Reichert Manual Lensometer ML1 User manual

ML1
Manual Lensometer®
User’s Guide

2 15110-101 Rev. D
©2013 AMETEK, Inc.
Reichert, Reichert Technologies and Lensometer are registered trademarks of Reichert, Inc.
AMETEK is a registered trademark of AMETEK, Inc.
All other trademarks are property of their respective owners.
The information contained in this document was accurate at time of publication. Specifications subject to change without notice.
Reichert, Inc. reserves the right to make changes in the product described in this manual without notice and without incorporating those
changes in any products already sold.
ISO 9001/13485 Certified – Reichert products are designed and manufactured under quality processes meeting
ISO 9001/13485 requirements.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic,
mechanical, recording, or otherwise, without the prior written permission of Reichert, Inc.
Caution: Federal law restricts this device to sale by or on the order of a licensed physician. Rx only.

3
15110-101 Rev. D
Table of Contents.............................................................................................3
Warnings and Cautions ...................................................................................4
Symbol Information..........................................................................................6
Introduction......................................................................................................7
Indications for Use.....................................................................................7
Contraindications.......................................................................................7
Instrument Setup.............................................................................................8
Unpacking..................................................................................................8
Parts Diagram............................................................................................9
Battery Installation...................................................................................10
Tilt Angle..................................................................................................10
Focus the Eyepiece................................................................................. 11
Holding a Lens for Measurement ............................................................12
Lens Analysis...........................................................................................12
Reading in Plus Cylinder....................................................................13
Reading in Minus Cylinder..................................................................13
Lens Centering & Marking.......................................................................13
Measuring Prism Power ..........................................................................14
Measuring Bifocal Additions ....................................................................15
Fused Bifocal Lens.............................................................................15
One-Piece Bifocal Lens......................................................................15
Measuring Contact Lenses......................................................................15
Cleaning & Maintenance ...............................................................................16
Cleaning ..................................................................................................16
Exterior Cleaning................................................................................16
Lens Cleaning.....................................................................................16
InkpadRelling ........................................................................................16
Target Centering......................................................................................17
Removing Prism Compensator...........................................................17
Target Centering Procedure ...............................................................17
Troubleshooting.............................................................................................18
Specications ................................................................................................19
Warranty........................................................................................................20
Table of Contents

