PDC spas Accent User manual

Navigaon
Navigang the enre menu structure is done with 2 or 3 buons on the control panel.
The Accent control panel has separate WARM (Up) and COOL (Down) temperature buons.
The MENU Buon is used to choose the various menus and navigate each secon.
The two Temperature buons (Warm and Cool) are used to select and change the parameter.
Typical use of the Temperature buons allows changing the Set Temperature while the numbers are ashing in the LCD.
The menus can be exited by pressing the Menu key repeatedly to return to main screen.
Waing a few seconds in the Main Menu will allow the display to revert to the Main Screen. Most changes are not saved
unless Menu/Select is pressed.
Sample Navigaon Flow
Power-up Screens
Each me the System powers up, a series of numbers is displayed.
Aer the startup sequence of numbers, the system will enter Priming Mode.
Eect. 6/2019
Accent™ Control Operaon Guide

Preparaon and Filling
Fill the spa to its correct operang level. Be sure to open all valves and jets in the plumbing system before filling to allow as
much air as possible to escape from the plumbing and the control system during the filling process.
Aer turning the power on at the main power panel, the top-side panel will go through specific sequences. These sequences
are normal and display a variety of informaon regarding the configuraon of the hot tub control.
Priming Mode – M019*
This mode will last for 4-5 minutes or you can manually exit the priming mode aer the pump has primed.
Regardless of whether the priming mode ends automacally or you manually exit the priming mode, the system will automa-
cally return to normal heang and filtering at the end of the priming mode. During the priming mode, the heater is disabled to
allow the priming process to be completed without the possibility of energizing the heater under low-ow or no-ow condi-
ons. Nothing comes on automacally, but the pump can be energized by pushing the “Jet” buons.
Priming the Pumps
As soon as the above display appears on the panel, push the “Jet” buon once to start pump in low-speed and then again to
switch to high-speed. The pump will now be running in high-speed to facilitate priming. If the pump has not primed aer 2
minutes, and water is not owing from the jets in the spa, do not allow the pump to connue to run. Turn o the pump and
repeat the process. Note: Turning the power o and back on again will iniate a new pump priming session. Somemes mo-
mentarily turning the pump o and on will help it to prime. Do not do this more than 5 mes. If the pump will not prime, shut
o the power to the spa and call for service.
Important: A pump should not be allowed to run without priming for more than 2 minutes. Under NO circumstances should a
pump be allowed to run without priming beyond the end of the 4-5 minute priming mode. Doing so may cause damage to the
pump and cause the system to energize the heater and go into an overheat condion.
Exing Priming Mode
You can manually exit Priming Mode by pressing a “Temp” buon (Up or Down). Note that if you do not manually exit the
priming mode as described above, the priming mode will be automacally terminated aer 4-5 minutes. Be sure that the pump
has been primed by this me.
Once the system has exited Priming Mode, the top-side panel will momentarily display the set temperature but the display will
not show the temperature yet, as shown below. This is because the system requires approximately 1 minute of water owing
through the heater to determine the water temperature and display it.
Vitality Series Fitness Spas / LifeStyle Series Hot Tubs
Pump
Press the “Jets ” buon once to turn pump on or o, and to shi between low and high-speeds.
If le running, the pump will turn o aer a me-out period. The pump low-speed will me out aer 30 minutes. The high-speed
will me out aer 15 minutes.
If the spa is in Ready Mode, pump low may also acvate for at least 1 minute every 30 minutes to detect the spa temperature
(polling) and then to heat to the set temperature if needed. When the low-speed turns on automacally, it cannot be deacvated
from the panel, however the high speed may be started.

