RST TUF-2000S User manual

Ultrasonic Flow Meter User Manual
Type: TUF-2000S

TUF-2000S User Manual
Contents
1. Introduction....................................................................................................................................4
1.1 Preface
............................................................................................................................. .....4
1.2 Features
................................................................................................................................4
1.3 Flow measurement principle
..............................................................................................4
1.4 Optional transducer
............................................................................................................5
1.5 Typical application
..............................................................................................................5
1.6 Product Identification
.........................................................................................................5
1.7 Specifications
.......................................................................................................................5
2. Installation and Measurement ........................................................................................................7
2.1 Wiring diagram
...................................................................................................................7
2.2 Keypad
.................................................................................................................................7
2.3 Menu Windows
....................................................................................................................8
2.4 Steps to Configure the Parameters
....................................................................................9
2.5 Transducers Mounting Allocation
...................................................................................10
2.6 Transducers Installation
................................................................................................... 11
2.6.1 Wiring diagram of transducer ...................................................................................12
2.6.2 Transducers Spacing .................................................................................................12
2.6.3 V-method Installation ...............................................................................................12
2.6.4 Z-method Installation................................................................................................13
2.6.5 W-method Installation ..............................................................................................13
2.6.6 N-method Installation ...............................................................................................14
2.7 Installation Checkup
.........................................................................................................14
2.7.1 Signal Strength..........................................................................................................14
2.7.2 Signal quality ............................................................................................................15
2.7.3 Time Ratio between the Measured Total Transit Time and the Calculated Time ....15
3. Menu Window Details .................................................................................................................15
3.1 Menu Windows Arrangement
..........................................................................................15
3.2 Menu Window Details
......................................................................................................16
4. How To.........................................................................................................................................26
4.1 How to judge if the instrument works properly
.............................................................26
4.2 How to judge the liquid flowing direction
......................................................................26
4.3 How to change between units systems
.............................................................................26
4.4 How to select a required flow rate unit
...........................................................................27
4.5 How to use the totalizer multiplier
..................................................................................27
4.6 How to open or shut the totalizers
...................................................................................27
2

TUF-2000S User Manual
4.7 How to reset the totalizers
................................................................................................27
4.8 How to restore the flow meter with default setups
........................................................27
4.9 How to use the damper
.....................................................................................................27
4.10 How to use the zero-cutoff function
..............................................................................28
4.11 How to setup a zero point
...............................................................................................28
4.12 How to get a scale factor for calibration
.......................................................................28
4.13 How to use the system locker
.........................................................................................28
4.14 How to use 4-20mA current loop output
.......................................................................28
4.15 How to use the Frequency Output
.................................................................................29
4.16 How to use the Totalizer Pulse Output
..........................................................................30
4.17 How to produce an alarm signal
....................................................................................30
4.18 How to use the built-in Buzzer
.......................................................................................31
4.19 How to use the OCT output
...........................................................................................31
4.20 How to modify the built-in calendar
.............................................................................31
4.21 How to view the Date Totalizers
....................................................................................31
4.22 How to use the Working Timer
......................................................................................32
4.23 How to use the manual totalizer
....................................................................................32
4.24 How to know how long the battery will last
.................................................................32
4.25 How to check the ESN and other minor details
...........................................................32
4.26 How to use the data logger for scheduled output
.........................................................32
4.27 How to output analogue voltage signal
.........................................................................32
4.28 How to adjust the LCD display
.....................................................................................33
4.29 How to use RS232/RS485?
.............................................................................................33
4.30 How to use automatic amending function for offline compensation
..........................33
4.31 How to use batch controller
...........................................................................................33
4.32 How to adjust the analogue output
...............................................................................33
4.33 How to solidify the parameters
......................................................................................34
4.34 How to enter the parameters of user-type-transducer
................................................34
4.35 How to use the circular display function
......................................................................34
4.36 How to enter into the linearity correcting? How to enter into the data?
...................34
4.37 How to save / restore frequently-used pipe parameters
..............................................35
5. Troubleshooting ...........................................................................................................................36
5.1 Power-on Error Displays and Counter-Measures
..........................................................36
5.2 Error Code and Counter-Measures
.................................................................................36
5.3 Other Problems and Solutions
.........................................................................................37
3

