Scanivalve DSA3217-PTP Series Operation manual

DSA3217-PTP/3218-PTP



iv
DSA3200-PTP
iv
Product
Preface
PREFACE
The WARNING! symbol indicates that danger of injury for persons and the environment and/or
considerable damage (mortal danger, danger of injury) will occur if the respecve safety precau-
ons are not taken.
The CAUTION ! symbol indicates danger for the system and material if the respecve safety precau-
ons are not taken.
The ESD note symbol indicates that proper precauons for handling Electrostac Sensive Devices
needs to be taken when performing the related operaon. This includes the use of grounded work
surfaces and personal wrist straps to prevent damage to sensive electronic components.
Scanivalve Corporaon, Liberty Lake, Washing-
ton, hereaer referred to as Seller, warrants to the Buyer
and the rst end user that its products will be free from
defects in workmanship and materi al for a period of twelve
(12) months from date of delivery. Wrien noce of any
claimed defect must be received by Seller within thirty
(30) days aer such defect is rst discovered. The claimed
defecve product must be returned by prepaid transporta-
on to Seller within ninety (90) days aer the defect is rst
discovered. Seller’s obligaons under this Warranty are
limited to repairing or replacing, at its opon, any product
or component part thereof that is proven to be other than
as herein warranted.
Surface transportaon charges covering any
repaired or replacement product or component part shall
be at Seller’s expense; however, inspecon, tesng and
return transportaon charges covering any product or com-
ponent part returned and redelivered, which proves
not to be defecve, shall be at the expense of Buyer or the
end user, whichever has returned such product or compo-
nent part.
This Warranty does not extend to any Seller product
or component part thereof which has been subjected to
misuse, accident or improper installaon, maintenance or
applicaon; or to any product or component part thereof
which has been repaired or altered outside of Seller’s facili-
es unless authorized in wri ng by Seller, or unless such
installaon, repair or alteraon is performed by Seller; or
to any labor charges whatsoever, whether for removal and/
or reinstallaon of the defecve pro duct or component
part or otherwise, except for Seller’s labor charges for
repair or replacement in accordance with the Warran ty.
Any repaired or replacement product or component part
thereof provided by Seller under this Warranty shall, upon
redelivery to Buyer, be warranted for the unexpired poron
of the original product warranty.

v
DSA3200-PTP
v
Product Preface
v
THIS WARRANTY IS IN LIEU OF AND EXCLUDES ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, ARISING BY
OPERATION OF LAW OR OTHERWISE, INCLUDING THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE, AND IN NO EVENT SHALL
SELLER BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES.
In the event of a failure:
1) Nofy Scanivalve Corporaon, Customer Service
De partment. Include model number and serial number.
On receipt of this informaon, service data or shipping
instrucons will be forwarded. This may be transacted by
telephone or e-mail.
2) On receipt of shipping instrucons, forward the
pro duct, transportaon prepaid. Repairs will be made and
the product returned.
3) All shipments should be made via “Best Way”.
The product should be shipped in the original packing con-
tainer or wrapped in protecve material and surrounded by
a minimum of four (4) inches of a shock absorbing material.
Scanivalve is a registered trademark of Scanivalve Corpora-
on.
All other brand and product names are trademarks or regis-
tered trademarks of their respecve companies.
If the product must be shipped, whether being returned
to Scanivalve or relocated to another locaon it must be
packaged properly to minimize the risk of damage. The
recommended method of packing is to place the instru-
ment in a container, surrounded on all sides with at lease
four inches of shock aenuang material such as Styrofoam
peanuts.
Please note that the product specicaons and other
informaon contained in this manual are subject to change
without noce. Scanivalve Corporaon makes an eort and
strives to provide complete and current informaon for the
proper use of the equipment. If there are any quesons
regarding this manual or the proper use of the equipment,
contact Scanivalve Corporaon.
If there are any quesons or concerns regarding any Sca-
nivalve product please do not hesitate to contact us at the
following:
Scanivalve Corp.
1722 N. Madson Street
Liberty Lake, WA 99019
Telephone: (800)935-5151 (509)891-9970
Fax: (509)891-9481
scanco@scanivalve.com
www.scanivalve.com
Scanivalve Corporaon is an ISO9001:2015 cered
company.

