18
displayed is the most recent measurement saved. The 3 digits displayed under Memory indicates the
memory location.
Use the and arrows to scroll through the saved measurements. By pressing and holding either
arrows for 2 seconds, the memory will scroll at a rate of 5 memory locations per second. By holding
the key for an additional 5 seconds, the auto-scroll rate is increased to 25 memory locations per
second.
To indicate that the memory is almost full, the word MEMORY will flash. If the memory is full, the
words MEMORY and FULL is displayed and will be flashing together. The user will not be able to save
anymore measurements until at least 1 data point is erased from memory.
Repeated presses of will cycle through MIN, MAX, AVG before returning to RECORD. The
Minimum, Maximum and Average values displayed are calculated based on the entire memory.
Erase Measurement from Internal Memory
You can erase a single measurement from the memory by using the CLEAR button. When in RECALL
mode, scroll through the saved measurements using the UP/DOWN arrows. To clear a selected
measurement, press once. The remaining data in memory is indexed down. Subsequent data
in memory fills cleared locations.
To erase the entire memory at once, please refer to the FUNCTION section below.
05 FUNCTION
The resistance meter allows the user to access some internal settings and provide some
customization. The section below describes what these are and how to set them.
Null Adjustment for Lead Resistance
To NULL lead resistance means to compensate for, or to subtract lead resistance from the
measurement so that only the item or circuit resistance being measured is displayed. This is especially
important when making low resistance measurements less than 1.0 or 2.0 ohms. Prostat leads have a
combined resistance of approximately <0.35 ohms.
To adjust the resistance meter to compensate for lead resistance, connect the leads to the
instrument and then short the leads together using the supplied shunt (Prostat 800S), and follow
these steps:
1. Press once to enter Null Adjustment function. LEAdS will be displayed.
2. Press again. Short will displayed. The <2 LED will also be illuminated.