Rish Insu20 User manual

RISH
Insu20
MW
50V
100V
250V
500V1KV
V
7
TEST ON
OFF
ESC
STORE
V MW
W
fused
600V CAT II / 300V CAT III
6
5
4
3
2
1
600V
3
(1) Liquid Crystal Display (11) Symbols for displaying the
(2) ON/OFF push button selected functions
(3) Push button (12) Display for selected function
(4) Push button (13) Display for the unit of
(5) TEST Push button measured quantity
(6) Function selector switch (14) Over range Indication
(7) Terminal sockets (15) Pointer forAnalog indication
(8) External Power supply jack (Optional)
(9) Symbol for (16) Scale forAnalog Indication
“CONTINUOUSLY ON” (17) Activated stop watch Indicator
(10) Digital display with indication (18) Zero adjust indicator
of decimal point and polarity (19) Low battery indicator
(20) Buzzer indication
Contents Page
1 Safety Features and Precautions 4
2 Initial Start-Up 5
3 Selecting Measuring Functions 6
4 Liquid Crystal Display 6
4.1 Digital Display 6
4.2 Analog Display 6
4.3 Backlit 6
5 Insulation Resistance Measurement (MW) 7
5.1 Preparation for measurement 7
5.2 Insulation resistance measurement 7
5.2.1 Manual Insulation resistance measurement 7
5.2.2 Continuous measurement for a pre-selectable time 8
5.3 Conclusion of measurement and discharging 8
6 Additional Features 8
6.1 Pre-selectable limit checks (Go / No-Go option) 9
6.2 Storing the measured resistance value 9
6.3 Minimum value and Maximum value “MIN/MAX” 9
storage facility
7 Continuity Testing and Resistance Measurement 11
7.1 ContinuityTesting 11
7.2 Low Ohm Measurement 11
7.3 Zero adjustment 11
8 Voltage Measurement (VAC/DC) 12
9 Default Live Circuit Warning 12
10 Stop Watch 13
11 Menu Structure 13
11.1 Description of Menu items 14
11.1.1 Setting the GoNoGo values 14
11.1.2 Setting the Timer values 14
11.1.3 To see the stored values 14
12 Specifications 15
13 Maintenance 18
13.1 Battery 18
13.2 Operation with mainsAdapter (Optional) 18
13.3 Fuses 19
13.4 Housing 19
MIN MAX 400
100kM 600
10kM VAC VDC
ON
200
1MG
800
1kM
F
A
G
M
k
100k
1000
10MG
8
L
L
ZERO
8
(optional)
MIN MAX 400
100kM 600
10kM VAC VDC
ON
200
1MG
800
1kM
F
A
G
M
k
100k
1000
10MG
8
L
L
ZERO
20
17
9 19 10 11 12
13
18 14 15 16
CE