vi
DSA3200-PTP
vi
Table of Contents
[INTENTIONALLY LEFT BLANK]

vii
DSA3200-PTP
vii
Table of Contents
Warnings, Cauons and Notes iv
Warranty iv
Trademarks ® and Copyrights © v
Packaging for Shipment v
Important Noce v
Contact Informaon v
DSA3217-PTP 1
DSA3218-PTP 2
General Descripon 5
Module Overview 5
DAS3217-PTP Conguraons 5
DSA3217-PTP/16Px 5
DSA3217-PTP/16Px - Dual Range 6
DSA3217-PTP/8DPx - Dierenal 6
DSA3217-PTP/16Px - Absolute 6
DSA3217-PTP/16Px - Individual Reference 7
DSA3218-PTP Conguraons 8
DSA3218-PTP/16Px 8
DSA3218-PTP/16Px - Dual Range 8
DSA3218-PTP/8DPx - Dierenal 8
DSA3218-PTP/16Px - Absolute 9
DSA3218-PTP Heater Opon 9
DSA3218-PTP Cooling Kit Opon 9
Environmental Consideraons 10
Unpack & Inventory 11
Mounng 11
Warm-up 11
Communicaons 11
Serial Communicaons 11
Ethernet Communicaons 12
Client/Host Opons 12
PC - TCP/IP 12
PC - UDP 13
PC - ScanTel 13
PC - LabVIEW Conguraon Ulity 13
PC - LabVIEW Development Kit 13
PC - HyperTerminal 13
Zero Oset Calibraon 13

viii
DSA3200-PTP
viii
Table of Contents
Quick Zero Jumper 14
Scanning 14
Scanning With An External Trigger 14
Hardware Trigger 14
Soware Trigger 14
Freon Applicaon Operaons 15
Pneumac Connecons 16
Making 1/16” Tubulaon Connecons 16
Pressure Measurement (Px) Ports 17
Calibraon (CAL) Port 17
Reference (REF) Port 17
Control 1 (CLT1) Port 17
Control 2 (CTL2) Port 17
Control Supply (CTL SUPPLY or CTL) 17
Return (RTN) Port 17
Purge Control (PRGCTL) Port 18
Purge Supply (PRG) Port 18
Power Requirements 18
Calibraon Valve 18
DSA-PTP Control and Conguraon 20
DSA-PTP Commands 20
Command List 21
Calibrate Barometric Pressure 21
Clear 21
Calibrate Zero 22
Delete 23
Digital Out 23
Error 23
Insert 24
List All 25
List All Ranges 26
List Calibrate 27
List Delta 27
LIST FTP Sengs (LIST FTP) 28
List Gain 28
List Idencaon 29
LIST IP Sengs (LIST IP) 29
List Master 30
List MAXn Pressures 31
List MINn Pressures 31
List Negave Points 32
List Oset 32
List PTP Variables 33
List Scan 33
List Temperature 34
List Zero 34
Save 34

ix
DSA3200-PTP
ix
Table of Contents
Scan 35
Scan Trigger 36
Set 36
Status 36
Stop 37
Version 37
Set PTP Time 37
Get PTP Time 38
Get UTC Oset 38
Conguraon Variables 39
Scan Variables (LIST S) 39
Average (AVG) 39
Binary Data (BIN) 39
Unit Conversion Factor (CVTUNIT) 39
Engineering Units (EU) 40
Data Format (FORMAT) 40
Frames Per Scan (FPS) 40
Page (PAGE) 41
Period (PERIOD) 41
QPKTS 41
Data Sample Source (SIM) 41
Time Stamp (TIME) 42
Scan Unit (UNITSCAN) 42
Enable External Trigger (XSCANTRIG) 42
Zero Correcon (ZC) 43
IP Variables (LIST IP) 44
IP Address (IPADD) 44
Subnet Mask (SUBNET) 44
MAC Address (MAC) 44
LOGIN (LOGIN) 44
Password (PASSWORD) 45
Login 1(LOGIN1) 45
Password 1(PASSWORD1) 45
Anonymous Logon (ALLOWANON) 45
Applicaon (APP) 45
Set Gateway (GW) 46
PTP Variables (LIST PTP) 47
PTP Enable (PTPEN) 47
PTP Stascs (STAT) 47
Scan Start Time (SST) 47
Scan Start Date (SSD) 48
UTC Oset (UTCOFFSET) 48
FTP Variables (LIST FTP) 49
FTP User Name (UESRFTP) 49
FTP Password (PASSFTP) 49
Enable FTP (ENFTP) 49
FTP File Path (PATHFTP) 49
FTP IP Address(IPFTP) 49
FTP File Name(FILEFTP) 50
Pressure Variables (LIST B, LIST H, LIST L and LIST N) 51
Absolute Enable (ABSn) 51
Maximum Pressure (MAXn) 51
Minimum Pressure (MINn) 51

