Troubleshooting Alarm Situations
Disc/Sense:
There may be a leak or disconnect in the circuit or proximal pressure
sense line or the proximal pressure sense line is pinched or occluded.
High Pressure:
The circuit pressure may have exceeded the high pressure setting-check
to see if the tubing is crimped. Check the tidal volume setting, this may
have been increased.
Low Pressure:
If the peak inspiratory pressure is below the Low Pressure Alarm
setting-check all connections
When using the LTV with an open circuit (SIP and Puff), we are using
settings to avoid nuisance alarms.
High pressure is set to 50-90 cmh2O, Low pressure is turned off and the
apnea delay is set to 60 sec.
Filters:
On the left side of the ventilator there is a cooling fan with a thin gray
filter over it. At the other end there is a foam filter (off white in color).
Cleaning:
Replace all the supplies listed above once a week or as needed.
Both filters need to be cleaned as needed, checked weekly. Take them
out, clean with warm soapy water, let air dry thoroughly before putting
them back in the machine.
Battery Operation:
The internal battery will be 90% charged within 8 hours when it is
connected to an appropriate power source.