Sadelco MiniMax 800 User manual

This document contains information proprietary to
Sadelco, Inc. The information in this document is
not to be used or duplicated in any manner without
prior written approval from Sadelco.
6DGHOFR ,QF
75 West Forest Avenue
nglewood, NJ 07631
USA
800-569-6299
tel: 201-569-3323
fax: 201-569-6285
www.sadelco.com
e-mail: [email protected]
Manual Part No. ST1003A
rev. 08/03/01

217(176
K YPAD SUMMARY AND S TUPS 4
INTRODUCTION 5
SP CIFICATIONS 6
INITIAL STARTUP 8
BATT RY OP RATION AND CHARGING 8
AUTO-CALIBRATION 9
SIGNAL L V L M ASUR M NT 10
General Procedure 10
Autoranging Attenuator 10
Channel Plan Selection 10
Tuning by Channel Number 11
Automatic Delta PA Calculation 11
Tuning by Frequency 11
Tuning Using the UP/DOWN Keys 11
Switching Between Channel and Frequency 12
Tuning Sub Band, UHF, and Special Channels 12
USING TH MMDS POW R OUT FACILITY 13
HUM M ASUR M NT 13
VISUAL CARRI R TO NOIS RATIO 14
DIGITAL CHANN L M ASUR M NT 15
FAVORIT CHANN LS 16
General Information 16
Programming Favorite Channels and Frequencies 16
rasing Favorite Channels 17
Recalling Favorite Channels 17
RROR COD S 17
WARRANTY POLICY 18
APP NDIX A: CHANN L TABL S 19
3

.(<3$'6800$5<$1'6(7836
Tune a Channel Number n (n=ch number), CH/FRQ
Tune a Frequency n (n=frq number with a decimal point), CH/FRQ
Tune a Digital Channel* From any Channel, Press DIG
Verify Current Channel Plan PLAN ( SC/CLR to return to tuning)
Select a New Channel Plan PLAN, n (n=plan number), NT or
PLAN, UP/DOWN, NT
Program a Favorite Channel tune channel, FAV, n (n=fav slot number), NT
Recall a Favorite Channel n (n=fav slot number), FAV
rase a Favorite Channel n (n=fav slot number), FAV, decimal, SC/CLR
rase All Favorite Channels 0, FAV, decimal, SC/CLR
Adjust the Volume VOL, UP or DOWN, SC/CLR
Beep ON/OFF VOL, decimal, SC/CLR
Disable Auto Shutoff decimal, 00, NT
Auto Shutoff 5 minutes decimal, 05, NT
Auto Shutoff 10 minutes decimal, 10, NT
View Software Date & Rev decimal, 80, NT
View Serial Number decimal, 81, NT
View Last Factory Cal Date decimal, 82, NT
Disable Power-Out Key decimal, 98, NT
(Minimax-M Only)
Select dBmV or dBuV decimal, 99, NT
* Minimax-M Models Only - During digital channel measurement, use microammeter
scales only for relative peaking. Use L D readout for absolute value of signal level.
4

,1752'8 7,21
The Minimax and Minimax-M hand-held meters incorporate state of the art technology
to make signal level measurement faster and easier than ever before. Auto-calibration,
using a built-in white noise source, corrects every measurement for frequency response,
attenuator setting, and temperature. Keypad tuning, an autoranging attenuator and dig-
ital readouts allow novice users to take accurate, foolproof readings with minimal train-
ing.
The full keypad provides instant access to all channels and frequencies. Dedicated keys
for all functions eliminate the need to look up hidden commands or menus. Advanced
features include C/N, Hum, Favorite Channel Programming, and ▲Picture-Audio.
Minimax-M Models have been tailored for the MMDS installer. An extra battery provides
switchable 18 V power at the F connector for powering the downconverter and a
microammmeter makes antenna alignment easy. A special channel plan designed for
MMDS is standard on these models.
All models are programmed with the most commonly used channel plans. Additional
standard and custom plans are also available.
5

