Schmalz FS-5-D-2NA-K User manual

Flow Sensor FS
Assembly Instructions
WWW.SCHMALZ.COM EN-US · 30.30.01.02533 · 00 · 11/20

Note
The Operating instructions were originally written in German. Store in a safe place for future reference.
Subject to technical changes without notice. No responsibility is taken for printing or other types of er-
rors.
Published by
© J.SchmalzGmbH, 11/20
This document is protected by copyright. J.SchmalzGmbH retains the rights established thereby. Repro-
duction of the contents, in full or in part, is only permitted within the limits of the legal provisions of
copyright law. Any modifications to or abridgments of the document are prohibited without explicit writ-
ten agreement from J.SchmalzGmbH.
J. Schmalz GmbH · Johannes-Schmalz-Str. 1 · 72293 Glatten, Germany · T: +49 7443 2403-0
2 / 42 EN-US · 30.30.01.02533 · 00 · 11/20

Contents
EN-US · 30.30.01.02533 · 00 · 11/20 3 / 42
Contents
1 Important Information ...................................................................................................................................5
1.1 Note on Using this Document ............................................................................................................ 5
1.2 The technical documentation is part of the product ........................................................................ 5
1.3 Type Plate ............................................................................................................................................ 5
1.4 Warnings in This Document................................................................................................................ 5
1.5 Symbol.................................................................................................................................................. 6
2 Fundamental Safety Instructions...................................................................................................................7
2.1 Intended use ........................................................................................................................................ 7
2.2 Non-Intended Use ............................................................................................................................... 7
2.3 Safety Instructions ............................................................................................................................... 7
2.4 Personnel Qualification....................................................................................................................... 7
2.5 Modifications to the Product.............................................................................................................. 7
3 Product Description ........................................................................................................................................8
3.1 Designation code................................................................................................................................. 8
3.2 Variants and Type Key ........................................................................................................................ 8
3.3 Design of the flow sensor ................................................................................................................... 9
3.4 Display and Operating Element in Detail .......................................................................................... 9
3.5 Function Instuction............................................................................................................................ 11
3.5.1 Function Setting Mode..............................................................................................................11
3.5.2 Measurement............................................................................................................................. 13
4 Technical Data ...............................................................................................................................................14
4.1 General Parameters........................................................................................................................... 14
4.2 Dimensions......................................................................................................................................... 16
5 Installation.....................................................................................................................................................17
5.1 Installation Instructions..................................................................................................................... 17
5.2 Mounting Bracket / Optional Parts .................................................................................................. 18
5.3 Piping ................................................................................................................................................. 19
5.4 Electrical Connection......................................................................................................................... 20
6 Operation Instructions..................................................................................................................................22
6.1 Function Selection Mode .................................................................................................................. 22
6.2 OUT1 Setting Selection ..................................................................................................................... 22
6.3 OUT2 Setting Selection ..................................................................................................................... 26
6.4 LCD Display Color Selection .............................................................................................................. 26
6.5 Response Time Selection................................................................................................................... 27
6.6 Display Refresh Time Selection......................................................................................................... 28
6.7 Unit Selection .................................................................................................................................... 29
6.8 Flow Reference Standard Selection.................................................................................................. 29
6.9 Analog Output Selection .................................................................................................................. 30
6.10 Accumulated Value Hold Selection .................................................................................................. 30
6.11 Flow Sensor Display Mode Selection................................................................................................ 31
6.12 Power-Save Mode Selection ............................................................................................................. 31
6.13 External Input Selection.................................................................................................................... 31
6.14 Fine Adjustment Setting ................................................................................................................... 32

Contents
4 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
6.15 Forced Output Function.................................................................................................................... 33
6.16 Reset to the Default Setting ............................................................................................................. 34
6.17 Pressure Zero Adjustment Function ................................................................................................. 34
6.18 Instantaneous Flow Zero Adjustment Function .............................................................................. 34
6.19 Reset Accumulated Flow Function ................................................................................................... 35
6.20 Peak Value Display ............................................................................................................................ 35
6.21 Bottom Value Display........................................................................................................................ 35
6.22 Key Lock / Unlock Mode.................................................................................................................... 36
7 Warranty........................................................................................................................................................37
8 Troubleshooting............................................................................................................................................38
8.1 Specific Internal Voltage drop.......................................................................................................... 38
8.2 Error Code Instruction....................................................................................................................... 38
9 Maintenance Precautions .............................................................................................................................39
10 EC Declaration of Conformity ......................................................................................................................40
11 Notes..............................................................................................................................................................41

