Installation Instructions
Sartorius
YDI01C-WP (Option A6|A7)
Installing a Digital Interface in Place of an A/D Converter in
Combics Indicators, Models CIS1..|CIS2..|CIS3.. | CIS 1..|CIS 2..|CIS 3..
98647-003-78
The YDI01C-WP digital interface built into
the Combics indicator enables connection
of any of the following:
– Digital weighing platform
(e.g. IS weighing platform) using the
YCC02-R12F6 connecting cable (Option
M41: 12-contact round female connector;
length: 6 m; open cable ends), or
– Digital weighing instrument (e.g., A, P,
ME, EA, EB, CP balance/scale) using the
YCC02-D25F6 connecting cable (Option
M42: 25-pin D-Sub male connector;
length: 6 m; open cable ends), or
– Digital weighing instrument (e.g. QA,
QC, FB, FC scale) using the YCC02-R12M6
connecting cable (Option M43: 12-pin
round male connector; length: 6 m;
open cable ends).
Note: To have devices communicate with the Combics indicator as plat-
forms and scales through the SBI protocol, configure the operating menu
as follows:
Baud rate: 9600 Handshake mode: Hardware, 1 character after CTS
Parity: Odd Number of data bits: 7
Stop bits: 1
Printing: Automatic without stability
Identification: With ID code (22 characters)
To have devices communicate with the Combics indicator as platforms and
scales over the xBPI protocol, configure the operating menu as follows:
Setting: xBPI protocol
Interface: RS-232 or RS-485
!Refer to the relevant operating instructions for descriptions of the
operating menus.
The menu setting in the Combics indicator depends on the particular
platform to be connected:
Setting: SBI or xBPI protocol
Interface: RS-232 or RS-485
Example: For Combics 1 and 2, this setting is configured in the Setup
menu; for Combics 3, in the menu under Device parameters
WP1
RS-232 or RS-485
SBI: Standard o IS-485
SBI: Trade ADC-485
o IS-232
ADC-232
!Always make sure the equip-
ment is disconnected from
power before performing any
installation, maintenance or
repair work.
!Work that affects IP protec-
tion must be carried out with
extreme care.
!Any installation work that
does not conform to the
instructions in this manual
will void all claims under the
manufacturer’s warranty.
!Do not operate the YDI01C-
WP in hazardous areas.
§Place the Combics indicator
on a soft surface and remove
the four cap nuts from the
front panel.
§Remove the front panel
Installing t e YDI01C-WP PCB
§Remove the A/D converter
PCB
– Unplug the male connector
– Remove the 3 screws
– Remove the A/D converter
PCB
§Plug in the YDI01C-WP PCB
– Use the 2 screws to fasten
the PCB
– Plug in the male connector