Reed Instruments R4500SD User manual

Data Logging
Hot Wire
Thermo-
Anemometer
Instruction
Manual
R4500SD
REED Instruments

2
Table of Contents
Introduction ................................................................................................ 3
Product Quality........................................................................................... 3
Safety ......................................................................................................... 4
Features...................................................................................................... 4
Included...................................................................................................... 4
Specications..........................................................................................5-6
Instrument Description............................................................................... 7
Display Description .................................................................................... 8
Hot Wire Probe Description........................................................................ 9
Operating Instructions...........................................................................9-11
Power ON/OFF ...................................................................................... 9
Selecting Measurement Modes ............................................................. 9
Air Velocity/Temperature Measurement............................................9-10
Changing the Air Velocity Unit of Measure .......................................... 10
Type K/J Thermocouple Measurement Mode...................................... 10
Data Hold ............................................................................................. 11
Recording Maximum and Minimum Readings..................................... 11
Backlight .............................................................................................. 11
Setup Mode.........................................................................................12-15
Setting the Time and Date (dAtE) ....................................................... 12
Setting Data Decimal Format (dEC).................................................... 13
Enabling/Disabling Auto Power Off (PoFF)......................................... 13
Enabling/Disabling the Beeper (bEEP) ............................................... 13
Selecting Thermocouple Type (tYPE) ................................................. 14
Selecting the Temperature Unit of Measure (t-CF)............................. 14
Setting the Data Logging Sampling Rate (SP-t) ................................. 14
Formatting the SD Card (Sd F) ......................................................14-15
Verify Set Time and Date .................................................................... 15
continued...
REED Instruments

3
Verify Sampling Rate .......................................................................... 15
Data Logging.......................................................................................15-17
Data Recording Modes....................................................................... 15
Manual Data Logging Mode ..........................................................15-16
Automatic Data Logging Mode .......................................................... 16
Transferring Data from the SD Memory Card to a Computer........16-17
Data Stream Output ............................................................................17-18
3.5mm Terminal to RS-232 Serial Connection ................................... 18
System Reset ........................................................................................... 18
Battery Replacement................................................................................ 19
Applications.............................................................................................. 19
Accessories and Replacement Parts ..................................................19-20
Frequently Asked Questions (FAQ's)........................................................ 20
Product Care ............................................................................................ 20
Product Warranty ..................................................................................... 21
Product Disposal and Recycling .............................................................. 21
Product Support....................................................................................... 21
Introduction
Thank you for purchasing your REED R4500SD Data Logging Hot Wire
Thermo-Anemometer. Please read the following instructions carefully
before using your instrument. By following the steps outlined in this
manual your meter will provide years of reliable service.
Product Quality
This product has been manufactured in an ISO9001 facility and has been
calibrated during the manufacturing process to meet the stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
REED Instruments

4
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
Features
• Measuresairvelocity(fpm,m/s,kph,mphandknots)andairtemperature
• ContacttemperaturemeasurementwithtypeJ/Kthermocoupleoption
• Thermistorsensorforfastairtemperaturereadings
• Hotwireprobeforpreciselowairvelocitymeasurements
• Real-timedataloggerwithintegratedSDmemorycard
• Userselectablesamplingratefrom1to3600seconds
• Easy-to-readbacklitLCDdisplay
• DataholdandMin/Maxfunctions
• Tripodmountforlong-termmonitoring
• Lowbatteryindicatorandautoshutoff
Included
• TelescopeHotWireProbe
• HardCarryingCase
• Batteries
REED Instruments