4 15110-101 Rev. D
Reichert Technologies (Reichert) is not responsible for the safety and reliability of this instrument when:
Assembly,disassembly,repair,ormodicationismadebyunauthorizeddealersorpersons.•
Instrument is not used in accordance with this User’s Guide.•
WARNING: AN INSTRUCTION THAT DRAWS ATTENTION TO THE RISK OF INJURY OR DEATH.
WARNING: UNITED STATES FEDERAL LAWAND EUROPEAN REGULATIONS REQUIRE THAT THIS
DEVICE BE PURCHASED ONLY BYA PHYSICIAN OR A PERSON ACTING ON BEHALF OF A PHYSICIAN.
WARNING: THIS INSTRUMENT SHOULD BE USED IN STRICT ACCORDANCE WITH THE
INSTRUCTIONS OUTLINED IN THIS USER’S GUIDE. THE SAFETY OF THE OPERATOR AND THE
PERFORMANCE OF THE INSTRUMENT CANNOT BE GUARANTEED IF USED INA MANNER NOT
SPECIFIED BY REICHERT TECHNOLOGIES.
WARNING: DO NOT REPAIR OR SERVICE THIS INSTRUMENT WITHOUTAUTHORIZATION FROM
THE MANUFACTURER. ANY REPAIR OR SERVICE TO THIS INSTRUMENT MUST BE PERFORMED BY
EXPERIENCED PERSONNEL OR DEALERS WHO ARE TRAINED BY REICHERT SO THAT CORRECT
OPERATION OF THIS INSTRUMENT IS MAINTAINED.
WARNING: MODIFICATIONS TO THIS INSTRUMENT IS NOT ALLOWED. ANY MODIFICATION TO THIS
UNIT MUST BE AUTHORIZED BY REICHERT SO THAT CORRECT OPERATION IS MAINTAINED.
WARNING: IF THIS INSTRUMENT IS MODIFIED, APPROPRIATE INSPECTION AND TESTING MUST BE
CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT.
WARNING: THE INTERNAL CIRCUITRY OF THIS INSTRUMENT CONTAINS ELECTROSTATIC
DISCHARGE SENSITIVE DEVICES (ESDS) THAT MAY BE SENSITIVE TO STATIC CHARGES PRODUCED
BY THE HUMAN BODY. DO NOT REMOVE THE COVERS WITHOUT TAKING PROPER PRECAUTIONS.
WARNING: THE EQUIPMENT OR SYSTEM SHOULD NOT BE USED ADJACENT TO OR STACKED WITH
OTHER EQUIPMENTAND THAT IFADJACENT OR STACKED USE IS NECESSARY, THE EQUIPMENT
OR SYSTEM SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE CONFIGURATION IN
WHICH IT WILL BE USED.
WARNING: THIS INSTRUMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE
ANESTHETIC MIXTURES, SUCH AS OXYGEN OR NITROUS OXIDE.
WARNING: THE BATTERY SHOULD ONLY BE REPLACED WITH THE BATTERY SPECIFIED IN THIS
MANUAL. USE OF ANOTHER BATTERY MAY CAUSE FIRE OR AN EXPLOSION.
WARNING: DO NOT PLACE A SHORTING DEVICE BETWEEN THE BATTERY TERMINALS, OR ALLOW
THE BATTERY TO BECOME WET. MISUSE OR IMPROPER DISPOSAL OF THIS BATTERY MAY
CAUSE IT TO BECOME VERY HOT, IGNITE OR EXPLODE. DAMAGE TO THIS UNITAND/OR SERIOUS
PERSONAL INJURY MAY RESULT.
WARNING: DO NOT RECHARGE THE BATTERIES. THE BATTERY IS NOT DESIGNED TO BE
CHARGED BYANY ELECTRICAL SOURCE. CHARGING COULD GENERATE GAS AND INTERNAL
SHORT-CIRCUITING, LEADING TO DISTORTION, LEAKAGE, OVERHEATING, EXPLOSION OR FIRE.
WARNING: DO NOT EXPOSE THE BATTERIES TO TEMPERATURES ABOVE 140ºF, DISASSEMBLE THE
BATTERIES, OR DAMAGE TO THIS UNITAND/OR SERIOUS PERSONAL INJURY MAY RESULT.
WARNING: NEVER ALLOW LIQUID LEAKING FROM THE BATTERY TO GET IN YOUR EYES OR MOUTH
AS THIS LIQUID COULD CAUSE SERIOUS PERSONAL INJURY. IF IT COMES IN CONTACT WITH
YOUR EYES OR MOUTH, FLUSH THEM IMMEDIATELY WITH PLENTY OF WATER AND CONSULTA
PHYSICIAN.
Warnings and Cautions

5
15110-101 Rev. D
Warnings and Cautions(continued)
WARNING: ALWAYS KEEP BATTERIES OUT OF THE REACH OF INFANTS AND YOUNG CHILDREN TO
PREVENT THEM FROM BEING SWALLOWED. IF SWALLOWED, CONSULT A PHYSICIAN IMMEDIATELY.
CAUTION: AN INSTRUCTION THAT DRAWS ATTENTION TO THE RISK OF DAMAGE TO THE PRODUCT.
CAUTION: THE INTERNAL CIRCUITRY OF THE INSTRUMENT CONTAINS ELECTROSTATIC
DISCHARGE SENSITIVE DEVICES (ESDS) THAT MAY BE SENSITIVE TO STATIC CHARGES PRODUCED
BY THE HUMAN BODY. DO NOT REMOVE THE COVERS WITHOUT TAKING PROPER ESDS
PRECAUTIONS.
CAUTION: DO NOT USE SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANY PART OF THIS
INSTRUMENTAS DAMAGE TO THE UNIT MAY OCCUR. SEE MAINTENANCE SECTION FOR DETAILED
CLEANING INSTRUCTION.
CAUTION: MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING EMC
AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO THE EMC INFORMATION
PROVIDED IN THEACCOMPANYING DOCUMENTS.
CAUTION: PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL
ELECTRICAL EQUIPMENT.
CAUTION: THIS INSTRUMENT IS NOT TO BE USED NEAR HIGH-FREQUENCY EMITTING SURGICAL
EQUIPMENT.