TUF-2000S User Manual
1. Introduction
1.1 Preface
Welcome to use new generation transit-time ultrasonic flow meter, please read the user manual
carefully before using. The wall-mount ultrasonic flow meter is designed to be installed in a fixed
location for long-term flow measurement.
1.2 Features
◆
◆
◆
◆
◆
Linearity: 0.5%, Repeatability: 0.2%, Accuracy:±1%
Easy to operate.
Several type transducers for selection, measuring pipe size is from DN15mm to DN6000mm
Adopt low voltage, multi-pulse technology to improve accuracy, useful life and reliability.
Powerful Recording Function, record the totalizer data of the last 64 days/64 monthes/5 years.
1.3 Flow measurement principle
The TUF-2000S ultrasonic flow meter is designed to measure the fluid velocity of liquid within a closed
conduit. The transducers are a non-contacting, clamp-on type, which will provide benefits of
non-fouling operation and easy installation.
The TUF-2000S transit-time flow meter utilizes two transducers that function as both ultrasonic
transmitters and receivers. The transducers are clamped on the outside of a closed pipe at a specific
distance from each other. The transducers can be mounted in V-method where the sound transverses
the pipe twice, or W-method where the sound transverses the pipe four times, or in Z-method where
the transducers are mounted on opposite sides of the pipe and the sound crosses the pipe once. This
selection of the mounting method depends on pipe and liquid characteristics. The flow meter operates
by alternately transmitting and receiving a frequency modulated burst of sound energy between the two
transducers and measuring the transit time that it takes for sound to travel between the two transducers.
The difference in the transit time measured is directly and exactly related to the velocity of the liquid in
the pipe, show as follows:
Downstream transducer
V
MD
T
sin2
T
up
T
down
Tup
Tdown
θ
flow
4
spacing
Upstream transducer

TUF-2000S User Manual
Where
θ is the include angle to the flow direction
M is the travel times of the ultrasonic beam
D is the pipe diameter
Tup is the time for the beam from upstream transducer to the downstream one
Tdown is the time for the beam from downstream transducer to the upstream one
∆T=Tup –Tdown
1.4 Optional transducer
◆
◆
◆
◆
◆
TS-1
clamp-on type transducer, pipe size from DN15-100mm
TM-1
clamp-on type transducer, pipe size from DN50-1000mm
TL-1
clamp-on type transducer, pipe size from DN300-6000mm
HTS-1
clamp-on type transducer, pipe size from DN15-100mm
HTM-1
clamp-on type transducer, pipe size from DN50-1000mm
1.5 Typical application
The wall-mounting flow meter can be applied to a wide range of pipe flow measurements. Applicable
liquids include pure liquids as well as liquid with small quantity of tiny particles.
Examples are:
★
Water (hot water, chilled water, city water, sea water, waste water, etc.);
★
Sewage with small particle content;
★
Oil (crude oil, lubricating oil, diesel oil, fuel oil, etc.);
★
Chemicals (alcohol, acids, etc.);
★
Plant effluent;
★
Beverage, liquid food;
★
Ultra-pure liquids;
★
Solvents and other liquids
1.6 Product Identification
Each set of the flow meter has a unique product identification number or ESN (electronic serial
number) written into the software that can only be modified with a special tool by the
manufacturer. In case of any hardware failure, please provide this number which is located on menu
window
M61
when contacting the manufacturer.
1.7 Specifications
Items
Specifications
Main unit Accuracy
Repeatability
Principle
Better than ±1%
Better than 0.2%
Transit-time measuring principle
5

