RBCBioscience MagCore HF48 User manual

TEL : +886-2-8912-1200
FAX : +886-2-8912-1300
15F, No.15, Qiaohe Rd,Zhonghe Dist,
New Taipei City 235029, Taiwan
http://www.rbcbioscience.com

Engineering maintenance manual
Ver. 2022-5
www.rbcbioscience.com
15F, No.15,QiaoheRd,Zhonghe Dist,New TaipeiCity 235029,Taiwan

1
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Handle the power cord carefully.
Hold the plug when unplugging the cord.
Keep the set free from moisture, liquid, and dust.
Unplug the power cord if the system will not be used
again in a long time.
Do not obstruct the ventilation space.
Do not let foreign objects in the set.
Do not let insecticides, benzene, and thinner come in
contact with the set.
Do not ever disassemble or modify the set in any
way.
Avoid high temperatures.
Leave space for sufficient heat dissipation when installed on a rack.
CAUTIONS
Risk Of Electric Shock Do Not Open
Safety Precautions
Before use
WARNINGS and CAUTIONS stated below are intended to protect individual from injury and minimize damage to machines. To ensure safe operation, please follow them carefully.
CAUTION:
To reduce the risk of electric shock, do not remove the cover. Refer to qualified service
technician for inner part servicing.
CAUTION:
1. Handle the power supply cord carefully
Do not damage or twist the power supply cord. Damaged or twisted cords may
cause electric shock or error when used. When you remove it from the outlets, be
sure to remove by holding the plug and not by pulling the cord.
2. Do not open the top cover
In order to prevent electric shock, do not open the top cover.
3. Do not place anything inside
Do not place metal objects inside the MagCore®System and wipe away all spilled
liquid to minimize risk of system error and electric shock.
Note On Use:
1. Read Instructions – All the safety and operating instructions should be read before operating
the instrument.
2. Keep the Instructions – Please keep the safety and operating instruction manual for future
reference.
3. Cautions and Warnings – All warning signs should be attached on the product and shown
in the user's manual.
4. Follow Instructions – All operating and use instructions should be followed.
5. Cleaning – Unplug the product before cleaning. Use only 75% of EtOH to clean the surface of
instrument.
6. Attachments – Please do not attach anything on the instruments.
7. Liquid and Moisture – Do not use this product near water and avoid moist areas - bath tub,
wash bowl, kitchen sink, laundry, swimming pool or in a moist basement.
8. Proper Installation – Do not place this product on an unstable cart, stand, tripod, bracket, or
table. The product may fall and cause serious injury to users and damage to the product. Use
only with a cart, stand, tripod, bracket, or table recommended by the manufacturer, or sold
with the product. Any mounting of the product should follow the manufacturer’s instructions,
and should use a mounting tool recommended by the manufacturer.
9. Ventilation – Slots and openings in the cabinet are designed for ventilation, which is required
for safe operation and minimizing the chance of overheating. The product should be placed
on a hard surface to ensure proper ventilation. This product should not be placed in a built-in
installation such as a bookcase or rack unless proper ventilation is ensured or instructed by the
manufacturer.
10.Power Sources – This product should be operated only from the type of power source
indicated on the label. If you are not sure of the type of power supply in your lab, consult your
local distributor or power companies. For products intended to operate from battery to other
sources, refer to the operating instructions.
11.Grounding or Polarization – This product may be equipped with a polarized alternating-
current plug. This plug will fit into the power outlet in one way only. If you are unable to insert
the plug fully into the outlet, try reversing the plug. If the plug should still fail to fit, contact your
electrician to replace your obsolete outlet. Do not defeat the safety purpose of the polarized
plug.
12.Power-Cord Protection – Power-supply cords should be routed so that they are not likely
to be walked on or pinched by items placed upon or against them. Pay extra attention to the
plugs, convenience receptacles, and the connecting point where the cord extends out of the
product.
13.Lightning – Unplug the product if it will not be used again for a long period or in the case of a
heavy storm.
14.Overloading – Do not overload wall outlets, extension cords, or integral convenience
receptacles as this can result in a risk of fire or electric shock.
15.Object and Liquid Entry – Never push objects of any kind into this product as they may touch
dangerous voltage points or short-out parts that could result in a fire or electric shock. Never
spill liquid of any kind on the product.
16.Servicing – Do not attempt to repair this product due to the risk of exposure to dangerous
voltage or other hazards. Refer all servicing to qualified service technicians.
17.Damage Requiring Service – Unplug this product from the wall outlet and refer servicing to
qualified service technicians under the following conditions:
a) When the power-supply cord or plug is damaged,
b) If liquid has been spilled, or objects have fallen into the product,
c) If the product has been exposed to rain or water,
d) If the product does not operate normally following the operating instructions. Adjust only
controls covered by the operating instructions as an improper adjustment of other controls
may result in damage and will often require extensive work by a qualified technician to restore
the product to its normal operation,
e) If the product has been dropped or damaged in any way, and
f)When the product exhibits a distinct change in performance
– this indicates a need for service.
18.Replace Parts – When replacement is required, be sure the service technician uses
replacement parts specified by the manufacturer or those with the same features as the
original part. Unauthorized substitutes may result in fire, electric shock, or other hazards.
19.Safety Check – Upon completion of any service or repairs to this product, ask the service
technician to perform safety checks to confirm that the product is in proper operating
condition.
20.Heat – The product should be situated away from heat sources such as radiators, heat
registers, stoves, or other products that produce heat.
General Warning
Warning: Biological Hazards
Warning: High Temperature
Warning: Electricity
SAFETY INSTRUCTIONS