Repair,PartsReplacementandCalibration
After opening the instrument, live parts may be exposed. Therefore, the
instrument must be disconnected from the measuring circuit prior to
openingits casefor repair, replacementofparts orcalibration. Ifrepairor
calibrationcannotbeavoided,unlesstheinstrumentisopenandlive, the
work must be performed by a qualified person, who understands the
dangerinvolved.
FaultsandExtraordinaryStress
When it is realized that the safe operation is no longer possible, take the
instrumentoutofserviceandsecureitagainstaccidentaluse.
Safeoperationisnolongerpossible-
nWhentheinstrumentshowsobvioussignofdamage,
nWhentheinstrumentnolongerfunctionscorrectly,
nAfteraprolongedstorageunderadverseconditions,
nDuetoseverestressduetotransportation.
2InitialStart-Up
Battery
Fit the instrument with the six 1.5V cells (IEC LR6 type) provided along
withtheinstrument.
Beforeyouusetheinstrumentforthe first timeorafterstoragerefer
tosection''3.MaintenanceBattery''
MainsAdapterforexternalpowersupply(Optional)
MainsAdapter (Output : DC 9V, 500mA) can be connected for long term
measurement(withoutusageofinternalbatteries).
IsolatedAdapterhastobeused.
Donotusenon-isolatedAdapter.
Connect the jack to the instrument and Adapter to the mains power
supply.
SwitchingtheInstrumentON
InstrumentcanbeswitchedON:
1. BypressingtheON/OFFbutton.
2. Bychangingthefunctionselectorswitchfromcurrentposition
todesiredfunction.
Switch-on is acknowledged by a sound signal. The instrument is now
readyforoperation.
Attention!
Beforeopeningdisconnect theinstrumentfromthe
measuringcircuitandensurehighvoltageisnotavailable
onleads.
Automaticturn-off
Your instrument turn off automatically, if any of the keys or the function
selector switch have not been activated for about 10 minutes in
insulationresistancerangeand5minutesinotherranges.
HowtopreventautomaticTURN-OFF
In order to prevent automatic “TURN-OFF” select “CONTINUOUSLY
ON” mode For this push TEST push button and the “ON-OFF”
pushbuttontogetherwhenswitchingONtheinstrument.
The symbol is displayed on the LCD to indicate this
“CONTINUOUSLYON”mode.
SwitchingtheInstrumentOff
Press“ON/OFF”pushbuttonforalongertime.
5
ON
1) SafetyFeaturesandPrecautions
You have selected an instrument which provides you with a high level of
safety.
The RISH Insu 20 Analog-Digital Insulation Tester is manufactured in
compliance with safety regulations IEC 61010-1/EN 61010-1/VDE 0413
part1.Whenusedforitsintended purpose,safetyof theoperator,aswell
as that of the instrument, is assured. In case of incorrect use or careless
handling,thesafetyofbothuserandinstrumentisnotassured.
Tomaintainthesafeandproperconditionoftheinstr mentandtoensure
its safe operation, it is absolutely necessary to carefully and completely
read these operating instructions before using the instrument. These
instructionsmustbefollowedinallrespects.
Observethefollowingsafetyprecautions:
* The instrument must be operated only by persons who
understand the danger of shock hazards and know how to
apply safety precautions. Shock hazards exist anywhere,
where voltages of more than 30 V may occur.
* Do not work alone in shock hazardous environment while carrying
out measurement.
* The maximum allowable voltage between any terminal sockets and
earth is equal to 600 V for voltage and 1000V for Insulation
resistancemeasurement.
* Verify that the test leads are in good condition, e.g. no cracked
insulation,noopencircuitsintheleadsorconnectors.
* Measurementsundermoistambientconditionsarenotpermitted.
* Do not exceed the permissible overload limits of the measuring
ranges.Limitvaluescanbefoundinthetablein“12.Specification”.
SafetyPrecautionsforInsulationResistance
Measurement
* The circuit under test must be switched off, de-energized,
isolated before Insulation or Continuity tests are made.
* The TEST button must not be pressed while connecting the
test leads or while changing ranges.
* The Voltage warning does not function while in STOP WATCH
mode or in MENU structure.
* During an Insulation test, connections must not be touched.
* After insulation tests, capacitive circuits must be allowed
to discharge before disconnecting test leads.
* Test leads and crocodile clips must be in good condition,
clean and with no broken or cracked insulation.
Meaningofsymbolsontheinstrument
u
4
Warning concerning a point of danger
(Attention: observe documentation)
Earth
Continuous, doubled or
reinforced insulation
CE EU Confirmity Mark

