4) Temperature unit
Hold press the button to switch unit between ℃/℉.
5) Calibration
Before first use, please use storage solution to activate electrode. Please
refer to safety use & care for more details.
(1)Two point auto-calibration
SA1:PH=6.86/ 6.86
SA2:PH=4.00/ 4.00
Automatic calibration --- automatically identify the calibration solution
in sequence of 6.86-4.00.
1) Power on pH meter, hold & press the button , the LCD will
display [CAL] and flash [SA1] at the same time.
2) Putting sensor into PH=6.86 solution (the glass bulb should be
submerged completely), short press , the ADC value (a 3 digital
number) will be displayed.
3) Wait for the ADC value to stabilize (stir gently for a few seconds)
and register this value by short pressing the again (or wait for 3
minutes to auto-calibrate this point). Remove the sensor from this
solution and clean bulb with distilled water at least 1 minute, then wipe
it gently with a soft tissue paper. 4) When the LCD flashes [SA2], put
the sensor into the PH=4.00 solution. Short press , the ADC
value will be displayed. Wait for the ADC value to stabilize and register
it by short pressing the again (or wait for 3 minutes to auto-
calibrate). When complete the calibration it will display [End]. Wait for