Parker Balston NitroFlow60 User manual

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
1
1-800-343-4048
NitroFlow60
and NitroFlow60D
Installation, Operation, and Maintenance Manual

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
2
1-800-343-4048
Symbol Description
Caution, refer to accompanying documents
for explanation.
Refer to the caution note indicated for
explanation.
Caution, risk of electric shock.
Refer to the warning note indicated for
explanation.
1
1
Explanation of Warning Symbols

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
3
1-800-343-4048
These instructions must be thoroughly read and
understood before installing and operating this product.
Any modification of the product will void the warranty.
Failure to operate this product in accordance with the
instructions set forth in this manual and other safety governing
bodies could jeopardize the safety of the operator and void the
safety certication of this product. Retain these instructions for
future reference.
If you have any questions, please call the Technical services
department at 1-800-343-4048, 8AM to 5PM Eastern Time or
e-mail at balstontechsupport@parker.com. For other locations,
please contact your local representative.
Please refer to page 2 for an explanation of the caution/warning
symbols used throughout this manual.
Installation, Operation, and Maintenance Manual
Parker Balston
Nitroflow60 Generator Series
Figure 1 Model NitroFlow60 & NitroFlow60D Nitrogen Generator

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
4
1-800-343-4048
The NitroFlow60 Nitrogen Generator System Series are available in two models (Figure 1):
(1) NitroFlow60 Nitrogen Generator System(No Dry Air output)
(2) NitroFlow60D Nitrogen Generator System(Dry Air output)
The integrated system is engineered to produce pure nitrogen gas and clean, dry -40°F dew point
air gas from compressed air delivered by a carefully matched scroll compressor. The system is
housed in two fully integrated cabinets: the lower cabinet produces the compressed air using an
oil-less rotary scroll compressor. The upper cabinet removes contaminants and separates the
compressed air into the two gases.
General Description
The generator is certied to the electrical safety requirements as specied by the IEC, UL,
and CSA standards. These units bear the CSA marking on the product. Product supplied
internationally also carries the CE mark. The product meets EMC compliance. Internal pressure
receivers are CRN and ASTM certied or CE certied.
Regulatory
Compliance
The NitroFlow60 generator contains the necessary components to produce reliable, oil-free
compressed air, pure nitrogen, and dry air gases. The system consists of ve functional
technologies: (1) compressed air, (2) liquid water removal, (3) preltration, (4) oxygen/carbon
dioxide/water vapor removal, and (5) nal ltration. The nitrogen generator ow schematic (gure
2) shows the airow through the intake lter, compressor pump, aftercooler, receiver tank, and
check valve. The compressed air then ows through a water separator and preltration where
additional water (liquid and vapor) is removed. After ltration, the compressed air ow passes
through a 3-way pneumatic valve, an activated carbon prelter, nitrogen module, pressure
regulator, and post carbon lter, providing the nitrogen gas. The second ow path passes through
a back pressure regulator to maintain a 100% duty cycle of the compressor. The third ow path
passes through a dryer membrane providing dry air. Dry air available on NitroFlow60D series only.
Integrated System
The oil-less rotary scroll compressor is based on the theory of scroll compression. It offers high
reliability and quiet operation. The scroll set consists of two identical spirals; offset 180° with
respect to the other so the scrolls mesh. One scroll is orbited around the xed scroll, trapping and
compressing gas pockets as the pockets move to the center of the xed scroll. The compressed
gas is discharged from the pump through the center outlet.
The air-cooled aftercooler on the scroll compressor is a series of cooling ns and a high output
fan. The cooling features allow the scroll compressor to operate at lower temperatures and
extend bearing, tip seal and grease life. The aftercooler cools the temperature of the discharge
air. The temperature of the discharge air is lowered, producing trapped water in the tank. The
water is then discharged through the automatic electric drain valve and collected in a disposal
container.
Compressor Pump
Compressor
After Cooler
Figure 2, Nitrogen Generator Flow for NitroFlow60 and NitroFlow60D