x
DSA3200-PTP
x
Table of Contents
Negave Calibraon Points (NEGPTSn) 51
Calibraon Variables (LIST C) 52
Channel 1-8 Number of Negave Calibraon Points (NEGPTSL) 52
Channel 9-16 Number of Negave Calibraon Points (NEGTPSH) 52
Channel 1-8 Maximum Pressure (PMAXL) 52
Channel 9-16 Maximum Pressure (PMAXH) 52
Channel 1-8 Minimum Pressure (PMINL) 53
Channel 9-16 Minimum Pressure (PMINH) 53
Error Limit Variables (LIST LIMIT) 54
High Temperature Error Limit (THI) 54
Low Temperature Error Limit (TLOW) 54
High Pressure Error Limit (PHI) 54
Low Pressure Error Limit (PLOW) 54
Idencaon Variables (LIST I) 55
Serial Echo (ECHO) 55
Model (MODEL) 55
Ethernet Port (PORT) 55
Ethernet Host (HOST) 56
Pressure Deltas (LIST D) 56
Deltas (DELTAn) 56
Zero Osets (LIST Z) 56
Zero Oset (ZEROn) 56
DSA-PTP Packet Denions 57
Packet Types 57
Host to DSA-PTP Command Packet 57
Status Packet 57
Scan Raw Packet 57
Scan EU Packet 58
Scan Raw with Time Packet 58
Scan EU with Time Packet 58
Scan EU Packet with FPT - Floang Point Temperature (SET EU 2) 58
Scan RAW/EU with PTP Time (SET TIME 3) 59
ASCII Packet 59
ASCII Packet Examples 60
Network Protocols Supported 60
DSA3200-PTP Boot Parameters 60
Changing Boot Parameters (including IP address) 60
DSA3200-PTP Buer Descripon 61
DSA-PTP Network Conguraons 61
Isolated Network 61
Integrated with System Network 61
Sub-Network with Gateway 61
DSA-PTP Network Addressing 61
Calibraon 62
Full Calibraons 62
Field Calibraons 62
Zero Oset Calibraon 62
Special Procedures 63
Internet Explorer Setup 63

xi
DSA3200-PTP
xi
Table of Contents
DSA-PTP Firmware Installaon (Windows 7 & 10) 63
DSA-PTP Coecient Upload 64
DSA3200-PTP Sensor Replacement 65
DSA3200-PTP General Checkout Procedure 68
Appendix A: Temperature Compensated Pressure Conversion 71
Appendix B: Engineering Unit Conversion Constants 72
Appendix C: DSA3200-PTP Error List 73
Boot/Inializaon Errors 73
Set Conguraon Variable Errors 73
Command Errors 75
Communicaon Errors 76
Appendix D: Soware Change Log 78

xii
DSA3200-PTP
xii
Section 1: Specifications
[INTENTIONALLY LEFT BLANK]

