
Adjustment
Coarse Focus
Power Outlet
Base
Light Source
Fine Focus
2.1 General Description
The VM-500 digital stereo microscope with autofocusing and built-in measurement capabilities has
0.7X ~ 4.5X electronic magnification with a zoom ratio of 6.5:1. High-resolution photo and video
recordings are easy to obtain (includes 22-inch Full HD Widescreen LCD monitor or better) The VM-
500 digital stereo microscope is ideal for low magnification documentation of metallographic specimens
prior to polishing and grinding. It is also very useful for measuring cracks, weld penetration depth, and
other topographical features.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
3
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
2.2 Features and Benefits
The VM-500 is a high-definition Auto-Focusing Digital Stereo Microscope that is capable of
capturing optimal images of targeted areas. The high-definition camera communicates
automatically with the included monitor for better viewing than traditional microscopes. The built-in
image processing software has easy-to-use mouse controls that make the image capturing
process easier and hassle-free. Use the screen image capture capability to create detailed
pictures for reporting results.
High-definition close-up
of a screen crack taken
on VM-500

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
4
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
2.3 Technical Specifications
Optical Objective: 0.7 X ~ 4.5X Horizontal zoom lens
Zoom ratio: 6.5:1
Camera Parameters: 2 million pixels (1920 x 1080)
Image size: 1/2.86-inch
Frames per second: 60 fps
Camera Function: White balance
Brightness control
Digital noise reduction
Photograph and video recording.
HDMI/USB Output
USB storage
OSD: Comprehensive digital UI design
Wireless USB mouse/keyboard combination
Measuring Function: Measuring software can measure image in the field of view
Illumination: Bottom: Adjustable LED Illumination
Surface: Adjustable LED Illumination
Microscope Stand: Z-axis Travel: 150 mm Coarse/fine lifting system
Electrical Parameters 110~240 V; 50~60 Hz
Working Temperature: 41 - 104°F (5 - 40°C)
Maximum Relative Humidity: 88°F (31°C) - 80%
93°F (34°C) - 70%
99°F (37°C) - 60%
104°F (40°C) - 50%
Environmental Information: Voltage fluctuation: ±10%
Pollution level: 2 (according to IEC664)
Classification of installation voltage: II (IEC664)

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
5
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
3.0 Shipping, Unpacking, and Installation
3.1 Shipping The VM-500 is packaged in a custom box with securing foam and a
horizontal crossbar. The foam and box may be discarded after the
removal of the machine.
! Caution: Heavy sensitive electronic equipment. Take care to
avoid bodily injury and damage to the unit.
3.2 Unpacking When moving box, lift from bottom.
Measures WxHxD: 13.5 x 24 x 13-inch (343 x 610 x 330 mm)
Weight: 35 lbs (16 kg)
3.3 Installation The VM-500 should be installed on a flat and sturdy surface. Extra
care should be taken to ensure the surface is secure enough to
reduce any vibrations or wobbling. Excess outside motion can
reduce image quality.
Locate a power outlet within range specified under Electrical
Requirements in the previous section titled Technical
Specifications. Plug in the supplied power cable to the wall outlet
and secure the other end into the back of the VM-500.
Place the monitor next to the VM-500. Follow the instructions
located inside the monitor box to ensure proper installation.
Plug the HDMI into the back of the monitor and then secure into
the back of the VM-500.
Install the two-pronged power cable into the back of the VM-500.
Plug in the wireless keyboard USB receiver into the back of the
unit.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
6
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
Backside Connections
Back Side
Bottom
HDMI Port
USB Port (Keyboard and Mouse)
Circular Power Plug
USB Port (Memory)
HDMI Port
Two Prong Power