6 15110-101 Rev. D
Symbol Information
The following symbols appear on the instrument.
Consult Instructions for Use symbol indicating important operating
and maintenance instructions that are included in this User’s Guide
Caution Symbol indicating important information and maintenance
instructions that are included in the User’s Guide
TypeBProductClassication
Class I Equipment, Continuous Operation
Protective Earth
ON / OFF
DC Volts
Date of Manufacture
REF Catalog Number
SN Serial Number
Waste of Electrical and Electronic Equipment
Compliance to Medical Device Directive 93/42/EEC
Fragile Contents in Shipping Container - handle with care
Keep Dry - Package shall be kept away from rain
AuthorizedRepresentativeinEuropeanCommunity
This Way Up - Indicates the correct upright position of package

7
15110-101 Rev. D
Introduction
Congratulations on the purchase of your new ML1 Manual Lensometer®.
The ML1 will provide you with fast, accurate and reliable measurements of eyeglass lenses and contact
lensesformanyyears.Theinstrumentutilizesaninnovativebattery-powered,LEDilluminationsystem
to enable clear and precise measurements of single, multifocal, and progressive lenses.
This User’s Guide is designed as a training and reference manual for operation, maintenance and
troubleshooting. We recommend that you read it carefully prior to use and follow the instructions in the
guide to ensure optimum performance of your new instrument.
Please retain this manual for future reference and to share with other users. Additional copies can
beobtainedfromyourauthorizedReichertdealerorfromtheReichertCustomerServicedepartment.
Contact information is provided at the end of this guide.
Indications for Use
The ML1 is a battery powered lensmeter that is intended for measuring spectacle lenses as well as
contact lenses.
Contraindications
None.

8 15110-101 Rev. D
Instrument Setup(continued)
Unpacking
Open the top of the shipping box and remove the1. User’s Guide from the top. Refer to Figure SU-01.
Remove the Styrofoam Box and set on a flat2. surface. Refer to Figure SU-02.
Cut or remove the Tape that is at the seam of3. the Styrofoam Box. Refer to Figure SU-02.
CAUTION: THERE IS ONLY A THIN AMOUNT OF
STYROFOAM BETWEEN THE INSTRUMENTAND THE
BOX. BE CAREFUL NOT TO CUT INTO THE TAPE
AND STYROFOAM TOO DEEP IF USING A KNIFE OR
SCISSORS, OR THE ML1 MAY BECOME SCRATCHED OR
DAMAGED.
Lift off the top part of the Styrofoam Box.4. Refer to Figure SU-03.
Remove the ML1 Lensometer and stand it5. upright on a flat surface.
Remove the accessories. Refer to Figure SU-03.6.
Accessories Included:
2 AAA Batteries (P/N 13950000-902)• Dust Cover (P/N 15110-016)• Contact Lens Holder (P/N 15110-017)• User’s Guide (P/N 15110-101)• (Removed in step 1)
Carefully remove the plastic bag that was7. protecting the ML1 during shipping.
Refer to Figure SU-03.
Remove the Rubber Band, Styrofoam Block,8. and package of Desiccant that was holding the
Marking Pens upright during transport.
Refer to Figure SU-04.
Keep the packaging in case transportation of9. the ML1 is required in the future.
Figure SU-01 User’s Guide
Styrofoam
Box
Cardboard Box
User’s
Guide
Figure SU-02 Styrofoam Box
Styrofoam Box
Tape
Figure SU-03 ML1 and Accessories
2 ‘AAA’
Batteries Contact
Lens Holder
Dust
Cover
ML1
Figure SU-04 Rubber Band & Styrofoam
Rubber
Band
Styrofoam
Block Desiccant