0,1,0$;0,1,0$;063( ,), $7,216
FREQUENCY RANGE
MINIMAX 600 & M600 47 - 600 MHz
MINIMAX 800 & M800 5 - 872 MHz
STANDARD CHANNEL PLANS
MINIMAX 600, 800: IA, HRC, IRC, AIR (VHF, UHF, FM)
MINIMAX M600, M800: IA, HRC, IRC, AIR (VHF, UHF, FM), MMDS
Custom channel plans available on request.
TUNING INCREMENTS:
Frequency Mode: 125 KHz
Channel Mode: 6, 7, or 8 MHz
FREQUENCY ACCURACY: +/- 20 KHz
IF BANDWIDTH: 280 KHz at 3 dB points
ADJACENT CHANNEL REJECTION: 40 dB or better
FM SELECTIVITY:
Will distinguish 2 FM signals of similar amplitude 300 KHz apart
MEASUREMENT RANGE:
MINIMAX Models: -30 to +50 dBmV (+30 to +110 dBuV)
MINIMAX-M Models: -20 to +40 dBmV (+40 to +100 dBuV)
AMPLITUDE ACCURACY: Includes frequency response, log linearity and attenuator
+/- 1.0 dB at 770F (+/- 0.5 dB typical)
+/- 2.0 dB from 0 - 1200F (+/- 1.0 dB typical)
AUTO-CALIBRATION:
ach measurement computer corrected for level, attenuator setting and temperature.
Internal Calibrator: +/- 0.25 dB, 4.5 - 872 MHz White Noise Source
AUTORANGING ATTENUATOR:
Range: MINIMAX: 40 dB, MINIMAX-M: 20 dB
Input: 75 ohm, user replaceable F81 connector
BNC available by special order
DISPLAY:
All Models
Digital LCD: Indicates channel, frequency, and signal level in dBmV (switchable to
dBuV)
MINIMAX-M Model only:
Analog Microammeter: Indicates real time signal level for easy antenna orientation,
+/- 1.5 dB
6

0,1,0$;0,1,0$;063( ,), $7,216
RQWLQXHG
FEATURES:
Power Out: xtra large battery provides switchable 18V at F connector (MINIMAX-M
Models only)
C/N: Range >50 dB with external preselector, Accuracy +/- 2 dB
Hum: 0.5 - 5%, +/- 0.3%, detects all hum up to 600 Hz
Favorite Channels: 9 User Programmable Channels
▲PA: Calculates and displays difference between Picture and Audio carriers of same
channel .
Audio: Adjustable volume, waterproof mylar speaker
Illumination: Switchable lights for both LCD and Microammeter
Battery Test: Shows battery voltage and elapsed time
Power Supply: Replaceable nicad packs provides > 16 hrs of normal intermittent use
lectronic Shutoff: User adjustable - 5 min, 10 min., continuous
Digital Channel Measurement: Measures average digital power, Accuracy +/-1.8 dB
DIMENSIONS: MINIMAX: 8.5”L x 4” W x 2.5” D, 21 cm x 10 cm x 6 cm
MINIMAX-M: 8.5”L x 4” W x 3.0” D, 21 cm x 10 cm x 7.6 cm
WEIGHT: MINIMAX: 3.1 lbs, 1.4 Kg, including case and batteries
MINIMAX-M: 4.5 lbs, 2.0 Kg, including case and batteries
ACCESSORIES SUPPLIED:
MINIMAX Models:
Water resistant cordura nylon case Part No. CAS 002
AC 110 V Charger/Adaptor Part No. T60
Nicad Batteries Part No. BAT02/BAT03
Operator’s Manual Part No. ST1003A
MINIMAX-M Models:
Water resistant cordura nylon case Part No. CAS 003
AC 110 V Charger/Adaptor Part No. T60
Nicad Batteries Part No. BAT05
Operator’s Manual Part No. ST1003A
OPTIONS:
Model SH2 Strand Hook
Model T60 220 V Charger/Adaptor
Specifications Subject to change without notice.
7

