
ProcessSensing.com 3
Loop Barrier Oxygen Analyzer Quick Start Guide
First span calibration and measurement
A. Gas connection
NOTE: The calibration span gas should be 50...80% of the required
measurement range or one range above. E.g. For a 1 ppm
measurement, span gas should be 5...8 ppm or 50...80 ppm.
1. Connect your span gas line to either of the ports on your analyzer.
2. Ensure the flow rate is at 1...2 SCFH and allow the span gas to flow for 2...3
minutes. This will purge the system.
3. Continue to section B and follow the procedure relevant to your analyzer
model.
B. First span calibration
GPR-1500
The GPR-1500 is delivered without the sensor installed to preserve its
operational life. To install the sensor:
1. Apply power to your analyzer (refer to Figure 4 on page 5).
2. Using the two latches, open the front window.
3. To open the sensor housing, loosen the star wheel then disengage the top
sensor housing by turning it 90° counter-clockwise. Refer to ‘b’ in Figure 2 on
page 4.
4. Remove the sensor from its packaging, remove the shorting flags and
immediately place in the bottom sensor housing (refer to Figure 1 on page 4
for guidance).
5. Now re-connect the top sensor housing and secure using the star wheel.
6. Use the two latches to close the front window on your analyzer.
7. Now press and use and to navigate to
Calibration > Span Calibrate.
8. Use and to enter the span gas value. Ensure the reading has stabilized
before continuing.
NOTE: When a Span or Zero Cal starts, only "Abort" with is shown
until the reading is stable, then "Accept" with appears.
9. Use to Accept, and to Abort. If you selected Accept, your analyzer is
now calibrated.
GPR-2500
The GPR-2500 is delivered with the sensor installed:
1. Apply power to your analyzer (refer to Figure 4 on page 5).
2. Now press and use and to navigate to
Calibration > Span Calibrate.
3. Use and to enter the span gas value. Ensure the reading has stabilized
before continuing.
NOTE: When a Span or Zero Cal starts, only "Abort" with is shown
until the reading is stable, then "Accept" with appears.
4. Use to Accept, and to Abort. If you selected Accept, your analyzer is
now calibrated.