Phenix Technologies KVM300 User manual

`
KVM300
AC KILOVOLTMETER
MODEL NO. KVM300
VERSION 2.1
Copyright © Phenix Technologies, Inc.
JPG/SLU
MAY, 2008
REV 9/4/’14 nab

KVM300
TABLE OF CONTENTS
Description Section Number
DANGER / WARNINGS 1
TECHNICAL SPECIFICATIONS 2
UNCRATING / MECHANICAL SET-UP 3
ELECTRICAL SET-UP / OPERATIONAL NOTES 4
OPERATING INSTRUCTIONS 5
CALIBRATION 6
MECHANICAL MAINTENANCE 7
STORAGE OF EQUIPMENT 8
TROUBLESHOOTING 9
RECOMMENDED SPARE PARTS 10
PARTS ORDERING INFORMATION 11
RETURNED MATERIAL 12
CIRCUIT DIAGRAM SYMBOLS 13
ELECTRICAL SCHEMATICS 14
CUSTOMER COMMENTS / SUGGESTIONS 15
APPENDIX
TEST REPORT A-1

KVM300
1-1
SECTION 1: DANGER / WARNINGS
WARNING !!
Complete Grounding of this unit is necessary for
the safe operation of this equipment. Disconnect
inputs before ungrounding this equipment

`
KVM300
1-2
DANGER / WARNINGS
GENERAL SAFETY PRECAUTIONS
HIGH VOLTAGE
This equipment is capable of providing POTENTIALLY LETHAL VOLTAGES!
Improper operation or test practices may result in injury or death to the
operator or surrounding personnel.
The operation of High Voltage test equipment should only be performed by
personnel familiar with HIGH VOLTAGE testing and safety procedures. The
operator of this equipment must be aware of all hazards associated with High
Voltage testing. The operator is responsible for himself and others in close
proximity of the testing area.
Some General Safety Practices for working with High Voltage Test Equipment
have been listed below for your reference.
Become familiar with your instrument before performing an actual test
Know your work area, check that all circuits are de-energized and locked
out.
Never work alone; always work with another qualified worker.
Mark off entire work area with barriers and warning tape.
Make all personnel aware of your testing activities.
Be aware of dangerous conditions that may arise from energizing a test
specimen.
Never modify test equipment; modifications to equipment could introduce
an unknown hazard or hinder a designed-in safety feature.
DO NOT operate damaged equipment. Remove power, and do not use
the equipment until safe operation can be verified by service-trained
personnel.
Phenix Technologies, Inc. assumes no liability for unsafe or improper use of test
equipment.

`
KVM300
2-1
SECTION 2: TECHNICAL SPECIFICATIONS
AC/DC KILOVOLTMETER
KVM300
KVM 300 DIVIDER
HIGH VOLTAGE AC INPUT:
300 KVAC maximum
HIGH VOLTAGE DC INPUT:
300 KVDC maximum
HIGH VOLTAGE PK INPUT:
300 KV peak max.
DIVIDER CAPACITANCE:
67 pF
DIVIDER RESISTANCE:
1,200 M
DIVIDER RATIO:
10,000:1
INSTRUMENTATION: AC / BATTERY POWERED
POWER ADAPTOR:
INPUT: 100-240 Volts AC, 0.4 AMPS, 47-63 HZ
OUTPUT: +18 VDC, 0.8 AMP
INPUT JACK:
+18 VDC, 0.8 AMP
BATTERY PACK:
9.6 Volt Ni-MH 3200 mA hr
VOLTAGE MEASUREMENT:
Voltage is measured by means of Peak Responding, RMS or AVG
Circuitry. Display is 4 ½ Digit LED with adjustable backlighting
ACCURACY: 1% of Reading 3 counts from 10% -100% of Range
FUNCTION SELECTIONS:
1. Peak / 2 (Peak responding calibrated to RMS Value)
2. Peak (Peak responding calibrated to Peak Value)
3. RMS (True RMS)
4. AVG (True Average Value, not RMS equivalent)
5. Peak DC (Peak responding including ripple peak
6. Ripple Voltage (RMS value of a DC Ripple voltage)
FREQUENCY RESPONSE:
(SINUSOIDAL WAVEFORM)
Peak / 2, Peak, --Response: DC, 5-1000Hz
RMS, Ripple V --Response: 20 - 1000Hz
AVG --Response: DC, 20-1000Hz
SETTLING TIME:
Peak / 2, Peak --Up to 30 seconds
RMS, Ripple V --Up to 15 seconds
AVG --Up to 5 seconds
*Note: The lower the frequency the longer the settling time.
DIMENSIONS:
KVM 300
Controls 7” H x 12” W x 14” D 11 lbs.
HV 63” H x 26” W x 26” D 80 lbs.