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
5
1-800-343-4048
The NitroFlow60 generator series incorporates a water separator to remove liquid water from the
compressed air before it reaches the prelters.The liquids are collected in the supplied disposal
container. Refer to the included installation and service instructions, 2F101H, for maintenance
instructions.
Two stages of high efciency coalescing preltration are incorporated into the design of the
NitroFlow60 to remove water and particulate contamination to 0.01 micron. The generator is
equipped with oat drain valves to automatically discharge any liquids which have accumulated in
the lter bowls. The liquids are collected in the supplied disposal container. These lters protect
the membrane and associated components from contamination.
A 3-way pneumatic valve is located after the prelters to release any leftover liquid water that may
be present on startup.
An activated carbon module is located before the nitrogen module to remove any chlorinated
hydrocarbon contaminates that may exist in the room air. A particulate lter is located after this
carbon module to remove any carbon contaminates
The NitroFlow60 generator has a hollow ber membrane module which converts compressed
air into high purity nitrogen. The membrane module has no moving parts assuring years of
maintenance-free operation.
The water vapor in the compressed air passes through the hollow bers of the membrane for
removal, resulting in dry air. A small portion of the dry air is redirected along the bers to sweep out
additional water vapor laden air and vented to atmosphere in the cabinet. A dew point of -40°F is
achieved at a pressure of 100 psig and ow rate of 4 slpm. Available with NitroFlow60D only.
Prefilters
Water Separator
Inlet Carbon
Module and Filter
3-way Pneumatic Valve
Nitrogen Module
Dryer Membrane
There is one nal activated carbon module and particle lter located at the outlet gas that protects
the gas purity from environmental contaminants to assure medical/research grade is entering the
instrument.
There are two diagnostic indicator lights on the Parker Balston NitroFlow60 System:
High Temperature LED (red): The “High Temperature” LED indicator illuminates when the
compressor pump has reached its maximum temperature. The compressor will shut down by a
thermal switch opening, preventing power to the motor. Compressor restarts automatically as
soon as the over-temperature condition is corrected.
Power On LED (Green): This indicates the compressor is operating.
Outlet Carbon
Module and Filter
Diagnostics
Installation
All installation, operation, and maintenance activities for the NitroFlow60 Generator System
Series (NitroFlow60 and NitroFlow60D) should be performed by trained personnel using
reasonable care to avoid injury or damage.
The NitroFlow60 system has casters and is packed in a wooden shipping container. The
installation kit, consisting of plastic tubing, tube ttings, and a plug is packed in a box inside
the wooden container. Remove the wire L-shaped fasteners from the sides of the container
(see Figure 3). Do not attempt to lift the generator from the container. Remove and locate the
plywood ramp (see Figure 4). Release the locked casters. Move the generator down the ramp.
Move the generator and installation kit to its installation location.
Figure 3 Figure 4
Unpacking

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
6
1-800-343-4048
Location of the generator with proper ventilation is critical. Install the generator in a clean, well
ventilated area. To provide adequate ventilation and servicing for the generator, a minimum
clearance distance of 30 inches in front and in the rear is required (see Figure 5). Fresh intake
ventilation for the generator enters from the bottom and sides, while warm exhaust air exits from
the right, rear side. A minimum of 3 inches clearance is required on both sides of the generator
(see Figure 6).
For the NA Model: The NitroFlow60 Generator Series requires one electrical power outlet:
a dedicated grounded 208 to 254VAC, 60Hz, single phase, minimum 30 Amp circuit breaker
with a NEMA L6-30R receptacle (see Figure 7). Parker requires purchasing a step-down
transformer if your facility voltage is 255 or higher or a step-up transformer if your voltage
decreases below 208VAC. See the accessory section on page 17. Main supply voltage
uctuations must be within ±10% of the nominal main supply voltage.
For the JA and WD Models: The NitroFlow60 Generator Series, rated at 230V, 50 Hz, single
phase, 13A fully loaded installation requires the following electrical setup:
1The supply line should be dedicated to the generator. No other equipment should operate
using the same line.
2The service should be capable of carrying 20 amperes. The wire gauge should be sized
accordingly.
3The service should be protected by a minimum 20 amperes circuit breaker. The value is based
on the application/equipment requirements as well as the electrical components’ tolerance
and uncontrollable external variables.
The generator is for indoor use only. Do not install outdoors. The ambient temperature of
the air surrounding the generator should not exceed 90°F (32°C) or be below 60°F (15°C). If the
generator is located in a totally enclosed room without ventilation, an exhaust fan with access to
outside air must be installed. Do not locate the generator where hot exhaust air from other heat
generating sources may be drawn into the generator. The area should be free of excessive dust,
toxic or ammable gases, solvent fumes, and moisture. An exhaust vent attachment is available
to route exhaust away from generator(see Figure 5). The diameter of the exhaust vent ducting
must be six inches. An in-line duct fan, capable of exhausting 226 CFM, is required for ducting
greater than 15 feet. See Accessory Parts table on Page 17 for details.
The Parker Balston NitroFlow60 Generator Series creates minimal noise during operation. The
noise generated is about 49dB at one meter. Periodically there is an air discharge noise from the
drain port eliminating accumulated uids in the prelters and receiver tank. The noise and heat
generated by the unit should be considered when selecting an installation location. A silencer
kit is included for the drain discharge noise. There is also a pump “burp” noise lasting a few
seconds each time the generator is shut down. This is the compressor depressurizing the pump
during shutdown.
Figure 5 Figure 6
Location
Utilities