2
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Specification
Model HF48 (under development)
Processing
Capacity
Max .48 samples per batch
System Feature Special Cartridge rack, T-Rack, Tube Rack design for easy installation
Power Supply Voltage: AC 200V-220V; Frequency: 50/60Hz;
Dimension W1000 x D800 x H1600 (mm)
W39.37 X D31.49X H62.99(inches)
Net Weight 250kg / 551.25Lb
Operation parameters
Operation environment
Applications
Whole Blood (200/400ul), Viral Nucleic Acid (DNA/RNA), Tissue Genomic DNA, Plant Genomic
DNA Bacteria Kit, Total RNA Kit
Processing Capacity 1-48 samples per batch
Sample Volume 200/400/1200 μl
Elution Volume 60/100/150/200/300 μl
Purity O.D. 1.8 -2.0 A260/280 ratio
Temperatures allowed during
transportation, storage, and packaging 15° C-35° C
Temperatures allowed during operation 18° C -300C
Pollution Degree Level 2

3
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Table of content
Safety Precautions…………………………………………………………………………………………………… 1
Specification/Operation
environment/Operation parameters/
Application…………………………………………………………………………………………………… 2
Package content…………………………………………………………………………………………………… 5
Unpacking…………………………………………………………………………………………………… 5
Attachments…………………………………………………………………………………………………… 6
Installation…………………………………………………………………………………………………… 7
Before installation…………………………………………………………………………………………………… 7
System guidance…………………………………………………………………………………………………… 8
Installations…………………………………………………………………………………………………… 9
Touch panel and user interface…………………………………………………………………………………………………… 10
Test Home Value…………………………………………………………………………………………………… 11
Accessory Installation…………………………………………………………………………………………………… 13
Emergency stop and Warning…………………………………………………………………………………………………… 14
Start…………………………………………………………………………………………………… 15
Load program…………………………………………………………………………………………………… 16
Save program…………………………………………………………………………………………………… 17
Delete program…………………………………………………………………………………………………… 17
Easy maintenance…………………………………………………………………………………………………… 18
Replace O-Rings…………………………………………………………………………………………………… 19
Clear piercing needles…………………………………………………………………………………………………… 19
Piston test…………………………………………………………………………………………………… 20
Rejection Tip…………………………………………………………………………………………………… 21
Heater test…………………………………………………………………………………………………… 21
UV sterilization…………………………………………………………………………………………………… 22
Replace UV lamp…………………………………………………………………………………………………… 22
Moving test…………………………………………………………………………………………………… 23
DC motor operate…………………………………………………………………………………………………… 23
Settings…………………………………………………………………………………………………… 24
System time settings…………………………………………………………………………………………………… 24
System status…………………………………………………………………………………………………… 25
Export error history files…………………………………………………………………………………………………… 25
Others…………………………………………………………………………………………………… 26
Replace fuse…………………………………………………………………………………………………… 26
Advanced maintenance…………………………………………………………………………………………………… 28
Open maintenance board…………………………………………………………………………………………………… 29
Axial direction…………………………………………………………………………………………………… 30
Main components…………………………………………………………………………………………………… 31
Sensor distribution and replacement…………………………………………………………………………………………………… 32
Measure the interval between Y axis
and costal plate…………………………………………………………………………………………………… 34
Adjust the interval between Y axis and
costal plate…………………………………………………………………………………………………… 37

