
2.4 Glossary of terms
• Backlight: The LCD display backlight has three
user-adjustable parameters. The brightness, the
contrast, and the length of inactivity which triggers
automatic shutdown can be individually set within
available limits.
• Error handling: The manner in which various
XMT1000 outputs respond to automatically generated
system errors can be set by the user by selecting the
options from a drop-down list.
• Frequency output: In frequency mode, the digital
output generates a pulse with a frequency that is
proportional to the magnitude of the measured
flow parameter (e.g., 10 Hz = 1ft3/hr).
• HART: HART is a serial communication protocol
used for industrial automation. Its main advantage
is that it can communicate over 4-20 mA analog
instrumentation current loops over the pair of wires
used by the analog only host system.
• K-factor: To calibrate the XMT1000 against a
reference value, the raw flow velocity readings
may need to have a correction factor applied.
This K-factor can be a single constant or a table of
K-factor vs. velocity values.
• Loop powered: Loop powered analog devices
use the power provided by the meter itself rather
than by an external power supply. This results
in a simple 2-wire connection between the meter
and the external device.
• Mass flow: mass flow is the mass of fluid passing
the measurement point in a given period of time.
This value is calculated by the meter from the
measured flow velocity and the programmed pipe
and fluid parameters.
• Modbus: Modbus is a serial communication
protocol developed by Modicon® for use with its
programmable logic controllers. It is a method
used for transmitting information over serial lines
between electronic devices.
• Pulse output: In pulse/totalizer mode, the digital
output generates a pulse with a width corresponding
to one unit of the measured flow parameter
(e.g., 1 pulse = 1 ft3).
• Reynolds correction: When turned on, a factor
based on the kinematic viscosity and the flow rate
of the fluid is applied to all measurements. Reynolds
correction should be On for most applications.
• Serial port settings: When the XMT1000 serial
port is connected to an external serial device,
communication between the two devices only
occur if the serial port settings of the two devices
match. These settings include: baud rate, data bits,
stop bits, and parity.
• Special transducer: If your transducer has no
number engraved on the body, select SPECIAL
in the transducer number menu.
• Tw: The TW parameter is the time the transducer
signal spends traveling through the transducer
body and the transducer cable. This value must
be subtracted from the total signal transit time
to calculate the actual signal transit time only
through the fluid.
• Totalizer: The totalizer accumulates a running
total of the amount of fluid which passes the
measurement point between specified start and
stop times.
• Tracking window: The tracking window is used
to detect the receive signal when you are unsure
of the fluid sound speed. For the XMT1000 flow
transmitter, the tracking Window is always ON.
• Zero cutoff: When the measured flow rate is
below the zero cutoff value, the display is forced
to 0.00. This is to avoid rapid fluctuations in the
reading whenever the flow rate is close to zero.
11