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
7
1-800-343-4048
System Comments Region
NitroFlow60NA 230V-60Hz, cordset has U.S. plug North America
NitroFlow60DNA 230V-60Hz, cordset has U.S. plug North America
NitroFlow60WD 230V-50Hz, cordset matched to region Rest of world
NitroFlow60DWD 230V-50Hz, cordset matched to region Rest of world
NitroFlow60JA200 230V-50Hz, cordset has a JA PSE plug Japan
NitroFlow60DJA200 230V-50Hz, cordset has a JA PSE plug Japan
The generator drain tube is a black nylon ¼” tubing connecting the drain outlet to the provided
disposal container (see Figure 8). A blue silencer is provided for the discharge end of the tubing
to reduce the discharge noise. Periodically check the built in strainer to ensure it has not clogged.
The pre-lter and water separator drain tubes are also a black nylon ¼” tubing connecting the
drain outlets to the same provided disposal container (see Figure 9). Blue silencers are provided
for each drain line.
The two outlet gas ports consist of two ¼” NPT bulkhead ttings (NitroFlow60D only. See Figure 10).
The non dry-air version consists of one ¼” NPT bulkhead tting and a cap on the second outlet port.
Parker Balston recommends minimum 3/8” PFA plastic tubing with a minimum pressure rating of
125 psig (8.6 bar). Check all piping connections for leaks using a bubble-type leak detecting
solution.
Drains
Outlet Gases
Figure 8
There are six plastic tubing connections for the NitroFlow60 generator: (1) nitrogen outlet gas
connection; (1) dry air outlet gas connection (NitroFlow60D only); (4) drain connections. Press
the tubing into the quick-connect ttings. Assure tubing is fully inserted (about ½”). Tubing is
provided in the installation kit. For remote locations, additional length tubing is provided.
Piping
The following table describes the necessary cordset based on region. Do not replace the provided
cordset with an inadequately rated cordset. The plug on the cordset is the mains disconnect.
The electrical schematic for the generator is shown in Figure 34 on page 22.
Notes
- “D” designation indicates additional dry air supply for chip cube requirements
- JA-200 versions include an external step-up transformer and gages indicating pressure in kPa
- WD(World Design) requires assignment of appropriate cordset, see table below
A03-0430 NA.............................................................30A, 250V
A03-0344 EU, IND, UK (ROW 230V,50Hz)...............32A, 250V
A03-0345 AU.............................................................32A, 250V
A03-0407 JA..............................................................30A, 250V
A03-0347 CHINA.......................................................32A, 250V
Cordset Rating
Figure 7, NEMA L6-30R 30AMP/250V
Receptacle for Model: NA (North America)

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
8
1-800-343-4048
A03-0286 voltage step-down transformer - If your facility supply voltage is 255V or higher,
then the use of this transformer is required.
A03-0281 voltage step-up transformer - If your facility supply voltage is 195V or lower, then
the use of this transformer is required.
Recommended
Accessories
Figure 11 Figure 12
Check that the power cord of the generator is properly connected to the generator receptacle.
Rotate clockwise to lock. Check that the power cord of the generator is connected to the proper
power supply outlet.
During normal operation, the generator’s power switch is illuminated green. The compressor
pressure gauge indicates a pressure between 115 to 120 psig (7.9 to 8.2 barg). If the pressure
drops below 115 psi, then the nitrogen ow is too high. Maximum nitrogen pressure should
not exceed 100 psi at the maximum ow rate of 60 slpm. The compressor operates on a
100% duty cycle (pump on continuously).
The generator has its own automatic water drain system. The 3 lters located on the back of
the generator are equipped with pneumatic drains that open when the water level reaches a
predetermined point. Compressor drain operate every 10 minutes, opens for 2 seconds; the two
internal lter bowls also drain every 10 minutes.
The generator cabinet has two LED indicators (see Figure 13): Power ON, and High Temperature.
The green LED remains illuminated during normal operation. The other LED indicator,
“High Temperature” remains unlit if the compressor is functioning properly. Red illumination
of the High Temperature LED signals a faulty condition and warns the operator that the unit
requires servicing.
Rotate, clockwise, the Power On (see Figure 11) on the front panel of the generator. The
generator will start and the green LED will illuminate. The pressure gauge indicates compressor
pressure. The pressure slowly increases to a maximum set pressure of 125 psig
(8.6 barg). Full pressurization will occur within 10 minutes.
Rotate the nitrogen regulator knob to set the desired outlet pressure as indicated by the nitrogen
output gauge (see Figure 12)
Start-up
Operation
Diagnostics
Operation
Figure 9 Figure 10
POWER

