Prostat PAR-809C User manual

User Manual
PAR-809C
VARIABLE RESISTANCE REFERENCE


PROSTAT® PAR809C VARIABLE RESISTANCE REFERENCE
Sec on Topic Page
I. Introduc on 4
II. Cau ons & Warnings 4
III. Inspec on & Prepara on for use 4
IV. Applica ons 5
V General Opera ons 5
VI. Maintenance 9
General Specifi ca ons 10
Table of Contents
Copyright © 2011 by Prostat® Corpora on. All rights reserved. Printed in the United States of America. No part of this
manual may be used or reproduced in any manner whatsoever without wri en permission. For informa on contact
Prostat Corpora on, 1072 Tower Lane, Bensenville, IL 60106 USA
Prostat is the registered trademark of Prostat® Corpora on

I. Introduc on
A. The Prostat PAR-809C Variable Resistance Refer-
ence is a precision wide range resistance instru-
ment used to measure the resistance range per-
formance of wrist strap testers, footwear testers
and other resistance indica ng meters, fi xtures
and instruments.
1. The device under test’s resistance perfor-
mance range is indicated by the device’s
Pass/Fail lights when the PAR-809C is posi-
oned to a selected resistance.
2. Every PAR-809C is individually inspected
and measured for precise resistance of each
reference posi on before leaving the factory
and recorded.
B. The PAR-809C is intended as a resistance reference for DC circuits of 100 volts or less.
1. The factory resistance references are documented at 10 and 100 volts DC.
2. Inspec on and re-cer fi ca on of resistance reference values should be conducted annually.
II. Cau ons & Warnings
A. Do not use the PAR-809C on circuits exceeding 100 volts DC.
B. As with any electrical device, use proper electrical precau ons to avoid personnel shock.
C. Only qualifi ed instrument repair personnel should open or repair the PAR-809C.
D. Do not store or use in damp environments.
E. PAR-809C Equipment and Accessories
1. One PAR-809C Variable Resistance Reference Instrument
2. One 9 volt DC ba ery used to power posi on indica on lights.
3. Two (2) 24 – 30 inch test leads equipped with male banana plugs.
III. Inspec on & Prepara on for use
A. Install provided 9 volt ba ery by sliding ba ery com-
partment open and fi rmly snapping ba ery terminals
to the appropriate connectors. Ba ery can only be
connected in one way. Re-close the ba ery compart-
ment.
B. Turn ON the ON/OFF slide switch located at the side
of the unit: vary the rota ng posi on knob to insure
all indica ng lights are func onal.
4Rev. B / January 2012
PAR-809C Variable Resistance Reference

1. Slide ON/OFF switch to OFF posi on un l unit is intended for use.
2. NOTE that the ba ery provides power to indica ng lights only.
a. The ba ery does not aff ect the func on of the instrument as a resistance reference.
b. The posi on lights are provided as a convenience for the opera on of the unit during use.
IV. Applica ons
A. The PAR-809C is a precision resistance reference constructed using 1 percent tolerance resistors to
assure accuracy. It is designed to func onally test the resistance range and performance of a va-
riety of equipment during rou ne audits, including: in-plant wrist strap checker PASS/FAIL range,
footwear checker PASS/FAIL range, and other resistance indica ng devices.
B. Used in conjunc on with the Prostat PRS-812 or PRS-801 Resistance Meters, the PAR-809C func-
onally tests performance of the Prostat PWA-805 Wrist Strap Auditor. Calibra on of resistance
meters is easily confi rmed using the PAR-809C.
C. Proper use of the PAR-809C insures accurate, func onal performance confi rma on of cri cal ESD
control tools.
V. General Opera ons
A. Confi rming Wrist Strap Tester Resistance Performance Range
1. Plug the two supplied test leads into the PAR-809C receptacles.
2. Plug one PAR-809C test lead into the wrist strap tester’s Wrist strap Ground Cord receptacle.
3. Rotate the PAR-809C’s Posi on Selector knob to Posi on 1, the lowest resistance reference.
4. Turn ON the PAR-809c’S posi on indica ng lights by sliding the ON/OFF switch located on the
side of the instrument to the ON posi on.
5. Insure the wrist strap device under test is ON and func onal.
6. Using the second PAR-809C test lead with metal alligator clip installed, press the Push to Test
bu on or plate on the wrist strap tester.
a. The wrist strap tester PASS or REJECT/FAIL indica on should actuate upon contact with the
PAR-809C test lead alligator clip to the wrist strap tester’s push to test bu on or plate.
b. Note the reference resistance value and the response of the wrist strap device under test.
7. Vary the PAR-809C reference resistance by rota ng the Posi on Selector Knob on the face of
the PAR-809C to the next posi on and repeat step 6.
5
Rev. B / January 2012
PAR-809C Variable Resistance Reference