5
Specifications
Type: Hot Wire
Measuring Ranges: fpm: 40 to 4921
m/s:0.2to25
km/h:0.7to90
mph: 0.5 to 55.9
knots: 0.4 to 48.6
Air Temperature: 32 to 122°F (0 to 50°C)
TypeKTemperature: -148.0to2372.0°F(-100.0to1300°C)
TypeJTemperature: -148.0to2192.0°F(-100.0to1200°C)
Accuracy: Air Velocity: ±(5% rdg + 20 fpm,
0.2m/s,0.3km/h,0.2mph,0.2knot)
Air Temperature: ±0.8°C (1.5°F)
Contact Temperature: ≥-58°F (-50°C)
±(0.4% + 1°F (0.5°C))
≤-58°F (-50°C) ±(0.4% + 1.8°F (1°C))
Resolution: fpm: 1
m/s,km/h,mph,knot:0.1
Temperature: 0.1°F/°C
Telescope Probe Length: 11 to 37" (280 to 940mm)
Response Time: <1 second
Display: 5-digit, Dual LCD Display
Backlit Display: Yes
Data Hold: Yes
Min: Yes
Max: Yes
Datalogging Capabilities: Yes
Real-Time Clock and
Date Stamp: Yes
Sampling Rate: Yes (1, 2, 5, 10, 30, 60, 120, 300,
600, 1800, 3600 seconds)
continued...
REED Instruments

6
ExternalMemory: Yes,expandableupto16GBwith
SD card (optional)
AutoShut-off: Yes(after10minutes/off)
KickStand: Yes
Tripod Mountable: Yes
Low Battery Indicator: Yes
PowerSupply: 6xAAbatteriesorACAdapter(optional)
Data Output: Yes (RS-232)
Product Certications: CE
Operating Temperature: 32 to 122°F (0 to 50°C)
Storage Temperature: 14 to 140°F (-10 to 60°C)
Operating Humidity Range: 10 to 80%
Dimensions: 8x3x1.5"(203x76x38mm)
Weight: 1.13lb (515g)
REED Instruments

7
Instrument Description
2
3
Hot Wire Anemometer
SD Series Data Logger
R4500SD
4
5
6
7
8
9
10
11
12
13
14
15
16
17
POWER
ESC
FUNC.
HOLD
NEXT
REC
ENTER
SET
TIME
CHECK
SAMPLING
CHECK
LOG
18
1.
POWER/Backlight/ESCButton
2.
SET/Down/TIME Button
3. SD Card Slot
4. LCD Display
5. HOLD/FUNCTION/
NEXT Button
6. REC/ENTERButton
7.
LOG/Up/SAMPLING
CHECK Button
8. RS-232OutputJack
9. Reset Pin
10. Power Adapter Input
11. Tripod Mounting Screw
12. Battery Cover
13.Kickstand
14.TypeK/JThermocouple
InputJack
15.
Telescopic Probe Input Socket
16.
Telescopic Probe Sensing Head
17. Telescopic Probe Handle
18.
Telescopic Probe Plug
REED Instruments

8
Display Description
3
6
Km/h
mph
K
J
Knot
FPMm/S
F
1. Measurement Reading Indicator
2. Date&TimeValues/MeasurementModesIndicator
3. Date/TimeStamp
4. TemperatureUnitofMeasure
5. UnitofMeasure
6. MaximumandMinimumIndicators
7. Record Mode Indicator
8. Data Hold Indicator
9. Low Battery Indicator
REED Instruments

9
Hot Wire Probe Description
1. Wind Direction Marker
2. Air Velocity Sensor
3. Protective Cover in the
Open position
4. Protective Cover in the
Closed position
5. Temperature Sensor
Operating Instructions
Power ON/OFF
Turn the meter on by pressing the POWER button. To turn the meter off,
press and hold the POWER button for 2 seconds.
Note: Thismetercanbepoweredbyeithersix(6)"AA"batteriesorAC
adapter (sold separately).
Selecting Measurement Modes
The R4500SD offers 2 types of measurement modes:
• AirVelocity/Temperaturemeasurement
• TypeK/JThermometermeasurement
1. Whenthemeterison,pressandholdtheFUNCTIONbuttonto
togglebetween"An"forairvelocity/temperatureand"tP"forType
K/JThermometermeasurementmodesofoperation.
2. ReleasetheFUNCTIONbuttontoconrmselection.
Air Velocity/Temperature Measurement
When the Telescopic Probe is not in use, slide the Sensor Cover over the
Sensor Head. Before using the Telescopic Probe, slide the Sensor
Cover down.
continued...
3
REED Instruments