If unable to turn o low speed, note the pump will connue to operate on low speed unl set temp is reached. It will automacal-
ly turn o when temp is reached.
Freeze Protecon
If the temperature sensors within the heater detect a low enough temperature, then the pump will automacally acvate to
provide freeze protecon. The pump will run either connuously or periodically depending on condions.
Clean-up Cycle
When a pump is turned on by a buon press, a clean-up cycle begins 30 minutes aer the pump is turned o or mes out. The
pump and the ozone generator will run for 30 minutes or more, depending on the system.
Temperature and Temp
Adjusng the Set Temperature
The panel has Up and Down buons (Temperature buons), pressing Up or Down will cause the temperature to ash. Pressing
a temperature buon again will adjust the set temperature in the direcon indicated on the buon. When the LCD stops ashing,
the spa will heat to the new set temperature when required. Spa temperature can be set between 80°F and 104°F.
Press-and-Hold
If a Temperature buon is pressed and held when the temperature is ashing, the temperature will connue to change unl the
buon is released.
NOTE: Spa will not reach set temp if unit is in Low Range. Check sengs to be sure unit is in High Range to reach temps above
80°F.
Dual Temperature Ranges
This system incorporates two temperature range sengs with independent set temperatures. The High Range designated in the
display by an “up” arrow, and the Low Range designated in the display by a “down” arrow.
These ranges can be used for various reasons, with a common use being a “ready to use” seng vs. a “vacaon” seng. The
Ranges are chosen using the menu structure below. Each range maintains its own set temperature as programmed by the user.
This way, when a range is chosen, the spa will heat to the set temperature associated with that range. For example: High range
might be set between 80°F and 104°F. Low range might be set between 50°F and 99°F. Freeze protecon is acve in either range.
1) Press menu key TEMP will be displayed.
2) Use Up or Down arrow to change from High to Low range. A small Up arrow or Down arrow icon will be displayed on the
screen.
NOTE: Spa will not reach set temp if unit is in Low Range. Check sengs to be sure unit is in High Range to reach temps above
80°F.
Mode, Ready and Rest
In order for the spa to heat, a pump needs to circulate water through the heater. The pump that performs this funcon is known
as the “heater pump.”
READY Mode will circulate water periodically, using pump Low, in order to maintain a constant water temperature, heat as needed,
and refresh the temperature display. This is known as “polling.”
REST Mode will only allow heang during programmed filter cycles. Since polling does not occur, the temperature display may not
show a current temperature unl the pump has been running for a minute or two.
Ready-in-Rest Mode
READY/REST appears in the display if the spa is in Rest Mode and Jet 1 is pressed. It is assumed that the spa is being used and will
heat to set temperature. While Pump High can be turned on and o, pump Low will run unl set temperature is reached, or 1 hour
has passed. Aer 1 hour, the System will revert to Rest Mode. This mode can also be reset by entering the Mode Menu and
changing the Mode.

Show and Set Time of Day
Seng the me-of-day can be important for determining filtraon mes and other background features. When in the TIME menu,
SET TIME will ash on the display if no me-of-day is set in the memory.
Seng Time:
1) Repeatedly press menu key unl set me or me appears on screen.
2) Press up or down arrow to adjust hour.
3) Press menu key to switch to minutes.
4) Press up or down key to adjust minutes.
5) Press menu key to exit.
If power is interrupted to the system, Time-of-Day is not stored. The system will sll operate and all other user sengs will be
stored. If filter cycles are required to run at a parcular me of day, reseng the clock will return the filter mes to the actual
programmed periods.
When the system starts up, it defaults to 12:00 Noon, so another way to get filter mes back to normal is to start up the spa at
noon on any given day. SET TIME will sll ash in the TIME Menu unl the me is actually set, but since the spa started at noon,
the filter cycles will run as programmed.
Restricng
The control can be restricted to prevent unwanted use or temperature adjustments. Locking the panel prevents the controller from
being used, but all automac funcons are sll acve.
Locking the Temperature allows Jets and other features to be used, but the Set Temperature and other programmed sengs
cannot be adjusted. Temperature Lock allows access to a reduced selecon of menu items. These include Set Temperature, FLIP,
LOCK, UTIL, INFO and FALT LOG. Unlock sequence may be used from any screen that may be displayed on a restricted panel.
Lock Panel
1) Press menu key unl Lock appears.
2) Press up key, temp will be displayed.
3) Press menu buon, panel will be displayed.
4) Press up key twice, ON will be displayed.
5) Press menu buon to lock panel.
Hold
Hold—M037*
Hold Mode is used to disable the pumps during service funcons like cleaning or replacing the filter. Hold Mode will last for 1 hour
unless the mode is exited manually.
Set Hold:
1) Press menu key repeatedly unl hold is displayed.
2) Press up arrow to enter hold.
3) Press down arrow to exit hold.
Adjusng
Main Filtraon
Filter cycles are set using a start me and a duraon. Start me is indicated by an “A” (AM) or “P” (PM) in the boom right corner of
Lock Temp
1) Press menu key unl Lock appears.
2) Press up key for temp lock.
3) Press temp buon again.
4) Press temp buon to ON.
5) Press menu buon to exit.
Unlock Panel
1) Press menu.
2) Press and hold up key.
3) Press menu buon twice to unlock.
Unlock Temp
1) Press temp buon unl Lock appears.
2) Press up key, temp will be displayed.
3) Press and hold up key buon.
4) Press menu buon twice.