Measurement
Period
500ms
Display
LCD with backlight, display accumulated flow/heat, instantaneous
flow/heat, velocity, time etc.
Output
Analogue output: 4-20mA or 0-20mA current output. Impedance 0
∼
1k
. Accuracy 0.1%.
OCT output: Frequency signal (1~9999HZ)
Relay output: over 20 source signal (no signal, reverse flow etc.)
RS485 serial port
Input
Three analogue input
Three-wire PT100 resistor input (optional)
Other functions
Automatically record the totaliser data of the last 64 days / 64
months / 5 years;
The power-on time and corresponding flow rate of the last 64 power
on and off events. Allow manual or automatic flow loss
compensation
The instrument working status of the last 64 days
pipe
material
Steel, stainless steel, cast iron, cement pipe, copper, PVC,
aluminum, FRP etc. Liner is allowed
Size
15-6000mm
Straight pipe
section
In the upstream it must be beyond 10D, in the downstream it must
be beyond 5D, in the upstream the length must be beyond 30D from
the access of the pump. (D stands for pipe diameter)
Liquid
Types
Water, sea water, industrial sewage, acid & alkali liquid, alcohol,
beer, all kinds of oils which can transmit ultrasonic single uniform
liquid
Temperature
Standard: -30˚C - 90˚C
,
High-temperature
:
-30˚C - 160˚C
Turbidity
Less than 10000ppm, with a little bubble
Flow Direction
Bi-directional measuring, net flow/heat measuring
Environment
Temperature
Main Unit: -30˚C - 80˚C
Transducer: -40
℃
-110
℃
, Temperature transducer: select on
enquiry
Humidity
Main Unit: 85% RH
Transducer: water-immersible, water depth less than 3m
Cable
Twisted Pair Line, standard length of 20m, can be extended to 500m (not
recommended); Contact the manufacturer for longer cable requirement.
RS-485 interface, transmission distance up to 1000m
Power Supply
AC220V or DC24V
Power
Consumption
Less than 1.5W
Protocols
MODBUS, M-BUS, Fuji extended protocol and other factory protocol
TUF-2000S User Manual
6

TUF-2000S User Manual
2. Installation and Measurement
2.1 Wiring diagram
2.2 Keypad
The keypad for the operation of the flow meter is as
shown by the right picture.
Keys 0 - 9 and . are keys to enter numbers
Key ▲/+ is the going UP key, when the user wants to go to the
upper menu window. It also works as the “+” key when entering
numbers
Key ▼/- is the going DOWN key, when the user wants to go
7
4
1
0
8
5
2
·
9
6
3
MENU
/+
/
-
ENT
down-sided menu window. It also works as the “–” key when
entering numbers.
Key ◄ is backspace key, when the user wants go left or wants backspace the left character that is
located to the left of the cursor.
7

TUF-2000S User Manual
Key ENT is the ENTER key for any inputting or selections.
Key MENU is the key for the direct menu window jump over. Whenever the user wants to proceed to a
certain menu window, the user can press this key followed by 2-digit numbers.
The MENU key is shortened as the ‘M’ key afterward when referring to the menu windows.
The
The
ON key is for the power on.
OFF key is for the power off.
2.3 Menu Windows
The user interface of this flow meter comprises about 100 different menu windows that are numbered
by M00, M01, M02 … M99.
There are 2 methods to enter certain menu window:
1. Direct going/entering. The user can press the MENU key followed by two-digit number keys. For
example, the menu window M11 is for the entering of pipe outer diameter. The display will go to the
M11 menu window after the user presses MENU 1 1 .
2. Pressing ▲/+ and ▼/- keys. Each time of the ▲/+ key pressing will proceed to the
lower-numbered menu window. For example, if the current window is on M12, the display will go to the
number M11 window after pressing the ▲/+ key.
There are three different types of menu windows:
1. Menu windows for number entering, like M11 for the entering of pipe outer diameter.
2. Menu windows for option selection/selecting options, like M14 for the selection of pipe materials.
3. Displaying windows only, like M00 to display Velocity, Flow Rate etc.
For number entering windows, the user can directly press the starting digit key when the user is going
to modify the value. For example, when the current window is on M11, and the user is going to enter
219.2345 as the pip e outer diameter, the user can get the numbers entered by pressing the following
serial keys: 2 1 9 . 2 3 4 5 ENT.
For the option selection windows, the user should first press the ENT key to a selection modification
mode and then select the relevant options by pressing the ▲/+ and ▼/- keys or the digit keys to select
the option with a number antecedent to the option. In the end, the ENT key must be pressed to make
the selection. For example, with menu window M14 for the selection of pipe material selection, (the
MENU 1 4 should be pressed first to enter this menu window if the current menu window is on a
different window. The pipe material is stainless steel which has a number “1” antecedent to “stainless
steel” on the display, the user should first press the ENT key to enter into a selection modification mode,
then either make the selection by pressing the ▲/+ and ▼/- keys to make the cursor on the line that
displays “1. Stainless Steel”, or make the selection by pressing the 1 key directly.
Generally, the ENT key must be pressed to enter a modification mode. If the “Locked M47 Open’
message is indicated on the lowest line of the LCD display, it means the modification operations is
locked out. In such cases, the user should go to M47 to have the instrument unlocked first before any
8