1
DSA3200-PTP
1
Section 1: Specifications
DSA3200
DSA3217-PTP
Size (WxHxD) 2.86” x 3.40” x 7.08”
(7.26cm x 8.64cm x 17.98cm)
Weight 6.4lbs
(2.91kg)
Pressure Inputs: 16 (Standard)
8 (True Dierenal)
Input Fings:
Pressure Inputs .063” tubulaons
Control Inputs .063” tubulaons
Full Scale Ranges (standard)
Dierenal ±5”WC, ±10”WC,
±20”WC, 1 , 2.5, 5, 15,
30, 50, 100, 250, 500,
750 psid
(1.25, 2.5, 5, 7, 17, 35,
100, 210, 345, 700, 1725,
3500, 5170 kPa)
Absolute 15, 30, 50, 100 and
250 psia
(100, 210, 345, 700 and
1725 kPa)
Accuracy (aer calibraon)*
Dierenal
5 inH2O ±0.40% FS
10 inH2O ±0.20% FS
20 inH2O ±0.20% FS
1.0 psid ±0.12% FS
2.5 psid ±0.08% FS
5 to 500 psid ±0.05% FS
501 to 750 psid ±0.08% FS
Absolute (With CALB performed)
15 psia ±0.05% FS
30 psia ±0.05% FS
50 psia ±0.05% FS
100 psia ±0.05% FS
250 psia ±0.05% FS
*Some combinaons of pressure ranges in dual-range
modules may dictate a reducon in accuracy
Overpressure Capacity (No damage)
5 inH2O 2 psi
10 inH2O 2 psi
20 inH2O 2 psi
1 psid 5 psi
2.5 to 499 psid 200%
500 psid 150%
750 psid 113%
Maximum Reference Pressure 250 psid (1750 kPa) or
sensor overpressure,
which ever is less
Power Requirements 20 to 36 Vdc at 8 W
I/O and mang connectors
Ethernet 100Base-T RJ45
Trigger (bulkhead) PT02A-8-2P
Trigger (mang) PT06A-8-2S-SR
Trigger (bulkhead) JTP02RE8-6P
Trigger (mang) JTO1RE8-6S-SR
Power (bulkhead) PTO2A-8-3P
Power (mang) PTO6A-8-3S-SR
Communicaons Protocol Ethernet IEEE802.3
Typical Data Acquision Rate*
TCP/IP ASCII 50 samples/chan/sec
UDP Binary 850 samples/chan/sec
*Maximum data output depends on network
Typical Communicaons Rate
Ethernet 100 Mbits/sec
RS232 9600 BAUD
(Conguraon only)
Operang Temperature 0 to 60 °C
Compensated Range 0 to 72 °C
Total Thermal Error Less than ±0.001% FS
Humidity
Operaon 5 to 95% RH, Non-Condensing
Storage 5 to 95% RH, Non-Condensing
Shock & Vibraon MIL-STD-810D Curve H
Shock 10G
Vibraon 10G
Acceleraon 10G
Media Gasses compable with
Silicon, Silicone,
Aluminum and Stainless
Steel
Precision Time Protocol IEEE-1588 PTPv2

2
DSA3200-PTP
2
Section 1: Specifications
DSA3218-PTP
Size (WxHxD) 4.08” x 6.97” x 9.0”
(10.36 cm x 17.70 cm x 22.86 cm)
Weight 9.78 lbs
(4.45 kg)
Pressure Inputs: 16 (Standard)
8 (True Dierenal)
Input Fings:
Pressure Inputs 1/8” SwageLok®
(1/16” and 1/4” oponal)
Control Inputs 1/8” SwageLok®
(1/16” and 1/4” oponal)
Full Scale Ranges
Dierenal ±5”WC, ±10”WC,
±20”WC, 1 , 2.5, 5, 15,
30, 50, 100, 250, 500,
750 psid
(1.25, 2.5, 5, 7, 17, 35,
100, 210, 345, 700, 1725,
3500, 5170 kPa)
Absolute 15, 30, 50, 100 and
250 psia
(100, 210, 345, 700 and
1725 kPa)
Accuracy (aer calibraon)*
Dierenal
5 inH2O ±0.40% FS
10 inH2O ±0.20% FS
20 inH2O ±0.20% FS
1.0 psid ±0.12% FS
2.5 psid ±0.08% FS
5 to 500 psid ±0.05% FS
501 to 750 psid ±0.08% FS
Absolute (With CALB performed)
15 psia ±0.05% FS
30 psia ±0.05% FS
50 psia ±0.05% FS
100 psia ±0.05% FS
250 psia ±0.05% FS
*Some combinaons of pressure ranges in dual-range modules
may dictate a reducon in accuracy
Overpressure Capacity (No damage)
5 inH2O 2psi
10 inH2O 2psi
20 inH2O 2psi
1.0 psid 5psi
2.5 to 499 psid 200%
500 psid 150%
750 psid 113%
Maximum Reference Pressure 250 psid (1750 kPa) or
sensor overpressure,
which ever is less
Power Requirements
No Heater 20 to 36 Vdc at 8 W
With Heater 20 to 36 Vdc at 40 W
(Drops to 8 W aer
module has reached
temperature)
I/O and mang connectors
Ethernet 100Base-T D-Code M12
Trigger (bulkhead) PT02A-8-2P
Trigger (mang) PT06A-8-2S-SR
Trigger (bulkhead) JTP02RE8-6P
Trigger (mang) JTO1RE8-6S-SR
Power (bulkhead) PTO2A-8-3P
Power (mang) PTO6A-8-3S-SR
Communicaons Protocol Ethernet IEEE802.3
Typical Data Acquision Rate*
TCP/IP ASCII 50 samples/chan/sec
UDP Binary 850 samples/chan/sec
*Maximum data output depends on network
Typical Communicaons Rate
Ethernet 100 Mbits/sec
RS232 9600 BAUD
(Conguraon only
Operang Temperature
Without Heater 0 to 60 °C
With Heater -55 to 60 °C
(Cooling kit recommended above 55 °C)
Compensated Range 0 to 72 °C
Total Thermal Error Less than ±0.001% FS
Humidity
Operaon 5 to 95% RH, Non-Condensing
Storage 5 to 95% RH, Non-Condensing
Shock & Vibraon MIL-STD-810D Curve H
Shock 10G
Vibraon 10G
Acceleraon 10G
Media Gasses compable with Silicon,
Silicone, Aluminum and Stainless
Steel
Precision Time Protocol IEEE-1588PTP v2