the display. Duraon has no “A” or “P” indicaon. Each seng can be adjusted in 15-minute increments. The panel calculates the
end me and displays it automacally.
Set Filter Cycle 1:
1) Repeatedly press menu key unl FLTR appears.
2) Press up arrow and BEGN will be displayed.
3) Press up arrow again and then use up or down arrow to select filter cycle start me.
4) Press menu key twice to exit beginning me.
5) Press up menu to select run hours then press up arrow or down arrow to adjust hour.
6) Press menu key twice to exit.
Filter Cycle 2
It is possible to overlap Filter Cycle 1 and Filter Cycle 2, which will shorten overall filtraon by the overlap amount. Filter Cycle 2
is OFF by default.
Set Filter Cycle 2:
1) Repeatedly press menu unl FLTR2 appears
2) Press up arrow.
3) Press up arrow again to turn on.
4) Repeat step for seng filter 1.
Purge Cycles
In order to maintain sanitary condions, pump will purge water from plumbing by running briey at the beginning of each filter
cycle.
If Filter Cycle 1 is set for 24 hours, enabling Filter Cycle 2 will iniate a purge when Filter Cycle 2 is programmed to begin.
Preference
F/C (Temp Display)
Change the temperature between Fahrenheit and Celsius.
12/24 (Time Display)
Change the clock between 12 hr. and 24 hr. display.
RE-MIN-DERS (Reminders)
Turn the reminder messages (like “Clean Filter”) On or O.
CLN-UP (Cleanup)
Cleanup Cycle Duraon is not always enabled, so it may not appear. When it is available, set the length of me Pump will run
aer each use. 0-4 hours are available.
Ulies
A ulies menu selecon may be accessed and is intended for qualified service techs only.
GFCI Test
The Ground Fault Circuit Interrupter (GFCI) or Residual Current Detector (RCD) is an important safety device and is required
equipment on a hot tub installaon. (The GFCI Test Feature is not available on CE rated systems.)
Once the GFCI is tripped manually, reset the GFCI and the spa will operate normally from that point.
Warning: The end-user must be trained how to properly reset the GFCI or RCD. If freezing condions exist, the GFCI or RCD should
be reset immediately or spa damage could result.

General
Priming Mode—M019
Each me the spa is powered up, it will enter Priming Mode. The purpose of Priming Mode is to allow the user to run each pump
and manually verify that the pump is primed (air is purged) and water is owing. This typically requires observing the output of the
pump and is generally not possible in normal operaon. Priming Mode lasts 4 minutes, but you can exit it earlier by pressing any
Temp buon. The heater is not allowed to run during Priming Mode.
Water Temperature is Unknown
Aer the pump has been running for 1 minute, the temperature will be displayed
Too Cold – Freeze Protecon
A potenal freeze condion has been detected, or the Aux Freeze Switch has closed, and pump is acvated. The pump is ON for
at least 4 minutes aer the potenal freeze condion has ended, or when the aux freeze switch opens. In some cases, pumps may
turn on and o and the heater may operate during Freeze Protecon. This is an operaonal message, not an error indicaon.
Water is too Hot (OHS) – M029
The system has detected a spa water temp of 110°F (43.3°C) or more, and spa funcons are disabled. System will auto reset when
the spa water temp is below 108°F (42.2°C). Check for extended pump operaon or high ambient temp.
Heater-Related
Heater Flow is Reduced (HFL) – M016
There may not be enough water ow through the heater to carry the heat away from the heang element. Heater start up will
begin again aer about 1 min. See “Flow Related Checks” below.
Heater Flow is Reduced (LF)* – M017
There is not enough water ow through the heater to carry the heat away from the heang element and the heater has been
disabled. See “Flow Related Checks” below. Aer the problem has been resolved, you must press any buon to reset and begin
heater start up.
Heater may be Dry (dr)* – M028
Possible dry heater, or not enough water in the heater to start it. The spa is shut down for 15 min. Press any buon to reset the
heater start-up. See “Flow Related Checks” below.

