Changes in elevation will also affect your initial
O2 reading. If the analyzer is calibrated for
sea level, and you bring it to 5000 feet above
sea level, you may notice the starting O2
concentration shows 17.9%. This is normal –
repeat the “initial setup” procedure so that the
instrument reads 20.9%.
UNDERSTANDING THE ADJUSTMENTS
The operating controls and adjustments
include power switch, pump switch, span, zero
and pump time adjustments:
SPAN: Use this on the side of the instrument
to adjust the analyzer to 20.9%. Press the
pump button, let the reading stabilize, then
adjust to 20.9%. You should do this once a
week, or if the instrument’s altitude has
changed. You should wait until the pump has
stopped, and the reading has stabilized, before
adjusting the span.
ZERO: This should generally not be touched.
See the section on checking the zero.
PUMP TIME: Use this to set the pump time. 5
seconds is ideal, but if your application,
package, or container has limited headspace,
you can set it to 3 seconds. Use a longer
pump time if the sensor is responding slow.