9
15110-101 Rev. D
Instrument Setup(continued)
Parts Diagram
ML1 Parts
On-Off Switch1. Table Lift Lever2. Lens Clamp Handle3. Marking Lever4. Focusable Eyepiece5. Prism Compensator6. Lens Clamp7. Lens Holder8. Power Wheel9. Axis Wheel10. Battery Compartment11. Tilt Lock Lever12. Lens Table13.
Figure SU-05 ML1 Parts Diagram
12
7
9
11
51086 7
4
3
2
13
Accessories
Dust Cover (P/N1. 15110-016)
Contact Lens Holder (P/N 15110-017)2. 2 “AAA” Batteries (P/N 13950000-902)3. Figure SU-06 MML1 & Accessories
1
2 3

10 15110-101 Rev. D
Instrument Setup(continued)
Battery Installation
Using a flathead screwdriver, remove the1. Screw that secures the Battery Compartment
to the ML1. Refer to Figure SU-07.
Open the Battery Compartment and install the2. two “AAA” Batteries. Refer to Figure SU-08.
Close the Battery Compartment and secure it3. with the screw. Refer to Figure SU-07.
Tilt Angle
The angle of the main body can be adjusted for
comfort or convenience.
Note: When measuring contact lenses, the ML1
needs to be fully upright.
Rotate the Tilt Lock Lever counter-clockwise1. to loosen the tilt mechanism.
Refer to Figure SU-08.
Move the main body up or down to the desired2. position.
Tighten the Tilt Lock Lever to secure the main3. body in the desired position.
Refer to Figure SU-08.
Turning the ML1 On and Off
To turn on the instrument, press the On-Off
Switch. Refer to Figure SU-09. A green LED
light will illuminate, indicating the unit is on.
Note: The ML1 will enter ‘sleep mode’ and turn
off automatically after approximately 5
minutes.
To turn the power off, press the On-Off Switch
again. Refer to Figure SU-09. The green LED
light will turn off, indicating that the power is off.
Figure SU-07 Battery Screw
Screw
Battery
Compartment
Figure SU-08 Install Batteries
Tilt Lock
Lever
“AAA” Batteries
Figure SU-09 On-Off Switch
On-Off
Switch

11
15110-101 Rev. D
Focus the Eyepiece
It is important to insure that the ML1 is focused
before use. Each user must focus the instrument
for their individual eye.
Note:
Failure to focus the ML1 could result in
inaccurate measurements.
Place a white sheet of paper at the position of1. the Lens Holder. This will provide enough light
to illuminate the reticle in the Eyepiece.
Turn the Eyepiece counterclockwise to the2. stop.
With both eyes open, look through the3. Eyepiece with your dominant eye.
Slowly turn the Eyepiece clockwise until the4. black reticle lines are in sharp focus.
Refer to Figure SU10.
Note: Do not go beyond that point.
Remove the paper.5. Press the ON-OFF Switch. The green LED6. light will illuminate and the internal LED will
turn on.
Note: The ML1 will enter ‘sleep mode’ and turn
off automatically after approximately 5
minutes.
Turn the Power Wheel until the illuminated7. cross-line target becomes clear.
The power scale should read “0”.8. Refer to Figure SU-11.
If it does not, repeat from step 1.9.
Note: If the power scale does not read “0”
without a lens in place, the instrument
needs calibration.
Figure SU-10 Reticle Lines
Figure SU-11 Zero Power
Instrument Setup(continued)

12 15110-101 Rev. D
Instructions For Use
Holding a Lens for Measurement
Place the concave, or back side of the lens1. against the Lens Holder.
Note: If the lens is edged, or “laid out” for cutting,
place the 180° line parallel to the Lens
Table.
Note: Be sure the center point of the lens is in the
center of the Lens Holder.
If the lens is mounted in a frame, raise the Lens2. Table using the Lens Table Lever until the Lens
Table supports the bottom of the frame.
Pull back on the Lens Clamp Handle and3. slowly allow the Lens Clamp to move forward to
contact the lens.
Note: Do not release the Handle until the Lens
Clamp makes contact with the lens.
Lens Analysis
With the lens in position, rotate both the Power1. Wheel and Axis Wheel to sharply focus the 3
narrow lines.
Note: Make certain that the last focusing
movement of the Power Wheel is topside
toward you.
If both the Sphere Lines (3 narrow lines) and2. Cylinder Lines (3 wide lines) come into focus
at the same time, the lens has sphere power
only and the sphere power can be read on the
Power Wheel. Refer to Figure IU-01.
If only the Sphere Lines or only the Cylinder Lines3. can be focused sharply, the lens has cylinder
power. Refer to Figures IU-02 and IU-03.
By turning the Power Wheel, you will be able to4. find a position at which the Sphere Lines are in
focus. Refer to Figure IU-02.
When the Sphere Lines are in focus, turning the5. Power Wheel only will bring the Cylinder Lines
into focus. Refer to Figure IU-03.
Note: The prescription of any cylinder lens can be
written in two forms; one with a plus cylinder
and the other with a minus cylinder.
Figure IU-01 Sphere Power Only
Sphere
Lines
Cylinder
Lines
Figure IU-02 Sphere Power Focused
Figure IU-03 Cylinder Power Focused