8. The resistance range posi ons where the device under test indicates PASS are those resistance
values accepted by the wrist strap tester.
a. Resistance posi ons below the PASS range should all be rejected by the device under test
as too low.
b. Resistance posi ons above the PASS range should all be rejected by the device under test
as too high.
B. Confi rming Foot Wear Tester Resistance Performance Range
1. Plug the two supplied test leads into the PAR-809C receptacles.
2. A ach an alligator clip to one PAR-809C test lead. A ach the alligator clip to the Foot Wear test
plate.
6Rev. B / January 2012
PAR-809C Variable Resistance Reference

3. Rotate the PAR-809C’s Posi on Selector knob to Posi on 1, the lowest resistance reference.
4. Turn ON the PAR-809C’S posi on indica ng lights by sliding the ON/OFF switch located on the
side of the instrument to the ON posi on.
5. Insure the Foot Wear device under test is ON and func onal.
6. Using the second PAR-809C test lead with a metal alligator clip installed, press the push to Test
Bu on or plate on the Footwear tester.
a. The Foot Wear PASS or REJECT/FAIL indica on should actuate upon contact with the PAR-
809C test lead alligator clip to the Foot Wear tester’s push to test bu on or plate.
b. Note the reference resistance value and the response of the Foot Wear device under test.
7. Vary the PAR-809C reference resistance by rota ng the Posi on Selector Knob on the face of
the PAR-809C to the next posi on and repeat step 6.
8. The resistance range posi ons where the device under test indicates PASS are those resistance
values accepted by the Foot Wear tester.
a. Resistance posi ons below the PASS range should all be rejected by the device under test
as too low.
b. Resistance posi ons above the PASS range should all be rejected by the device under test
as too high.
C. Confi rming Resistance Measurement Performance of analog
Megohmmeters (Prostat PRS-800, 3-M-701 or similar)
1. Check the BATTERY func on of the meter; replace bat-
teries if indica on is LOW.
2. Turn the meter to its OFF posi on.
3. Install the meter test leads in the meter lead receptacles.
7
Rev. B / January 2012
PAR-809C Variable Resistance Reference

4. Plug the two meter test lead banana plugs into the PAR-809C receptacles.
5. Rotate the PAR-809C’s Posi on Selector knob to Posi on 1, the lowest resistance reference.
6. Turn ON the PAR-809C’s posi on indica ng lights by sliding the ON/OFF switch located on the
side of the instrument to the ON posi on.
7. Turn the Meter switch to the con nuity posi on, push the test bu on on the meter and record
the measurement.
8. Vary the PAR-809C reference resistance by rota ng the Posi on Selector Knob on the face of
the PAR-809C to the next posi on; record measurement as in step 7.
9. Select the 10 V posi on on the Megohmmeters, push the Test Bu on on the Megohmmeter
and record measurement.
10. Vary the PAR-809C reference resistance by rota ng the posi on Selector Knob on the face of
the PAR-809C to successive posi ons, and record measurements as in step 9.
11. The indicated resistance of the Megohmmeter should be within 10 percent of the factory resis-
tance indicated on the PAR-809C Calibra on sheet for each posi on tested.
D. Confi rming Func onal Performance of the PWA-805 Wrist Strap Auditor
NOTE
The PRS-812 or PRS-801 is used for this func onal test procedure.
1. Check the BATTERY func on of the Resistance Meter; replace ba eries if indica on is LOW.
2. Turn the Resistance Meter to its OFF posi on
3. Install the meter test leads into the proper receptacles.
4. Plug the tow test lead banana plugs into the PWA-805 Meter receptacles
5. Plug the two supplies test leads into the PAR-809C receptacles
6. A ach an alligator clip (not provided) to one PAR-809C test lead. A ach the alligator clip to the
Right hand cuff cylinder of the PWA-805 Wrist Strap Auditor.
7. Install the second PAR-809C test lead banana plug in the PWA-805 Cord receptacle
8. Posi on the PWA-805 three way rocker switch to its middle posi on, marked “Cord & Cuff ”.
9. Rotate the PAR-809C’s posi on Selector knob to Posi on 1, the lowest resistance reference
10. Turn ON the PAR-809C posi on indica ng lights by sliding the ON/OFF switch located on the
side of the instrument to the ON posi on.
11. Turn the resistance meter on in the AUTO mode. Push the Test bu on on the meter and record
the measurement.
8Rev. B / January 2012
PAR-809C Variable Resistance Reference