4
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Measure the level height of X-axis
and costal plates at area A and area B …………………………………………………………………………………………………… 39
Horizontally adjust X-axis to make the
tolerance lower than 1mm…………………………………………………………………………………………………… 40
Horizontally adjust X-axis to make the
tolerance higher than 1mm…………………………………………………………………………………………………… 42
Replace component…………………………………………………………………………………………………… 45
Replace X-axis servo motor belt…………………………………………………………………………………………………… 45
Replace X-axis servo motor…………………………………………………………………………………………………… 49
Replace Y-axis motor…………………………………………………………………………………………………… 53
Replace Y-axis brake component…………………………………………………………………………………………………… 55
Replace Y-axis belt…………………………………………………………………………………………………… 56
Replace V-axis motor…………………………………………………………………………………………………… 57
Replace base elevator motor…………………………………………………………………………………………………… 59
Replace Shutter motor…………………………………………………………………………………………………… 63
Replace Shutter timing belt…………………………………………………………………………………………………… 66
Replace ejector rod module…………………………………………………………………………………………………… 70
Replace piston…………………………………………………………………………………………………… 72
Replace heater module…………………………………………………………………………………………………… 77
Replace the nitrogen cylinders of the
top cover…………………………………………………………………………………………………… 81
Engineering software…………………………………………………………………………………………………… 82
Software installation…………………………………………………………………………………………………… 83
Test Home Distance…………………………………………………………………………………………………… 84
Input Home Distance…………………………………………………………………………………………………… 86
Power panel…………………………………………………………………………………………………… 87

5
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
12
3 4
6
拆開上蓋 拆開上方三支固定桿
拆開前蓋後蓋
拆開下方固定桿 拆開下方固定塊
拆開側蓋
5
The packing materials are recyclable.
We suggest keeping the packing materials for
future uses.
Package content
Unpacking
Disassemble the top cover.
Disassemble the front cover and rear cover.
Remove the bottom fastening bars.
Remove the three fastening bars at the top.
Disassemble the side covers.
Remove the fastening blocks at the bottom

6
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Attachments
Cartridge Rack X 6
Cartridge X 20 Sample Tube X 50 Tip / Tip Holder Set X 50
Elution Carrier X 1
Cartridge Carrier B X 1 / Cartridge Carrier A X 1
Tip Carrier B X 1 / Tip Carrier A X 1
Power wire X 1 Lubricating oil X 1 O-Rings X 24
Please put the system test tools at a clear place and don’t lose them.
System test tools

7
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Before installation
Installation
Please find a place near the power outlet and reserve enough space for proper heat dissipation and system
installation.
Please make sure the instrument is on a flat surface.
Use the leveling glides to fine-tnue the elevation level.
!
System height
160 cm
System depth
80 cm
System width
100 cm
Distance to the wall
(left and right)
20 cm
Distance to the wall (rear)
40 cm

8
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
System guidance
Power switch and connection
Power switch (Red)
Light switch (Green)
Part chamber
Maintenance cover
Cooling fans
Piston module
Operation area B
Cabinets and electrical box
Operation area A
Shutter
Leveling Glides (4 legs)
Fixing Knob on the Top Cover
of Part Chamber
Status indicator
Viewing Window
Touch screen
Viewing Door
Feeding Door
Top cover of Part
Chamber

9
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Installation
Please make sure that the voltage of the
outlet is between 220V and 240V. Keep
your hands dry. Please also make sure
that the machine is turned off when you
insert the plug.
Please note the weight of the top cover of
part chamber. Two people are required to
lift it up.
Place the system in a suitable location and use the leveling glides to
fine-tune the elevation. Remember to tighten the screws.
Cut the fixation belt and close the top cover of part chamber.
Cut the fixation belts on the racks.
Plug in the power cord and turn on the main power.
Turn on the power and the system will start initializing setup
automatically.
Please set the local time (refer to page 24) for
the first installation, and perform the “Piston
Test” (page 20).
Please finish reading the operation manual
before operating.
Rotate and pull out the knobs on both sides of the top cover of part
chamber using coins.
Open the top cover of part chamber.
1 2
!
Next Step
Next Step
Next Step
Next Step
Next Step
Next Step