6
Insulation Resistance Measurement
ON
50V
100V
250V
500V1KV
V
R
MW
Note
During Insulation resistance measurement the symbol
flashesaslongasthehighvoltageispresentattheterminals
5.InsulationResistanceMeasurement(MW)
5.1.Preparationformeasurement
·Selecttherequired"TestVoltage(50V/100V/250V/500V/1000V)"
bychangingthe"FunctionSelectorSwitch".
·Connectthetestprobesfirmlytothecircuitundertest.
Note!
Insulation resistance can only be measured at voltage free
object. Hence the circuit under test must be switched off,
de-energized and isolated before carrying out any insulation
resistancemeasurement.
When measuring high-ohmic insulation resistance do not
touchthemeasurementcables.
If an interference voltage > 25V is present within the
system, insulation resistance measurement is disabled.
The instrument defaults to the default voltmeter and
displays the value of the interference voltage on the LCD.
Also the instrument indicates this additionally by an
acousticsignal.
HIGH-VOLTAGE
Do not touch the conductive ends of the test probes
after insulation resistance measurement has been
activatedattheinstrument.
A current with a value of 2 mA (limited by the instrument)
may flow over your body, and although this is not life
endangering,theelectricshockisdistinctlyperceptible.
Ifyouaretaking ameasurement atacapacitive deviceunder test,for
example a cable, it may be charged with as much as
1000V, depending upon the selected nominal voltage. Touching the
device under test after measurement may be life endangering in
suchcases!
5.2 Insulationresistancemeasurement
The instrument provides flexibility for both manual and
continuous (for a pre-selectable time) insulation - resistance
measurement.
5.2.1 ManualInsulationresistancemeasurement:
In manual Insulation resistance measurement, Press and hold "TEST"
keyuntilthedisplayhasstabilized.Insulationresistancemeasurementis
endedwhen the yellowTESTkey is released.Instrumentwillcontinue to
show the last reading for few seconds even after releasing the "TEST
key".
Note!
The instrument's batteries are depleted during insulation
resistance measurement. Only press and hold the TEST key
as long as it is necessary to take the reading in the manual
measurement mode. Continuous measurement as described
belowshouldbeperformed,ifabsolutelynecessary.
7
RISH
Insu20
400
100kM 600
10kM
ON
200
1MG
800
1kM
G
100k
1000
10MG
8
L
L
3. SelectingMeasuringFunctions
ThedesiredmeasuringfunctionisselectedwiththeFunctionselectorswitch.
4. LiquidCrystalDisplay
4.1 DigitalDisplay
Thedigitaldisplayshowsthemeasurementvalueanddecimalpoint.The
selectedmeasuringunitandfunctionaredisplayed.
4.2 AnalogDisplay
The analog indication gives the dynamic performance of a moving coil
mechanism. This is especially advantageous for the observation of
measurementvaluefluctuationsandforcalibrationprocedures.
4.3 BackLit
The instrument is provided with user selectable Back-lit for taking
measurements in poor lighting conditions / dark areas. Refer section 11
on page 14 for switching ON / OFF the backlit from the menu.
By pressing " " and "ON/OFF" key simultaneously
during power ON, the back-lit can be switched ON.
CE