`
KVM300
2-2
1. CONTROL POWER SWITCH: Depressing this switch will turn on or turn off the control power
of the KVM.
2. LOW BATTERY: Battery needs recharged when lamp is illuminated
3. INPUT JACK: External input / charging jack provides for battery charging or AC operation with
included power adapter.
4. INPUT / CHARGING LAMP: Illuminates when external input / charging power is present.
5. GROUND STUD: This Ground stud must be connected to ground for metering module ground.
6. BNC INPUT CONNECTOR: This BNC input connector connects the meter module to the
divider.
7. BACKLIGHT DIMMER: This dial adjusts the intensity of the backlight for use during low light
conditions.
8. METER: This 4 1/2 digit meter displays the measured value of the input voltage.
9. MEASUREMENT RANGE SELECTOR: This rotary selector changes between the low and
high range of the meter. Meter low range should not be used past 30kV. Meter may not be
accurate past 100% of reading and actual voltage may be much higher than indicated.
10. MEASURING FUNCTION SELECTOR: This rotary selector allows selection of the
measurement function of the KVM.
TECHNOLOGIES
PHENIX

KVM300
3-1
SECTION 3: UNCRATING/MECHANICAL SET-UP
Exercise care in removing shipping materials so as not to damage the unit.
Perform a visual inspection to determine if the unit was damaged in shipment. If there are any
signs of physical damage such as dents, scratches, or oil leaks, contact the factory before
proceeding.
NOTE
Read and understand all setup and operating instructions before use of the unit. Failure
to do so may cause damage to the unit and possibly void the warranty.

KVM300
4-1
SECTION 4: ELECTRICAL SET-UP / OPERATIONAL NOTES
Place the test set in the desired location. Divider base must be at ground level. High voltage
should enter divider from above at an angle of no greater than forty-five degrees from
vertical for highest accuracy readings.
Connect a ground of sufficient size from the source ground to the grounding post located at
the base of the divider. (A 10’ lead is supplied with the unit.)
CAUTION: The ground post on the divider base must be connected to the ground or
damage to the unit may result.
Connect the supplied coaxial cable between the low voltage arm located at the base of the
divider and the metering module. (A 25’ lead is supplied with the unit.)
NOTE: The coaxial lead is part of the calibration. Significantly changing the length of the meter
lead will affect the AC calibration.
CAUTION: The meter module has to be grounded for safe and proper operation.
Connect the ground post on the metering module to an appropriate ground (A 10’ lead is
supplied with the unit).
CAUTION: A ground should always be connected to the ground post on the metering
module.
Connect the source voltage to be measured to the top of the divider by an appropriate and
safe method.
CAUTION: Meter low range should not be used past 30kV. Meter may not be accurate past
100% of reading and actual voltage may be much higher than indicated.
Note: Unit is designed to measure voltage in reference to ground. Source voltage
and divider ground post must be ground referenced for unit to operate
properly.
WARNING:
THIS UNIT SHOULD ONLY BE OPERATED BY PERSONS KNOWLEDGEABLE OF HIGH
VOLTAGE TESTING AND SAFETY PROCEDURES. IMPROPER OPERATION MAY
RESULT IN INJURY OR DEATH.
ENSURE THAT UNIT TO BE TESTED IS DE-ENERGIZED AND DISCHARGED! ENSURE
THAT WORKING ENVIRONMENT IS SAFE AND FREE OF HAZARDS.