10
Note: Avoid touching the sensor with your ngers or with any tools, as
this may cause damage or incorrect readings.
1. Plug the Telescopic Probe into the Probe
Input Socket while the meter is off.
2. Once the meter is powered up,
select "Air" function (see Selecting
Measurement Modes for details).
3.
Hold the Probe Handle and face the
wind direction mark on the sensor head
towards the direction of the airow. The
top part of the display will indicate the air
velocity, and the lower part of the display
will indicate the air temperature value.
Changing the Air Velocity Unit of Measure
To change the air velocity unit of measure, press and hold the ENTER
button.Thedisplaywillscrollthrougheachunit(m/S,Km/h,mph,Knot,
& FPM). Release the ENTER button to select the displayed unit. The
selected unit will be saved as the new default unit of measure.
Type K/J Thermocouple Measurement Mode
1. Select the "tP" function (see Selecting Measurement Modes for details).
2. PlugtheoptionalTypeKorJThermocoupleProbeintotheTypeK/J
thermocouple input jack.
Note: Make sure to remove the telescopic probe before inserting the
thermocouple probe.
3. Thescreenwilldisplaya"K",conrming
that is congured by default to a Type
Kthermocoupleprobe.Themeasured
temperature will now be displayed on
the display screen.
Note:IfusingatypeJthermocoupleprobe,
see Selecting Thermocouple Type for details.
continued...
K
FPM
REED Instruments

11
Data Hold
1. While taking a measurement, press the HOLD button to freeze the
current readings on the display.
2. While in this mode a "HOLD" symbol will appear.
3. Press the HOLD button again to resume normal operation.
Note:WhentheDataHoldfeatureisactiveallbuttonsexceptthePOWER
button are disabled.
Recording Maximum and Minimum Readings
1. Press the REC button to enter recording mode as indicated by
"REC"ontheLCD.Themeterwillnowbeginrecordingmaximum
and minimum readings.
2. While in recording mode;
A) Press the RECbuttononceandthemaximumvaluewillappear
on the display as indicated by "REC MAX".
B) Press the REC button again and the minimum value will appear
on the display as indicated by "REC MIN".
C) Toexitrecordingmodeandresumenormaloperation,pressand
hold the REC button for two seconds.
When in recording mode the POWER button is disabled
and the meter cannot be turned off.
Backlight
After powering the meter ON, the LCD Backlight will turn on automatically.
PresstheBACKLIGHTbuttontoturntheLCDBacklightonoroff.
REED Instruments

12
Setup Mode
1. Press and hold the SET button for 2 seconds to enter Setup Mode.
2. Press the NEXT button continuously to scroll through the
following parameters.
Parameter Description
dAtE Set the time and date
dEC Setthedecimalformat(USA(20.00)orEuropean(20,00))
PoFF Enable or disable the auto-power off function
bEEP Turn the beeper on or off
tYPE Selectthermocoupletype(KorJ)
t-CF Select the temperature unit of measure (°C or °F)
SP-t Set the data logging sampling rate
Sd F Format the SD memory card
3. Once the appropriate parameter has been selected, follow the
instructions below.
Note:ThemeterwillautomaticallyexitoutoftheSetupmodeifnokeyis
pressed within 7 seconds.
Setting the Time and Date (dAtE)
1. Press the ENTER button when "dAtE" appears on the LCD.
2. Press the and buttons to adjust the year as indicated by "YY."
3. Press the ENTER button to conrm selection.
4. Repeat steps 1 and 2 for month, day, hour, minute and seconds as
indicatedby(mm/dd/h/m/s).
5. After each value has been selected and conrmed, it will automatically
skiptothenextparameter.
Note: At any time, you can press the ESCbuttontoexittheSetupmode
and resume normal operation. The internal clock will keep accurate time
when the meter is powered off. When new batteries are installed the clock
will have to be reset.
continued...
REED Instruments