TUF-2000S User Manual
further modification can be made.
2.4 Steps to Configure the Parameters
The following parameters need to be configured for a proper measurement:
(1) Pipe outer diameter
(2) Pipe wall thickness
(3) Pipe materials (for non-standard pipe materials*, the sound speed for the material must be
configured too)
*Standard pipe materials and standard liquids refer to those with the sound parameters that have
already been programmed into software of the flow meter, therefore there is no need to configure
them
(4) Liner material and its sound speed and thickness, if there is any liner.
(5) Liquid type (for non-standard liquids, the sound speed of the liquid is also needed)
(6) Transducer type adapted to the flow meter. Generally the Standard M1 clamp-on transducers will
be the selected option.
(7) Transducer mounting methods (the V-method or Z-method is the common option)
(8) Check up the Space displayed on M25 and install the transducers accordingly.
(9) Store the parameter setup
For standard pipe materials and standard liquids, the following detailed step-by-step setup is
recommended.
(1) Press keys MENU 1 1 to enter M11 window to input the digits for the pipe outer diameter,
and then press ENT key.
(2) Press key ▼/- to enter M12 window to input the digits for the pipe outer diameter and then press
ENT key.
(3) Press key ▼/- to enter M14 window, and press ENT key to enter the option selection mode. Use
keys ▲/+ and ▼/- to scroll up and down to the intended pipe material, and then press ENT key.
(4) Press key ▼/- to enter M16 window, press ENT key to enter the option selection mode, use
keys ▲/+ and ▼/- to scroll up and down to the liner material, and then press ENT key. Select “No
Liner”, if there is no liner.
(5) Press key ▼/- to enter M18 window, press ENT key to enter the liner thickness and then press
ENT key (if there is liner)
(6) Press key ▼/- to enter M20 window, press ENT key to enter the option selection mode, use keys
▲/+ and ▼/- to scroll up and down to the proper liquid, and then press ENT key.
(7) Press key ▼/- to enter M23 window, press ENT key to enter the option selection mode, use keys
▲/+ and ▼/- to scroll up and down to the proper transducer type, and then press ENT key.
(8) Press key ▼/ - to enter M24 window, press ENT key to enter the option selection mode, use keys
▲/+ and ▼/- to scroll up and down to the proper transducer mounting method, and then press
ENT key.
(9) Press key ▼/- to enter M25 window and check up the installation space.
(10) Press MENU 2 6 to store the parameter setup (refer to M26)
(11) Press MENU 9 0 to check up signal strength and quality, the bigger of the value the better.
Generally the signal strength should be better than 60.0, and signal quality should be better than 50.0.
9

a n d
T r a n s d u c e r P o s i t i o n
D i m e n s i o n
D i m e n s i o n
L up
x D iam eters
L dn
x D iam eters
L up L dn
10D
5D
L up L dn
10D
5D
L up L dn
10D
5D
L up L dn
L up L dn
L up L dn
12D
20D
20D
5D
5D
5D
L up L dn
30D
5D
TUF-2000S User Manual
(12) Press MENU
(13) Press MENU
(14) Press MENU
9
0
0
1
8
1
to check up time ratio, the ratio value should be in the range of 100
±
3%
to check up the working status, “R” means work well
to check up the measuring data.
Note: 1. For heat measurement, please connect PT100 which installed in water supply and water back
pipe to T1, TX1, T2, TX2, GND terminal.
2. After setting parameter, remember to store parameter in MENU 26, to avoid parameter lose
after turn off.
2.5 Transducers Mounting Allocation
The first step in the installation process is the selection of an optimum location in order to obtain a more
accurate measurement. For this to be completed effectively, a basic knowledge about the piping and its
plumbing system would be advisable.
An optimum location would be defined as a straight pipe length full of liquid that is to be measured. The
piping can be in vertical or horizontal position. The following table shows
P i p i n g
C o n f i g u r a t i o n
U p s t r e a m
D o w n s t r e a m
10

