PeriOptics Firefly User manual

Solutions in magnification
and portable LED illumination
Instructions for Use
Warnings
Light pod
Inspect light pod prior to use.
Do not use damaged light pod.
Turn the light pod off when not in use.
The light pod can get hot.
If the light pod is dropped or receives a mechanical
impact, ensure that the casing is not damaged.
The electrolyte contained within the lithium cells
can cause irritation to the respiratory tract, eyes
and skin. In addition, cell venting could result in air
contaminants, including corrosive or flammable
vapors. All precautions should be taken to limit
exposure to the electrolyte vapor.
Dispose of light pod in accordance with existing
state and local laws.
Do not use the light pod if ambient temperatures
exceed 30C (86F).
When light pod is in use, always keep the front of
the light at least 6 inches from skin and any surface
being illuminated.
Do not shine light into patient’s eyes.
Do not shine light onto combustible material.
Do not allow contact of the lamp to the patient.
Charging
When charging the light pod, inspect the charging
lights to ensure that charging is active.
Do not simultaneously touch the charging unit
(when plugged in) and the patient.
Use only Firefly charger unit to charge the light
pods. Li-ion batteries can be dangerous and even
explode if improperly charged. The light pods are
designed to be charged by the supplied charging
station only.
General
Check all components for burrs, nicks, and sharp
edges prior to use.
If unit is jarred or dropped, re-inspect prior to use.
Do not use the Firefly system or charge the battery
in an oxygen rich environment.
Do not store liquids above the charger, power
pack or any other Firefly component or accessory.
If a spill occurs into the charger, battery, or LED
lamp, discontinue use. Using the charger after
liquid is spilled into the unit could result in electrical
shock. Clean only as directed.
Do not modify this equipment without
authorization from the manufacturer.
Warranty
DenMat warrants its products to be free of
original defects in material and workmanship
when used for intended purposes. The Firefly
comes with a 1-year warranty.
Warranty Exclusions:
• Breakage or failure due to tampering, misuse,
neglect, accidents, modification or shipping.
• If the instrument is not used in accordance with
manufacturer’s recommendations or instructions.
• If repaired or serviced by anyone other
than DenMat or a DenMat authorized
representative.
Prompt inspection and reporting of missing
or damaged product should be reported to
shipping carrier and DenMat within 3 days
of receipt of product.
If DenMat products or any component thereof
is found to be defective or at variance with the
instrument or component(s) they will be replaced
at no cost to the purchaser. This warranty only
applies to products purchased new from DenMat
or its authorized distributors or representatives.
For any questions or problems, please
contact your local distributor or call
DenMat Customer Care at 800-433-6628
between the hours of 8am to 5pm (Pacific Time)
Monday-Friday. Outside the U.S., please call
1-805-347-7990 or visit denmat.com.
750194300 R:8 09/19CM
©2019 Den-Mat Holdings, LLC. All rights reserved. CaviCide and
CaviWipes are registered trademarks of Metrex Research, LLC.
Replacement Parts Part Number
Power Adapter LED5-BC-A
Battery Charger Unit LED5-BC
Light Pod LED5-B-A2
Mount LED5-MA1
denmat.com English
MDSS GmbH
Schigraben 41
30175 Hannover, Germany
Den-Mat Holdings, LLC
1017 W. Central Avenue
Lompoc, CA 93436 USA

Operating Instructions
1. Charging light pod batteries
a. Plug the power adapter (1) into a power outlet
and connect to battery charger unit (4).
b. A single blue light in the center of the charger
will indicate power to the charger.
c. Place a light pod (3) into the charger unit (4).
Make sure the lens is facing up to engage
charger terminals.
d. When the light pod(s) is inserted into the
charger, the appropriate charger indicator
lights will illuminate.
e. When all four charger lights are illuminated, the light pod is fully charged.
2. Mounting the light
Attach the Firefly mount to your loupes.
Universal Mount attachment to eyeglass frames
Open the clip and slide down over the top of the nose bridge. The padded arms of the clip
should grasp each side of the eyeglass lenses. Tighten thumb screw to secure clip
to frame.
There are many mount options for all glasses styles. Please contact DenMat Customer Care
with mount options and questions.
3. Clip the Firefly light pod to its mount
a. Remove the fully charged light pod from the Firefly charger.
b. Place the rear of the light pod on the mount. It will magnetically “click” into place.
4. Turn on light pod by rotating pod
Rotating the light pod on its mount will cycle through off-high power-low power-off settings.
5. Adjust the direction of the LED beam
To adjust the tension or the lamp head position, use the supplied phillips screwdriver to
tighten or loosen the two small link screws as needed.
5. Removing the Firefly light pod from the mount
To remove the firefly light pod from the magnetic mount twist the light to break the
magnetic seal, then pull to the left or right of the loupes and NOT up or down.
Device Description and Intended Use
The PeriOptix®Firefly Cordless Headlight System is a fully portable lighting system
worn by the operator and is intended to illuminate a working area. It can be worn on a
headband or attached to all types of loupes. The light pod(s) are internally powered
with a rechargeable lithium-ion battery that provides more than 60 minutes of bright,
focused light between charges.
What’s Included:
Cleaning Instructions
Disinfection with a low-alcohol (<25%) solution such as CaviCide®or CaviWipes®is
recommended. If using spray disinfectant, avoid spraying liquid directly on the unit.
Instead, spray onto a soft no-scratch cloth, and wipe the product as per disinfectant’s
directions for use. Clean all lens surfaces with the dry optical cleaning cloth provided with
the unit. Do not use a liquid cleaner on the lens. Should you have any questions, please call
DenMat Customer Care at 800-433-6628.
CAUTION
•DO NOT SUBMERGE PRODUCT OR SUBJECT TO SURFACE SATURATION.
Your product is not waterproof.
•DO NOT SUBMERGE IN ANY LIQUID OR ULTRASONIC CLEANER.
•DO NOT CLEAN VIA AN AUTOCLAVE, CHEMICLAVE, GLUTARALDEHYDE, IODOPHOR,
OR ANY OTHER STERILIZATION METHOD.
• NO MODIFICATION OF THIS EQUIPMENT IS ALLOWED.
• THIS UNIT CONTAINS A LITHIUM-ION BATTERY. AS A RESULT PLEASE BE SURE TO
NOT STORE THE UNIT UNCHARGED FOR AN EXTENDED PERIOD OF TIME AS THIS MAY
CAUSE THE BATTERY TO DISCHARGE AND REDUCE THE LIFETIME OF THE UNIT.
denmat.comEnglishdenmat.com English
110v / 120v
1
4
3
Carrying Case(2) Light PodsLight Pod Mount Battery Charger UnitPower Adapter
1 2 3 4 5
Operating Environment
• 10º – 30º C (50º – 86º F)
• 95% or less humidity
• Maximum barometric pressure 80kPa
• Maximum altitude of operation must be 2000m or less
Transport Environment
• 0º – 40º C (32º – 104º F)
• 95% or less humidity
• 70kPa – 106kPa
Symbols Glossary available at denmat.com/symbols.