,1,7,$/67$5783
Unpack and inspect the instrument for any visible defects or damage. Claims for ship-
ping damage must be made by the receiver. Make sure to save all packaging materials.
Pre and hold ON/OFF key. Make ure to hold the ON/OFF key for one econd
to allow the meter to lock on.
The LCD Window will briefly display the software rev and then PASS. The meter will per-
form an Autocalibration and tune to the last channel or frequency tuned before it was
shut off.
To adju t volume, Pre VOL, UP or DOWN, ESC/CLR.
%$77(5<23(5$7,21$1' +$5*,1*
MINIMAX models are powered by 15 AA high energy nicad batteries. The 15 cells are
split into one pack of 7 and one pack of 8. MINIMAX-M models are powered by 15 Sub-
C high energy nicad batteries (1 pack). The batteries in your new meter may not be fully
charged when you receive it.
Check the battery voltage before u ing your meter and recharge if nece ary.
Charge time is aproximately 14 hours.
Use only the proper Sadelco Charger:
AC Charger: T60 for 110V
T60 for 220V
The operating voltage range is approximately 20 Volts to 15 Volts. At 17 Volts, the meter
will beep and begin flashing a Lo Bat warning. Under average operating conditions, you
will have about 30 minutes left for both models. When the charger is plugged in, the dis-
play will show 24 V.
TO CHECK THE BATTERY VOLTAGE
Di connect the charger.
Minimax Model :
Press the BAT key. The battery voltage will be displayed on the left side of the display.
The Minimax will operate for a minimum of 3.5 hours continuously with a full charge.
(> 16 hours of normal intermittent use)
Minimax-M Model :
Press the HUM/BAT key. The Battery voltage will be displayed on the left side of the dis-
play. If this key is held down for 3 seconds, the HUM mode will be activated. To escape,
press SC/CLR.
With the power out on, the Minimax-M will operate for a minimum of 4 hours continuously
with a full charge (>16 hours of normal intermittent use), or 7 hours with the power out
off.
Use of lights, maximum volume, or extremely cold temperature will reduce operating
times.
8

%$77(5<23(5$7,21$1' +$5*,1*
RQWLQXHG
USING THE ELAPSED TIME INDICATOR
To help maximize battery power, we have included an elapsed time indicator which
appears on the right side of the LCD whenever battery voltage is displayed.
After fully charging the meter, re et the timer to zero by pre ing: decimal point, BAT.
Now, whenever the meter is used, the operating time is accumulated.
After every 30 minutes of elapsed time, the meter will beep and display the total elapsed
time and the battery voltage for 4 seconds.
AUTOMATIC SHUTOFF
To increase the effective battery life, all Minimax and Minimax-M models are equipped
with an automatic shutoff circuit. It is factory set at 5 minutes. Any keypad entry during
the 5 minutes resets the clock. To turn the meter on after an auto shutoff, press and hold
the ON/OFF key.
To deactivate the auto shutoff: Press decimal point, 00, NT
To change setting to 10 minutes: Press decimal point, 10, NT
To return to 5 minutes: Press decimal point, 05, NT
$872 $/,%5$7,21
Using Sadelco’s patented white noise calibration technology in conjunction with stored
correction tables, auto-calibrations are performed every two minutes of operation.
Auto-calibrations are also performed:
when the meter is turned on
when a new channel is tuned
when the autoranging attenuator switches pads in or out
when the internal temperature changes more than 8 degrees
You may notice a momentary delay and hear the clicking of the attenuator during an
Auto-calibration. The display is frozen during this time.
9

6,*1$//(9(/0($685(0(17
GENERAL PROCEDURE
1. Turn power ON.
2. Check battery.
3. Connect RF signal to input.
4. Select Channel or Frequency via keypad.
5. Select Picture or Audio carrier using ▲PA Key.
6. Read signal level on right side of LCD display.
(The microammeter also indicates level on MINIMAX-M models)
AUTORANGING ATTENUATOR
The autoranging attenuator eliminates the need to manually switch pads in or out to
bring the signal level in range. You will hear some clicking as the relays are switched in
and out. If the RF signal is above or below the meter’s range, the LCD will indicate this
by flashing Or (Over Range) or Ur (Under Range). On MINIMAX-M Models, the
microammeter will automatically change ranges as necessary. L Ds indicate the prop-
er scale to read. If the RF signal is above or below the meter’s range, the microamme-
ter will peg right or left respectively.
CHANNEL PLAN SELECTION
Refer to Appendix A for a list of standard channel plans programmed into your meter.
Additional standard and custom plans are available from the factory.
Pre the PLAN key to view the current channel plan on the LCD.
To elect another plan:
1. Press the PLAN key.
2. nter the number of the Plan or scroll to another plan using the UP/DOWN keys.
3. Press the NT key.
4. The meter will tune to Channel 2 Picture carrier of the new plan.
10