12. Vary the PAR-809C reference resistance by rota ng the Posi on Selector Knob on the face of
the PAR-809C to successive posi ons and record measurements as in step 11.
13. The indicated resistance of the resistance meter should be within 10 percent of the factory
resistance documented.
SUMMARY COMMENT
The PAR-809C is a wired test fi xture and switch assembly. There
are no resistors or electronic devices employed in this fi xture.
VI. Maintenance
A. Replace 9 volt ba ery at least once each year. Remove ba ery when device is in extended storage,
i.e. six months or greater
B. Periodically confi rm resistance values of each resistance posi on at 10 volts using precision wide
range ohmmeter. Contact Prostat Corpora on for calibra on Service if necessary.
C. Resistance measurement confi rma on should be conducted once each year using a precision wide
range ohmmeter traceable to NIST.
D. Wipe foreign materials from case, receptacles and controls using a so dry cloth.
9
Rev. B / January 2012
PAR-809C Variable Resistance Reference

PAR-809C Variable Resistance Reference Specifi ca ons
Resistance Reference & Posi on: Resistance in Ohms
Posi on #1: 1.2x104 Posi on #5: 1.5x106 Posi on #9: 3.7x107
Posi on #2: 1.0x105 Posi on #6: 9.5x106 Posi on #10: 9.5x107
Posi on #3: 7.6x105 Posi on #7: 1.05x107 Posi on #11: 1.05x108
Posi on #4: 8.45x105 Posi on #8: 3.3x107 Posi on #12: 1.0x109
Case Size (L x W x H): 4.0” x 3.0” x 2.0” (11.4 cm x 9.5 cm x 5.1 cm)
Controls: 12 Posi on Rotary Range selector knob selects reference resistance – Knob
Height 1.0”
ON/OFF Slide Switch controls 12 LED posi on indicator lights
Connec ons: 2 Banana receptacles at end of case (Used to connect device under test in
series with selected resistance)
Indica on: 12 LED Range & Posi on indicator lights
Accessories: 2 each 24” (min) output cables with standard male banana plugs at each
end.
Power: One 9 Volts ba ery (included) for indicator light opera on only. Ba ery
does not infl uence performance of the resistor references.
Weight: 10.2 oz
Accuracy: ±2% Tolerance
10 Rev. B / January 2012
PAR-809C Variable Resistance Reference


Specifi ca ons are subject to change without no ce.
All Prostat trademarks and trade names are the property of Prostat Corpora on.
All other trademarks and trade names are the property of their respec ve companies.
PROFESSIONAL STATIC CONTROL PRODUCTS
Prostat Corpora on
630-238-8883 • Fax: 630-238-9717 • 1-855-STATIC1 • www.prostatcorp.com
Other manuals for PAR-809C
1
Table of contents
Popular Controllers manuals by other brands

Andover Controls
Andover Controls ACX 780 installation guide

TOPP
TOPP ACK4 Installation and use instructions

Eaton
Eaton FD120 manual

fadini
fadini Elpro S40 manual

CD Automation
CD Automation STE Series user manual

National Instruments
National Instruments CompactRIO cRIO-9066 Operating instructions and specifications