8
5.2.2 Continuous measurement for a pre-selectable time :
In normal course, the insulation test terminates when the test key
is released and the measured insulation resistance value remains
on display for few seconds after the test key is released.
With the Pre-selectable measurement time feature, the
insulationtestcontinuesforthepre-determinedtime.
Pre-selectabletime:10sec-5min
Refer to the “11. Menu structure” section for how to set the value for
pre-selectable time. Once the time is set it is applicable to all the 5
ranges of Insulation measurement.
nSelect the required value of the time for which the test is
carriedout.
nBrieflyPresstheTESTkey.Theinsulationresistancetestcarries
onforthetimeintervalselected.
nThe test terminates automatically after the selected time has
elapsed.
nThe display then flashes the value of measurement at
theenduntilthe Keyispressed.
5.3 Conclusionofmeasurementanddischarging
After the measurement has been completed, any residual voltage is
displayed which may be caused by capacitive circuits. Contact with the
device under test must be maintained. The reduction in voltage can be
directlyobservedattheLCD.
Do not disconnect the device under test until voltage has dropped
below25volts!!
Note:Automaticdischarge!
The capacitive circuits are automatically discharged when
TEST button is released following an insulation resistance test.
This auto discharge facility eliminates the need for the user to
dischargethecircuitundertestmanually.
6 AdditionalFeatures
AdditionalfeaturesavailableinInsulationmeasurementranges.
1. Pre-selectablelimitchecks(Go/Nogooption)
2. Storingthemeasuredresistancevalue
3. MinimumandMaximumvalue“MIN/MAX”storage
to be
6.1 Pre-selectablelimitchecks(Go/No-Gooption)
An acoustic signal can be generated when the measured value of
resistancefallsbelowanadjustablenoGovalue.
A separate noGo value can be set for each range of insulation
resistancemeasurementaswellasforOhms,Continuityranges.
Refer to the “11. Menu structure” section for how to set the value
forthenoGovalue.
6.2 Storingthemeasuredresistancevalue
lTostorethevalueofresistancepress“STORE”keywhenthe
resistance measurement is in progress. This will store
the resistance measured at the instant the key is pressed.
lThis way maximum 10 readings per range can be stored.
lAfter 10 values are stored for the selected range the
th
11 value when stored will over-write the first value i.e. at any
given point of time only ten readings per range can be
stored.
lThe readings are stored in a volatile memory. Therefore
when the instrument is switched off, the readings are lost.
Recallingthestoredvalues
The stored values can be recalled and viewed in the menu
structure.
Refer to the “11. Menu Structure” section for how to view the
storedvalue.
6.3 Minimum value and Maximum value “MIN/MAX” storage
facility (for all measuring ranges)
With the MIN/MAX function, you can hold the minimum and the
maximum measured value which was applied to the input of the
instrument after activating MIN/MAX. The most important application is
determination of minimum and the maximum value for long term
monitoringofmeasuredquantities.
The function MIN/MAX can be activated in all measuring ranges.
MIN/MAX does not influence analog indication, you can still read the
actualmeasuredvalue.
Apply the measured quantity to instrument prior to activating MIN/MAX
function.
TheMIN/MAXfunctionisswitchedoffwhen.
1. Keyispressedforapproximatelyfor1Sec. OR
2. Thenfunctionselectorswitchisoperated OR
3. WhentheinstrumentisturnedOFFandONagain
9
Function
MIN/MAX Measured Value
MIN and MAX
Meter reaction
Display
Measured value Digital MIN/MAX Sound
signal
1. Activate
and store
2. Store
and display
3. Return
to 1.
Reset
short
short
short
short
long
Stored
Stored continued
in background
new MIN and MAX
values are displayed
Same as 1.
Stored values are
not cleared
Cleared
Actual measured value
Stored MIN value
Stored MAX value
Same as 1.
Cleared
MIN/MAX
flash
MIN flash
MAX flash
Same as 1.
Cleared
1x
1x
1x
1x
2x

10
Continuity Test
-
MW
50V
100V
250V
500V1KV
V
+
-
MW
50V
100V
250V
500V1KV
V
+
7. ContinuityTestingandResistanceMeasurement
Attention!
Be absolutely certain that the device under test
isvoltage-free.
7.1 ContinuityTesting
* Verifythatdeviceundertestiselectricallydead.
* Setthefunctionselectorswitchto“ ”position.
* IfrequirednullleadresistancebydoingZeroadjustmentbelow.
* Connectthedeviceundertestasshownindiagram.
* Themeasuredresistancevaluewillbedisplayedonthedisplay.
* The instrument generates a continuous sound signal if
themeasuredresistancevalueislessthanthepre-setted
“GoNoGO”value.Defaultvalueis5W
Refertothe“11.Menustructure”sectionforhowtosetthevalue
forthelimitvalue.
7.2 LowOhmMeasurement(W)
* Verifythatdeviceundertestiselectricallydead.
* Setthefunctionselectorswitchto“W”position.
* Ifrequirednullleadresistancebydoingzeroadjustmentbelow.
* Connectthedeviceundertestasshownindiagram
* Themeasuredresistancevaluewillbedisplayedonthedisplay.
* The instrument generates a continuous sound signal if the
measuredresistancevalueislessthanthepre-setted
“GoNoGO” value.Defaultvalueis0W
Refer to the “11. Menu structure” section for how to set the
valueforthelimitvalue.
7.3 Zeroadjustment
When measuring very low resistance values, you can eliminate
resistanceofleadsandcontactresistancebyzeroadjustment.
* Connect the test leads to instrument and connect the free ends of
theleadstooneanother.
* Waitforreadingtostabilize.
* Briefly press the “TEST” push button. The instrument
acknowledges zero adjustment by a sound signal and
the LCD shows “0.00” (+1digit) and the symbol “ZERO” (17) is
displayed. The resistance measured at the instant the push button
is pressed is used as reference value (0-9.99W). It is automatically
deductedfromthevaluesmeasuredthereafter.
* Zeroadjustmentcanbecleared.
Ø Bypressingthe“TEST”buttononceagain.
ØByswitchingtheinstrumentoff.
AlsozeroadjustmentisclearedwhenAutoshutOFFoperates.
AdditionalfeaturesavailableinOhmandContinuityranges.
1. Pre-selectablelimitchecks(Go/Nogooption)
2. Storingthemeasuredresistancevalue.
3. MinimumandMaximumvalue“MIN/MAX”storage.
11
RISH
Insu20
400
100kM 600
10kM
200
1MG
800
1kM
100k
1000
10MG
8
L
L
RISH
Insu20
400
100kM 600
10kM
200
1MG
800
1kM
100k
1000
10MG
8
L
L
CE
CE

