9
MRM10E
5. Messages
MRM-10E
When turning the equipment On using the On/Off switch, this introductory message appears briefly. During
that time, the equipment carries out functional checking.
WAIT...
This message appears each time the equipment needs to adjust any parameter to optimize the readings.
PRESS START
The equipment is ready to start a measurement, thus the operator has to press the Start key.
LOW CURRENT
It shows that the test current is not enough to carry out the reading. It appears at the beginning of each test
and it keeps on being there up to the operator rotates the test current control clockwise, as necessary for the
test current to be higher than 10 % of the nominal current in the scale. The inappropriate connection of the
cables may cause a difficult circulation of test current. If this message keeps on being displayed, please check
that the current cables are connected appropriately.
OVERRANGE
It indicates that the measured resistance is higher than the maximum value readable in the selected range.
H
Indicates that the value is the one retained in the memory when pressing the hold key.
BAT
It shows that the battery is discharged. It is necessary to recharge the battery.
6. Some Notes About Accuracy
To obtain the specified accuracy, the operator has to adjust the test current to a value higher than the 80% of
the nominal value. If it is necessary, it is possible to use a lower current, but by doing this the accuracy will be
affected.
MRM-10E has an auto-compensation system that automatically eliminates the error produced by internal
offset. Thus, it is not necessary to carry out measurements by reversing the polarity to compute the average
value. Nevertheless, if the operator suspects that there is a difference of temperature between the contact
points that may generate thermoelectric voltages, it is necessary to carry out two measurements by reversing
the current cables and so, the circulation sense of the current through the resistance under measurement. The
resistance value to be measured will be the average between the values in one sense and in the contrary
(direct and inverse current).