PRO-TEK D705 User manual



True-RMS Multimeter
2
Warning. Read before use
To avoid possible electrical shock or personal injury, follow these instructions and guidelines.
▶ Make sure that the Meter should be used only as specied in this manual
▶ Stop using the Meter or test leads if they look damaged, or if the Meter is not operating
properly
▶ Make sure that only the rated voltage should be applied to the Meter, as marked on meter
▶ Replace the battery immediately as soon as the low battery indicator( ) is displayed
▶ Disconnect the test lead from test point before changing the meter function and range
▶ Do not measure voltage out of specication(over 1,000V)
▶ Verify the meter’s operation by measuring a known voltage
▶ Use caution with voltages above 20V ac rms, 42 V ac peak, or 60 V dc.
▶ Disconnect circuit power and discharge all high voltage capacitors before testing resistance,
continuity, diodes, or capacitance.
▶ Remove the test leads before opening the case or battery cover.
▶ The Meter is for indoor use only(IPX0)
▶ Use the Meter with test probe(Model # GTP01)provided by GSI.
▶ The protection provided by the Meter may be impaired if the Meter is used ina manner not
specied by the manufacturer.

User Manual
3
International Symbols

True-RMS Multimeter
4
“Warning” and “Caution” Statements
A “ Warning” identies hazardous conditions and actions that could result in bodily harm or
death.
A “Caution” identies conditions and actions that could damage the Meter, the equipment under test,
or cause permanent loss of data.
Contact Information
GS Instruments Co., Ltd
1385-14, Nam-Gu, Juan-Dong, Incheon 402-200, Rep. of Korea
Visit GSI website : www.gsinstrument.com
For Technical Assistance
Call to 82.32.870-5656

User Manual
5
Table of Contents
Safety Information(Read before Use)...........................................................................................2
Warning Statement.....................................................................................................................4
Contact Information....................................................................................................................4
Introduction..............................................................................................................................6
Overview...................................................................................................................................6
Nominal Specications................................................................................................................8
Features....................................................................................................................................9
General Specications...............................................................................................................10
Terminals................................................................................................................................12
Rotary Switch Positions.............................................................................................................13
Display symbols........................................................................................................................16
Detailed Specications...............................................................................................................18
Making basic measurements(How to use)...................................................................................23
Cleaning..................................................................................................................................29
Testing the Fuses.....................................................................................................................30
Replacing battery and fuses......................................................................................................30
Warranty.................................................................................................................................31

True-RMS Multimeter
6
Introduction
Overview
The Protek D705/D705T are true-RMS multimeters and battery-powered with a 5000-counts, 25
segments bar graph. These multimeters are fully compatible to CAT III and CAT IV IEC 61010
standards, which denes four overvoltage categories(CAT I to IV) based on the magnitude of danger
from transient impulses.
Measurement Category I:
This category is for measurements of voltages from specially protected secondary circuits. Such
voltage measurements include signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.
Measurement Category II:
This category refers to local-level electrical distribution, such as that provided by a standard wall
outlet or plug in loads (for example, 115 AC voltage for U.S. or 200 AC voltage for Europe). Examples
of Measurement Category II are measurements performed on household appliances, portable tools,
and similar modules.
Measurement Category III:
This category refers to measurements on hard-wired equipment in xed installations, distribution
boards, and circuit breakers. Other examples are wiring, including cables, bus bars, junction boxes,

User Manual
7
switches, socket outlets in the xed installation, and stationary motors with permanent connections
to xed installations.
Measurement Category IV:
This category refers to origin of installation or utility level measurements on primary over-current
protection devices and on ripple control units.
Rated Voltage IEC 61010-1 2nd Edition UL 61010B-1 (UL 31111-1)
CAT IV CAT III CAT II CAT III CAT II CAT I
150V 4,000V 2,500V 1,500V 2,500V 1,500V 800V
300V 6,000V 4,000V 2,500V 4,000V 2,500V 1,500V
600V 8,000V 6,000V 4,000V 6,000V 4,000V 2,500V
1,000V 12,000V 8,000V 6,000V 8,000V 6,000V 4,000V
Resistance 2 ohms 2 ohms 12 ohms 2 ohms 12 ohms 30 ohms

True-RMS Multimeter
8
Nominal Specications
Function Absolute Range or Description
AC Voltage, True-rms 1mV to 1000V (1kHz)
DC Voltage 1mV to 1000V
Continuity Beeper guaranteed on < 25Ω, Guaranteed off > 60Ω;
Detects opens or shorts of 1ms or longer
Resistance 0.1Ω to 50.00MΩ
Diode Test 1.6V
Capacitance 1nF to 4000uF
AC Current, True-rms 0.01mA to 10.00A (20.00A over-range for 30 seconds)
DC Current 0.01mA to 10.00A (20.00A over-range for 30 seconds)
Frequency 10Hz to 999.9kHz
Temperature(D705T) -40℃ to +400℃; -40℉ to +752℉
Basic dc voltage accuracy 0.1%
Basic ac voltage accuracy 1%

User Manual
9
Features
Feature Description
Digital Display 5,000 counts, 3-4/5 digit
Analog Bargraph display 25 segments
Backlight Automatically turns off after 1 minutes to save battery life
MIN MAX AVG Minimum, maximum and average reading memory
Manual or auto ranging In auto range, the meter selects the range with the best resolu-
tion for the present measurement value
Auto Power Off The meter shuts off automatically after 30 minutes if it is not
used.
Data Hold HOLD. Press this button to turn hold mode ON and OFF. When
the hold mode is activated, the meter beeps, freezes the display,
and displays the HOLD indicator on the LCD. HOLD mode freezes
the display for later view.
Range Hold RANGE. Use the RANGE button to manually select a range. Press
and hold RANGE button for two seconds to return the meter to
auto range mode. The meter is in auto range mode when the
AUTO indicator is on.