TUF-2000S User Manual
Examples of optimum locations.
Principles to selection of an optimum location
(1) Install the transducers on a longer length of the straight pipe. The longer the better, and make sure
that the pipe is completely full of liquid.
(2) Make sure that the temperature on the location does not exceed the range for the transducers.
Generally speaking, the closer to the room temperature, the better.
(3) Take the pipe fouling into consideration. Select a straight length of a relatively newer pipe. If the
condition is not satisfying, consider the fouling thickness as part of the liner for a better result.
(4) Some pipes have a kind of plastic liner, and between the outer pipe and the liner there may be a
certain thickness difference that will prevent the ultrasonic waves from direct traveling. Such conditions
will make the measurement very difficult. Whenever possible, try to avoid this kind of pipes. If impossible,
try our plug-in transducers that are installed permanently on the pipe by drilling holes on the pipe while
liquid is running inside.
2.6 Transducers Installation
The transducers used by the TUF-2000 series ultrasonic flow meter are made of piezoelectric crystals
both for transmitting and receiving ultrasonic signals through the wall of liquid piping system. The
measurement is realized by measuring the traveling time difference of the ultrasonic signals. Since the
difference is very small, the spacing and the alignment of the transducers are critical factors to the
accuracy of the measurement and the performance of the system. Meticulous care should be taken for
the installation of the transducers.
Steps to the installation of the transducers
(1) Locate an optimum position where the straight pipe length is sufficient, and where pipes are in a
favorable condition, e.g., newer pipes with no rust and ease of operation.
(2) Clean any dust and rust. For a better result, polishing the pipe with a sander is strongly
recommended.
(3) Apply adequate coupler to the spot where the transducers are to be installed and leave no gap
between the pipe surface and the transducers.
Extra care should be taken to avoid any sand or dust particles left between the pipe outer surface and
the transducers.
To avoid gas bubbles inside the upper part of the pipe, the transducers should be installed horizontally
by the side of the pipe.
11

TUF-2000S User Manual
2.6.1 Wiring diagram of transducer
2.6.2 Transducers Spacing
The spacing value shown on menu window M25 refers to the distance of inner spacing between the two
transducers. The actual transducers spacing should be as close as possible to the spacing value.
2.6.3 V-method Installation
V-method installation is the most widely mode for daily measurement with pipe inner diameters ranging
from 15 mm to 200 mm. It is also called reflective mode.
12

TUF-2000S User Manual
2.6.4 Z-method Installation
Z-method is commonly used when the pipe diameter is above 200mm.
2.6.5 W-method Installation
W-method is usually used on pipes with a diameter from 15mm to 50mm.
13

TUF-2000S User Manual
2.6.6 N-method Installation
Rarely used method.
2.7 Installation Checkup
Through the checkup of the installation, one can: check the receiving signal strength, the signal quality
Q value, the traveling time difference of the signals, the estimated liquid speed, the measured traveling
time of the signals and the calculated traveling time ratio. Therefore, optimum measurement result and
longer running time of the instrument can be achieved.
2.7.1 Signal Strength
Signal strength indicates the amplitude of receiving ultrasonic signals by a 3-digit number. [00.0] means
there is no signal detected and [99.9] refers to the maximum signal strength that can be received.
Although the instrument works well if the signal strength ranges from 50.0 to 99.9, stronger signal
strength should be pursued, because a stronger signal means a better result. The following methods
are recommended to obtain stronger signals:
(1) Relocate a more favorable location, if the current location is not good enough for a stable and
reliable flow reading, or if the signal strength is lower than 60.0.
(2) Try to polish the outer surface of the pipe, and apply more coupler to increase the signal strength.
(3) Adjust the transducers both vertically and horizontally while checking the varying signal strength,
stop at the highest position, and then check the transducers spacing to make sure the transducers
spacing is the same as what the M25 shows.
14

TUF-2000S User Manual
2.7.2 Signal quality
Signal quality is indicated as the Q value in the instrument. A higher Q value would mean a higher
Signal and Noise Ratio (short for SNR), and accordingly a higher degree of accuracy would be
achieved. Under normal pipe condition, the Q value is in the range 60.0-90.0, the higher the better.
Causes for a lower Q value could be:
(1) Interference of other instruments and devices such as a powerful transverter working nearby. Try to
relocate the flow meter to a new place where the interference can be reduced.
(2) Bad sonic coupling for the transducers with the pipe. Try to apply more coupler or clean the surface,
etc.
(3) Pipes are difficult to be measured. Relocation is recommended.
2.7.3 Time Ratio between the Measured Total Transit Time and the Calculated
Time
This ratio would be used to check the transducer installation. If the pipe parameters are entered
correctly and the transducers are installed properly, the value for this ratio should be in the range of
100±3. If this range is exceeded, the user should check:
(1) If the pipe parameters are correctly entered.
(2) If the actual spacing of the transducers is right and the same as what the window M25 shows.
(3) If the transducers are installed properly in the right directions.
(4) If the mounting location is good and if the pipe has changed shape or if there is too much fouling
inside the pipes
(5) Other poor condition
3. Menu Window Details
3.1 Menu Windows Arrangement
M00~M09 windows for the display of the flow rate, velocity, date time, totalizers, battery voltage and
estimated working hours for the battery.
M10~M29 windows for entering the pipe parameter.
M30~M38 windows for flow rate unit selections and totalizer unit selections.
M40~M49 windows for response time, zeroing, calibration and modification password setup.
M50~M53 windows for the built-in logger
15