12
8.VoltageMeasurement(VAC/DC)
You can use this instrument for voltage measurement from 25V to600V
ACorDC.
* Keepthefunctionselectorswitchto“V ”positionand
applytheinput.
* Theinstrumentautomaticallydisplayswhethertheinput
isACorDCalongwiththevalueoftheexternalvoltage.
* Ifthevalueofvoltageis<25V(ACorDC),thedisplay
“025V”flashes.Ifthevalueof thevoltageis>600V(AC
orDC),acontinuosacousticsignalisgenerated.
* WhenreverseDCisapplieditdisplays“-dc”
AdditionalfeaturesavailableinInsulationmeasurementranges.
1. MiniumandMaximumvalue“MIN/MAX”storage
Voltage Measurement
9. DefaultLiveCircuitWarning
When more than 25V (AC or DC) is applied to the terminals the
instrument defaults to voltmeter on all switch positions. In addition the
beeperwillsoundonallswitchposition(excepttheV position).
When external voltage>25 V is present, Insulation Ohm and Continuity
Testsareinhibited.
When external voltage <25 V is present, only the Insulation Tests are
available!!
Note:Thedefaultlivecircuitwarningdoesnotoperate
(i)wheninSTOPWATCHmode.
(ii)wheninMENUstructure.
+(-)/~
-(+)/~
2 92
MW
50V
100V
250V
500V1KV
V
10 StopWatch
Thisfunctionallowsyoutomeasureelapsedtimeuptoonehour.
Tousestop-watch
1.SetthefunctionselectorswitchtoContinuity" "position.
2.Press“TEST”key.Thiswillshowthestopwatchfrom00:00
3.Press keyatanytimetoRESETthestopwatch.
4.Press keyatanytimetoSTART orSTOPthestopwatch.
5.Toquitstopwatchandgobacktobasicdisplay,Press“TEST”key.
11 MenuStructure
StructureofcommandmenuforInsulationmeasurement:
Structure of command menu for Resistance & Continuity measurement
Basic
Display
nnnn
mm:ss
xxxx
OFF/ON
RISH
Insu20
400
100kM 600
10kM
200
1MG
800
1kM
100k
1000
10MG
8
L
L
VAC VDC
Basic
Display
nnnn
xxxx
xxxx
OFF/ON
Basic
Display
OFF/ON
13
Structure of command menu for Voltate measurement
= Backlit off
= Backlit on
= Time
= Recall
= Minutes
= Seconds
= Value for noGo
= Strored value
mm
ss
nnnn
xxxx
OFF
ON
Esc
= Confirmation of
selected menu item
via ENTER key
= Select next menu item
in the flow direction
= Go back to previous
level of menu
CE