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
9
1-800-343-4048
Figure 13 Figure 14
If the compressed air pressure decreases less than 90 psig (6.2 barg), the nitrogen output purity
level will decrease. Once pressure is restored, purity level will be restored.
The compressor output pressure gauge should read 115 psig to 120 psig during normal
operation.
See the Troubleshooting section of this manual on pages 17-20 to resolve any abnormal or faulty
conditions.
If the electrical power is interrupted, the generator will start automatically when the power
returns.
For normal or emergency shutdown, pull out the plug on the cordset to disconnect the mains
power (see Figure 1). Full depressurization takes 1 hour. Do not remove cover while system
is pressurized.
Maintenance tasks for the NitroFlow60 and NitroFlow60D Generator should be performed by
trained personnel familiar with the service and safety precautions of electromechanical devices
to avoid injury or damage. Safety risks that may affect the service personnel are identied with
the necessary protective measures described.
Parker highly recommends service be performed by a Parker trained technician only. Call
the factory about PMP, the Preventative Maintenance Program. The service personnel will
verify the safe state of the generator after the maintenance or repair is completed.
Prior to servicing the generator, turn off the system and disconnect the cordset from the power
supply. Allow the system a minimum of 1 hour to depressurize. See shutdown section above.
To ensure product performance and reliability, periodic preventative maintenance must be
performed according to the following schedules (refer to the replacement parts section for
the part numbers when ordering); Schedule 1 is servicing the compressor and Schedule 2 is
servicing the generator.
Pressure Interruption
Power Interruption
Shutdown
Maintenance
The compressor has an hour meter to record the accumulated operating time of the compressor
for preventative maintenance scheduling of the serviceable components. The meter indicates
00000.0 hours. The hour meter is not re-settable.
POWER
POWER

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
10
1-800-343-4048
Schedule 1 (refer to gures 15 and 16)
Replace Quantity Hours¹
Intake Filter 2 8700 hours
Belts 2 8700 hours
Tip Seal Set 1 8700 hours
Grease 1 8700 hours
Pump 1 30,000 hours
Discharge Pipe Sleeve 1 8700 hours
Schedule 2 (refer to gure 17)
Replace Quantity Hours¹
Filter Element
(P/N 100-18-DX) 1 8700 hours
Filter Element
(P/N 100-18-BX) 1 8700 hours
Filter Silencer
(P/N 75117) 1 8700 hours
Filter Element
(P/N 100-09-BX) 2 8700 hours
Carbon Module
(P/N B04-0525) 2 8700 hours
Schedule 1
Compressor
Figure 16
Figure 15
Figure 17, Maintenance Items
Schedule 2
Nitroflow60 and
Nitroflow60D
Generator
¹ Frequency hours for NitroFlow60 and NitroFlow60D are recorded by an hour meter.
Note: The meter indicates 00000.0 hours. The hour meter does not reset.

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
11
1-800-343-4048
If necessary, wipe the generator with a clean, dry cloth on an as-needed basis. Do not use
water, aerosols or other cleaning agents on the unit. Use of any liquid detergent to clean
the unit could pose an electrical hazard.
Figure 18
Compressor Inlet Tube
Figure 19
Figure 20
Cleaning
Schedule 1 Instructions
1Remove the upper side panel (see Figure 15).
2 By hand, push-in and rotate one intake lter cap to remove. Undo latch to remove second
intake lter cap (see Figure 20).
3 Replace the old lter elements with a new one, reverse steps to reassemble.
Intake Filter
1Remove the top panel on the unit by removing two screws at the rear with a phillips head
screwdriver and pulling top panel to the rear.
2Remove the generator cabinet on the unit by removing two screws with a phillips head
screwdriver.
a. LOOSEN (do not remove) three hex head screws located at the front of the cabinet
(gure 18).
b. Remove compressed air inlet tube from generator cabinet (see Figure 19).
c. Pull cabinet to the rear and lift off of compressor cabinet. Be cautious of exterior pre-lters.
3 Check belt tension (see gure 16) using a belt tension gauge or the deection method.
Tension gauge should indicate minimum 45 lbs. If using the deection method, 3.0-3.38 lbs
force at mid span should yield 7/32” belt deection.
4When making adjustments to the belts, make sure that pulley misalignment is no more than
1/16” and that both pulleys are not misaligned by more than 1/16” also (see Figure 21)
5Adjust tension if required by adjusting motor bracket. Replace belts if needed.
Belts