TUNING BY CHANNEL NUMBER
1. Press the desired channel number. Leading zeros such as 02 are not necessary.
2. Press CH/FRQ.
By not using a decimal point, the meter automatically tunes to either the Picture or Audio
Carrier of the channel number that was entered.The display will indicate the channel
number followed by a P for Picture or an A for Audio and the signal level.
xample: CH 2P -8.7
If an invalid channel number is selected, the meter defaults to the previously tuned
channel or frequency.
AUTOMATIC ▲PA CALCULATION
To automatically toggle between Picture and Audio carriers of the same channel and cal-
culate the ▲(difference) between them:
1. Tune to a Picture carrier.
2. Press the ▲PA key to switch to the Audio carrier.
3. Press the ▲PA key again to calculate and display the ▲PA.
4. Press the ▲PA key again to return to the Picture carrier.
Note: the ▲PA function can not be activated in the digital channel mode.
TUNING BY FREQUENCY
1. nter the desired frequency number including a decimal point.
2. Press CH/FRQ.
The meter will tune and display frequencies in 125 KHz steps. If the entered frequency
does not fall on one of the 125 KHz steps, the meter will tune to and display the closest
step. Offsets up to +/-40 KHz are read with negligible error.
You must enter a decimal point even for frequencies that are whole numbers such as
54.00 MHz. The decimal point tells the meter to tune to a frequency instead of a chan-
nel. You do not have to enter zeros after the decimal point.
TUNING USING THE UP/DOWN KEYS
The UP/DOWN keys function in all three tuning modes: Channel, Frequency, and
Favorite Channel.
In the Channel Mode, each press of the UP or DOWN key initiates a 6 MHz shift to the
next carrier (7 MHz or 8 MHz in some uropean Tables).
In the Frequency Mode, each press of the UP or DOWN key initiates a 125 KHz shift.
11

TUNING USING THE UP/DOWN KEYS (Continued)
In the Favorite Channel Mode, each press of the Up or Down key initiates a jump to the
next programmed channel or frequency.
In all modes, when the top or bottom of the tuning range is reached, it will loop around
continuously.
SWITCHING BETWEEN CHANNEL AND FREQUENCY
When tuned to a channel, pressing the CH/FRQ key will display the frequency of that
channel and switch the meter to the Frequency Mode.
When tuned to a frequency, pressing the CH/FRQ key will display the channel number
of that frequency and switch the meter to the Channel Mode. If no channel number
exists for a particular frequency, the meter will tune to either the closest picture or audio
carrier depending on which mode the ▲PA is in.
TUNING SUB BAND, UHF, AND SPECIAL CHANNELS
Some channel plans contain more than one channel with the same channel number. In
order to provide direct keypad access to all channels, we have added an additional num-
ber in front of the channel number in some cases. Refer to Appendix A for complete
channel listings.
Sub Band channel , T07 - T13 can be acce ed directly by entering 707 - 713.
In the MMDS Channel Plan, UHF Channel 14 - 70 are elected directly by entering
914 - 970.
In the AIR Channel Plan, UHF Channel are elected by entering 14 - 70.
For example:
The MMDS Channel Plan contains:
Channel Name Channel No. Keypad Entry Frequency
MDS1 Channel 14 14 123.25 MHz
UHF14 Channel 14 914 471.25 MHz
A1 Channel 24 24 223.25 MHz
UHF24 Channel 24 924 531.25 MHz
The IA Channel Plan contains:
Channel Name Channel No. Keypad Entry Frequency
T07 Channel 7 707 7.00 MHz
VHF7 Channel 7 7 175.25 MHz
T13 Channel 13 713 43.00 MHz
VHF13 Channel 13 13 211.25 MHz
12