Important Information
EN-US · 30.30.01.02533 · 00 · 11/20 5 / 42
1Important Information
1.1Note on Using this Document
J.SchmalzGmbH is generally referred to as Schmalz in these Operating instructions.
These Operating instructions contain important notes and information about the different operating
phases of the product:
• Transport, storage, start of operations and decommissioning
• Safe operation, required maintenance, rectification of any faults
The Operating instructions describe the product at the time of delivery by Schmalz.
1.2The technical documentation is part of the product
1. For problem-free and safe operation, follow the instructions in the documents.
2. Keep the technical documentation in close proximity to the product. The documentation must be ac-
cessible to personnel at all times.
3. Pass on the technical documentation to subsequent users.
ðFailure to follow the instructions in these Operating instructions may result in injuries!
ðSchmalz is not liable for damage or malfunctions that result from failure to heed these instructions.
If you still have questions after reading the technical documentation, contact Schmalz Service at:
www.schmalz.com/services
1.3Type Plate
The type plate (1) is printed to the Flow Sensor and must always be clearly legible.
It includes the following information:
• Name
• Article/part number
• Pressure Range
• Flow Range
• Wire Color and Function
• Lot No
• QR code
• CE label
1
1.4Warnings in This Document
Warnings warn against hazards that may occur when handling the product. The signal word indicates the
level of danger.
Signal word Meaning
WARNING Indicates a medium-risk hazard that could result in death or serious injury if
not avoided.
CAUTION Indicates a low-risk hazard that could result in minor or moderate injury if
not avoided.
NOTE Indicates a danger that leads to property damage.

Important Information
6 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
1.5Symbol
This symbol indicates useful and important information.
üThis symbol represents a prerequisite that must be met prior to an operational step.
4This symbol represents an action to be performed.
ðThis symbol represents the result of an action.
Actions that consist of more than one step are numbered:
1. First action to be performed.
2. Second action to be performed.

Fundamental Safety Instructions
EN-US · 30.30.01.02533 · 00 · 11/20 7 / 42
2Fundamental Safety Instructions
2.1Intended use
The Vacuum/pressure switch is used for measuring and displaying folw or pressure and outputs it propor-
tionally as an electrical voltage from 1 to 5V on an analog output.
This device has been designed, developed and constructed solely for industrial and commercial use. Pri-
vate use is excluded.
The Vacuum/pressure switch is built in accordance with the latest standards of technology and is shipped
in safe condition. However, hazards can arise during use.
Intended use includes the observance of the technical data and the installation and operating instructions
in this manual.
2.2Non-Intended Use
Schmalz accepts no liability for damages caused by the use of the Vacuum/pressure switch for purposes
other than those described under "Intended Use."
Non-intended use includes the following:
• Use in potentially explosive atmospheres
2.3Safety Instructions
WARNING
Use of flammable fluids and/or permeable fluids
They may cause fire, explosion or corrosion.
4Do not use flammable fluids and/or permeable fluids.
2.4Personnel Qualification
Unqualified personnel cannot recognize dangers and are therefore exposed to higher risks!
1. Only instruct qualified personnel to perform the tasks described in these operating instructions.
2. The product may only be operated by persons who have undergone appropriate training.
3. Electrical work and installations may only be carried out by qualified electrical specialists.
4. Assembly and maintenance work must only be carried out by qualified personnel.
2.5Modifications to the Product
Schmalz assumes no liability for consequences of modifications over which it has no control:
1. The product must be operated only in its original condition as delivered.
2. Use only original spare parts from Schmalz.
3. The product must be operated only in perfect condition.

Product Description
8 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
3Product Description
3.1Designation code
The breakdown of the item designation (e.g. FS-10-D-2PA-K) is as follows:
Feature Variants
Type FS
Flow Rate Range
05: 5L/min
10: 10L/min
50: 50L/min
100: 100L/min
200: 200L/min
Display D
Output Specification 2PA: 2x PNP + 1x 1...5 V
2NA: 2x NPN + 1x 1...5 V
Electrical connection K: Cable
3.2Variants and Type Key
Part no.
10.06.04...
Designation Size Connection Pneumatic
connection
...00001 FS-5-D-2NA-K 0 ... 5 L/min 2 NPN output + Analog output 1 ~ 5 V Thread G1/8“
...00002 FS-5-D-2PA-K 0 ... 5 L/min 2 PNP output + Analog output 1 ~ 5 V Thread G1/8“
...00003 FS-10-D-2NA-K 0 ... 10 L/min 2 NPN output + Analog output 1 ~ 5 V Thread G1/8“
...00004 FS-10-D-2PA-K 0 ... 10 L/min 2 PNP output + Analog output 1 ~ 5 V Thread G1/8“
...00005 FS-50-D-2NA-K 0 ... 50 L/min 2 NPN output + Analog output 1 ~ 5 V Thread G1/8“
...00006 FS-50-D-2PA-K 0 ... 50 L/min 2 PNP output + Analog output 1 ~ 5 V Thread G1/8“
...00007 FS-100-D-2NA-K 0 ... 100 L/min 2 NPN output + Analog output 1 ~ 5 V Thread G1/4“
...00008 FS-100-D-2PA-K 0 ... 100 L/min 2 PNP output + Analog output 1 ~ 5 V Thread G1/4“
...00009 FS-200-D-2NA-K 0 ... 200 L/min 2 NPN output + Analog output 1 ~ 5 V Thread G1/4“
...00010 FS-200-D-2PA-K 0 ... 200 L/min 2 PNP output + Analog output 1 ~ 5 V Thread G1/4“