10
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Touch panel and user interface
Start
Maintenance
Load program
Setup Status
2012.06.07
12:12:12
━ Panel
━ Main menu
MagCore®HF48 provides 8.9 inches full-color touch
screen and user-friendly interface. You can monitor
the progress of extractions with the status and time
displayed on the screen.
2012.06.07
12:12:12
Start
Maintenance
Loadprogram
Setup Status
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
DNase I Treatment
Sample Volume
Elution Volume
Please select Cartridge Code
Elution Buffer
Optical Detection
AAB B E
D
Start
Set basic setting parameters and operations
Load program
Load the frequent-used programs stored by users
Maintenance and other test functions
Basic maitenance, UV sterilization, and system tests
System time
System status
Basic system information, serial number, error
messages
Setup
System time setup
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
DNase I Treatment
Sample Volume
Elution Volume
Please select Cartridge Code
Elution Buffer
Optical Detection
AAB B E
D
A A
A
BE
B
DA
B
B E
Not detected
Normal
Front door
Front door is open
Not detected
Not detected
Not detected
Not detected
Not detected
Cartridge Carrier B
Tip Carrier B
Cartridge Carrier A
Tip Carrier A
Elution Carrier
訊息指示
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
DNase I Treatment
Sample Volume
Elution Volume
Please select Cartridge Code
Elution Buffer
Optical Detection
AAB B E
D
Menu
Operating Functions
Message
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
DNase I Treatment
Sample Volume
Elution Volume
Please select Cartridge Code
Elution Buffer
Optical Detection
AAB B E
D
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
DNase I Treatment
Sample Volume
Elution Volume
Please select Cartridge Code
Elution Buffer
Optical Detection
AAB B E
D

11
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Test home value
After installation, it is necessary to make sure that the position of modules and axes
have not been shifted during shipping. Please check if the values conform to the given
values in the original factory document.
1. Please make sure that the instrument is turned off.
2. Disconnect the brake from the instrument and install the electronic brake thread.
3. Install Z-axis calibration tool on Operating Area A and manually rotate the Y-axis
belt anti-clockwise to lower the Y-axis till it reaches the home calibration tool
and correctly insert the
calibration point in the
calibration tool.
6 拆開X軸伺服馬達4顆螺絲
5 拆開X軸時規輪側擋片3顆螺絲及時規輪上2顆固定螺絲,並將X軸皮帶及時規輪脫離馬達軸心
4. Repeatedly move the Y-axis up and down to make sure whether the calibration point
can insert into or separate from the Z-axis calibration tool without any difficulties.
5. Prepare a computer and connect it to the instrument. Turn on the instrument.
Rotate the Y-axis belt anti-
clockwise to lower the
Y-axis.
18良距離
Put the Z-axis calibration tools at the disposition
holes on the costal plates of the Cartridge Racks.

12
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Test Tip uptake and Test for Tip Leakage
Please refer to "Piston test" in Simple maintenance.
6. Read the current home distance of the machine and record the X-axis and Y-axis
values.
7. Repeat from Step 3 for Operating Area B.
8. Compare current home distance with the home distance in the factory document. If
it fails to conform to the home distance in the factory document, please perform the
machine calibration process.
9. Disconnect the electronic brake thread and reconnect the brake to the instrument.
MagCore HF48 System Teaching Data Sheet
SerialNo.J48M000123456 Date10/25/2013(mm/dd/yyyy)
UISoftware48‐20131023‐3‐B‐NTUPLCSoftware20130717‐01
Inspector
NoSensorValue
1YA‐axis112.21
2YB‐axis112.19
3XA‐axis516.05
4XB‐axis206.58
Note:Pleasestorethisdatasheetbecausethemaintenancepersonusesit.

13
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Accessory installation
B
A
Tip
Sample
Install Elution Tube
Warning:
DO NOT use other types of centrifuge tubes to replace the Sample Tubes provided.
Poor extraction results or error may occur due to tubes having different heights.
Warning:
Carrier of Operating area A can only
be used in area A; Carrier of area B
can only be used in area B. Incorrect
use may cause system error.
Install CartridgeInstall Tips and Sample Tubes
B
A
Insert Tips and Sample Tubes into the slots in the Tip Carriers. Then
place the Tip Carriers into Operating Area A and B as illustrated.
Insert Cartridge into Cartridge Rack. Then place the entire set
of Cartridge Carrier in the slots in Operating Area A and B as
illustrated.
Place Elution Tubes into their corresponding positions and place
Elution Carrier in the elution slot.
E
Elution A
Elution B
Elution Area
Operation Area A
Operation Area B