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
12
1-800-343-4048
1 Remove the plastic dust cap from the pump (see gure 22).
2 Rotate the compressor pulley until the grease tting is visible in the dust cap hole.
3 Using a grease gun extension adapter to engage the grease tting, supply the proper amount
of grease according to the grease delivery Table 1, gure 22.
Note: Pump grease gun before feeding to eliminate air from the needle adapter. Each pump
equals 0.65 grams of grease. The volume of grease is less after the 2nd pump since some
grease remains in the extension adapter.
4Replace the plastic dust cap.
Figure 21, belt diagram
Grease Orbital
Scroll Compressor
Bearings
Table 1 - Grease Delivery SLAE03E
Bearing 1st Pump 2nd Pump
Orbit Scroll Bearing 5 times 4 times
Pin Crank Bearing 4 times 4 times
(Orbit Scroll side)
Figure 22 Grease and Orbit Scroll Bearing

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
13
1-800-343-4048
1 Remove the six housing nuts (see Figure 22).
2Remove the stationary scroll housing.
3Remove the old tip seals from both the orbiting and stationary scroll housings.
4Remove the old dust seals from both housings.
5Blow off any dust from both housings.
6Install the new high pressure (shorter one) tip seal starting from the center of the scroll,
outward inside the seal channel.
Note: The side and bottom lip notches of the tip seal face INWARD and DOWNWARD in the
channel.
7 Install the new low pressure (longer one) tip seal in the same way. Note: Make sure there is
no gap between the high and low pressure seals.
Note: Make sure there is no gap between the high and low pressure seals.
Caution! Do not attempt to remove the orbit scroll from the housing.
8After installing half of the low pressure seal, carefully remove the seal from the channel to
make sure the seal is properly locking onto the channel indentations located just past the high
pressure seal.
Note: The indentations are machined into the seal channel to prevent the low pressure seal
from moving.
9Blow off any dust caused by removing the seal from the channel.
10 Install the low pressure seal completely.
Note: Make sure the side and bottom lip notches are facing INWARD and DOWN into the
seal channel.
Note: The lip notches must not be distorted in the seal or torn off.
11 Install backup tube in the dust seal channel.
Replace Tip Seal
1 Remove the side service panel (see gure 15).
2Remove the fan duct.
3 Remove the nuts, bolts, and then the xed scroll.
4 Grease three pin crank bearings according to Table 1 and gure 23.
5 Replace xed scroll and fan duct. Torque bolts, initially 17 inch-lbs., then nally 175 inch-lbs.
Grease Pin Crank
Bearing
Figure 23
Note: The grease tting located in the center of the pin crank bearing feeds only the orbit scroll
side bearing. Use a needle adapter to supply grease to the housing side bearing. Pump grease
gun to eliminate any air in the needle adapter. Hold grease gun for 5 to 10 seconds after feeding
grease to prevent grease blowback from the grease tting.

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
14
1-800-343-4048
Fuse Replacement
Schedule 2
Instructions
Coalescing Filters
12 Place dust seal over the backup tube.
Caution! The backup tube must meet at the bottom of the housing in the six o’clock
position. The dust seal must meet on the right side of the housing at the three o’clock
position.
13 Install the stationary scroll housing onto the orbiting scroll housing and reassemble the unit.
14 Tighten the pulley bolt (M8) to 265 in -lbs, using a torque wrench.
15 Tighten the six housing nuts to 265 in-lbs, using torque wrench.
There are three fuses located in the compressor cabinet (see gure 24). Before servicing the
fuses, turn the generator off and disconnect the power cord from the power supply.
To access the fuses, remove the upper side panel (see Figure 15), and locate the fuse blocks.
Both lines 1 and 2 are fused separately. Remove both line fuses with a small screwdriver.
Replace either one or both fuses as required. To maintain safety, use only fuses of the size
and type specied in the Specications section of this manual.
These instructions describe the replacing of the two coalescing pre-lters, the two carbon modules,
and two particulate lters in the NitroFlow60 Generator System Series.
Generator must be set to “OFF” and pressure gage must read zero before lter replacement.
1 The replacement elements are B2104N-1B1-DX (rst stage) and B2104N-0B1-BX (second
stage). It is very important to install the proper grade lter into the proper housing. Take extra
care to install in the proper sequence (see gure 25). From the rear, DX on left, BX on right.
2Twist the bowl loose from the head of the assembly, counter clockwise, and remove carefully
so as not to damage or lose the o-ring (see gure 26).
3 Unscrew the element retainer at the base of the element, and replace the spent element with
a new one (see gure 27).
4Reverse the steps to reassemble. Make sure the o-ring is sealed in the groove before
reinstalling the bowl.
Generator must be set to “OFF” and pressure gage must read zero before module replacement.
Figure 24
Figure 26 Figure 27
Reset
Button
DXWater
Separator
BX
Figure 25