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
7
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
4.0 Steps for Observation
1. Turn on VM-500 using main power switch on back of unit.
2. Adjust light source using Surface and Contour Illumination switches.
3. Place sample onto the stage. Secure specimen using Sample Clips if needed.
4. Use Coarse Adjustment knob to find general focus.
5. Use Fine Adjustment knob to find final focus. Adjust light again if needed.
6. Zoom in using the Magnification switch.
7. Fine Adjustment might be needed again to focus sample.
8. Observe sample.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
8
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
5.0 Adjustment Features
5.1 Illumination Adjustment
5.2 Specimen Placement
The sample clips are for xing the object being observed, and can eecvely prevent vibraons in the
environment from moving the image. Steps for securing specimen:
1. Loosen both spring screws.
2. Place sample on stage and posion in desired locaon.
3. Posion both sample clip arms onto sample.
4. Tighten sample clip arms using spring screws.
TIP: Use gloves to avoid leaving nger prints
on samples when adjusng posion on stage.
The illumination adjustment knob is located on the stem of the VM-500, near the backside of the
machine. See illustration on page 6 for a diagram.
Adjust light levels by turning the knob. Adjust light as necessary for best picture quality.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
9
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
5.3 Adjusting Focus
Do not use force movement, knobs should operate without applying excessive force.
!
NOTE: Ensure Coarse Adjustment Knob
Tension Ring is tensioned
(VM-100 ships already tensioned
for ease of use)
1. After placing sample on stage, move
the Coarse Adjustment knob until the
object is in focus on the LCD monitor.
2. If the object is too far or too close,
adjust the Lens Height Adjustment
screw and repeat step 1
IMPORTANT:
Knob is securely fastened to avoid slipping on the
Column. Slipping can cause a collision with the
stage causing serious damage to the VM
the Limit Fixing Screw to help avoid this.
5.3.1 Auto Focus
The VM-500 has an automatic focusing feature. This is enabled inside the side menu screen.
See the following section titled “Interface” for more information on this feature’s activation.
1. Put the microscope into Autofocus mode. (AF option under the Control Tab)
2. Place sample onto the stage.
3. Use the Coarse Focus knob to get the sample into view (it may still be slightly blurry at this
point).
4. Allow the VM-500 to find a focus spot.
5. Move the sample as needed for observation.
5.3.2 Manual Focus
The VM-500 allows manual focusing as well. This feature uses the coarse focus knob in
conjunction with the MF slider under the control tab.
1. Put the microscope into Manual mode. (MF option under the Control Tab)
2. Place sample onto the stage.
3. Use the Coarse Focus knob to adjust the sample into view.
4. Use the MF slider to fine focus the image.
5. Move sample and adjust as necessary.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
10
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
6.0 Image Software Operation
The VM-500 is provided with an HD monitor for use
with the microscope. Many of the features are
accessible through the user interface. The following
sections will provide information on these features.
6.1 Activating the User Interface
Move the mouse to the upper left corner of the
screen to have the user interface appear.
6.1.2 Locking the User Interface
The Focus Settings are located under the Control
Tab. Use these to adjust how the focus behaves.
Locked User Interface Unlocked User Interface
6.2 Control Settings
6.2.1 Focus Settings
Move the mouse to the upper left corner of the
screen to have the user interface appear.
Click AF to enter autofocus mode. This mode allows
the VM-500 to automatically focus on the image. The
Click F mode is a one-click autofocus. This mode
works similar to the AF mode. However, the
microscope will only focus once a click is registered.
The MF mode is for manual focusing. Use this to
manually control the focus of the machine. There is a
slider located above these buttons which is used to
fine-tune the focus for MF mode.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
11
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
6.2.2 Auto Focus Region of Interest
The VM-500 has a region of interest feature for the autofocus mode. Use this to select a spot
on the sample that requires the focusing. This is useful if the sample surface is uneven, or if
the region of interest is near the edge of a part.
To activate ROI, click on the “ROI” toggle shown in the above image. Use the
dropdown to adjust the size of the ROI. Adjusting the Hor and Ver sliders will adjust the
horizontal and vertical box boundaries respectively. Select Set to save the parameters.
Click anywhere on the image to select the ROI. A red box will appear indicating where
the camera will now try to focus.
6.2.3 Exposure Settings
Use these settings to adjust the Exposure, Brightness, Gain, and Shutter for the camera.
Adjusting the sliders will modify the values. AE is the auto exposure feature. This will allow
the computer to automatically select the most optimal exposure for the image. Changing
this to ME (manual exposure) will allow the user to adjust the exposure as needed.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
12
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
6.2.4 Color Menu
Use these settings to adjust the White Balance and Color for the camera.
AWB is auto white balance. This setting will allow the computer to select the most optimal
white balance for the image.
MWB is manual white balance. This setting will allow the user to manually adjust the white
balance of the image. Each of the color sliders will add a color filter to the image.
6.2.5 Function Menu
Use these settings to adjust Contrast, Sharpness, Saturation and Ezoom.
Value Range Default Value
Contrast 0-15 0
Sharp 0-15 4
Saturation 0-254 128
Ezoom 0-50 0
The default values and ranges for these settings are as follows:

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
13
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
6.2.6 Crosshair Tool
This tool allows the user to place a crosshair overlay on the image. The Cross Line tool can be used
to place several grid lines across the image. Click and drag to move the cross lines around the
screen. Double click the line to set its color. Click Hide to remove these lines.
6.3 Image and Video Capturing
6.3.2 Storage Device
Ensure that there is a USB slotted into the VM-500. This allows the operator to save the images and
videos taken on the VM-500. The USB indicator will display when there is a USB detected.
USB indicator
6.3.3 Setting the File Path
The line labeled “Photo Path” is where the images and videos will be saved on the storage device.
Edit this to change directory paths.
6.3.4 Changing Image Format
The dropdown menu shows the exportable file formats available.
6.3.5 Image Naming
The operator can use the default naming scheme to save the images and videos. There is also an
option to use custom naming each time. Select Custom and a prompt will show each time image
naming is needed.

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲ INSTRUCTION MANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
14
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
VM-500 Auto Focus Digital
Stereo Microscope
NO. Problems Cause Solutions
1Poor visibility
Image is blurry
Dirty Lens
Surface dust or dirty
Focus is not correct
Clean the lens (see manual
and maintenance)
Cleaning surfaces
Adjust the focus
2Head moves when
adjusting focus
Focus wheel is too
loose
Appropriately lock Coarse
Adjustment Knob Tension
Ring
3Power light is not
working
Power supply
connection is bad Check the power connections
4No image on the
monitor VGA cable Check the VGA cable
connection
5Incorrect image zoom Monitor resolution
settings
Moving the mouse pointer to
the edge of the underside of
the monitor (middle) button,
select the appropriate
resolution.
6Sample not displaying
on monitor
Samples are not
placed correctly on
the stage/sample
surface is not vertical
to the lens
Place samples in view on
loading plate and/or use
sample clips .
7Hard to turn focus
knobs
Tension becomes
very tight
Relax Coarse Adjustment
Knob Tension Ring
7.0 Trouble Shooting
Use this page to help with some issues that an operator could be experiencing. If this page is
not helping solve an issue, call a PACE Technologies technical assistant.