86,1*7+(00'632:(5287)$ ,/,7<
All MINIMAX-M models come equipped with an extra large built-in battery that provides
a switchable 18V DC power supply out of the F connector.
This feature is used to power MMDS antenna mounted downconverters during installa-
tion. Do not activate thi feature when connected to a CATV or MATV y tem.
To activate, Press the DOWN/CONV key. PWR OUT L D lights, 18V is present at the F
connector.
To deactivate, Press the DOWN/CONV key again
DISABLING MMDS POWER OUT
To prevent accidental activation when connected to a CATV or MATV system:
Press decimal point, 98, NT.
Repeat to enable Power Out Key.
+800($685(0(17
On MINIMAX-M models, the HUM function shares a key with the BAT check. This
requires the operator to hold the key down for 3 seconds to activate HUM. During this
3 seconds, the battery condition is displayed. On the MINIMAX models, the HUM key is
separate so it needs only to be pressed to activate HUM.
Measurement of HUM (low frequency disturbance) can be made on any active audio car-
rier or unmodulated CW signal. A minimum signal level of -20 dBmV for the Minimax and
-11 dBmV for the Minimax-M is required for Hum measurement.
To mea ure HUM:
1. Tune to the Picture or Audio carrier of the channel you wish to measure or tune to
the frequency of the CW signal.
2. Press HUM (Hold for 3 seconds on MINIMAX-M models).
3. HUM will be displayed in %. (If Hum is >6%, the display will flash 6%).
4. Press SC/CLR to return to signal level measurement mode.
Note: If you tune to a picture carrier using a channel number, the meter automatically
switches to the corresponding audio carrier before making the HUM measurement. If
you tune to a picture carrier using a frequency number, the meter remains at that fre-
quency and attempts to measure HUM. This feature allows HUM measurement of a CW
signal at any frequency but will give incorrect HUM measurements if the entered fre-
quency is an active picture carrier.
Note: Hum cannot be activated in the digital channel measurement mode.
13
▲
▲

▲
9,68$/ $55,(57212,6(5$7,2
Noise measurements on active cable systems are effected by many factors such as car-
rier sidebands, harmonic products, and spurious beats.
Ideal measurements are made by using a band pass filter to isolate the channel being
tested and by shutting off the carrier before making the noise measurement. Refer to the
NCTA recommended practices for guidelines on making the best measurements under
various conditions.
Use of a low noise preamplifier is recommended when signal levels are below +20
dBmV.
Method One
This is the preferred method when conditions permit.
1. Connect a band pass filter between the test point and the meter.
2. Tune to the desired picture carrier using channel number or frequency.
3. Press the C/N key. The meter stores the carrier level.
4. If you tuned using a channel number the meter will switch to the corresponding fre-
quency and begin to flash. If you tuned using a frequency, the meter will remain in
the frequency mode and begin to flash.
5. Shut off the carrier.
6. Press the NT key. The meter reads the noise level, adds a correction factor, cal-
culates and displays the C/N ratio.
7. Press SC/CLR to return to the signal level mode.
Method Two
If you cannot turn off the carrier.
1. Connect a band pass filter between the test point and the meter.
2. Tune to the desired picture carrier using channel number or frequency.
3. Press the C/N key. The meter stores the carrier level.
4. If you tuned using a channel number the meter will switch to the corresponding fre-
quency and begin to flash. If you tuned using a frequency, the meter will remain in
the frequency mode and begin to flash.
5. Use the UP/DOWN keys or number keys and CHN/FRQ to search for the lowest
noise level near the channel. This is often 1.5 MHz below the carrier but will
depend on the shape of your filter.
6. When you have found the lowest noise reading, press NT. The meter reads the
noise level, adds a correction factor, calculates and displays the C/N ratio.
7. Press SC/CLR to return to the signal level mode.
14