13
Setting Data Decimal Format (dEC)
Numeric formats vary in different countries. By default the meter is set
to bASIC mode where a decimal point is used to separate units, (i.e.
20.00). The European format uses a comma (i.e. 20,00) to separate units.
To change this setting, follow steps 1 and 2 when the "dEC" parameter
appears on the LCD.
1. Press the and buttons to select between bASIC and Euro.
2. Press the ENTER button to conrm selection and skip to
nextparameter.
Note: At any time, you can press the ESCbuttontoexittheSetupmode
and resume normal operation.
Enabling/Disabling Auto Power Off (PoFF)
Follow steps 1 and 2 when "PoFF" appears on the LCD.
1.
Press the and buttons to select between ON (enabled) or OFF
(disabled). With the Auto Power OFF feature enabled, the meter will
automatically switch OFF after 10 minutes of inactivity to preserve
battery life.
2. Press the ENTER button to conrm selection and skip to
nextparameter.
Note: At any time, you can press the ESCbuttontoexittheSetupmode
and resume normal operation.
Enabling/Disabling the Beeper (bEEP)
Follow steps 1 and 2 when "bEEP" appears on the LCD.
1.
Press the and buttons to select between ON (enabled) or
OFF (disabled).
2. Press the ENTER button to conrm selection and skip to
nextparameter.
Note: At any time, you can press the ESCbuttontoexittheSetupmode
and resume normal operation.
continued...
REED Instruments

14
Selecting Thermocouple Type (tYPE)
Follow steps 1 and 2 when "tYPE" appears on the LCD.
1. Press the and buttonstoselectbetweentypeKandJ.
2. Press the ENTER button to conrm selection and skip to
nextparameter.
Note: At any time, you can press the ESCbuttontoexittheSetupmode
and resume normal operation.
Selecting the Temperature Unit of Measure (t-CF)
Follow steps 1 and 2 when "t-CF" appears on the LCD.
1. Press the and buttons to select between °C and °F.
2. Press the ENTER button to conrm selection and skip to
nextparameter.
Note: At any time, you can press the ESCbuttontoexittheSetupmode
and resume normal operation.
Setting the Data Logging Sampling Rate (SP-t)
Follow steps 1 and 2 when "SP-t" appears on the LCD.
1. Press the and buttons to to adjust the sampling rate between 0,
1, 2, 5, 10, 30, 60, 120, 300, 600, 1800, and 3600 seconds.
2. Press the ENTER button to conrm selection and skip to
nextparameter.
Note: At any time, you can press the ESCbuttontoexittheSetupmode
and resume normal operation.
Formatting the SD Card (Sd F)
Follow steps 1 through 5 when "Sd F" appears on the LCD.
1.
Press the and buttons to select "YES" to format the card. Select
"NO" to abort.
2. Press the ENTER button to conrm selection.
3. Press the ENTER button again to re-conrm.
4. The meter will format the SD card and automatically return to the
Setup menu when formatting is complete.
5. Press the ESCbuttontoexittheSetupmodeandresume
normal operation.
REED Instruments

15
Note: It is recommended that new SD cards should be formatted prior to
rst use. If the formatting process fails, the meter will display -E-.
Verify Set Time and Date
During normal operation press the TIMECHECK button to display the
year, month, day, hour and minutes.
Verify Sampling Rate
During normal operation, press the SAMPLINGCHECK button to view the
selected sampling rate.
Data Logging
Data Recording Modes
Manual Data Logging: Press the LOG button to manually log up to 99
readings on a SD memory card (see Manual Data Logging Mode for details).
Automatic Data Logging: Setup the meter parameters in order to
automatically log data on a SD memory card. The number of data points
is limited by the size of the memory card.
Note:
Itisrecommendedtouseaclass4SDHCmemorycardbetween1Gb
and16Gb.InserttheSDmemorycardintheslotatthebottomofthemeter.
The card must be inserted with the label side facing the rear of the meter.
Manual Data Logging Mode
In manual mode, the LOG button is pressed to manually log a reading on
the inserted SD card at any time. In order to setup the meter for manual
data logging, follow steps 1 through 8 below.
1. Set the data logging sampling rate to "0" seconds (see Setting the
Data Logging Sampling Rate section for details).
2. Press the REC button and "REC" will appear on the LCD.
3. The meter will also display P-X (X = memory position number
between 1 and 99).
4. Press the SET button to enter the data memory positions selection.
5. Press the and buttons to select one of the 99 data memory
positions in which to record.
6. Press the ENTER button to conrm selection.
7. Press the LOG button to save a reading to memory.
continued...
REED Instruments