True-RMS Multimeter
10
Safety Rating IEC 1010 CAT IV 600V, CAT III 1000V
Low battery indicator 2.4V
Battery type 2 x 1.5V AAA
General Specications
Accuracy is specied for 1 year after calibration, at operating temperatures of +18℃ to +28℃,
with relative humidity of 0% to 90%.
Maximum voltage between any terminal and
earth ground
1000V DC or AC RMS
Surge protection 1KV peak per IEC1010
Fuse for mA inputs 440mA, 1000V Fast Fuse
Fuse for A input 11A, 1000V Fast Fuse
Display Digital : 5,000 counts
Bar Graph : 25segments
Frequency : 9,999 counts
Capacitance : 4,000 counts

User Manual
11
Altitude Operating : 2000m; Storage : 12000m
Temperature Operating: +10℃ to +35℃
Storage: 0℃ to + 40℃
Electromagnetic compatibility In an RF eld of 3 V/M, accuracy – specied
except in temperature : specied accuracy ±
5℃, ± 9℉
Relative humidity 0% to 90% @ 0℃ to 35℃
0% to 70% @ 36℃ to +50℃
Size, with holster(LXWXH) 89 x 180 x 48
Weight 350g
Safety Compliance IEC1010 to 1000V Overvoltage Category III,
600V Overvoltage Category IV
Certications UL(E346942), CE

True-RMS Multimeter
12
Terminals
Item Description
1 Input terminal for AC and DC milliamp measurements to 400mA and frequency measurements.
2Input terminal for AC and DC current measurements to 10 A and frequency measurements.
3Input terminal for voltage, continuity, resistance, diode, capacitance, frequency, and temperature
(D705T only) measurements.
4 Common (return) terminal for all measurements.

User Manual
13
Rotary Switch Position
Switch
Position Measurement Function
AC voltage from 1mV to 1000V
DC voltage from 1mV to 1000V
Frequency from 1 Hz to 999.9kHz
Ohms from 0.1 Ω to 50MΩ
Beeper guaranteed on < 25Ω, Guaranteed off > 60Ω;
Detects opens or shorts of 1ms or longer

True-RMS Multimeter
14
Diode test. Displays OL above 1.6 V
Farads from 1nF to 4000 μF.
Temperature -40℃ to + 400℃
Temperature -40℃ to +752℉
AC uA from 30.0 uA to 4000uA
DC uA from 0.1uA to 4000uA
AC mA from 3.00 mA to 400mA
DC mA from 0.01mA to 400mA
AC A from 0.300 A to 10A
DC A from 0.001A to 10A

User Manual
15
Function Button
Button Function
RANGE Set measurement range such as DCV, ACV, ohm
REL Set default as “0”
HOLD Press Shortly: Hold
Press down until the backlight is on
MAX/MIN MAX: Maximum Reading memory
MIN: Minimum Reading memory
SELECT Change unit of measurement

True-RMS Multimeter
16
Display symbols
No. Symbol Meaning
1Relative
2Negative reading
3Continuity Test
4Diode Test
5Low battery. Replace battery.

User Manual
17
6Auto Hold is enabled.
7MIN MAX enabled.
8Direct current, alternating current
9~13 Measurement units
14 Analog display bar graph

True-RMS Multimeter
18
Detailed Specications
DC Voltage
Range Resolution Accuracy
500 mV 0.1 mV ± (0.09 % + 2 d)
5 V 0.001 V ± (0.09 % + 5 d)
50 V 0.01 V ± (0.09 % + 2 d)
500 V 0.1 V ± (0.09 % + 2 d)
1000 V 1 V ± (0.09 % + 2 d)
AC Voltage
Range Resolution Accuracy
(45 Hz ~ 500 Hz) (500 Hz ~ 1 KHz)
500 mV 0.1 mV ± (0.8 % + 5 d) True RMS ± (1.0 % + 5 d) True RMS
5 V 0.001 V ± (0.8 % + 5 d) True RMS ± (1.0 % + 5 d) True RMS
50 V 0.01 V ± (0.8 % + 5 d) True RMS ± (1.0 % + 5 d) True RMS
500 V 0.1 V ± (0.8 % + 5 d) True RMS ± (1.0 % + 5 d) True RMS

User Manual
19
1000 V 1 V ± (0.8 % + 5 d) True RMS ± (1.0 % + 5 d) True RMS
DC Current
Range Resolution Accuracy
500 uA 0.1 μA ± (1.0 % + 3 d)
5000 uA 1 μA ± (1.0 % + 3 d)
50 mA 10 μA ± (1.0 % + 3 d)
400 mA 100 μA ± (1.0 % + 3 d)
5A 1 mA ± (1.0 % + 5 d)
10A 10 mA ± (1.0 % + 3 d)
AC Current
Range Resolution Accuracy
(45 Hz ~ 500 Hz) (500 Hz ~ 1 KHz)
500 uA 0.1 μA ± (0.8 % + 5 d) True RMS ± (2.0 % + 5 d) True RMS
5000 uA 1 μA ± (0.8 % + 5 d) True RMS ± (2.0 % + 5 d) True RMS
50 mA 10 μA ± (0.8 % + 5 d) True RMS ± (2.0 % + 5 d) True RMS
400 mA 100 μA ± (0.8 % + 5 d) True RMS ± (2.0 % + 5 d) True RMS
This manual suits for next models
1
Table of contents
Languages:
Other PRO-TEK Multimeter manuals