Patterson 043-2047 User manual

Operation Manual
OM-MK038E
1. Specifications
2. Installation/Removal of Handpiece onto/from Tubing Connector
3. Installation/Removal of Prophy/Contra Angle or Bur (Attachment)
onto/from Handpiece
・The Patterson Hygiene Handpiece is intended for use by dental
professionals only.
・Read this Operation Manual carefully before using the handpiece. Be
sure to follow all the operating instructions, including all cautions and
warnings.
・Keep this Operation Manual for future reference.
Caution before use
!
・Never operate the handpiece if it rattles after connection to the tubing
connector or if any abnormality is sensed. In such a case, consult the
Patterson Service Center.
・Never attempt to forcibly connect the handpiece to the tubing
connector when the screw-in of the handpiece to the tubing connector is
too tight. Never use pliers or any other tools for the installation.
Caution
!
・Never install a damaged attachment onto the handpiece such as a
worn-off bur, bur or prophy/contra angle with a bent spindle. Such
attachments may not only injure patients, but critically damage the
handpiece.
・Never install a non-ISO standard bur or prophy/contra angle with a
non-ISO standard spindle.
・Never open/close the chuck without installing the attachment onto the
handpiece, as the chuck holding power may be prematurely reduced.
Caution
!
Back-end configuration Applicable to ISO standard 4-hole connector
Nose configuration Applicable to U-type prophy/contra angles
Chuck configuration Applicable to ISO standard 2.35 dia. spindle
Maximum speed 5,500rpm at drive-air of 2.0 kgf/㎠(28psi)
6,000rpm at drive-air of 2.5 kgf/㎠(36psi)
Note:
The maximum speed may vary slightly depending on the type of tubing
connected to the handpiece.
●Installation:
1. Make sure that the tubing connector is ISO standard 4-hole.
2. Align the air/water inlet ports at the handpiece back end with the
corresponding holes of the tubing connector.
3. Holding the swivel ring of the handpiece with fingers, tighten the
nut of the tubing connector by turning it clockwise.
●Removal: Holding the swivel ring, remove the handpiece from the
tubing connector by turning the nut counterclockwise.
●Removal of Bur Blank: Holding the motor, twist the Bur Lock Ring
in the direction indicated by the black arrow until the chuck clicks open
and pull the bur blank out.
●Installation of Prophy/Contra Angle:
1. Make sure that the chuck is open.
2. Insert the spindle into the chuck all the way until the back end of the
prophy/contra angle contacts the shoulder of the nose of the handpiece,
aligning the slit at the back end of the prophy/contra angle with the
protuberance of the nose of the handpiece.
3. Holding the motor, twist the nose in the direction indicated by the
white arrow until the chuck clicks shut.
4. Before operation, pull the attachment to make sure that it is securely
installed onto the handpiece.
●Installation of Bur:
1. Make sure that the chuck is open.
2. Insert the bur spindle into the chuck all the way until the spindle
touches the back of the chuck.
3. Place the nose of the handpiece upward so that the bur inserted does
not slide down, then twist the nose in the direction indicated by the
white arrow until the chuck clicks shut.
4. Before operation, pull the attachment to make sure that it is securely
installed onto the handpiece.
Swivel RingSwivel Ring
001
043-2047 Hygiene Handpiece
4. Operation
●Set the air pressure gauge for 2.0 kgf/㎠(28psi).
●Before operation, be sure to make a trial run and make sure:
1. The drive-air is not leaking from the connecting part of the handpiece
back end and tubing connector.
2. Changes in the rotating direction and rotating speed can be manipulated
with the control ring as shown below:
*When the ring is turned in "F" direction, the attachment installed onto
the handpiece rotates clockwise.
*When the ring is turned in "R" direction, the attachment installed onto
the handpiece rotates counterclockwise.
*The rotation is stopped at the center position between "F" and "R".
●Select the rotating direction and the speed by manipulating the ring
according to the treatment purposes.