14
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Emergency stop and warning
Please read about how to stop the system in case of emergency.
Method 1 Method 2 Resume
Min
90
Processing
Stop
Save
Back
Load
Program
Heater B 68°C / 154.4 °F | Heater A 68°C / 154.4 °F
Heating
Test :
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200
µl
150
µl
---
AB
DA
B E
Min
90
Processing
Stop
Save
Back
Load
Program
Heater B 68°C / 154.4 °F | Heater A 68°C / 154.4 °F
Heating
Test :
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---
AAB B E
D
Warning?
Abort
Program Stopped.
Press ABORT to terminate the process or
CONTIUNE to run.
Contiune
The system will stop all operations
immediately if the front door is opened.
Select Continue to resume to previous
operations or Abort to end all operations.
Press Stop key to stop all operations immediately
in case of emergency.
Warning:
Please do not place your hands near the operation areas when the system is under
operation for safety reasons.

15
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Start
Press Start in the main window.
Select 24 or 48 test channels.
Select the Cartridge code.
Please confirm the number you selected and input all parameters.
Confirm that the icon does not appear in the status bar. If
the icon shows, operation is disabled. Please solve the error to
continue. (Refer to page 10 for Warning Signs).
After you finish inputting and press Start, a confirmation window
will show up. Please press Yes to operate the system if the
parameters are correct. Please press No to modify incorrect
parameters.
You may monitor the extraction progress and check the remaining
time on the screen. The system will display Complete page after
completion.
Start
Maintenance
Load program
Setup Status
2012.06.07
12:12:12
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
Cartridge Code
Dnase Treatmen
Elution Buffer
Sample Volume
Elution Volume
AB
DA
B E
Start
Cartridge Code
DNase I Treatment
Sample Volume
Elution Volume
Please select Optical Detrction
Optical Detection
Save
Back
Load
Program
101 MagCore Genomic DNA Whole Blood Kit (Speedy installation)
Start
Yes No
48
Test
Test :
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
Cartridge Code
Dnase Treatmen
Elution Buffer
Sample Volume
Elution Volume
AB
DA
B E
Please select Channel
Press START to Proceed
105 Plasma DNA Extraction Kit (1.2 ml)
200 µl
150 µl
Min
90
Processing
Stop
Save
Back
Load
Program
Heater B 68°C / 154.4 °F | Heater A 68°C / 154.4 °F
Heating
Test :
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---
AB
DA
B E
Min
90
Processing
Stop
Save
Back
Load
Program
Heater B 68°C / 154.4 °F | Heater A 68°C / 154.4 °F
Heating
Test :
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---
AB
DA
B E
System operation remaining time
Heater status
Current Operation Status
Parameters
Emergency Stop
Next Step
Next Step
Next Step
Next Step
Please do not open the viewing window
or feeding door during operation if there
is no emergency to prevent error or poor
extraction results.
Complete
Back
Start

16
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Press Load Programs to enter the Stored Programs page.
Or
Press Load Programs after pressing Start to enter the Stored
Programs page.
Please select one of your saved programs.
Check whether the parameters are correct and press Yes.
Please select Cartridge Code
Stored Programs
Please select Stored Program Item
Back
1 5
2 6
3 7
4 8
Stored Program 1
Stored Program 2
Stored Program 3
Stored Program 4
Stored Program 5
Stored Program 6
Stored Program 7
Stored Program 8
Stored Programs
Please select Stored Program Item
Back
1 5
2 6
3 7
4 8
Stored Program 1
Stored Program 2
Stored Program 3
Stored Program 4
Stored Program 5
Stored Program 6
Stored Program 7
Stored Program 8
Load Program?
Yes No
48
Test
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---
Start
Maintenance
Load program
Setup Status
2012.06.07
12:12:12
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
Cartridge Code
Dnase Treatmen
Elution Buffer
Sample Volume
Elution Volume
AB
DA
B E
Please select Channel
Next Step
Next Step
Load program
Load program
You can save frequently used programs with names you prefer for your convenience.