13
15110-101 Rev. D
Instructions For Use (continued)
Lens Analysis (continued)
Reading Plus Cylinder
Set the Power Wheel at the lower of the two1. focusing positions (toward the minus end of
the scale).
Rotate theAxis Wheel until the Sphere Lines are2. in focus and unbroken. Refer to Figure IU-02.
Read the sphere power and axis from the3. Power Wheel and Axis Wheel.
Rotate the Power Wheel until the Cylinder4. Lines are in focus. Refer to Figure IU-03.
The cylinder power is the difference between5. the first and second powers indicated on the
Power Wheel.
Refer to Figure IU-04 for a sample reading of6. plus cylinder lenses.
Reading Minus Cylinder
Set the Power Wheel to the higher of the two1. focusing positions (toward the plus end of the
scale).
Rotate theAxis Wheel until the sphere lines are2. in focus and unbroken. Refer to Figure IU-02.
Read the sphere power from the Power Wheel3. and the axis from the Axis Wheel.
Rotate the Power Wheel until the Cylinder4. Lines are in focus. Refer to Figure IU-03.
The minus cylinder power is the difference5. between the first and second powers indicated
on the Power Wheel.
Refer to Figure IU-04 for a sample reading of6. minus cylinder lenses.
Example of reading Plus Cylinder:
First reading: -3.00D X 110°
Second Reading: -2.50D
Direction: +
Rx: -3.00 +.5 X 110°
Example of reading in Minus Cylinder:
First reading; -2.50D X 20°
Second reading: -3.00D
Direction: -
Rx: -2.50 -0.50 X 20°
Figure IU-04 Sample Readings
Lens Centering and Marking
The optical center of a lens can be located and
marked in the following manner.
With the lens clamped in position against the1. Lens Holder, slightly release the pressure of
the Lens Clamp by pulling back somewhat on
the Lens Clamp Handle.
Move the lens so that the center of the target is2. located at the center of the reticle.
Refer to Figure IU-05.
Mark the lens by pressing the Marking Lever3. forward, gently and slowly, until the Marking
Pens come in contact with the lens.
Note: Avoid forceful operation which can cause
inaccurate marking.
Three ink dots will be placed on the lens.4.
Note: The center dot indicates the optical center.
Note: All three dots indicate the 180° line of the
lens. The axis is indicated on the Axis
Wheel.
When marking a lens for layout, turn the Axis5. Wheel to the axis specified on the prescription.
Focus the Sphere Lines by turning the Power6. Wheel and the lens.
Move the target to the center of the Eyepiece.7. Refer to Figure IU-05.
Mark the lens. The 3 dots will mark the optical8. center of the lens and indicate a line parallel to
the frame.
Figure IU-05 Target Centered

