HS-04 Operator’s Manual Chapter 3: Measuring Mode • 17
3. If CAL.RMDR is set to ON (factory setting) and a calibration is due, the screen that appears
next depends on how CAL.EXPD is set in User Mode (see page 55). The three possible
screens are described below. If a calibration is not due, the instrument displays the number of
days left until a calibration is due.
CAL.EXPD set to
CONFIRM (factory setting)
CAL.EXPD set to
CANT.USE
CAL.EXPD set to
NONE
LCD
Sound Buzzer sounds double pulsing
tone
Buzzer sounds double pulsing
tone
None
Action • Option A, Perform
calibration: Press and
release POWER MODE to
enter User Mode and
perform a calibration. The
instrument takes you
straight to the calibration
start screen in User Mode’s
GAS CAL\A-CAL(E-
CAL) item (if Password
Protection is set to On
using the 04 Series Setup
Program, you must enter a
password first). See page 43
for calibration instructions.
If the calibration is
successful, the screen above
will not appear again until
the unit is due for
calibration. If the calibration
is not successful, the screen
above will again appear in
the startup sequence.
• Option B, Bypass message:
To continue without
performing a calibration,
press and release AIR.
The HS-04 cannot be used
until a successful calibration is
performed. Press and release
POWER MODE to enter User
Mode and perform a calibra-
tion. The instrument takes you
straight to the calibration start
screen in User Mode’s GAS
CAL/A-CAL (E-CAL) item
(if Password Protection is set
to On using the 04 Series
Setup Program, you must
enter a password first). If you
don’t press POWER MODE,
the instrument automatically
goes to the calibration start
screen after 6 seconds (if
Password Protection is set to
On using the 04 Series Setup
Program, you must enter a
password first). See page 43
for calibration instructions.
If the calibration is successful,
the screen above will not
appear again until the unit is
due for calibration. If the cali-
bration is not successful, the
screen above will again appear
in the startup sequence.
• Option A, Perform
calibration: If you want to
enter User Mode and
perform a calibration, press
and release POWER
MODE. The instrument
takes you straight to the
calibration start screen in
User Mode’s GAS CAL/A-
CAL (E-CAL) item (if
Password Protection is set
to On using the 04 Series
Setup Program, you must
enter a password first).
• Option B, Bypass message:
To continue without
performing a calibration,
wait a few seconds for the
instrument to continue with
its startup sequence.