16
Note: If a card is not inserted or the card is defective, the meter will ash
"CArD -E-". In this case, power the meter OFF and try again with another
SD memory card or verify that the card is correctly inserted.
8. Toexitmanualdataloggingmode,pressandholdtheREC button
2 seconds to resume normal operation.
Automatic Data Logging Mode
In automatic mode the desired data logging sampling rate is set to 1, 2, 5, 10,
30, 60, 120, 300, 600, 1800 or 3600 seconds prior to recording
(see Setting
the Data Logging Sampling Rate section for details).
In order to setup the
meter for automatic data logging, follow steps 1 through 5 below.
1. To begin a data logging session, press the REC button and "REC"
will appear on the LCD.
2. Press the LOG button. The meter will scan for a SD memory card.
Note: If a card is not inserted or the card is defective, the meter will ash
"CArD -E-" indenitely. In this case, power the meter OFF and try again
with another SD memory card or verify that the card is correctly inserted.
3. The "REC" icon will appear on the LCD and will continuously ash
while in a data logging session.
4. To pause the data logging session press the LOG button and "REC"
will stop ashing. To resume press the LOG button again.
5. To end the current data logging session pause the data logging
session and hold the REC button until "REC" disappears.
Transferring Data from the SD Memory Card to a Computer
1. While the meter is powered Off remove the SD memory card from the
SD card slot.
2. Insert the SD memory card into the computer.
continued...
REED Instruments

17
3. Openthele(s)withExcel.Seesamplebelow:
Place Date Time Value Unit Value Unit
1 19-07-10 8:14:05 0.6 m/S 21.6 AMTemp C
2 19-07-10 8:14:08 0.5 m/S 21.6 AMTemp C
3 19-07-10 8:14:10 15 m/S 27 AMTemp C
4 19-07-10 8:14:12 0.4 m/S 21.6 AMTemp C
5 19-07-10 8:14:14 0.4 m/S 21.6 AMTemp C
6 19-07-10 8:14:16 10 m/S 30 AMTemp C
7 19-07-10 8:14:18 0.3 m/S 21.6 AMTemp C
8 19-07-10 8:14:20 0.1 m/S 21.6 AMTemp C
9 19-07-10 8:14:22 13 m/S 25 AMTemp C
10 19-07-10 8:14:24 1.1 m/S 21.4 AMTemp C
11 19-07-10 8:14:26 2.8 m/S 21.3 AMTemp C
12 19-07-10 8:14:28 3.7 m/S 21.3 AMTemp C
TherawdatacanbeusedtocreateagraphinExcel.Seesamplebelow:
0
5
10
15
20
25
30
35
8:14:05 8:14:08 8:14:10 8:14:12 8:14:14 8:14:16 8:14:18 8:14:20 8:14:22 8:14:24 8:14:26 8:14:28
19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10
1 2 3 4 5 6 7 8 9 10 11 12
Value
Unit
Value
Unit
Data Stream Output
UsingTerminalapplication,adatastreamcanbeviewedfromtheRS-232
output,connectRS-232/USBcablebetweentheproductandterminal/PC
and use the following setting to view the data stream.
Terminal Settings:
• Bitspersecond:9600
• Databits:8
• Parity:None
• Stopbits:1
continued...
REED Instruments