17
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
You may save up to 8 programs with names you prefer for you
convenience.
Please press Save key to enter Stored Programs page after all
parameters have been selected on Start page.
Please click to modify.
Confirm if the parameters are correct.
Enter the name you want to save and press Enter.
Store programs
Stored Programs
Please select Stored Program Item
Back
1 5
2 6
3 7
4 8
Stored Program 1
Stored Program 2
Stored Program 3
Stored Program 4
Stored Program 5
Stored Program 6
Stored Program 7
Stored Program 8
Press Enter to send
Store Program 1
Cancel
1 2 3 4 5 6 7 8 9 0 -
Q W E R T Y U I O P
Backspace
Enter
A S D F G H J K L
Z X C V B N M
Stored Programs
Please select Stored Program Item
Back
1 5
2 6
3 7
4 8
Stored Program 1
Stored Program 2
Stored Program 3
Stored Program 4
Stored Program 5
Stored Program 6
Stored Program 7
Stored Program 8
Save Program?
Yes No
48
Test
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
Cartridge Code
Dnase Treatmen
Elution Buffer
Sample Volume
Elution Volume
AB
DA
B E
Please select Channel
You may delete your saved programs as well.
Please press Load Programs to enter Stored Programs page after
entering Start page.
Please select an item you want to delete by pressing X.
Press Yes to delete the program.
Delete programs
Stored Programs
Please select Stored Program Item
Back
1 5
2 6
3 7
4 8
Stored Program 1
Stored Program 2
Stored Program 3
Stored Program 4
Stored Program 5
Stored Program 6
Stored Program 7
Stored Program 8
Stored Programs
Please select Stored Program Item
Back
1 5
2 6
3 7
4 8
Stored Program 1
Stored Program 2
Stored Program 3
Stored Program 4
Stored Program 5
Stored Program 6
Stored Program 7
Stored Program 8
Delete Program?
Yes No
48
Test
Cartridge Code :
Remove DNA
Sample Volumn :
Elution Volumn :
Elution Buffer :
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---
Start
Stop
Save
Back
Load
Program
Start
24 Test 48 Test
Cartridge Code
Dnase Treatmen
Elution Buffer
Sample Volume
Elution Volume
AB
DA
B E
Please select Channel
Next Step Next Step
Next Step
Next Step
Next Step

18
RBCBioscience
IssuedBy:RBCBioscienceCorp. MagCoreHF48 Service Manual Revised:February,2014 Ver. 2014-2
Copy new programs into your USB Flash Drive.
*Confirm if“48ExtractorUpdateProgram” is successfully saved in
your USB Flash Drive.
Open the door of viewing window, insert your USB Flash Drive into
the USB slot on the back of the control panel, and press Export key
to export data to your USB Flash Drive.
Select Update Program Files.
Press Yes to start the update and the window will close
automatically after the update is complete.
Maintenance
Please select Item
Back
UV Sterilization Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate Update Program Files
Maintenance
Please select Item
Back
UV Sterilization Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate Update Program Files
Maintenance
Please select Item
Back
UV Sterilization Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate Update Program Files
Easy maintenance includes the maintenance items which can be done by users. These items include
sterilization, cleaning, system tests, error message report, etc. Users must perform these items regularly to
ensure that the system can function normally.
Update program
UV sterilization
UV sterilization
Piston test
Piston draining and leakage test
Moving test
Moving Test for all axes
Heater test
Temperature test for heater
Rejection Tip
Return tips to original positions.
Update program
Update programs
O-Ring and piercing needles
Replace O-Ring and clean piercing needles
DC Motor Operate
Raise or lower the protective cover on Elution
area and motor in Operating area A and B
Maintenance
Easy maintenance
Please DO NOT modify anything in
48ExtractorUpdateProgram.
Update Program Files?
Yes No
Other manuals for MagCore HF48
2
Table of contents
Popular Scrubber manuals by other brands

Hover Scrubber
Hover Scrubber Deluxe instructions

Pacific
Pacific SPE-2.5 Parts & operating manual

Nilfisk-Advance
Nilfisk-Advance 56390926(850SC) Instructions for use

Windsor
Windsor chariot iScrub cs20 operating instructions

Pullman Holt
Pullman Holt GB1400E Operation & care instructions

Ghibli & Wirbel
Ghibli & Wirbel POWER EXTRA 11 Use and maintenance

Husqvarna
Husqvarna PG 540 Operator's manual

Mytee
Mytee Speedster 1000HV User and parts manual

Renfert
Renfert SILENT powerCAM EC Translation of the original instructions for use

Nilfisk-Advance
Nilfisk-Advance BA 600S Service manual

ulsonix
ulsonix TOPCLEAN 1600 user manual

Nobles
Nobles Speed Scrub 2701 Quick Ship Pack Operator and parts manual