3
3
Section 1: Specifications
DSA3200-PTP

4
DSA3200-PTP
4
Section 1: Specifications

5
DSA3200-PTP
5
Section 2: Introduction
The Digital Sensor Array (DSA) is a stand-alone electronic
pressure scanner which can accept up to 16 pneumac
inputs. The module is specically designed for use in appli-
caons where long calibraon intervals, simple operaon
and high accuracy are required. Each DSA-PTP incorporates
16 individual, temperature compensated, piezoresisve
pressure sensors, an A/D converter and a microproces-
sor. Each pressure sensor is characterized over pressure
and temperature. This informaon is stored in EEPROM
memory and enables the module to directly output cor-
rected data in engineering units.
The sensors within each DSA3200-PTP series module are
arranged in groups of eight (8). Each block of eight (8)
sensors has its own calibraon valve. This calibraon valve
has four (4) modes of operaon and allows the DSA-PTP
module to perform measurement, calibraon and purge
funcons. The four (4) modes are:
• Measure
• Calibrate
• Isolate
• Purge
The calibraon valve mode is selected by applying 90-120
psi control pressure conguraons unique for each
operaon mode. The DSA-PTP calibraon valve ulizes
“Normally Px” valve logic where the valve defaults to
the measurement mode when no control pressures are
applied.
The DSA3200-PTP series is available in two basic congu-
raons; the DSA3217-PTP and the DSA3218-PTP. Both
conguraons oer 16 pressure inputs, both operate using
the same soware commands and both oer corrected
data output directly in engineering units. The dierence
between the modules is the packaging and the intended
applicaons. The DSA3217-PTP is enclosed in a stain-
less steel, splash resistant enclosure intended for most
laboratory, educaonal or controlled environments. The
DSA3218-PTP however, is enclosed in a stainless steel,
splash resistant, shock mounted enclosure with Swagelok®
ngs on all pneumac connecons. It is intended for
applicaons such as gas turbine tesng, ight tesng,
industrial applicaons and other more rugged environ-
ments.
In addion to the more rugged casing on the DSA3218-PTP
modules, addional opons of a heater or a cooling kit are
oered. The heater opon allows the module to be oper-
ated below the 0 °C temperature compensaon range of
the standard DSA3218-PTP or DSA3217-PTP modules. The
cooling kit is recommended for applicaons where the
ambient environment temperatures are likely to exceed 55
°C.
Both DSA3217-PTP and DSA3218-PTP modules are avail-
able with a variety of sensor conguraons. The standard
conguraon is 16 channels of the same pressure range,
however modules can be congured as dual-range units
with 8 channels of one pressure range and 8 channels of
another pressure range, ‘True-Dierenal’ measurement
modules which provide 8 channels of dierenal pressure
measurement, absolute pressure measurement modules
or units with individual reference ports for each sensor
(DSA3217-PTP only).
The DSA3217-PTP is available in 5 basic conguraons.
The basic DSA3217-PTP/16Px module provides 16 chan-
nels of the same ‘gauge’ pressure range sensors. These 16
pressure transducers are congured into two (2) groups of
eight (8) each. Each group of eight (8) transducers shares a
common calibraon valve and a common reference mani-
fold. Single range modules e the two reference manifolds
together and the two calibraon valves to provide the
user with a single reference and a single calibraon port to
connect to. Figure 2.1 depicts the top panel of a standard
DSA3217-PTP/16Px.