Menu
window No.
Function
M00
Display flow rate and NET totalizer
If the net totalizer is turned off, the net totalizer value shown on the screen is the value
prior to its turn off
Select all totalizer unit in menu M31
M01
Display flow rate, velocity
M02
Display date time and POS(positive) totalizer
If the positive totalizer is turned off, the positive totalizer value shown on the screen is the
value prior to its turn off
M03
Display flow rate and NEG(negative) totalizer
If the negative totalizer is turned off, the negative totalizer value shown on the screen is the
value prior to its turn off
M04
Display date and time, flow rate. The date and time setting method is found in MENU60
M05
Display energy rate(instantaneous Caloric)and total energy (Caloric)
M06
Display temperatures, inlet T1, outlet T2
M07
Display analog inputs, AI3/AI4, current value and its corresponding temperature or
pressure or liquid level value
M08
Display all the detailed error codes
Display working condition and system error codes. ‘R’ stands for normal; others refer to
Chapter 5 for details.
M09
Display today’s total NET flow
M10
Window for entering the outer perimeter of the pipe
If pipe outer diameter is known, skip this menu and go to Menu 11to enter the outer
diameter
M11
Window for entering the outer diameter of the pipe. Valid range:0 to 18000mm.
TUF-2000S User Manual
M60-M78 windows for time-keeper initialization, version and ESN information viewing and alarms.
M82 window for viewing date totalizer.
M90~M94 are diagnostic windows for a more accurate measurement.
M97~M99 are not windows but commands for the outputting of display copying and pipe parameter
setups.
M+0~M+8 are windows for some additional functions, including a scientific calculator, viewer on
records such as total working hours, turn-on and turn-off times, dates and times when the
flow meter has been turned on or turned off.
Other menu windows such as M88 have no functions, or functions were cancelled because they are
not applied to this version of the software.
The major reason why the menu windows are arranged in this way is that the software programmer hopes that
the menu window arrangement for this version can be compatibility with the previous versions of the flow meter
software. This will make it easier for the former version users with this flow meter series.
3.2 Menu Window Details
16

Note: you just need to enter either the outer diameter in M11 or the peripheral in M10
M12
Window for entering pipe wall thickness
You may skip the menu and enter inner diameter in M13 instead.
M13
Window for entering the inner diameter of the pipe
If pipe outer diameter and wall thickness are enter correctly, the inner diameter will be
calculated automatically, thus no need to change anything in the window
M14
Window for selecting pipe material
Standard pipe materials (no need to enter material sound speed ) include:
(0) carbon steel (1) stainless steel (2) cast iron (3) ductile iron (4) copper
(5) PVC (6) aluminum ,(7) asbestos (8) fiberglass
(9) other( need to enter material sound speed in M15)
M15
Window for entering the pipe material speed, only for non-standard pipe materials
M16
Window for selecting the liner material, select none for pipes without any liner.
Standard liner materials(no need to enter the liner sound speed) include:
(1) Tar Epoxy (2) Rubber (3) Mortar (4) Polypropylene (5) Polystryol
(6)Polystyrene (7) Polyester (8) Polyethylene (9) Ebonite (10) Teflon
(11) Other (need to enter liner sound speed in M17)
M17
Window for entering the non-standard liner material speed.
M18
Window for entering the liner thickness, if there is a liner
M19
Window for entering the ABS thickness of the inside wall of the pipe
M20
Window for selecting fluid type
For standard liquids(no need to enter fluid sound speed) include:
(0) Water (1) Sea Water (2) Kerosene
(3) Gasoline (4) Fuel oil (5) Crude Oil
(6) Propane at -45C (7) Butane at 0C
(8)Other liquids(need to enter sound speed in M21 and viscosity in M22) (9) Diesel Oil
(10)Caster Oil (11)Peanut Oil (12) #90 Gasoline (13) #93 Gasoline (14)
Alcohol (15) Hot water at 125C
M21
Window for entering the sound speed of non- standard liquid, used only when option item
8 ‘Other’ is selected in M20
M22
Window for entering the viscosity of the non-standard liquids, used only when option item
8 ‘Other’ is selected in M20
M23
Window for selecting transducer type, There are 22 types as following
0. Standard M (The middle size)
1. Insertion Type C
2. Standard S
3. User Type
4. Standard B
5. Insertion Type B(45)
6. Standrad L (The large size transducers)
7. JH-Polysonics
8. Standard-HS (small size transducer for Handheld flow meter)
TUF-2000S User Manual
17