9,68$/ $55,(57212,6(5$7,2
RQWLQXHG
Method Three
When you cannot shut off the carrier and you do not have a band pass filter.
1. Locate a channel that does not have another channel below it, or where there is a
band guard. Tune to the picture carrier of that channel using channel number or
frequency.
2. Press the C/N key. The meter stores the carrier level.
3. If you tuned using a channel number the meter will switch to the corresponding fre-
quency and begin to flash. If you tuned using a frequency, the meter will remain in
the frequency mode and begin to flash.
4. Use the UP/DOWN keys or number keys and CHN/FRQ to search for the lowest
noise level near the channel. This is often 1.5 MHz below the carrier.
5. When you have found the lowest noise reading, press NT. The meter reads the
noise level, adds a correction factor, calculates and displays the C/N ratio.
6. Press SC/CLR to return to the signal level mode.
',*,7$/ +$11(/0($685(0(1702'(
This Minimax is equipped with a digital channel measurement facility. It is activated by a
quick press of the CN/DIG key.
Note: If this key is held down, the Carrier to Noise function will be activated. To exit the
Carrier to Noise function, press SC/CLR.
To Mea ure a Digital Channel: From any channel, Pre (do not hold) “CN/DIG”.
The display will flash “digtL on”, then “di” followed by the channel number and the signal
level.
During digital channel measurements, the LCD readout is automatically corrected to
show the average power level of a 6 MHz digital channel.
The microammeter (Minimax-M models only) is not corrected during digital channel
reading and thus shows a lower, incorrect level. Use the microammeter only for peak-
ing during digital channel reading.
The Minimax remain in the Digital Mode until the CN/DIG key i pre ed again.
Once in the Digital Mode:
- Tune other digital channels using either the UP/DOWN or number keys followed by
CH.
- All Favorite Channels are measured as digital even if they were programmed in the
analog mode.
- The ▲PA and HUM functions cannot be activated.
- The Minimax remains in the digital mode after shutoff and restart.
To return to normal analog channel reading: Pre (do not hold) “CN/DIG”.
The display will flash “digtl OFF”
Note: To activate Carrier to Noise measurement, Press and Hold the CN/DIG key.
15

)$925,7( +$11(/352*5$00,1*
GENERAL INFORMATION
The MINIMAX can store up to 9 favorite channels in each channel plan. You may store
channel numbers and/or frequencies in any order and combination.
Favorite channels can be programmed in the normal analog channel mode or digital
channel measurement mode. Favorite channels are measured as either digital, if in the
digital mode, or analog, if in the normal analog mode.
To enter the Favorite Channel Mode, you must first program one or more favorite chan-
nels, and then recall one of them.
Once in the Favorite Channel Mode, pressing the UP/DOWN keys will switch in the next
programmed favorite channel. Any unprogrammed slots will be skipped and the unit will
loop from the highest to the lowest slot.
You can perform the C/N, Hum, CH/FRQ, and ▲PA functions using your favorite chan-
nels. You can also change the Optional User Settings for the Auto shutoff, Volume, Beep
ON/OFF, and dBmV - dBuV in the Favorite Channel Mode.
To exit the Favorite Channel Mode, press SC, tune a channel using a valid channel
number followed by CH/FRQ, or select a new channel plan.
Note: When in the digital channel measurement mode all favorite channels are meas-
ured as digital even if they were programmed in the normal analog mode.
PROGRAMMING FAVORITE CHANNELS AND FREQUENCIES
1. Tune the meter to the channel or frequency you wish to store.
2. Select the Favorite Slot (1 - 9) you wish to store it in, press FAV, then press NT.
xample: Store CH 2 Picture, IA (Table 1) into Favorite Slot #1.
Make sure the desired Channel Plan has been selected.
nter 2
nter CHN/FRQ
nter ▲PA to select Picture if necessary
nter FAV
nter 1
nter NT
xample: Store 67.250 MHz into Favorite Slot #2
nter 67.250
nter CH/FRQ
nter FAV
nter 2
nter NT
Repeat the above procedures to program additional Favorite Channels.
Note: Programming a Favorite Channel does not switch the meter into the Favorite
Channel Mode. This can only be done by recalling a previously programmed Favorite
Channel.
16