14
11.1DescriptionofMenuItems:
11.1.1SettingBacklitOFF/ON: (atallranges)
Basic OFF/ON
Display
1.NowscrollthroughON/OFFvaluesusingthe keyuntil
youcomeacrossrequiredvalue.
2.Tosavethisvaluepress key.Aftersaving,thiswilltake
youtobasicdisplay.
3.Togobackwithoutsavingpress"ESC" key.Thiswilltakeyouto
mainmenu" "
4.Togobacktobasicdisplayfrommainmenu" " use"ESC"key.
11.1.2SettingtheGoNoGovalues:
(atallrangesexceptV )
OperatingOrder:
Basic noGo nnnn
Display
Setting: nnnn=noGolimitvaluesinMW/GW(forInsulationranges)
nnnn=noGolimitvaluesinW(forOhmandContinuity)
1.Nowscrollthroughdifferentvaluesusingthe keyuntil
youcomeacrossrequiredvalue.
2.Tosavethisvaluepress key.Aftersaving,thiswilltake
youtobasicdisplay.
3.Togobackwithoutsavingpress"ESC" key.Thiswilltakeyouto
mainmenu"noGo"
4.Togobacktobasicdisplayfrommainmenu"noGo"use"ESC" key.
11.1.3SettingtheTimervalues
(OnlyatInsulationResistancemeasurementranges)
OperatingOrder:
Basic noGo ti mm:ss
display
Setting: mm=minutes, ss=seconds
1. Nowscrollthroughdifferentvaluesusingthe keyuntil
youcomeacrossrequiredvalue.
2. Tosavethisvaluepress key.aftersaving,thiswilltake
youtobasicdisplay.
3. Togobackwithoutsavingpress"ESC"key.Thiswilltakeyouto
mainmenu"ti"
4. Togobacktobasicdisplayfrommainmenu"ti"use"ESC"key.
11.1.4Toseethestoredvalues
(atallrangesexceptV )
OperatingOrder:
Basic noGo ti reCL nnnn
display
1. Youcannowseethefirststoredvalue.Thescalepointerisused
toindicatethenumber(outof10)ofthedisplayedstoredvalue.
2. Nowyoucanviewthesuccessivestoredvalueusingthe
key.Thescalepointeralsoincrementstoindicatethenumber
(outof10)ofthedisplayedstoredvalue.
3. Togobackpress"ESC"key.Thiswilltakeyouto
mainmenu"reCL"
4 Togobacktobasicdisplayfrommainmenu"reCL"use"ESC"key.
NoGo ti reCI
50V
100V
250V
500V
1000V
Ohm
Cont
V
1 MW
1 MW
5 MW
5 MW
500 MW
00 W
5 W
---
00:00 (Manual)
---
---
---
0.00
0.00
0.00
0.00
0.00
0.00
0.00
---
12.Specifications
Meas. Range Resolu- Accuracy Overload
function tion (..%ofrdg valueand
+..digit) duration
±
15
1) Forinsulationresistancerange, 2) ForlowOhms/continuityRanges
* Terminalvoltageonopen * Opencircuitvoltage
circuit(DC) 5V ±1Vd.c.
-0%to+30%ofratedvoltage * Leadresistance
* Shortcircuitcurrent<2mA compensation:0-9.99W
* Testcurrentonload1mAat
minimumpassvaluesof
insulationasspecified
inVDE0413Part1.
ReferenceConditions: 0
AmbientTemp. +23 C±2K
RelativeHumidity 45%...55%
BatteryVoltage 8V±0.1V
VoltageMeasurement AC(Sine),50/60Hz
Insulation
Resistance
1)
MW
U = 50 V,
N
100 V
0.01 MWto
0.99 MW
³1.0 MWto
9.9 MW
³10 MWto
999 MW
10 KW
( )
0.01 MW
100 KWto
( )
0.1MW
1 MW
± 3% ± 2
± 5% ± 2
± 30%
1200V rms,
10 sec
Insulation
Resistance
MW/G
500 V,
1000 V
1)
W
U = 250 V,
N
0.01 MWto
9.99 MW
³10.0 MWto
99.9 MW
³100 MWto
999 MW
10 KW
( )
0.01 MW
100 KWto
( )
0.1MW
1 MW
± 5% ± 2
± 5% ± 2
± 30%
1200V rms,
10 sec
2)
Low Ohms
W
0-9.99 W
³10W-99.9W
0.01 Wat
210 mA
0.1 W at
21 mA
± 3% ± 2
± 5% ± 2
1200V rms,
10 sec
2)
Continuity 0-9.99 W
³10W-99.9W
0.01 Wat
210 mA
0.1 W at
21 mA
± 3% ± 2
± 5% ± 2
1200V rms,
10 sec
V AC/DC 1 V ± 2% ± 3 1200V rms,
10 sec
³25V-450V
³450V-600V
1 V ± 3
Table for default values :
³1 GWto
2 GW10 MW± 30%