5. Cleaning and Drying
7. Sterilization
8. Guarantee
・Never install/remove the attachment onto/from the handpiece while the
handpiece is in operation.
・Never operate the handpiece without the attachment.
・Whenever any abnormalities in performance, sound level, heat level,
function, etc. are sensed, stop the operation and consult the Patterson
Service Center.
・Never keep the handpiece connected to the tubing connector.
Caution
!
・Do not use sprays other than PANA SPRAY Plus. Using sprays other
than PANA SPRAY Plus may cause the handpiece to overheat.
・Hold the spray can upright.
・Firmly hold the handpiece to prevent it slipping when spray pressure is
applied.
・Spray lubricant until it expels from the handpiece head.
Caution
!
!
Warning
!
●Autoclave is recommended for sterilization.
1. Before sterilization, clean the surface of the handpiece and lubricate the
handpiece.
2. Lubricate the handpiece and make a trial run.
3. Remove the attachment from the handpiece.
4. Put the handpiece into a sterilization pouch. Do not combine it with others.
5. Place the handpiece in the center or upper shelf of the autoclave chamber.
6. Autoclave the handpiece at 275°F or 135℃for 20 minutes.
Patterson guarantees the handpiece for 10 months subject to the observance
of maintenance instructions, cautions, warnings and normal use.
Notes:
●Patterson reserves the right to analyze and determine the cause of any failure.
●Guarantee is void if the breakage, accident or any other issues are caused by
the neglect of the daily maintenance, cautions, warnings and common sense
in use.
●Dismantling or reassembling by any third parties other than the Patterson
Service Center voids guarantee.
Never dismantle/alter the handpiece in any case.
・Never use dry heat sterilization, cold sterilization, etc.
・Never autoclave the handpiece together with others in a sterilization
pouch.
・Never place the handpiece at the bottom of autoclave chamber, as the
temperature at the bottom may rise above the actual setting
temperature.
Caution
●The handpiece must be cleaned and lubricated after each use and before
autoclaving using the following procedures:
1. Remove the attachment from the handpiece.
2. Remove the handpiece from the tubing connector.
Cleaning and Drying the Exterior
Wash the handpiece under running water following the procedures below.
(Water condition: ≤100°F, ≥3.5L/min; water should be of the same quality as
drinking water)
●Clean the external surfaces of the
handpiece with a soft bristled toothbrush
for 15 seconds or more.
●Clean around the bur insertion hole for 15
seconds or more, using an interdental
brush (wire thickness 0.7mm) that
conforms to size 4 in ISO 16409.
Under appropriate lighting (natural office
lighting, 500 lx or higher), inspect the
handpiece for blood or other debris. If any
visible debris remains, repeat the process
until the handpiece is visually clean.
After cleaning, if the residual moisture is
present, wipe off with dry cloth or blow it off with compressed air (≤50.75psi)
until there is no moisture in the interior and exterior.
When blowing off with compressed air, cover the handpiece with cloth to
prevent scattering of water. Proceed to 6: Cleaning and Lubrication.
6. Cleaning and Lubrication
Cleaning and Lubricating the Interior
●Attach a Tip Nozzle to the nozzle of the
PANA SPRAY Plus.
●Shake the can 3 or 4 times and insert the
Tip Nozzle into the drive air port while
holding the head of the handpiece with a
cloth etc.
●Hold the handpiece and spray for 2 or 3
seconds until oil comes out of the
handpiece head. Repeat lubrication until foreign material stops coming out of
the tip.
2021-11-00 CACB1100 N
15sec.
Pour une version en français du manuel d'utilisation, veuillez vous rendre sur le
site Web de Patterson: https://www.pattersondental.com/fr-CA
Manufactured for:
Fabriqué pour :
Patterson Dental Supply, Inc.
1031 Mendota Heights Road
Saint Paul, MN 55120
Table of contents
Other Patterson Dental Equipment manuals
Popular Dental Equipment manuals by other brands

KaVo
KaVo INTRA K-LUX 200 F Instructions for use

Luraco
Luraco IMD PRO-30K product manual

Renfert
Renfert Waxlectric light I instruction manual

Renfert
Renfert SILENT compact Quick reference card

Treedental
Treedental TR-CWL07 Operation and installation manual

DENTSPLY
DENTSPLY SmartLite Max Directions for use