Heater is Dry* – M027
There is not enough water in the heater to start it. The spa is shut down. Aer the problem has been resolved, you must press any
buon to reset and restart heater start up. See “Flow Related Checks” below.
Heater is too Hot (OHH)* – M030
One of the water temp sensors has detected 118°f (47.8°C) in the heater and the spa is shut down. You must clear the message
when water is below 108°f (42.2°C). See “Flow Related Checks” below.
A Reset Message may appear with other Messages
Some errors may require power to be removed and restored.
Flow-Related Checks
Check for low water level, sucon ow restricons, closed valves, trapped air, too many closed jets and pump prime. On some
systems, even when spa is shut down by an error condion, some equipment may occasionally turn on to connue monitoring
temperature or if freeze protecon is needed.
Sensor-Related
Sensors Balance is Poor – M015
The temperature sensors MAY be out of sync by 2°F or 3°F. Call for Service.
Sensors Balance is Poor* – M026
The temperature sensors ARE out of sync. The fault above has been established for at least 1 hour. Call for Service.
Sensor A Fault, Sensor B Fault – Sensor A: M031, Sensor B: M032
A temperature sensor or sensor circuit has failed. Call for Service.
Miscellaneous Messages
Communicaons error
The control panel is not receiving communicaon from the System. Call for Service.

System-Related
Memory Failure—Checksum Error* – M022
At Power-Up, the system has failed the Program Checksum Test. This indicates a problem with the firmware (operaon program)
and requires a service call.
Memory Warning—Persistent Memory Reset (Persistent Memory Error)* – M021
Contact your dealer or service organizaon if this message appears on more than one power-up.
Memory Failure—Clock Error* – M020
Contact your dealer or service organizaon.
Configuraon Error (Spa will not Start Up)
Contact your dealer or service organizaon.
The GFCI Failure (System Could Not Test the GFCI) – M036
(North America Only) May indicate an unsafe installaon. Contact your dealer or service organizaon.
A Pump Appears to be Stuck ON – M034
Water may be overheated. POWER DOWN THE SPA. DO NOT ENTER THE WATER. Contact your dealer or service organizaon.
Hot Fault – M035
A Pump Appears to have been Stuck ON when spa was last powered. POWER DOWN THE SPA. DO NOT ENTER THE WATER. Contact
your dealer or service organizaon.
Reminder
General maintenance helps.
Reminder Messages can be suppressed by using the Preferences Menu. Press a temp buon to reset a reminder message.
Clean the filter
May appear on a regular schedule, i.e. every 30 days. Clean the filter cartridge as instructed in the owner’s manual.

Test the GFCI (or RCD)
May appear on a regular schedule, i.e. every 65 days.
The GFCI or RCD is an important safety device and must be tested on a regular basis to verify its reliability. Every user should be
trained to safely test the GFCI or RCD associated with the hot tub installaon. A GFCI or RCD (European) will have a TEST and
RESET buon on it that allows a user to verify proper funcon.
WARNING: If freezing condions exist, a GFCI or RCD should be reset immediately or spa damage could result. The end user
should always be trained to test and reset the GFCI or RCD on a regular basis.
Change the water
May appear on a regular schedule, i.e. every 90 days. Change the water in the spa on regular basis to maintain proper chemical
balance and sanitary condions.
Change the Filter
May appear on a regular schedule, i.e. every 30 days. Filters should be replaced regularly to maintain proper spa funcon and
sanitary condions.
Check Ozone
May appear on a regular schedule, i.e. every 365 days. Check your ozone/UV generator to assure properly sanized water.
Service Check
May appear on a regular schedule, i.e. every 365 days. Have a service tech do a check-up on your spa.
Warning! Qualified Technician Required for Service and Installaon
Do not aempt service of this control system. Contact your dealer or service organizaon for assistance.
Follow all owner’s manual power connecon instrucons. Installaon must be performed by a licensed electrician
and all grounding connecons must be properly installed.
Table of contents