`
KVM300
5-1
SECTION 5: OPERATING INSTRUCTIONS
1) DIVIDER WITH METER MODULE OPERATION:
Make sure connections have been made as described in “Set-Up” section.
Switch control power switch to ON position.
Select the desired range for the voltmeter based on the expected input signal level.
Select desired mode for voltmeter with function switch.
Note: If the Ripple voltage measurement mode is to be selected, the Range selector must
be set for the required input voltage level. If ripple voltage is to be read on input
voltage over 30kV the range selector has to be on high range, regardless of the
expected ripple voltage level. To avoid damage to the unit ascertain the unit will
not over-range on the AVG or the DC modes before attempting to read the ripple
voltages. DO NOT attempt to read ripple voltage on a range lower than required to
measure the full input voltage.
Activate high voltage source.
Measure voltage as required. Meter low range should not be used past 30kV. Meter may
not be accurate past 100% of reading and actual voltage may be much higher than
indicated.
De-activate high voltage source and assure, as appropriate, that high voltage source is de-
energized and discharged.
When testing is complete, switch control power to OFF position.
After assuring that high voltage source and unit are de-energized and discharged, all leads
and connections may be disconnected and unit packaged for transport or storage.
Do not subject unit to Flashovers. Damage may occur. If meter becomes disabled from
Flashover or Transient condition, refer to Troubleshooting section.
Battery operational time will be extended if Backlight is kept at minimum setting.
Low battery lamp comes on at approximately 20% of charge. If unit is run too long after low
battery lamp comes on, it will shut down and must be recharged before it will turn on again.
WARNING
THIS UNIT SHOULD ONLY BE OPERATED BY PERSONS KNOWLEDGEABLE OF HIGH
VOLTAGE TESTING AND SAFETY PROCEDURES. IMPROPER OPERATION
MAY RESULT IN INJURY OR DEATH.

`
KVM300
5-2
OPERATING INSTRUCTIONS
2) STAND ALONE DIVIDER OPERATION:
Make sure connections have been made as described in “Set-Up” section, except that the
meter module will not be used or connected.
Connect a Voltmeter to the BNC output on the base of the divider with the 25’ coaxial cable.
NOTE: The divider has approximately 10,000 to 1 ratio. At 300kV input the output voltage will be
30V. Readings will correspond to the type of meter used. AVG will show AVG, RMS will show
RMS etc. The meter impedance needs to be 10MDC, and 1 M, ≤ 200pF AC for the highest
possible accuracy. Readings will correspond to the type of meter used. AVG will show AVG, RMS
will show RMS etc. Use DC range for DC signals (10MΩinput meter).
Select desired mode for voltmeter.
Activate high voltage source.
Measure voltage as required.
De-activate high voltage source and assure, as appropriate, that high voltage source is de-
energized and discharged.
After assuring that high voltage source and unit are de-energized and discharged, all leads
and connections may be disconnected and unit packaged for transport or storage.
3) STAND ALONE METER MODULE OPERATION:
It is not recommended to operate the meter head without the divider. The meter head is not a
stand alone voltmeter.

`
KVM300
6-1
SECTION 6: CALIBRATION
It is recommended that calibration be performed on a yearly basis.
CAUTIOIN: CALIBRATION SHOULD ONLY BE DONE BY PERSONS FAMILIAR WITH HIGH
VOLTAGE TESTING AND SAFETY PROCEDURES!
CALIBRATION PROCEDURES
Due to the complexity and the standards required to calibrate this instrument, Phenix
Technologies does not recommend customer calibration. For further information please contact
our Service Department.