18
41080100000035 = Upper Value, 3.5 m/s
Reading
Decimal Place 0-3
Positive (+) = 0,
Negative (-) = 1
°C= 01, °F = 02, m/s= 08, Knot= 09,
Km/h= 10, ft/min= 11, mile/h= 12
1 = Upper display value,
2 = Lower display value
4
41010100000221 = Canal T1, +22.1°C
Lecture
Position décimale 0-3
Positif (+) = 0, Negatif (-) = 1
°C = 01, °F = 02
Canal T1 = 1, T2 = 2, T3 = 3, T4 = 4
4
3.5mm Terminal to RS-232 Serial Connection
RS-232 Mono Jack
Pin 2 (RX)
Pin 4 (DTX)
Sleeve
Tip
System Reset
If the meter becomes unresponsive or if the display freezes, the
Reset
pin
can be used to reset the instrument.
1. UseapapercliporanysimilarsmallobjecttopresstheResetpin.
2. After pressing the
Reset
pin, power the meter back up by holding the
POWER button for 2 seconds.
3. If resetting the meter does not resolve the issue, please return the
meter to the nearest authorized REED distributor or authorized
Service Center for repair.
REED Instruments

19
Battery Replacement
When the low battery icon appears on the LCD, the batteries must
be replaced.
1. Remove the two (2) Phillips screws on the back of the meter directly
located above the top of the kickstand.
2. Remove the battery cover.
3. Replace6x"AA"batteries.
4. Secure the battery cover with the two (2) Phillips screws.
Applications
• HVAC/RandRefrigerationServicing
• DetectingFilterBlockage
• CondensersandChillers
Accessories and Replacement Parts
R4500SD-PROBE Replacement Probe for R4500SD
R2920 Surface Thermocouple Probe
R2930 Right Angle Thermocouple Probe
R2940Air/GasThermocoupleProbe
R2950 Immersion Thermocouple Probe
R2960 Needle Tip Thermocouple Probe
TP-01 Beaded Thermocouple
R1500 Tripod
RSD-ADP-NA Power Supply, 110V
RSD-ADP-EU Power Supply, 220V
CA-05A Soft Carrying Case
R8888DeluxeHardCarryingCase
SD-4GB4GBClass4SDHCMemoryCard
RSD-16GB16GBMicroSDMemoryCardw/Adapter
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.reedinstruments.com.
continued...
REED Instruments

20
Frequently Asked Questions (FAQ's)
I cannot turn off my meter, do you know why?
Often times, the meter cannot be turned off because it is in recording
mode("REC"mode).Makesuretoexitthatfunctionbyholdingdownthe
REC button before attempting to turn the meter off.
How long can I record for?
Battery life will depend on a number of factors:
-Quality of Battery
-Whether the beeper is ON or OFF
-Whether the backlight is ON or OFF
-Environmental Conditions during data logging
As a reference, using alkaline batteries with both the backlight and beeper
on while automatic data logging at a 2 second sampling rate; the batteries
lastedapprox.14hours.
Product Care
To keep your instrument in good working order we recommend the following:
• Storeyourproductinaclean,dryplace.
• Changethebatteryasneeded.
• Ifyourinstrumentisn'tbeingusedforaperiodofonemonthorlonger
please remove the battery.
• Cleanyourproductandaccessorieswithbiodegradablecleaner.Donot
spraythecleanerdirectlyontheinstrument.Useonexternalpartsonly.
REED Instruments
Table of contents
Other Reed Instruments Data Logger manuals

Reed Instruments
Reed Instruments R4000SD User manual

Reed Instruments
Reed Instruments R8100SD User manual

Reed Instruments
Reed Instruments R6050SD User manual

Reed Instruments
Reed Instruments R6030 User manual

Reed Instruments
Reed Instruments R3100SD User manual

Reed Instruments
Reed Instruments R5003 User manual

Reed Instruments
Reed Instruments R6250SD User manual

Reed Instruments
Reed Instruments R6020 User manual