16
0.01MW0.1MW0.5MW1.0MW10MW100MW
Insulation Resistance
Output Voltage ( V )
1300
1200
1100
1000
900
800
700
600
500
400
300
200
100
1000 V Range
500 V Range
250 V Range
0.1MW0.05MW1MW10MW
Insulation Resistance
Output Voltage ( V )
130
120
110
100
90
80
70
60
50
40
30
20
10
100 V Range
50 V Range
10MW
Required min. Value of Insulation Resistance
(Limiting Value)
Minimum Display Value
5MW
2MW
1MW
0.5MW
0.2MW
0.2MW0.5MW1MW2MW5MW
17
PowerSupply
Battery 6x1.5VcellsIECLR6
(Nickelcadmiumrechargeablecells
maybeused)
ServiceLife WithoutBacklitON:Typically2500x5sec.
Operationwithalkalinemanganesecells.
WithBacklitON:Typically1250x5seconds
Operationwithalkalinemanganesecells.
BatteryTest
Batterysymbol“ ”appearsontheLCDineitherofbelowconditions:
1. Batteryvoltageis<5.4V.Batteryreplacementisnecessary.
2. Foraspecificrangedependingonapplication
Ineither ofthe abovecondition,when thebattery symbolisdisplayed,no
measurementcanbedone,andthebatterysymbolflashes.
Note:Batterycellsshouldnotbeleftintheinstrumentwhichmayremain
unusedforextendedperiodoftime.
MainsAdapter(Optional)
Input : 100-240VAC,50/60Hz
Output : 12VDC,1000mA
Polarity :
AutoturnOFF
Instrument turns off automatically, if any of the keys or the function
selector switch have not been activated for about 10 minutes in
insulationresistancerangeand5minutesinotherranges.
CableSet 2
Cucross-sectionalarea1mm
Endmetallicpartsshouldpassterminal 4D11
TechnicaldataasperIEC61010/EN61010-2-031CATIII/1000V,
Air/creepage³16mm.
HighVoltagetest7.4kVACfor1min.
Breakingforceofthecable³36N,1min.
Fuse 500mA(F)/440VH.B.C.10kAmin(32mmx6mm)
ElectricalSafety
ProtectionClass IIperIEC61010-1/EN61010-1/VDE0411-1
Overvoltage
Category II III
NominalVoltage 600V 300V
Contaminationdegree 2 2
TestVoltage 3.7KV~perIEC61010-1/EN61010-1
EMC
IEC/EN61326 ElectromagneticCompatibility(EMC)
EnvironmentalConditions 0
NominalTemp.Range 0.....35 C
0 0
OperatingTemp. -20 C...+40 C(fullrange)
0 0
-20 C...+60 C(upto100MW)
0 0
StorageTemp -25 C...+65 C0
RelativeHumidity 90%RHat40 Cmax.
0
TemperatureCoefficient <0.1%per C
Display
LCD display field (65 mm x 30 mm) with analog indication and Digital
displayandwithdisplayofunitofmeasuredquantityandfunctions
Analog
Display LogarithmicScale
Digital
Display/Char.Height 7segmentdigits/12mm
NumberofDigits 3digitforMWandW
4digitforStopwatch
OverrangeIndication OL
Jack