14 15110-101 Rev. D
Instructions For Use(continued)
Measuring Prism Power
Prism power is indicated by the position of the center
of the target with respect to the Eyepiece reticle.
The innermost Prism Ring is 1 prism diopter, the
second Prism Ring is 2 prism diopters, and the third
Prism Ring is 3 prism diopters. Refer to Figure IU-06.
Note: The innermost tips of the scale lines
indicates 1/2 prism diopter.
Place a dot on the lens at the “optical center”1. position (the point at which the prismatic power
is to be measured).
Place the lens in the instrument with the dot in2. the center of the Lens Holder and the 180° line
parallel to the Lens Table.
Bring the target into focus.3.
Note: The target will not be centered on the reticle
if there is prism power. The target will
always be decentered in the direction of
the base of the prism. For example, if the
center of the target is above the center of
the reticle, base up prism is indicated.
Rotate the Prism Scale until the line of the scale4. passes through the center of the target. Refer
to Figure IU-07.
Note: The amount of prism power is indicated by
the displacement of the target center with
reference to the concentric circles of the
reticle.
Note: Powers of 1.00 to 4.00 prism diopters are
clearly indicated with numerical designations.
Small cross-lines on the Prism Scale indicate
1.5 prism diopters and 5 prism diopters at the
extreme ends of the Prism Scale.
Note: Figure IU-07 indicates a prism power of 3
diopters at an axis of 35°.
If the center of the target is below the center of5. the Eyepiece scale, add 180° to the axis scale
reading.
When measuring prism powers above 5.06. diopters, the Prism Compensator is used to
bring the center of the target into the center of
the Eyepiece scale.
The prism power and axis is then read from the7. Compensator Scales.
Figure IU-06 Prism Rings
Prism
Rings
Figure IU-07 3 Diopters of Prism

15
15110-101 Rev. D
Instructions For Use(continued)
Measuring Bifocal Additions
The bifocal addition equals the difference between
the powers measured through the distance and
reading portion of a bifocal lens with the segment
surface against the Lens Holder.
Fused Bifocal Lens
Place the front or convex surface of the lens1. against the Lens Holder.
Measure the power through the distance2. portion of the lens.
Measure the power through the reading portion.3.
Note: The add power is the difference between
these two readings.
One-Piece Bifocal Lens
Place the concave side of the lens against the1. Lens Holder.
Measure the power through the distance2. portion of the lens.
Measure the power through the reading portion.3.
Note: The add power is the difference between
these two readings.
Measuring Contact Lenses
Remove the spectacle Lens Holder.1. Refer to Figure IU-08.
Install the Contact Lens Holder in place of the2. Spectacle Lens Holder.
Note: The Contact Lens Holder has a smaller
aperture than the Spectacle Lens Holder.
Refer to Figure IU-09.
Loosen the Tilt Lock Lever and tilt the instrument3. to the full upright position. Refer to Figure IU-10.
Tighten the Tilt Lock Lever to secure the ML14. in the upright position.
Place the contact lens on the Contact Lens5. Holder.
Measure the contact lens as you would a6. spectacle lens.
Lens Holder
Removed
Figure IU-08 Spectacle Lens Holder Removed
Figure IU-09 Lens Holders
Spectacle
Lens Holder
Contact
Lens Holder
Smaller Aperture
Figure IU-10 Fully Upright

16 15110-101 Rev. D
Cleaning
Exterior Cleaning
CAUTION: DO NOT USE SOLVENTS OR STRONG
CLEANING SOLUTIONS ON ANY PART OF THIS
INSTRUMENTAS DAMAGE TO THE UNIT MAY OCCUR.
Clean the exterior surfaces with a soft cloth
moistened with a mild soap solution (1 cc of
liquid dish soap to one liter of clean, filtered water
(filtered below 5 microns)).
Note: Keep your ML1 clean. Use the Dust Cover
when the instrument is not in use.
Lens Cleaning
Clean the lenses using any commercial lens
cleaner.
Using a lint-free cotton swab lightly moistened
with a commercial lens cleaner, clean the lens
surfaces.
Note: Clean exterior surfaces only.
Note: Do not remove any lenses.
Ink Pad Refilling
CAUTION: DO NOT OVERFILLTHE INKWELL.
IT IS ONLY NECESSARY FOR THE PAD TO BE
SATURATED WITH INK. EXCESSIVE INK CAN
CAUSE DAMAGE TO THE INSTRUMENT.
Rotate the Marking Pens up so that the Inkwell1. Cover is accessible. Refer to Figure MM-01.
Remove the Inkwell Cover by pulling it straight2. up. Refer to Figure MM-02.
Saturate the Inkpad with ink.3. Install the Inkwell Cover.4.
Note: Replace the Inkpad when it becomes worn
or dried out.
Figure MM-01 Rotate Marking Pens Up
Inkwell
Cover
Figure MM-02 Remove and Rell Inkwell
Inkwell
Cover
Inkpad
Pull
Up
Cleaning & Maintenance