6
DSA3200-PTP
6
Section 2: Introduction
The DSA3217-PTP/16Px module is available as a dual-range
module, oering eight (8) channels of one pressure range
and eight (8) channels of a dierent pressure range. Like a
standard DSA3217-PTP, each group of eight (8) transducers
share a common calibraon valve and a common reference
manifold. However, in dual range modules the reference
manifolds and the calibraon valves are not ed together
and separate reference and calibraon ports are provided
on the top of the module. They are labeled CAL, CAL B, REF
and REF B. Figure 2.2 depicts the top panel of a dual range
DSA3217-PTP/16Px module.
For dierenal pressure measurement applicaons, the
DSA3217-PTP can be congured with eight (8) dieren-
al pressure transducers. They are arranged into two (2)
groups of four (4) transducers. Like the DSA3217-PTP/16Px,
each group of transducers has a dedicated calibraon valve
that allows the input pressures to be blocked o during any
calibraon, including a zero oset calibraon (CALZ). Figure
2.3 depicts the top panel of a DSA3217-PTP/8DPx module.
The DSA3217-PTP/16Px module can be used for absolute
pressure measurement when equipped with absolute pres-
sure transducers. A calibraon valve is provided for each
group of eight (8) sensors. All sixteen (16) transducers are
ed a single calibraon port, except in the cases of dual-
range absolute modules. Figure 2.4 depicts the top panel
of a DSA3217-PTP/16Px absolute pressure module.

7
DSA3200-PTP
7
Section 2: Introduction
Both the standard DSA3217-PTP/16Px and Dual Range
DSA3217-PTP/16Px are available with individual reference
ports for each sensor. A calibraon valve is provided for
each group of eight (8) sensors, however unlike the stan-
dard DSA3217-PTP/16Px modules a separate reference port
is provided for all sixteen (16) transducers. These indi-
vidual reference ports are not controlled by the calibraon
valve.
This means that any calibraon, including a zero oset
calibraon (CALZ) will have to be done with no pressure
applied to the module. The individual reference modica-
on is depicted in Figure 2.5 below.

8
DSA3200-PTP
8
Section 2: Introduction
The DSA3218-PTP is available in 4 basic pressure congura-
ons and has 2 oponal hardware conguraons.
The basic DSA3218-PTP/16Px module provides 16 chan-
nels of the same ‘gauge’ pressure range sensors. These 16
pressure transducers are congured into two (2) groups of
eight (8) each. Each group of eight (8) transducers shares a
common calibraon valve and a common reference mani-
fold. Single range modules e the two reference manifolds
together and the two calibraon valves to provide the
user with a single reference and a single calibraon port to
connect to. Figure 2.8 depicts the top panel of a standard
DSA3217-PTP/16Px.
The DSA3218-PTP/16Px module is available as a dual-range
module, oering eight (8) channels of one pressure range
and eight (8) channels of a dierent pressure range. Like a
standard DSA3218-PTP, each group of eight (8) transducers
share a common calibraon valve and a common reference
manifold. However, in dual range modules the reference
manifolds and the calibraon valves are not ed together
and separate reference and calibraon ports are provided
on the top of the module. They are labeled CAL A, CAL B,
REF A and REF B. Figure 2.9 depicts the top panel of a dual
range DSA3218-PTP/16Px module.
For dierenal pressure measurement applicaons, the
DSA3218-PTP can be congured with eight (8) dieren-
al pressure transducers. They are arranged into two (2)
groups of four (4) transducers. Like the DSA3218-PTP/16Px,
each group of transducers has a dedicated calibraon valve.
Figure 2.10 depicts the top panel of a DSA3218-PTP/8DPx
module.
-
This manual suits for next models
5
Table of contents
Other Scanivalve Scanner manuals
Popular Scanner manuals by other brands

Symbol
Symbol LS 5700 Product reference guide

ScanSnap
ScanSnap P3PC-7032-01XA Getting started

Design Technology
Design Technology Tiny-Scan 301 Quick start instructions

Avision
Avision BT-1604B user manual

Canon
Canon DR 7090C - imageFORMULA - Document Scanner reference guide

SICK
SICK microScan3 Core I/O AIDA operating instructions