Schaller Messtechnik humimeter.com PMSA User manual

User manual
Moisture analyser for single paper sheets
PMSA
version 1.3_en
© Schaller GmbH
2021
page 2 version 1.3_en
Design of device
Infrared temperature
sensor
Lid push button
Measuring plate
LED status bar
USB connection
Reset Button
Serial number
version 1.3_en page 3
General information
The PMSA (paper moisture single sheet analyzer) is a measuring
system for determining the water content of single sheets of paper.
The system consists of a moisture sensor and an analyzing
software for a PC.
Software start-up
The delivery contains an USB flash drive with the analysing
software PMSA. By double-clicking the file „PMSA_setup.exe“ the
installer is started. With the installation, the USB driver (FTDI) is
installed automatically. Your PC needs that driver for PMSA
connection. After that the software can be started.
Hardware start-up
The measuring device has to be connected to your PC via USB
mini B connector. The driver for the virtual COM port has been
installed together with the software and should be ready for use.
Setting the correct COM port
Clicking on the button „Settings“(USB connector symbol), a
window for selecting the COM port opens.
After choosing the correct COM port,
the connection between hardware
and software can be tested by
clicking the button „Test“. If the test
was successful, this is shown by a
popup window and by the LED status
bar shining in blue. Then click the
“Save” button.
page 4 version 1.3_en
Measuring procedure
After the connection has been established as described, the
measurement can be started.
First of all choose the corresponding characteristic curve for the
paper to measure. Clicking the arrow in the field Characteristic, a
list with all available calibration curves opens. Select the desired
curve by clicking on it.
For creating a new characteristic curve for your paper there is
another manual available.
Now press the button Start. The device effects a blank measure-
ment, executing a self-check and adjusting to the surrounding
version 1.3_en page 5
conditions. This procedure ensures long-time stable measurement
values. For this blank measurement, no paper sheet must be
situated in the device!
Upon successful completion of the blank measurement, put the
paper sheet into the device and press OK.
Now the measuring procedure is
started. The ongoing measure-
ment is shown by the blue LED
blinking. When the measurement
is completed, a window with the
measuring results appears,
offering the possibility of adding
additional information for meas-
ured paper.
page 6 version 1.3_en
If desired, the popup window „Meas-
urement successful“ can be hidden.
For that, remove the respective tick in
the tab „Common“ of the sensor
settings.
NOTE:
In this mode the measuring data is
NOT SAVED in the archive!!
Database / Archive
In this window all saved measuring values are viewed. For each
measurement a separate identification number is allocated.
version 1.3_en page 7
Filter
The filter functions allow an easy retrieval of measurements in the
database. If you leave archive, all filters will be reset.
Date filter:
With that function you can chose a period. All saved measured in that
period will be shown.
For selection of that time, you have to use two arrays:
In the first field enter the start date.
In the second field the end date is entered.
page 8 version 1.3_en
Filter by entered data:
Additionally, the filter by specific data like entered paper type, charge
number or grammage is possible. For that, click the button with the
magnifying glass, select one of the filter functions and enter the desired
search term.
The date filter and filter by entered data can also be combined.
By clicking on the fields ID, charge number ecc., the list is sorted in
ascending or descending order.
version 1.3_en page 9
Further database functions
All data in the database can be deleted, edited or exported.
Delete measurements:
Clicking the waste bin button, the selected data series are deleted
irretrievably.
The selection functions are the same as in Excel. For selecting a
measuring series, click the arrow field in front of the field ID, for
selecting a certain sector, keep the Shift button pressed. For the
selection of single measuring series, keep the Strg button pressed.
Editing measuring data:
page 10 version 1.3_en
Using the pencil button, saved measurements can be modified. Just
select the desired measuring series and click the pencil button.
Now paper type and charge number can be modified. Confirm by
clicking the Save button.
Print measurements:
Clicking on the printer symbol, all measuring series shown in the list
below are printed out. If a filter is set, only the filtered data is printed.
version 1.3_en page 11
Export measurements:
Clicking the Export button, the export of CSV, HTML or XML formats is
possible. All measuring data shown in the list below is exported. If a
filter is set, only the filtered data is exported.
Determination of reference water content
The PMSA determines the water content, which means that it
calculates the moisture referred to the total mass (EN ISO 287).
100% 
−
=MnMtMn
WG
Exemption from liability
For misreading’s and wrong measurements and of this resulting
damage we refuse any liability. This is a device for quick determi-
nation of moisture. The moisture depends on multiple conditions
and multiple materials. Therefore, we recommend a plausibility
check of the measuring results. Each device includes a serial
number and the guarantee stamp. If those are broken, no claims
for guarantee can be made. In case of a faulty device, please
contact Schaller GmbH (www.humimeter.com) or your dealer.
Mn: Mass of the sample before drying
Mt: Mass of the dried sample
%WG: Calculated water content
page 12 version 1.3_en
Device maintenance instructions
•Please note that the moisture sensor is a sensitive measuring
instrument and handle it with care.
•The white sensor plate (measuring plate) most not be exposed to
mechanical stress. Heavy mechanical load may cause the break-
ing of the sensor plate. In this case, return the sensor to Schaller
GmbH. The breaking of the sensor plate is not subject to warranty.
•If the device is transported, secure the lid to prevent unintentional
opening.
•We recommend to regularly save all measuring data in the data-
base.
•Furthermore we recommend to regularly save the database with
eventual added or modified calibration curves.
Technical data
Grammage range
30 to 800 g/m²
Paper size
minimum 180x180mm
Paper thickness
up to 0.8mm (higher range on request)
Resolution
0.01% water content; 0.1°C temperature
Measuring range
1 to 20 % water content (depending of
characheristics curve)
Operating temperature
+10°C to +40°C / 50°F to 104°F
Storage temperature
-20°C to +60°C / -4°F to 140°F
Temperature compensation
automatic
Menu languages
English, German
Power supply
Power supply via PC 5VDC (USB
connector)
Dimensions
280 x 200 x 135 mm
Weight
4.8 kg
Protection class
IP 30
Scope of supply
moisture sensor, PC analysing software,
USB connection cable for PC

