16 Rev. C / February 2, 2015
PFM-100 Field Meter Set
4. Select the range you wish to use for auding the ionizer’s decay performance. NOTE: Most
accurate decay me measurements are obtained using the analog output connected to a PGA-
710 Autoanalysis System or an X-Y ploer, PDT-740B Timer or digital Mulmeter.
a. Select the V/inch range on the PFM-711A Field Meter and zero the unit, ±2 volts to ground.
This will provide the most precise voltage indicaon. However, the instrument meter indi-
caon is dampened so that me of decay will be slower than actual decay me.
b. Select the kV/inch range on the PFM-711A Field Meter and zero (0) the unit to ground. This
will provide good indicaon and result in a faster decay me because the meter dampening
is less pronounced. The instrument meter indicaon is dampened in this range so that me
of decay will be slower than actual decay me.
c. For accurate decay mes that are documented, perform this procedure with the analog
output of the PFM-711A Field Meter connected to the input of a PGA-710 Autoanalysis Sys-
tem or an X-Y ploer as described below. Aer charng the charge and decay process, cal-
culate the actual decay me based on the chart speed of the ploer. (See Step 11, below).
5. Posion the sensing surface of the CPM-720A Charge Plate perpendicular to the ionizer’s air
ow, approximately 12 inches from its front air outlet.
a. Hold the instrument combinaon so that the CPM-720A Charge Plate is in the airow;
minimize the airow to the PFM-711A’s meter case.
b. Make sure good contact is made with the PFM-711A’s ground snap.
6. Allow the PFM-711A/CPM-720A combinaon to stabilize in the ion air ow for a minimum of
30-45 seconds and note the ion oset voltage.
7. Install the Charging Wand in the front of the PCS-730 Charging Source. Ground the PCS-730
Charging Source by making direct contact with the metal snap located on the back of its case.
a. To charge the Wand to >+1.0kV push the three posion rocker switch toward the front of
the unit.
b. To charge the Wand to >-1.0kV, push the three posion rocker switch toward the back of
the unit.
8. Touch the wand to the CPM-720A’s sensing (upper) plate and note the voltage indicated on the
PFM-711A meter.
9. Once the desired voltage level is achieved, disconnue
the charging process and remove the PCS-730 Charging
Source from the air path.
10. Observe the voltage drop indicated on the PFM-711A
meter. Voltage should decay over me to the inial oset
point indicated during the balance test procedure out-
lined above.
11. To obtain an accurate measurement of decay me from
1,000 volts to 100 volts, use the following procedure em-
ploying an X-Y ploer for documentaon and calculaon purposes.