pickeringtest.com
LXI Modular Reed Relay Matrix 65-223
ISSUE J APR 2023
yModular Matrix Design With 8 Y Axis Connections
yUser Configurable for X Dimensions
by Plug-in addition, up to 768x8
ySix Analog Buses
ySwitch up to 1 A
yUses High Quality Pickering
Ruthenium Reed Relays
yBuilt-In Scan List Sequence Stores With
Triggering Capability
ySupported by
BIRST
™ &
eBIRST
™ Test Tools
yFully Compliant to 1.5 LXI Standard
y3 Year Warranty
The 65-223 is a modular matrix platform that provides a large
reed relay matrix solution with 1 A current carrying capacity on all
Y paths.
Matrices are created by populating the 65-200 chassis with
plug-in modules that provide access to the X and Y axis on 200-
pin LFH connectors. The chassis is capable of supporting matrices
with a Y axis size of 8 and X axis sizes up to 768 in increments of
64. Users can specify as many or as few plug-in modules as they
require and can field upgrade the chassis to extend the matrix.
The chassis supports up to 6 plug-in modules. These can be with
or without direct Y bus access and are loaded into the chassis
via the front panel, greatly improving flexibility and ease of
maintenance.
Analog Buses
Each plug-in is a single or dual 64x8 sub matrix with the ability to
connect its Y axis to one of six analog buses. This provides a high
degree of flexibility in configuring the matrix. The analog buses
can be used to maximize bandwidth by disconnecting unused stub
matrices from the used bus, or it can be used to divide the 65-223
into six independent matrices whose size can be set by the user
in increments of 64 X connections. For example, six independent
banks of 128x8 matrix.
Simplified Diagram For 65-223
Configured as a 768x8 Matrix
Built-In Relay Self-Test -
BIRST
The BIRST facility provides a quick and simple way of finding
relay failures. No test equipment is required, simply un-plug the
user connectors, launch the BIRST application and the tool will
run a diagnostic test that will find all relays with faulty contacts.
For more information go to: pickeringtest.com/birst
Supported by
eBIRST
The matrix is also supported by our eBIRST test tools. These tools
simplify switching fault-finding by quickly testing the system and
graphically identifying the faulty relay.
For more information go to: pickeringtest.com/ebirst
Cooling for the 65-200 is provided by rear fans and front air
intakes ensuring no wasted space when the chassis is rack
mounted. The cooling system is adaptive, ensuring acoustic noise
is minimised when used in quiet laboratory environments and low
load conditions.
Configuring the Matrix
To select the parts that you need to create a matrix simply:
• Specify a 65-200-002 Modular Matrix Chassis.
• Add Matrix Plug-in Modules with direct Y access as required.
• Add Matrix Plug-in Modules without direct Y access as
required.
Both styles of plug-in module have options for single or dual 64x8
configurations, with six analog buses.
Pickering’s Range of LXI Modular Reed Relay Matrices
Y-Axis X Range Analog Buses Model No.
4 128 to 1536 12 or 6 65-221
8 64 to 768 6 65-223
16 32 to 384 3 65-225
32 32 to 192 1 65-227
X768
X765
X766
X767
Y2
Y4
Y1
Y3
X2
X4
X8
X1
X3
X5
X6
X7
X9
X10
Y6
Y8
Y5
Y7
MATRIX
CROSSPOINT
Front Panel Connector
This product is based on the obsolete Molex LFH series connector
that has been superseded by a Pickering commissioned form, fit,
function equivalent. The new connector series is 100% compatible
with the Molex connectors allowing either gender of Pickering
connector to mate with the corresponding Molex part without
issue.