9. Standard-HM (middle size transducer for Handheld flow meter)
10. Standard-M1 (middle size transducer #1)
11. Standard-S1 (small size transducer #1)
12. Standard-L1 (large size transducer #1)
13. PI-Type
14. FS410 (middle size transducer for FUJI flow meter)
15. FS510 (large size transducer for FUJI flow meter)
16. Clamp-on TM-1 (Middle size transducer for Taosonics Instrument)
17. Insertion TC-1 (for Taosonic Instrument)
18. Calmp-on TS-1 (small size for Taosonics Instrument)
19. Reserved
20. Clamp-on TL-1 (For Taosonics Instrument)
21. Insertion TLC-2 (For Taosonics Instrument)
If the user-type-transducer is selected, you need enter additional 4 user-type-wedge
parameters that describe the user transducers.
If the PI-type transducer is selected, you need enter additional 4 PI-type transducer
parameters that describe the PI-type transducers
M24
Window for selecting the transducer mounting methods
Four methods can be selected:
(0) V-method (1) Z-method (2) N-method (3) W-method
M25
Display the transducer mounting spacing or distance
M26
(1) A switch for the parameters in flash memory will be loaded when power is turned on.
The default option is that the parameters will be loaded. If this switch is not turned on,
the system will try to use the parameters in the system RAM, if these parameters are
ok, otherwise the system will load the parameters in flash memory
(2) Function to store the current parameters into the flash memory, so that these parameters
will be solidified and will be loaded as the default parameters every time when power
is turned on.
M27
Entry to store to or restore from the internal Flash memory, as many as 9 different pipe
parameter configurations
To save or load the current setup parameter, use the going up or going down keys to
change the address number, press ‘ENT’ key, and use going down or going up keys to
select to save to or load from the memory.
M28
Entry to determine whether or not to hold (or to keep) the last good value when poor signal
condition occurs. YES is the default setup.
M29
Entry to setup empty signal threshold. When the signal is less than this threshold, the pipe
is regarded as empty pipe, and the flow meter will not totalize flow.
This is based on the fact that, for most occasions, when pipe is empty, the transducer
would still receive signal, just smaller than normal, As a result, The flow meter would
show normal operation, which is not correct.
Make sure that the entered value must be less than the normal signal strength.
When much noisy signals are received, to make sure the flow meter will not incorrectly
totalize flow, there is also a ‘Q’ threshold should be entered in M.5
M30
Window for selecting unit system. The conversion English to Metric or vice versa will not
affect the unit for totalizers.
M31
Window for selecting flow rate unit system.
TUF-2000S User Manual
18