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
15
1-800-343-4048
Carbon Modules
Particulate Filters
There are two carbon modules that require changing. Please note that the carbon modules will
only mount to the brackets in one orientation (see gure 28).
1Remove the top panel from the NitroFlow60 Generator Series by removing two screws with a
Phillips head screwdriver.
2 Using a 1” wrench, remove the tie nut at the base of the bowl (see gure 30).
3Twist the bowl loose from the head of the assembly and remove carefully as not to damage or
lose the o-ring (see gure 31).
4 Unscrew the element retainer at the base of the element, and replace the spent element with
a new one (see gure 32).
5Reverse the steps to re-assemble. Make sure the o-ring is sealed in the groove before
reinstalling the bowl. Hand tighten.
Figure 28
Figure 30 Figure 31
Figure 32
Figure 29
1Remove the top panel from the NitroFlow60 Generator Series by removing two screws with a
phillips head screwdriver and pulling top panel to the rear.
2 Disconnect the plastic tubing from the press ttings on the inlet and outlet ports of the module.
To remove the tubing from the press tting, push the tubing into the tting, hold the collar of
the tting back, and pull the tubing out of the tting.
3 Release the hook and loop strap that clamps the module to the unit (see gure 29). Slide the
module out of the bracket.
4 Replace with a new carbon module, noting the correct orientation (see gure 28).
5Slide the new module into the bracket fully and re-attach the hook and loop strap.
6 Re-insert inlet and outlet tubing. Use a soap or a leak detecting solution to check for leaks
when pressurizing (re-starting) the unit.
Generator must be set to “OFF” and pressure gage must read zero before lter replacement.

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
16
1-800-343-4048
System Specifications NitroFlow60 NitroFlow60D
CSA Certication Standard CAN/CSA C22.2 No. 61010-1 CAN/CSA C22.2 No. 61010-1
IEC Standard IEC 61010-1 IEC 61010-1
CENELEC Standard EN 61010-1 EN 61010-1
UL Standard UL 61010-1 UL 61010-1
IEC 61010 Installation Category II Category II
IEC 61010 Pollution Degree 2 Degree 2
EMC Compliance EN61326-1/CISPR 11 EN61326-1/CISPR 11
Nitrogen (N2) Purity 97% 97%
Nitrogen Max. Flow/Pressure 60 LPM/100 PSIG (6.9 Barg) 60 LPM/100 PSIG (6.9 Barg)
Dry Air Purity - -40° F (-40°C) ATM Dew Point
Dry Air Max. Flow/Pressure - 4 SLPM/100PSIG (6.9 Barg)
Suspended Liquids None None
Particles > 0.01 Micron None None
Phthalates None None
Sound Level 49 dB(A) at 1 Meter 49 dB(A) at 1 Meter
Min./Max. Operating Pressure 115/120 PSIG (7.9/8.2 Barg) 115/120 PSIG (7.9/8.2 Barg)
Flow Capacity See Pressure/Flow Charts See Pressure/Flow Charts
Inlet Air Consumption 160 SLPM (5.7 scfm ) 160 SLPM (5.7 scfm)
Min./Max. Ambient Temp. 60°F/90°F (16°C-32°C) 60°F/90°F (16°C-32°C)
Max. Ambient Relative Humidity 80% 80%
Altitude 2000M 2000M
Heat Dissipation 10246 BTU/Hour 10246 BTU/Hour
Electrical Requirements¹ - 60 Hz 208-254VAC, 60 Hz, 1 Phase, 16A 208-254VAC, 60 Hz, 1 Phase, 16A
Electrical Requirements¹ - 50 Hz 230VAC, 50 Hz, 1 Phase, 13A 230VAC, 50 Hz, 1 Phase, 13A
Power Consumption - 60 Hz 16A @ 230VAC, 3680Watts 16A @ 230VAC, 3680Watts
Power Consumption - 50 Hz 13A @ 230VAC, 2990Watts 13A @ 230VAC, 2990Watts
Start-Up Time 10 Minutes 10 Minutes
Compressor Outlet - Tank Drain 1/4” Tube 1/4” Tube
Ports - Filter Drains (2 ) 1/4” Tube 1/4” Tube
Ports - Water Seperator 1/4” Tube 1/4” Tube
Ports - Nitrogen Outlet 1/4” NPT 1/4” NPT
Ports - Dry Air Outlet - 1/4” NPT
Fuses - Primary FU1, FU2 10.3 x 38.1mm, Type T, 600V, 1.25A 10.3 x 38.1mm, Type T, 600V, 1.25A
Fuses - Secondary FU3 6.3 x 31.7mm, Type T, 250V, 1.25A 6.3 x 31.7mm, Type T, 250V, 1.25A
Dimensions 33.5”L x 20.75”W x 45.25”H 33.5”L x 20.75”W x 45.25”H
Product Weight 407 lbs (185 kgs) 407 lbs (185 kgs)
Shipping Weight 632 lbs (287 kgs) 632 lbs (287 kgs)
¹ Main supply voltage uctiations not to exceed ± 10% of nominal voltage.
2 Motor change installed in compressor Serial Number (H)8/8/2012-5284488-610 and after; NITROFLOW60 Serial Number, N260-0019 and after.
System Specifications
Cautions 1The generator should be installed in an area with adequate ventilation to reduce the
ammability of the oxygen-rich permeate stream. The system should not be located in an
area where the permeate stream poses the risk of explosion or combustion
2Nitrogen is nontoxic and largely inert. It can act as a simple asphyxiant by displacing
oxygen in air. Inhalation of nitrogen in excessive concentrations can result in
unconsciousness without any warning symptoms such as dizziness and fatigue.
3The maximum operating pressure of the system is 125 psig(8.6 barg). Operating the
generator at pressures above 125 psig(8.6 barg) will result in damage to the pump and
membrane.
4The drain lines should be piped away to the provided collection container (open to
atmospheric pressure) to avoid any possible re-entrainment of liquid in the air which feeds
the generator.
5 Changes in inlet pressure or outlet ow demand will alter the gas purity.