17
15110-101 Rev. D
Target Centering
Removing Prism Compensator
It is best if the target centering is done without the
Prism Compensator in place.
Using a flathead screwdriver, remove the 31. Screws in the Prism Compensator.
Refer to Figures MM-03 and MM-04.
Note: There is a Screw on the top of the Prism
Compensator, and one on each side. Only
two Screws are visible in Figures MM-03
and MM-04.
Remove the Prism Compensator and set aside.2.
Target Centering Procedure
The target must be centered on the Eyepiece scale
when the Power Wheel is set to “0” and there is no
lens in place for measurement.
If the target is not centered, rotate the Axis Wheel
while observing the target. If the center of the target
remains stationary, proceed with the target centering
procedure.
Note: If the center of the target moves with respect
to the Eyepiece scale, the instrument should
be serviced.
Focus the Eyepiece to obtain the clearest image1. of the Eyepiece scale.
Turn the Power Wheel to focus the target.2. Using a flathead screwdriver, adjust the 3 Screws3. to bring the center of the target into the center of
the eye piece scale. Refer to Figure MM-05.
Note: It is necessary to loosen one Screw before
tightening another Screw in order to move
the target image.
Note: All 3 Screws should be tight when centering
is complete.
Figure MM-03 Prism Compensator Screws
Screw
Figure MM-04 Side Screws
Screw
Figure MM-05 Centering Screws
Centering
Screws
Cleaning & Maintenance(continued)

18 15110-101 Rev. D
Troubleshooting
Chart of Common Errors
Problem Cause Solution
Does not read “0” without a lens in
place Eyepiece not focused for user. Focus Eyepiece.
Needs calibration. Have instrument serviced.
High powers off same amount in
the same direction, i.e +12.00 reads
+12.25, -12.00 reads -11.75.
Vertex distance setting incorrect.
Pads may be installed on the
Lens Holder.
Remove any pads.
Needs calibration. Have instrument serviced.
High powers off in different directions
or different amounts, i.e. +12.00
reads +12.50, -12.00 reads -12.25.
Needs calibration. Have instrument serviced.
Not marking center of lens. Prism Compensator not at “0”. Set Prism Compensator to “0”.
Target not centered. Center the target.
Marker out of alignment. Have instrument serviced.
Target not centered with no lens in
place, target center remains in place
when Axis Wheel is rotated.
Prism compensator not set to “0”. Set the prism compensator to “0”.
Target not centered. Center the target.
Target not centered, target center
moves when the Axis Wheel is
rotated.
Target not mechanically
centered. Have instrument serviced.
Axis reading incorrect Target, Axis Wheel or table not
aligned. Have instrument serviced.
No Power Instrument in “sleep” mode. Press On-Off Switch.
Dead batteries. Replace batteries.
Batteries installed incorrectly. Install Batteries with negative
side toward the spring.
Electronic problem. Have instrument serviced.

19
15110-101 Rev. D
Specications
Catalog Number 15110
Physical Dimensions
Size:Weight, unpacked: 11 lbs. ( 4.9 Kg)
Height: 18.7 in. (47.5 cm)
Width: 5.1 in. (12.9 cm)
Depth: 12.9 in. (32.8 cm)
Electrical
Voltage: 3 VDC (2 AAA batteries at 1.5 VDC each)
Vertex Power Prismatic Power
Range -25 to +25 diopters Range 5 prism diopters
Steps 0.125 diopters up to +/- 3 diopters
0.25 diopters beyond +/- 3 diopters Step 0.5 prism diopters up to 2 diopters,
1 diopter beyond 2 diopters
Cylindrical Axis Prism Compensator
Range 1 to 180 degrees Diopter Range 15 prism diopters
Step 1 degree Graduations 1 prism diopter steps
Angle scale 0 degrees to 180 degrees
Graduations 5 degrees
Tilting Angle 30 degrees to 90 degrees Target American Crossline
Light Source LED (Light Emitting Diode) Lens Diameters 24 mm to 90 mm
Eyepiece Focusing +7D to -10 D
Disposal
Thisproductdoesnotgenerateanyenvironmentallyhazardousresidues.Attheendofitsproductservice
life, follow your local laws and ordinances regarding the proper disposal of this equipment.