`
KVM300
7-1
SECTION 7: MECHANICAL MAINTENANCE
SURFACE
All surfaces will provide adequate protection against the elements in normal use. It is
recommended that the finish be wiped down with an ordinary household cleaner for longer life
and for proper electrical operation of the unit. Also inspect all fabrication joints for oil leakage. If
a leak is found, consult the Phenix Technologies’ Service Department.

`
KVM300
8-1
SECTION 8: STORAGE OF EQUIPMENT
If the equipment will be stored for a prolonged period, the following precautions are
recommended.
The equipment should be covered and kept in a warm, dry environment (95% maximum
humidity, 5 to 50 degrees C).
Prior to placing the equipment back into operation, all aspects of the maintenance schedule
should be strictly adhered to.

`
KVM300
9-1
SECTION 9: TROUBLESHOOTING
METERING MODULE
Unit will not turn on: Battery may be discharged. Recharge battery
Unit blanked out during testing because of transient discharge occurrence and won’t come
back on by cycling the power switch: Carefully remove front panel, and unplug the battery
pack for approximately 30 seconds, then retry power switch.
Unit does not measure correctly after flashover or transient discharge occurrence: U5
possibly damaged. Replace with LT1055 OP Amp.

`
KVM300
10-1
SECTION 10: RECOMMENDED SPARE PARTS
Phenix Technologies recommends that the customer purchase and stock the following parts for
normal maintenance of the unit. The recommended quantity should be sufficient to support the
unit during normal operation.
If the unit will be operated at an isolated site for an extended period or will be subjected to
unusual stresses, a larger quantity of parts should be stocked as spares. In such a case, contact
your Phenix Technologies’ sales representative for a recommendation.
Current prices may be obtained by contacting the Parts Ordering Department at Phenix
Technologies.
Part Number
Description
Quantity
30050003
25’ COAXIAL CABLE
1
30080007
10’ GND CABLE
1
1864003
INPUT POWER SWITCH-SW2
1
1603605
F1, 5A
1
1863037
RANGE / MODE SWITCH
1
1506407
VOLTMETER-KNS DMO-782
1
1761087
BACKLIGHT POTENTIOMETER
1
1077167
MODULAR POWER CORD
1
1795107
U5, LT1055 OP AMP
2
1590006
15 VDC POWER / CHARGING SUPPLY
1
1590008
BATTERY PACK
1

`
KVM300
11-1
SECTION 11: PARTS ORDERING INFORMATION
Replacement parts are available from Phenix Technologies, Inc.
Changes to Phenix Technologies' products are sometimes made to accommodate improved
components as they become available, and to give you the benefit of the latest technical
improvements developed in our Engineering Department. It is, therefore, important when
ordering parts to include the serial number of the unit as well as the part number of the
replacement part.
When your purchase order is received at our office, a representative of Phenix Technologies will
contact you to confirm the current price of the part being ordered. If a part you order has been
replaced with a new or improved part, an Applications Engineer will contact you concerning any
change in part number.
Send orders for replacement parts to:
Service Department
Phenix Technologies, Inc.
75 Speicher Drive
Accident, Maryland 21520

`
KVM300
12-1
SECTION 12: RETURNED MATERIAL
If for any reason it should become necessary to return this equipment to the factory, the Service
Department of Phenix Technologies, Inc. must be given the following information:
Name Plate Information
Model Number
Serial Number
Reason for Return
Cause of Defect
If Phenix Technologies, Inc. deems return of the part appropriate; it will then issue an
"Authorization for Return."
If return is not deemed advisable, other inspection arrangements will be made.
NOTE: Material received at this plant without the proper authorization shall be held as
"Customer's Property" with no service until such time as the proper steps have been taken.
Your cooperation is requested in order to ensure prompt service.