Flow rate can be in
3
0. Cubic meter short for (m )
1. Liter (l)
2. USA gallon (gal)
3. Imperial Gallon (igl)
4. Million USA gallon (mgl)
5. Cubic feet (cf)
6. USA liquid barrel (bal)
7. Oil barrel (ob)
The flow unit in terms of time can be per day, per hour, per minute or per second. So there
are 32 different flow rate units in total for selection.
M32
Window for selecting the totalizers unit. Available units are the same as those in M31
M33
Window for setting the totalizer multiplying factor
The multiplying factor ranges from 0.001 to 10000. Factory default is 1
M34
Turn on or turn off the NET totalizer
M35
Turn on or turn off the POS (positive) totalizer
M36
Turn on or turn off the NEG(negative) totalizer
M37
(1) Totalizer reset
(2) Restore the factory default settings parameters. Press the dot key followed by the
backspace key. Attention, It is recommended to make note on the parameters before doing
the restoration
M38
Manual totalizer used for easier calibration. Press a key to start and press a key to stop the
manual totalizer.
M39
Interface Language selection.
The selection could also be changed automatically by the system, if English LCD display
is used as the display device.
M3.
Setup for local segmental LCD display. Enter 0 or 1 for the non-auto-scan mode; Enter
2~39 for the auto-scan mode. In the auto-scan mode the display will automatically scan
displaying from 00 to the entered number of the local segmental LCD display.
M40
Flow rate damper for a stable value. The damping parameter ranges form 0 to 999 seconds.
0 means there is no damping. Factory default is 10 seconds
M41
Low flow rate (or zero flow rate) cut-off to avoid invalid accumulation.
M42
Zero calibration/Zero point setup. Make sure the liquid in the pipe is not running while
doing the setup.
M43
Clear the zero point value, and restore the solidified zero point value.
M44
Set up a flow bias. Generally this value should be 0.
M45
Flow rate scale factor. The default value is ‘1’.
Keep this value as ‘1’, when no calibration has been made.
M46
Networks address identification number. Any integer can be entered except 13(0DH,
carriage return), 10 (0AH, line feeding), 42 (2AH), 38, 65535.
Every set of the instrument in a network environment should have a unique IDN. Please
refer to the chapter for communication.
M47
System locker to avoid modification of the system parameters.
If password is forgotten, you could send a command ‘LOCK0’ to the serial input to unlock.
Or you can write 0 to REGISTER49-50 under MODBUS protocol.
TUF-2000S User Manual
19

M48
Entry to linearity correcting data inputs. By using of this function, the non-linearity of flow
meter will be corrected. Correcting data shall be obtained by careful calibration.
M49
Displays the input contents for the serial port.
By checking the displays, you can know if the communication is ok.
M50
Switches for the built-in data logger. There are as many as 22 different items can be
chosen. To turn this function, select ‘YES’ the system will ask for selecting the items.
There are 22 items available. Turn on all those items you want to output
M51
Window to setup the time of scheduled output function (data logger, or Thermo-printer).
This includes start time, time interval and how many times of output. When a number great
than 8000 entered for the times of output, It means the output will be keeping always. The
minimum time interval is 1 second and the maximum is 24 hours.
M52
Data logging direction control.
(1) If ‘Send to RS485’ is selected, all the data produced by the data logger will be
transmitted out through the RS-232/RS485 interface
(2) If ‘To the internal serial BUS is selected, the data will be transmitted to the internal
serial bus which allows a thermal printer, or a 4-20mA analog output module, to be
connected to it.
M53
Display analog inputs, AI5, current value and its corresponding temperature or pressure or
liquid level value.
M54
Pulse width setup for the OCT (OCT1) output. Minimum is 6 mS, maximum is 1000 mS
M55
Select analog output (4-20mA current loop, or CL) mode. Available options:
(0) 4-20mA output mode (setup the output range from 4-20mA)
(1) 0-20mA output mode (setup the output range from 4-20mA, This mode can only be
used with Version-15 flow meter)
(2) Serial port controls 0-20mA
(3) 4-20mA corresponding fluid sound speed
(4) 20-4-20mA mode
(5) 0-4-20mA mode (can only be used with Version-15 flow meter)
(6)20-0-20mA mode(can only be used with Version-15 flow meter)
(7) 4-20mA corresponding flow velocity
(8)4-20mA corresponding heat flow rate
M56
4mA or 0mA output value,
Set the value which corresponds to 4mA or 0mA output current (4mA or 0mA is
determined by the setting in M55)
M57
20mA output value,
Set the value which corresponds to 20mA output current
M58
Current loop verification
Check if the current loop is calibrated correctly.
M59
Display the present output current of current loop circuit.
M60
Setup system date and time. Press ENT for modification. Use the dot key to skip the digits
that need no modification.
M61
Display Version information and Electronic Serial Number (ESN) that is unique for each
flow meter.
The users may employ the ESN for instrumentation management
M62
RS-232/RS485 setup. All the devices connected with flow meter should have matched
serial configuration.
TUF-2000S User Manual
20
Table of contents
Popular Measuring Instrument manuals by other brands

American Innovations
American Innovations Bullhorn RM 4100 Series user guide

Siemens
Siemens SQL36E110 Mounting instructions

inbody
inbody BSM 370 user manual

MONARCH INSTRUMENT
MONARCH INSTRUMENT Nova-Pro 100 User manual and installation guide

Saunia
Saunia 46304 Installation and operation guide

Rice Lake
Rice Lake MSI Dyna-Clamp Operation manual