20 15110-101 Rev. D
Reichert Warranty
This product is warranted by Reichert Technologies against defective material and workmanship under normal
useforaperiodofoneyearfromthedateofinvoicetotheoriginalpurchaser.(Anauthorizeddealershallnot
be considered an original purchaser.) Under this warranty, Reichert’s sole obligation is to repair or replace the
defective part or product at Reichert’s discretion.
This warranty applies to new products and does not apply to a product that has been tampered with, altered
in any way, misused, damaged by accident or negligence, or which has had the serial number removed,
altered or effaced. Nor shall this warranty be extended to a product installed or operated in a manner not in
accordance with the applicable Reichert instruction manual, nor to a product which has been sold, serviced,
installedorrepairedotherthanbyaReichertfactory,TechnicalServiceCenter,orauthorizedReichertDealer.
Lamps, bulbs, charts, cards and other expendable items are not covered by this warranty.
All claims under this warranty must be in writing and directed to the Reichert factory, Technical Service
Center,orauthorizedinstrumentdealermakingtheoriginalsaleandmustbeaccompaniedbyacopyofthe
purchaser’s invoice.
This warranty is in lieu of all other warranties implied or expressed. All implied warranties of merchantability
ortnessforaparticularuseareherebydisclaimed.Norepresentativeorotherpersonisauthorizedtomake
any other obligations for Reichert. Reichert shall not be liable for any special, incidental, or consequent
damages for any negligence, breach of warranty, strict liability or any other damages resulting from or relating
to design, manufacture, sale, use or handling of the product.
PATENT WARRANTY
Ifnotiedpromptlyinwritingofanyactionbroughtagainstthepurchaserbasedonaclaimthattheinstrument
infringes a U.S. Patent, Reichert will defend such action at its expense and will pay costs and damages
awarded in any such action, provided that Reichert shall have sole control of the defense of any such action
with information and assistance (at Reichert’s expense) for such defense, and of all negotiation for the
settlement and compromise thereof.
PRODUCT CHANGES
Reichert reserves the right to make changes in design or to make additions to or improvements in its products
without obligation to add such to products previously manufactured.
CLAIMS FOR SHORTAGES
We use extreme care in selection, checking, rechecking and packing to eliminate the possibility of error. If any
shipping errors are discovered:
1. Carefully go through the packing materials to be sure nothing was
inadvertently overlooked when the unit was unpacked.
2. Call the dealer you purchased the product from and report the
shortage. The materials are packed at the factory and none
should be missing if the box has never been opened.
3. Claimsmustbeledwithin30daysofpurchase.
CLAIMS FOR DAMAGES IN TRANSIT
Our shipping responsibility ceases with the safe delivery in good condition to the transportation company.
Claims for loss or damage in transit should be made promptly and directly to the transportation company.
If, upon delivery, the outside of the packing case shows evidence of rough handling or damage, the
transportation company’s agent should be requested to make a “Received in Bad Order” notation on the
delivery receipt. If within 48 hours of delivery, concealed damage is noted upon unpacking the shipment and
no exterior evidence of rough handling is apparent, the transportation company should be requested to make
out a “Bad Order” report. This procedure is necessary in order for the dealer to maintain the right of recovery
from the carrier.
Table of contents
Other Reichert Microscope manuals
Popular Microscope manuals by other brands

ThermoFisher Scientific
ThermoFisher Scientific DXR xi user guide

KERN
KERN OBN-14 User instructions

scope
scope LSU VL-7EX II user guide

VISIO-TEC
VISIO-TEC IWT user manual

Ken A Vision
Ken A Vision Video Flex® except 2000 series manual

Cypress Semiconductor
Cypress Semiconductor CapSense CY8CKIT-022 Guide