ERASING FAVORITE CHANNELS
To change the frequency or channel in a particular Favorite Channel Slot, just reprogram
over the old information.
To erase a Favorite Channel completely, enter the number of the slot (1-9), FAV, decimal
point, CLR.
To erase all Favorite Channels in a channel table at once enter 0, FAV, decimal point,
CLR.
RECALLING FAVORITE CHANNELS
nter the Favorite Channel number (1-9)
nter FAV
You may now use the UP/DOWN keys to recall other favorite channels or you may recall
them by repeating the above procedure with a new FAV channel number.
(5525 2'(6
1. F---Err
This code appears if the user tries to recall a Favorite Channel Number which has
not been programmed. Meter defaults to previously tuned channel.
2. Error Chn
This code appears if the user tries to switch to Channel from a Frequency which
has no associated channel. Meter defaults to previously tuned Frequency.
3. Error tbl
This code appears if the meter cannot access the previously selected channel
table. Minimax will default to Table 1- IA. Minimax-M will default to Table 5 dS1.
User must reselect desired channel plan if other than default plan.
4. Ur - Under Range
This code appears if the signal level is too low (or nonexistent) to measure.
5. Or - Over Range
This code appears if the signal level is too high to measure. An external attenua-
tor pad can be used to bring the signal level into range.
6. Error -Hu
This code appears if the signal level is too low (below -20 dBmV for Minimax and
below -11 dBmV for Minimax-M) to measure hum.
7. Error +Hu
This code appears if the signal level is too high (above +50 dBmV for Minimax and
above +40 dBmV for Minimax-M) to measure hum.
17
▲

▲
▲
▲
▲▲
▲
▲
:$55$17<32/, <
Sadelco warrants all instruments against defects in material and workmanship for a peri-
od of twelve months after shipment. Sadelco will repair or replace any assembly or com-
ponent (except batteries) found to be defective under normal use during this period.
Sadelco’s obligation under this warranty is limited solely to repairing any such instrument
which, in Sadelco’s opinion, proves to be defective within the scope of the warranty when
returned to the factory. Transportation to the factory is to be prepaid by the purchaser.
Return transportation via UPS surface will be provided by Sadelco. Should the customer
request an alternate shipping method, the customer will be solely responsible for any
charges in excess of UPS surface rates.
This warranty does not apply to any products repaired or altered by any persons not
authorized by Sadelco, or not in accordance with instructions furnished by Sadelco. If
the instrument is defective as a result of misuse, improper repair, or abnormal conditions
of operation, the warranty will become void. In doubtful cases we will contact the pur-
chaser for repair authorization. Please notify us if an estimate is required.
Sadelco assumes no responsibility for its products being used in a hazardous manner
either alone or in conjunction with other equipment. Sadelco assumes no liability for
secondary damages and, in any event, Sadelco’s liability for breach of warranty under
any contract or otherwise shall not exceed the purchase price of the specific product
shipped and against which a claim is made.
An RMA number is not required to return either warranty or repair units to Sadelco.
However, Sadelco assumes no liability for any product in transit to Sadelco. Please
make sure to include a description of the problem or symptom along with your complete
company name, address, telephone, fax, and contact person.
For Warranty or Repair service return to:
Sadelco, Inc.
Service Department
75 West Forest Avenue
nglewood, NJ 07631
USA
Tel: 800-569-6299 (USA only)
Tel: 201-569-3323
Fax: 201-569-6285
www.sadelco.com
e-mail: [email protected]
18

$33(1',;$
+$11(/7$%/(6
19
▲

$33(1',;$ +$11(/3/$16
Listed below are the standard channel plans for various MINIMAX and MINIMAX-M
models. Not all models contain all plans. Models with limited frequency ranges contain
only the appropriate portions of the plans.
Since some plans have more than one channel with the same channel number, the first
column lists the keypad entry for direct access to a channel and the second column lists
the channel number as it is displayed on the LCD.
Additional standard and custom channel plans are available from the factory.
Plan 1 - EIA (NTSC, NCTA)
Standard USA cable channel plan
Plan 2 - HRC
Picture and audio carriers are offset 1.25 MHz lower than IA Plan 1.
xception Channels 5 & 6 which are 0.75 MHz higher. Channel 1 added.
Plan 3 - IRC
Same as IA Plan 1 except channels 5 & 6 which are offset 2 MHz higher.
Channel 1 added.
Plan 4 - AIR
Standard USA over the air VHF & UHF channels
Plan 5 - MMDS1 (di play indicate DS1)
Standard USA over the air VHF & UHF, MMDS channels 24 - 54 (A1 - G4), MDS1 chan-
nel 14, and MDS2 channel 99.
Plan 6 - Cu tom
Plan 7 - Cu tom
Plan 8 - Cu tom
Plan 9 - Cu tom