Other Schaller Messtechnik Measuring Instrument manuals

Schaller Messtechnik Humimeter BMA User manual

Schaller Messtechnik

Schaller Messtechnik Humimeter BMA User manual

Schaller Messtechnik humimeter FS1 User manual

Schaller Messtechnik

Schaller Messtechnik humimeter FS1 User manual

Schaller Messtechnik Humimeter AW3 User manual

Schaller Messtechnik

Schaller Messtechnik Humimeter AW3 User manual

Schaller Messtechnik Humimeter BM1 User manual

Schaller Messtechnik

Schaller Messtechnik Humimeter BM1 User manual

Schaller Messtechnik humimeter GE1 User manual

Schaller Messtechnik

Schaller Messtechnik humimeter GE1 User manual

Popular Measuring Instrument manuals by other brands

Mirion Technologies SOR Quick user guide

Mirion Technologies

Mirion Technologies SOR Quick user guide

IFM PG243 operating instructions

IFM

IFM PG243 operating instructions

PCB Piezotronics ICP 352C66 Installation and operating manual

PCB Piezotronics

PCB Piezotronics ICP 352C66 Installation and operating manual

Sontex Superstatic 440 Installation and manual

Sontex

Sontex Superstatic 440 Installation and manual

Gas Data GFM 436 user manual

Gas Data

Gas Data GFM 436 user manual

GW Instek GRF-1300 user manual

GW Instek

GW Instek GRF-1300 user manual

koban KP-05 instruction manual

koban

koban KP-05 instruction manual

Actel Silicon Explorer quick start

Actel

Actel Silicon Explorer quick start

Laserworld RGB-1.0 manual

Laserworld

Laserworld RGB-1.0 manual

Interacoustics AC40 Instructions for use

Interacoustics

Interacoustics AC40 Instructions for use

IFM SV0527 operating instructions

IFM

IFM SV0527 operating instructions

RND 365-00002 user manual

RND

RND 365-00002 user manual

Ohm-Labs 100 Series instruction manual

Ohm-Labs

Ohm-Labs 100 Series instruction manual

Bender RCMA421H manual

Bender

Bender RCMA421H manual

MAHA LPS 3000 operating instructions

MAHA

MAHA LPS 3000 operating instructions

RDS Technology Master Driers MMM 100 Installation, Calibration and Operation

RDS Technology

RDS Technology Master Driers MMM 100 Installation, Calibration and Operation

BRUEL & KJAER 2240 instruction manual

BRUEL & KJAER

BRUEL & KJAER 2240 instruction manual

Etymotic Research ER-200 user guide

Etymotic Research

Etymotic Research ER-200 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.