70Xi Sample Injection Port – Quick Installation Guide
Page 1 of 2 Customer Service & Support: 604-241-9568
customer_service@phase-technology.com
1. Remove the ollowing items rom the 70xi analyzer box:
a) 1 70xi analyzer
b) 1 accessory box:
i) 1 AC power cable
ii) 1 dryer bottle
iii) 1 box labeled “Sample Injection Port”:
1 Sample Injection Port Assembly
1 plastic bag containing:
4 mounting screws
1 Sample IN Tubing (with 2 Swage ittings)
1 Sample OUT Tubing (with 1 Swage itting)
2. The analyzer should be set up:
on a lat, vibration- ree sur ace
in a well-ventilated location
3. Remove the caps on the back o the analyzer. Keep them
with the shipping boxes and related packing materials.
2 red plastic caps that cover the “PURGE IN” and “PURGE
OUT” ittings
2 brass caps that cover the “SAMPLE IN” and “SAMPLE
OUT” ittings
4. Connect the dryer bottle to the back o the analyzer:
a) Attach the dryer bottle to the
back o the analyzer by itting
the two mounting bracket holes
over the two mounting screws.
b) Separate the quick-connect
ittings on the dryer bottle. Press
the latch on the emale itting
and pull them apart.
c) Attach the quick-connect ittings
to the “PURGE IN” and “PURGE
OUT” ittings on the back o the
analyzer. A “click” sound
con irms that the ittings have
snapped into place.
5. Connect the AC power cable to the back o the analyzer.
Plug the power cables into a grounded AC outlet that is a
dedicated circuit. I the power source is unstable or noisy,
use a power surge protector, equipped with a line ilter.
6. Mount the Sample Injection Port Assembly onto the
analyzer cover with the 4 mounting screws, using a Philips
screw driver. Make sure the stainless steel tubing o the
Sample Injection Port Assembly point to the back o the
analyzer. Rotate the tubing i needed.
7. Install the Sample IN Tubing:
a) Connect the stainless steel
itting to the Sample Injection
Port Assembly, with two 7/16”
wrenches.
b) Connect the brass itting to the
port labeled “SAMPLE IN” at the
back o the analyzer, with a
7/16” wrench.
8. Install the Sample OUT Tubing:
a) Connect the brass itting to the
port labeled “SAMPLE OUT” at
the back o the analyzer, with a
7/16” wrench.
b) Put the ree end into a user-supplied waste bottle.
Ensure the end o this tubing stays above the e luent
level in the waste bottle at all times (i.e. tubing end
should never be submerged into the e luent).