manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Phase Technologies
  6. •
  7. Analytical Instrument
  8. •
  9. Phase Technologies 70Xi Installation manual

Phase Technologies 70Xi Installation manual

PhaseTechnology70XiFuelAutosamplerAnalyzerSystem–QuickInstallationandStartUpGuide
Page 1 of 2 Customer Service & Support: 604-241-9568
customer_service@phase-technology.com
1. Remove the following items from the 70Xi analyzer box and
the autosampler box:
70Xi analyzer box:
1 70Xi analyzer
1 Accessory box:
- 1 AC power cable
- 1 dryer bottle
Autosampler box
1 Autosampler
1 Accessory box:
- 1 AC power cable
- 1 RS232 cable
- 1 Teflon tubing
- 2 sample racks
2. The analyzer and autosampler should be set up:
on a flat, vibration-free surface
in a well-ventilated location
3. Put the autosampler on the right hand side of the analyzer,
with the AC power cable connector facing the back.
4. Remove the caps on the back of the analyzer. Keep them
with the shipping boxes and related packing materials.
2 red plastic caps that cover the “PURGE IN” and “PURGE
OUT” fittings
2 brass caps that cover the “SAMPLE IN” and “SAMPLE
OUT” fittings
5. Connect the dryer bottle to the back of the analyzer:
a) Attach the dryer bottle to the
back of the analyzer by fitting
the two mounting bracket
holes over the two mounting
screws.
b) Separate the quick-connect
fittings on the dryer bottle.
Press the latch on the female
fitting and pull them apart.
c) Attach the quick-connect
fittings to the “PURGE IN” and
“PURGE OUT” fittings on the
back of the analyzer. A “click”
sound confirms that the
fittings have snapped into
place.
6. Connect the RS232 cable from the “AUTOSAMPLER” port on
the back of the analyzer to the “HOST” port of the
autosampler.
“AUTOSAMPLER” port on
back of the analyzer
“HOST” port of
the autosampler
7. Connect the 2 sample tubings with a 7/16” wrench:
a) Connect the sample transfer tube
(i.e. tubing connected to the
autosampler probe) to the “SAMPLE
IN” port on the back of the
analyzer.
b) Connect the supplied 1/8” Teflon
tube from the analyzer “SAMPLE
OUT” port to a user-supplied waste
bottle. Ensure the end of this tubing
stays above the effluent level in the
waste bottle at all times (i.e. tubing
end should never be submerged
into the effluent).
8. Load the two sample racks to the autosampler. Make sure
the 4 feet of each rack sit inside the square cutouts of the
autosampler bottom tray.
8 square cutouts on the
bottom tray
2 sample racks on the
bottom tray
9. Connect the AC power cables:
a) Connect the AC power
cable to the back of the
analyzer
b) Connect the AC power
cable to the back of
the autosampler
Plug the power cables into a grounded AC outlet that is a
dedicated circuit. If the power source is unstable or noisy,
use a power surge protector, equipped with a line filter.
PhaseTechnology70XiFuelAutosamplerAnalyzerSystem–QuickInstallationandStartUpGuide
Page 2 of 2 Customer Service & Support: 604-241-9568
customer_service@phase-technology.com
10. Power up the analyzer:
Press the round button at the bottom right hand side at
the front of the analyzer.
The analyzer will perform a set of self tests to verify its
main components, including the optics system and
cooling system, are operating correctly.
If a problem is detected, the self tests will stop and a
message will be displayed on the bottom of the screen.
If any component fails the self tests, press the
INTELLINOSTICS button for further information
and troubleshooting steps.
If the pour point function is available, the analyzer will
automatically perform an auto-purge process after the
self tests.
11. Power up the autosampler:
Switch the power control above the power cable
connector to the “I” position.
The autosampler will perform a series of self-tests,
moving its arm to pre-defined locations.
Warning: The arm moves and rotates at a
moderately fast speed during its self-tests.
The tip of the autosampler probe is sharp.
Caution: Do not place obstacles that would
block or impede the movement of the arm. If
the arm movement is impeded at any point
during the self-test causing it to stop or
stutter, switch off the autosampler and
power it on again to reinitialize.
12. After the self-tests are completed, the analyzer shows
the Vial Setup screen.
The analyzer returns to this screen whenever
the VIAL SETUP button is pressed.
The Sample Cup Temperature, Optics Light Signal, and
Run Count are displayed at the top of the screen.
A set of Favorites buttons is listed on the right hand
side of the screen.
The main area in the middle of the screen is used for
programming test settings for any vial position. The
rows of circles represent the physical vial positions
available on the two vial racks of the autosampler.
The ACCESS button on the lower left hand
corner is used for changing user access levels,
resetting the analyzer, and shutdown.
13. To start a test:
a) Fill a vial with sample all the way to the neck.
Insert a septa inside the vial cap (Teflon side
facing the sample), and screw the vial cap onto
the vial. Place the vial into any available position
on the sample rack.
b) To program a vial using a predefined test setting,
simply drag a suitable favorite button to the vial
position where the corresponding test sample has
been placed on the sample rack, e.g. the diagram
below shows the “Diesel” favorite button can be
dragged from its starting position on the right and
be dropped onto the Vial position in row E and
column 5.
c) Repeat the drag & drop procedure if more
than one vial needs to be programmed.
After all vials have been programmed,
press the START TEST button to start the
test sequence.

Popular Analytical Instrument manuals by other brands

HUBITOOLS HU23100 quick guide

HUBITOOLS

HUBITOOLS HU23100 quick guide

Teledyne 311PC instruction manual

Teledyne

Teledyne 311PC instruction manual

Teledyne 3300TA operating instructions

Teledyne

Teledyne 3300TA operating instructions

PST PST-QSG-3003-01 quick start guide

PST

PST PST-QSG-3003-01 quick start guide

VOLTCRAFT BS-350XIPSD operating instructions

VOLTCRAFT

VOLTCRAFT BS-350XIPSD operating instructions

Jct JES-301E1 operating manual

Jct

Jct JES-301E1 operating manual

A&D Vibratory spoon AD-1651 instruction manual

A&D

A&D Vibratory spoon AD-1651 instruction manual

Shinkawa VM-5 instruction manual

Shinkawa

Shinkawa VM-5 instruction manual

InterTest iShot imaging MZ-5 FS Operation and service manual

InterTest

InterTest iShot imaging MZ-5 FS Operation and service manual

Leica Geosystems DISTO D810 quick start

Leica Geosystems

Leica Geosystems DISTO D810 quick start

Fluke 414D user manual

Fluke

Fluke 414D user manual

DeLaval Herd Navigator precision analytics 100 Instruction book

DeLaval

DeLaval Herd Navigator precision analytics 100 Instruction book

Ferret CFWF50A instruction manual

Ferret

Ferret CFWF50A instruction manual

Sensopart SmartRange FR 90 ILA operating manual

Sensopart

Sensopart SmartRange FR 90 ILA operating manual

MarMonix MLD-705 user manual

MarMonix

MarMonix MLD-705 user manual

Freescale Semiconductor CodeWarrior TAP user guide

Freescale Semiconductor

Freescale Semiconductor CodeWarrior TAP user guide

Metrohm 845 Eluent Synthesizer manual

Metrohm

Metrohm 845 Eluent Synthesizer manual

Vividia ME Series user manual

Vividia

Vividia ME Series user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.