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
17
1-800-343-4048
NitroFlow60 Compressor Replacement Parts Part Number
8,700/17,400 Hour Maintenance Kit 1(1) MKNITROFLOW60-1
17,400 Hour Maintenance Kit 2(2) MKNITROFLOW60-2
26,100 Hour Maintenance Kit 3(3) MKNITROFLOW60-3
V-Belts Kit 4(4) MKNITROFLOW60-4
Primary Fuses A03-0324 (Figure 24)
Secondary Fuses A03-0323 (Figure 24)
NitroFlow60 Maintenance Parts Part Number
8,700 Hour Maintenance Kit 5(5) MKNITROFLOW60-5
Accessory Parts Part Number
Voltage Step-Down Transformer A03-0412
Voltage Booster Step-Up Transformer A03-0411
Exhaust Ventilation Kit B04-0575
Bottle Drain Kit B04-0540
In-line Duct Fan A03-0378
(1) Kit 1 includes: Intake Filters, V-Belts, Tip Seal Set, Discharge Pipe Sleeve, Grease, and Grease Gun
(2) Kit 2 is an Auto Drain Valve Assembly
(3) Kit 3 includes: Compressor Pump, Intake Filters, and V-Belts, and Check Valve
(4) Kit 4 includes: (2) V-Belts
(5) Kit 5 includes: (2) Coalescing Filters, (4) Filter Silencer, (2) Particulate Filters, (2) Carbon Modules
Replacement Parts
Troubleshooting Disconnect the electrical power and depressurize the generator before
attempting any troubleshooting activities according to the shut down steps in
the operation section. Only trained personnel using reasonable care should
perform any troubleshooting activities.
Determine whether the trouble is with the compressor or the gas generator.
Disconnect the plastic tubing from the compressor tting (see Figure 33). Once
removed, place supplied plug into tting. Check that the compressed air source
builds full pressure and cycles off within two minutes.
If the compressor air supply is conrmed, remove lower side panel and relieve tank
pressure using auto tank drain test button. Proceed to the NItroFlow60 generator
Troubleshooting section (page 20), and begin troubleshooting the NitroFlow60
generator. Otherwise, refer to the Compressor Troubleshooting section (pages 18,
19), and begin troubleshooting the compressor.
Remove Tube from Compressor Fitting
Figure 33

