ROLA-CHEM GLOBE FLOWCELL User manual

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 1 of 8
GLOBE FLOWCELL
Operating Manual

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 2 of 8
Table of Contents
Section title Page
1. Safety Instructions 2
2. Description 2
3. Installation 2
4. Use of Flowcell 5
5. Maintenance 5
6. Troubleshooting 6
7. Parts List 7
8. Warranty and Service 8
1. SAFETY
See controller manual for additional safety instructions.
1.1 WARNING! Do not shut down circulating and control system immediately after using
pool, spa or hot tub. Systems must operate until water chemistry is stable. Systems must
be restarted before water chemistry is out of NSPI recommended ranges or water
chemistry must be manually corrected.
1.2 WARNING! Removing power from circulating pump must also remove power from the
controller. Provide a properly located outlet controlled by the circulating pump circuit.
1.3 WARNING! Use a device such as a flow switch to remove power from the controller in
the event of pool pump or circulation failure. It is not safe to automatically dispensed
chemicals into a circulating system that is not running.
2. Description
The Flowcell is designed to be used with a controller as part of a chemical automation
system, which continuously monitors and maintains pH balance and sanitizer level in a
swimming pool or spa.
3. Installation
3.1 Mount Flowcell
Pre-mark hole locations for Flowcell mounting screws using the Flowcell. The Flowcell
should not be above the controller unit or other equipment or electrical devices. Flowcell
holes will accept #10 screws.

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 3 of 8
FILTER & SHUT-OFF VALVE
(OPTIONAL)
RECOMMENDED GROUNDING STUD
BOTTOM VIEW
ORP PROBE
pH PROBE
PROBE FITTING
FLOW SWITCH LEADS
FLOAT
(INSIDE CELL)
SHUT-OFF VALVE
3.2 Install pH and ORP Probes
Special gold tipped probes are recommended for application with Salt Chlorine
Generators. Note: standard probes have platinum tips.
Probe ends should not be allowed to dry out. They are packaged within a jar to keep the
ends wetted during shipment and storage. Remove the probes from wetting jars by
unscrewing the jar cap. Remove the jar cap and grommet from the probe. (Save for
future probe storage.) Close left and right Flowcell valves, and open the center Flowcell
valve.
Remove nuts from probe fittings and slide onto probes, noting direction of nut thread.
Insert the pH probe into the right probe receptacle. Probe should slide freely through the
fitting and into the Flowcell until it bottoms out inside the Flowcell. Then raise the probe
about 1/8", hold in position and finger-tighten the nut onto the fitting. Use a wrench to
tighten the cap about another 1/4 turn.
DO NOT OVER TIGHTEN.
Repeat the procedure for the ORP probe in the left probe fitting.
3.3 Install Flowcell Connections to the Pool Water System
Turn off the pool's pump system during this procedure.

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 4 of 8
Check that sufficient tubing (supplied) is available for connecting Flowcell prior to drilling
holes. Additional tubing may be obtained –see parts list for part numbers.
Do not exceed a total of 40' of tubing.
Select installation points in the Pool Circulation System for the fitting that will provide flow
into the Flowcell. The inlet to the Flowcell should be at a higher pressure than outlet of the
Flowcell to provide flow though the Flowcell. Assuming the inlet connector will be located
upstream of the pool filter and the outlet connector will be located downstream from the
pool filter, proceed as follow:
Cut a length of black plastic tubing long enough to reach from the left (inlet)
Flowcell Valve to the installation point on the upstream side of the pool filter.
Install this piece of tubing into the left Valve of Flowcell. (Turn this Valve 'off.')
Cut a length of black plastic tubing long enough to reach from the right (outlet)
Flowcell fitting to the installation point on downstream side of the pool filter.
Install this piece of tubing into the right Valve of Flowcell. (Turn this Valve 'off.')
A length of tube can also be added to the center Flowcell Valve for convenience of water
sampling.
Drill and tap a 1/4" NPT thread into a straight section of water pipe at both installation
points.
Screw in 1/4" MNPT x TUBE fitting and tighten.
Note: use Teflon tape on pipe threads and DO NOT OVER TIGHTEN.
Connect the tubes coming from the Flowcell Valves into the fittings.
Open both the left and right Valves on the Flowcell (close the center Valve).
Turn on the Pool's pump system and check for water leaks. Fix leaks as required.
If flow is backwards, the Float will not move up. Reverse the tubes if needed.
Connect the two Probe cables to controller. The pH Probe cable connects to the mating
jack on the bottom of the controller labeled "pH." Push the connector onto the jack and
turn clockwise about 1/4 turn. The ORP Probe cable connects to the mating jack on the
bottom of the controller labeled "ORP."
3.4 Connect Flow Switch
Although the Float is an integral part of the flow switch, it is visible through Flowcell wall
and may also be used as a manual flow indicator. On Controllers with a flow switch
feature, connect the two lead wires as specified in the Controller manual.
3.5 Connect Grounding/Bonding Stud
The Flowcell is supplied with a titanium grounding/bonding stud. It is recommended that
this stud be used to ground/bond the Flowcell process water to earth ground.
Grounding/bonding the Flowcell helps to eliminate ground loop problems that could
potentially damage probes and/or cause erroneous readings.