13.3 Fuses
Application of high energy pulse during Insulation resistance
measurement or application of voltage during Continuity and Ohm
measurementwill blow the fuse. If the fuse blows, eliminate the causeof
theoverloadbeforeplacingtheinstrumentbackintooperation!
FuseTesting
* Setthefunctionselectorswitchtoanyinsulationmeasurementrange.
* Opencircuitthetestleads.
* Pressthe“Test”buttonuntil“OL”or“FUSE”appearsondisplay.
Ifdisplay=“OL”:FuseisOK.
Ifdisplay=“FUSE”:Fuseisblownandneedsreplacement.
Note : Once “FUSE” is displayed all the functions are inhibited.
ImmediatelyswitchtheInstrumentOFFandreplacethefuse.
ReplacementofFuses
* Opentheinstrumentasdescribedunderbatteryreplacement.
* Removethedefectivefuseandreplaceitwithanewfuse.
* Makesurethatthenewfusemakesgoodcontact.
Specification:
CurrentRating:500mA/500V-6.3mmx32mmwithabreakingcapacity
of10kA.
Recommendedfusemanufacturer:Siba
PartNo.:70-065-63/0.5A
Beabsolutely surethat only thespecified fusesareused !
The use of a fuse with different triggering characteristics,
adifferentnominalcurrentoradifferentbreakingcapacity
exposes the operator, the system and the measuring
instrument to danger. The use of repaired fuses or short-
circuitingofthefuseholderisprohibited.
13.4 Housing
No special maintenance is required for the housing. Excessive
contamination has an adverse effect on isolation and reduces input
resistance.Thesurfacemust be keptcleanfor this reason. Useaslightly
dampened cloth for cleaning. Oil free petrol is recommended for
cleaning.Avoid theusecleansers,abrasivesorsolvents.
19
MechanicalDesign
Protection Instrument:IP50
Forterminalsocket:IP20accordingto
DINVDE0470Part1/EN60529
Dimensions WxHxD
84mmx195mmx35mm
Weight 500g includingbattery
13 Maintenance
Attention!
Disconnect the instrument from the measuring circuit before
openingtheinstrumenttoreplacethebatteryorthefuse!
13.1 Battery
BeforeinitialStart-up,orafterlongstorageofyourinstrument,makesure
that no leakage has occurred at the instrument battery. Repeat this
inspectionatregularintervals.
Batterysymbol“ ”appearsontheLCDineitherofbelowconditions:
1. Batteryvoltageis<5.4V.Batteryreplacementisnecessary.
2. Foraspecificrangedependingonapplication
In either of above condition when the battery symbol is displayed, no
measurementcanbedone,andthebatterysymbolflashes.
ReplacingtheBatteries
* Lay the instrument on to its face, loosen the two screws at the rare
panel and lift out the housing base, starting at the bottom (a). The
housingbase and the housing top areheld together with snap hooks
atthetopofthefrontpanel.
* Remove all the cells from the battery holder and replace them with
newones.
* Important for reassembly : first lay the housing base into place such
as shown in diagram below. Push the housing base and housing
toptogether,firstatthebottomfront(a)andthenatthetopfront(b).
* Retightenthehousingbasewiththetwoscrews.
* Pleasedisposeoftheusedcellsproperly.
13.2 OperationwithMainsAdapter(Optional)
Use only Mains Adapter (Output : 12VDC, 1000mA) to power ON the
instrument.
During power supply via mains adapter in inserted batteries are cut off
electronicallyandthuscanbeleftinsidetheinstrument.
Note:ThereisnoneedofinternalbatterieswhenAC-DCAdapter
isconnectedtotheinstruments.
Ifre-chargeablebatteriesareusedtheymustbechargedoutside
theinstrument.
18
(b) (a)
Table of contents
Other Rish Test Equipment manuals
Popular Test Equipment manuals by other brands

Humboldt
Humboldt HM-5030 product manual

Myenergi
Myenergi EVSE Test Box Operation manual

JDS Uniphase
JDS Uniphase ANT-5 Result Documentation, Export and Printing Guide

Agilent Technologies
Agilent Technologies 16452A Operation and service manual

Hamron
Hamron 604-175 User instructions

LifeSafer
LifeSafer L 250 user guide