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
18
1-800-343-4048
POWER ON light (green)
does not appear
• Power not plugged in
supply outlet
• Power switch is not ON
• Blown fuse or tripped OFF
circuit breaker at customer
provided power supply
• Blown fuse at primary side
of transformer
• LED power switch has
failed
• Plug in power cord
• Switch POWER to ON
• Replace fuse or switch
disconnect ON
• Replace fuse on primary
side. Be sure to use the
same type and size
• Replace switch
Compressor Troubleshooting
Symptom Probable Cause Corrective Action
POWER ON light is on but
motor will not start
• Motor overload has tripped
• Wrong or low voltage
• Starter failed
• Motor failed
• Determine fault and reset
overload (Blue button)
• Check incoming power
supply and unit power rating
•
Replace contractor assembly
• Replace motor
Compressor is running but
will not make pressure
• Auto drain valve is open
continuously
• Drive belts broke or are too
loose
• Clogged intake lter
element
• Pressure relief valve has
opened
• Excessive tip seal wear
• Motor running wrong
direction
• Replace auto drain valve
• Replace drive belts and (or)
tighten to specication
• Replace intake lter element
per instructions
• Pressure switch needs
replacing or motor contacts
welded shut
• Replace tip seals per
instructions every 8,700 hrs
• Correct power connections
Excessive noise or vibration • Drive belts are loose
• Drive belt has separated or
at spot
• Cooling fan touching guard
• Motor bearing has failed
• Re-greasing procedure
not performed at 8,700 hr
intervals
• Pump damaged
• Intake lters loose on pump
• Tighten belts to specication
• Replace drive belt
• Adjust guard
• Replace motor
• Removal of stationary
scroll is required for pump
inspection
• Repair or replace pump
• Tighten intake lters

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
19
1-800-343-4048
Compressor Troubleshooting
Symptom Probable Cause Corrective Action
TEMPERATURE HIGH light
is on, Compressor shuts
down
• Room temp is above 90°F
• Inlet air duct is obstructed
• Cooling air fan not running
• Re-greasing procedure
not performed at 8,700 hr
intervals
• Aftercooler ns dirty, clogged
• Excessive tip seal wear
• Intake lter damaged
• Compressor is dirty
• Add ventilation or air
conditioning to room
• Remove obstruction or
reposition unit to allow for
cooling air
• Check electrical connection at
fan
• Replace cooling air fan
• Re-grease
• Clean aftercooler
• Replace tip seals per
instructions every 5,000 hrs
• Check, replace lter
• Clean unit
Compressor shuts down on
temperature malfunction
• Temperature switch has
unplugged
• Temperature switch has
failed
• Plug in temperature switch
• Replace temperature switch
Compressor turns on/off
rapidly
• Receiver tank has high
level of water
• Compressor check valve
has failed
• Defective pressure switch
• Replace electric tank drain
• Replace check valve
• Replace pressure switch
Safety valve blows off • Pressure switch has failed
to open
• Motor starter contacts
welded shut
• Replace pressure switch
• Replace motor starter
Motor overload has tripped • Motor has failed
• Improper wiring
• Wrong overload setting
• Low voltage
• Pump has failed
• Replace motor
• Check wiring
• Check overload setting
• Check incoming power supply
• Fix or replace pump

Bulletin TI-NITROFLOW60-G Installation, Operation and Maintenance Manual
NitroFlow60 and NitroFlow60D Generators
www.labgasgenerators.com
20
1-800-343-4048
No pressure
No or Low Flow
Low Output Pressure
Low Purity
Change in Purity
• Compressor is not ON
• Regulator is closed
• System leaks at tting
connections
• Internal leak
• External leak
• Compressor not at pressure
• Particulate lter plugged
• Flow demand greater than
60 LPM
• Flow demand greater than
60 LPM
• Nitrogen Module
Malfunction
• High relative humidity
• Water carryover due to
water seperator oat drain
malfunction
• Water carryover due to lter
oat drain malfunction
• Spent carbon modules
• Internal leak
• Water carryover due to
water seperator oat drain
malfunction
• Water carryover due to lter
oat drain malfunction
• High relative humidity
• Flow demand greater than
4 SLPM
• Turn on compressor
• Open pressure regulator
• Check for leaks at
connections
• Check for leaks
• Check for leaks
• See compressor section
• Replace lter
• Reduce ow demand
• Reduce ow demand
• Replace module
• Move to area where RH is
less than 80%
• Replace water seperator
oat drain
• Replace lter oat drain
• Replace carbon modules
• Check for leaks
• Replace water seperator
oat drain
• Replace lter oat drain
• Move to area where RH is
less than 80%
• Reduce ow demand
NitroFlow60 and NitroFlow60D Generator Troubleshooting
Symptom Probable Cause Corrective Action
NitroFlow60D Only
Other manuals for NitroFlow60
1
This manual suits for next models
1
Table of contents
Other Parker Balston Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Hitachi
Hitachi SEM S-4700 user manual

socorex
socorex CALIBRA 852 Series operating instructions

General Data Healthcare
General Data Healthcare TBS SHURMark SMC-A Operator's manual

Cardinal Health
Cardinal Health CHH05RESA manual

Koehler
Koehler K34710 Operation and instruction manual

Ecowell
Ecowell Homeguard manual