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 5 of 8
4. Use of Flowcell
4.1 Flowcell Inlet and Outlet Valves
4.1.1 Adjusting flow through Flowcell
Turn left Valve to full on position.
Close the right Valve and then open it until the Float rises to the top of its chamber.
The Float will not rise if flow is not sufficient or is backwards.
4.1.2 Removing Probes
The Valves allow the operator to remove Probes without shutting down the pool pump.
Close both left and right Valves before loosening the Probe fittings.
Whenever re-inserting Probes in Flowcell, open center Flowcell Valve before and
during insertion to relieve pressure, then close the center Valve after Probe insertion is
complete.
4.2 Sampling Water Using Sampling Valve
The center Valve on the Flowcell can be used to conveniently take water samples from
the pool water.
NOTE: Globe Flowcell valves should be closed when back flushing filter to prevent
damage to probes.
5. Maintenance
5.1 Cleaning
Periodically (approx. every month) remove Probes from Flowcell and clean them using
cotton swab or cloth. Clean more frequently as required (i.e. if bathers use large amounts
of oils and sun tan lotions in the pools).
Note: When using di-chlor or tri-chlor products clean probes every 2 to 3 weeks. Clean
Probes with dishwashing liquid and cotton swab or cloth, then soak end for three minutes
in a solution of muriatic acid diluted 20:1. Then rinse with water.
WARNING! Always add acid to water. Never add water to acid. (e.g. fill container with
water then add acid) Wear protective clothing, gloves and safety glasses.
5.2 Do not let Probe ends dry out
At year end shutdown, remove Probes from Flowcell and return to wetting jars.
If it is expected that Flowcell may go dry for a long period (several hours), remove them
from the Flowcell and submerse tips in water. Use pH buffer solution in wetting jar to
protect Probes. Drain the Flowcell to prevent freeze damage. Plug Probe fitting
openings in Flowcell to minimize contamination.

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 6 of 8
5.3 Test flow switch monthly
Close left-hand valve on Flowcell and confirm the flow light turns off. Then re-open valve
and confirm that light turns back on. A properly installed and functioning flow switch will
disconnect power to chemical feeders if there is no water flow in the circulation
plumbing.
5.4 Refer to Feeder and Controller manuals for their maintenance
6. Troubleshooting
6.1 If Controller doesn’t appear to be reading correctly:
6.1.1 Inspect Probes. See Cleaning in Maintenance section.
6.1.2 Check the pH Probe with 7.2 pH buffer solution.
The ORP Probe can be checked with a solution of 1 drop 5.25% bleach (i.e. household
bleach) and 1 pint water. This should provide a 1-2 ppm free chlorine solution (check
with test kit) Add one drop of chlorine at a time to adjust test mixture as required. Put
ORP Probe in mixture. The ORP reading should be 600 to 800 mV.
Note: It may take three to five minutes for Probes to respond to solution. If it takes
longer you should replace the Probes.
6.1.3 Adjust flowrate through Flowcell. Incorrect ORP/pH reading can occur if water is
flowing too fast in the Flowcell. Variable speed circulation pumps can cause Float to
indicate incorrectly. See Adjusting flow in Use of Flowcell section.
6.1.4 Verify that Probe ends are in the flow stream of the Flowcell.
6.2 If Controller flow light is not on:
6.2.1 Confirm that adequate flow is going through the Flowcell.
Open center Sampling Valve on the Flowcell. If light turns on, this may indicate a
blockage between right (outlet) Flowcell fitting and the installation point on downstream
side of the pool filter.
6.2.2 Confirm that Flow Switch float is free to move.
6.2.4 A filter on the inlet to the Flowcell can reduce contamination that can cause Flow Switch
float to stick.
NOTE: It is recommended that an inline filter be used to help keep the
Globe Flowcell (including probes) and flow switch free from debris. Debris may cause
obstruction in flow switch and prevent the switch from functioning.