`
KVM300
13-1
SECTION 13: CIRCUIT DIAGRAM SYMBOLS
Dispositif De Sur
Interrupteur Normalement
Interrupteur Normalement
Maintenu Ouvert
Interrupteur Normalement
Maintenu Ferme
Ouvert Momentanement
Ferme Momentanement
Interrupteur Normalement
DESCRIPTION
Condensateur
Unite d'amplificateur
Relais, Contacteur
Transformateur
Condensateur electrol
Transformateur de
Resistance Variable
Prise de Courant
Contact Normalement
Contact Normalement
Changement
Insrument Analogue
Diode Zener
Cable blinde
T Transformer
Device
Current Overload
Momentary Switch
Normally Open
Momentary Switch
Normally Closed
Normally Closed
Maintained Switch
Maintained Switch
Normally Closed
Normally Open
Normally Open
Thyristor
Zener
Diode
Analog Meter
Shielded Wire
Transistor
Changeover Contact
Relay Contact
Relay Contact
Connector
Terminal Block
DP
SW
SW
SW
SW
SCR
Z
D
TR
M
K
K
K
TB
X
Intensite
Diode
Transisteur
Thyristor
Borne
Ouvert
Ferme
Contact de
SYMBOLE ZU SCHEMA
SYMBOLES POUR SCHEMA DE CIRCUIT
CIRCUIT DIAGRAM SYMBOLS
Variable Resisitor
Resistor
Lamp, Indicator
Neon
Movistor
Motor
Inductor
Relay, Contactor
Circuit Breaker
Current Transfomer
Fuse
Electrolytic Capacitor
Bushing
Capacitor
Surge Arrestor
Amplifier
BSHG
LP
R
R
NE
MOV
MOT
CB
K
L
CT
F
C
SYMBOL
ARSR
C
REF
A
Tranversee
Parafoudre
Moteur
Parafoudre
Lampe
Resistance
Courant
Fusible
Interupteur
Self
DESCRIPTION
Parafoudre
Transformer
Oeffnungskontakt
Abgeschirmetes Kabel
UeberstromschutzEinheit
Schrittschalter
Druckschalter
(Schliesser)
Schrittshalter
(Oeffner)
(Schliesser)
(Oeffner)
Druckschalter
Analog Meter
Transistor
Thyristor
Diode
Zener
Losbare Klemme
Steckverbindung
Schlierskontakt
Umschaltkontakt
BEMENKUNG
Ueberspannungsableiter
Ueberspannungsableiter
Eleckrolytik kondensator
Stromtransformer
Durchfuehoung
Widerstand
Movistor
Meldeleuchte
Widerstand
Motor
Sicherung
Drossel, Spule
Relais, Schutz
Unterbrecher
Kondensator
Verstarker

`
KVM300
14-1
SECTION 14: SCHEMATICS AND DRAWINGS
Drawing Number
Description
1. 7916004
KVM300 SCHEMATIC

`
KVM300
15-1
SECTION 15: CUSTOMER COMMENTS/SUGGESTIONS
Phenix Technologies made significant efforts to ensure that the materials in this
Operator’s Manual are correct. If there are concerns or comments as you have
used this information, Phenix Technologies appreciates any feedback.
Unit Serial Number:
Sect
Page(s)
Comment
Please return to Phenix Technologies, Engineering Department, 75 Speicher
Drive, Accident, MD 21520 USA.
Table of contents
Other Phenix Technologies Measuring Instrument manuals

Phenix Technologies
Phenix Technologies MRM-10 User manual

Phenix Technologies
Phenix Technologies 400P User manual

Phenix Technologies
Phenix Technologies KVM50A User manual

Phenix Technologies
Phenix Technologies MRM-100 User manual

Phenix Technologies
Phenix Technologies KVM100 User manual

Phenix Technologies
Phenix Technologies MRM-100-V2 User manual

Phenix Technologies
Phenix Technologies MRM10E User manual

Phenix Technologies
Phenix Technologies KVM 100A User manual