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 7 of 8
7. Parts List
550591
550026
550027
550054
550089
550135
550136
PART NO.
2
4
3.3
ITEM NO.
1
pH PROBE- HEAVY DUTY, 8 FT CABLE
FITTING, 1/4 TUBE X 1/4 NPT
TUBING, PE, 1/4 OD x 40 FT
FILTER KIT, 3/8 TUBE (LARGE FILTER) (OPTION)
VALVE, SHUT-OFF, PVC-1/4 TUBE X 1/4 NPT
PROBE FITTING (qty 2)
ORP PROBE- HEAVY DUTY, 8 FT CABLE
DESCRIPTION
2
1
1 550209
FLOW SWITCH
5501833.2
FLOWCELL ASSEMBLY, 3/8 TUBE (INCLUDES 3.1 THRU 3.6)
5501803
FLOWCELL, CLEAR GLOBE
5501813.1
VALVE, SHUT-OFF, PVC-3/8 TUBE X 1/4 NPT (qty 3 )
71251903.4
3.5 550194 TUBING,PE, 3/8 OD x 40 FT
3.6 550193 FITTING,PP, 3/8 TUBE X 1/4 NPT (qty 2)
ITEMS USED ON FLOWCELLS WITH 1/4 TUBING
3.2
3.1
3.3
3.4
550591 FILTER KIT
(NOT TO SCALE)
3.6
4
550592 FILTER KIT
(NOT TO SCALE)
1 550201 ORP PROBE- GOLD TIP, 8 FT CABLE
ORP PROBE- HEAVY DUTY, 3 FT CABLE
5502082pH PROBE- HEAVY DUTY, 3 FT CABLE
5505924FILTER KIT, 3/8 TUBE (SMALL FILTER) (OPTION)
4

ROLA-CHEM® Flowcell PN 550182 7/11/2019 page: 8 of 8
8. Warranty and Service
Limited Warranty:
ROLA-CHEM® warrants the original purchaser that this unit is free from manufacturing
defects in material and workmanship from the date of the original purchase for one (1)
year.
If this unit fails within the one (1) year period, it will be repaired or replaced (ROLA-
CHEM® option) at no charge, when returned to ROLA-CHEM® with proof of purchase
receipt. This warranty does not apply to any product damage caused by improper use,
accident, misuse, improper line voltage, fire, flood, lightning, earthquake, other acts of
nature, or if product was altered or opened by anyone other than qualified ROLA-CHEM®
personnel.
All expendable items (tubing, etc.) are not covered under this warranty.
Heavy duty probes are warranted from the date of original purchase for one (1) year.
Standard duty probes are warranted from the date of original purchase for 90 days.
ROLA-CHEM®, under no circumstances, shall be liable for any consequential damages
directly or indirectly caused by this unit. Please observe all rules and regulations required
by state and local regulations, building codes, health codes, OSHA, etc.
Service Procedure: For operation assistance in the United States please call:
(800) 549-4473, Monday through Friday 8:00 am-5:00 pm Central Time or email:
[email protected]. Have model number and serial number available.
ROLA-CHEM® has specialized knowledge and equipment to test and diagnose your
product. Contact our service personnel as noted above for assistance. If you are directed
to ship the product to ROLA-CHEM® and given an RGA#, please note RGA# on the
package and ship freight pre-paid to:
ROLA-CHEM®
5858 Centerville Road
St. Paul, MN 55127-6804
www.rola-chem.com
Table of contents
Popular Measuring Instrument manuals by other brands

woodmizer
woodmizer Laser Sight manual

Aqualytic
Aqualytic AL 110 instruction manual

Delta OHM
Delta OHM HD21ABE manual

Kyoritsu Electrical Instruments Works, Ltd.
Kyoritsu Electrical Instruments Works, Ltd. 5201 instruction manual

KraftPowercon
KraftPowercon PRX3 manual

B meters
B meters HYDROCAL-M4 user manual