Product Description
EN-US · 30.30.01.02533 · 00 · 11/20 9 / 42
3.3Design of the flow sensor
12
3
4 5
6
8
9
1011
7
1 Lead Wire Connector 2 Connector
3SETTING BUTTON
Use for selection each mode and set value
confirmation
4 LCD Display
5UP-BUTTON
Use for mode selection, ON/OFF switching
and set value adjustment
6DOWN-BUTTON
Use for mode selection, ON/OFF switching
and set value adjustment
7 Identification Plate 8 Piping Port (Outlet side)
9 Through-Hole
Use to mount the product on a plate di-
rectly
10 Body
11 Piping Port (Inlet side)
3.4Display and Operating Element in Detail
1
2
345
6
7
8
9
10
1SETTING BUTTON 2 Pressure Symbol
3 Indicator for
Analog Signal, Output 1 and Output 2
4 Flow Symbol
5 Indicator for
Analog Signal, Output 1 and Output 2
6UP-BUTTON
7DOWN-BUTTON 8 Flow Unit Display Section
9 3 Color LCD Display 10 Pressure Unit Display Section

Product Description
10 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
Pressure definition with selected display unit depending on the vacuum switch version:
Selected unit Display definition for variant
VS-V...
Display definition for variant
VS-P10...
kPa 0.1 —
MPa — 0.001
kgf/cm20.001 0.01
bar 0.001 0.01
psi 0.01 0.1
inHg 0.1 —
mmHg 1 —

Product Description
EN-US · 30.30.01.02533 · 00 · 11/20 11 / 42
3.5Function Instuction
3.5.1Function Setting Mode
Function
Code
Item Default setting Explanation
[F-01]
[oVt1] OUT1 setting ovt 1
Select Output 1 corre-
sponding to flow sen-
sor or pressure sensor.
Set the flow rate or
pressure value to
switch ON/OFF.
[oVt1] OUT1 sensor
correspondence
[ ] OUT1 output
mode
xY5
[ovt1] OUT1 output
type
no
[Fl-1] OUT1 set value
input
50% of maximum measured flow rate
005 : 250 mL/min ' 010 : 500 mL/min
050 : 2.50 L/min ' 100 = 5.00 L/min
500 : 25.0 L/min ' 101 = 50.0 L/min
201 : 100 L/min
[Fx-1] OUT1 set value
input
60% of maximum measured flow rate
005 : 300 mL/min ' 010 : 600 mL/min
050 : 3.00 L/min ' 100 = 6.00 L/min
500 : 30.0 L/min ' 101 = 60.0 L/min
201 : 120 L/min
[F-02]
[oVt2] OUT2 setting ovt 2
Select Output 2 corre-
sponding to flow sen-
sor or pressure sensor.
Set the flow rate or
pressure value to
switch ON/OFF.
[oVt2] OUT2 sensor
correspondence
[ ] OUT2 output
mode
xY5
[ovt2] OUT2 output
type
no
[Fl-2] OUT2 set value
input
50% of maximum measured flow rate
005 : 250 mL/min ' 010 : 500 mL/min
050 : 2.50 L/min ' 100 = 5.00 L/min
500 : 25.0 L/min ' 101 = 50.0 L/min
201 : 100 L/min
[Fx-2] OUT2 set value
input
60% of maximum measured flow rate
005 : 300 mL/min ' 010 : 600 mL/min
050 : 3.00 L/min ' 100 = 6.00 L/min
500 : 30.0 L/min ' 101 = 60.0 L/min
201 : 120 L/min
[F-03] [CLor] LCD Display set-
ting
Clor Select back light color
and display mode.
[di5p] LCD Display cor-
responding to output
ovt1
[di5p] LCD Display
color selection
5og
[F-04] [rE5P] Response time
selection
rE5P Select the response
time for analog out-
put.
Pressure sensor: 2.5ms
~ 1500ms
Flow sensor: 50ms ~
1500ms
[rE5P] Flow/Pressure
sensor selection
[ ] Response time
selection
800(ms)

Product Description
12 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
[F-05] [vpda] Display refresh
time selection
vpda Display refresh cycle
can be set in 200ms,
500ms or 1000ms.
[vpda] Flow/Pressure
sensor selection
[vpda] Display refresh
time of flow sensor selec-
tion
500(ms)
[F-06] [vnit] Unit selection vnit Select the UNIT of
pressure / flow sensor.
[ ] Flow unit selec-
tion
[pre5]vnit
[F-07] [refe] Flow reference
standard selection
refe Select the flow value is
shown under standard
(ANR) or normal con-
dition (NOR).
[refe] Flow reference
standard selection
anr
[F-08] [ ang] Analog output
selection
ang Select the analog cor-
responding to pressure
or flow sensor.
[ ] Analog output
selection
[F-09] [eepr] Accumulated
value hold selection
eepr To save the last accu-
mulated flow value ev-
ery 2 or 5 minutes.
[eepr] Accumulated
value hold selection
off
[F-10] [ di5] Flow sensor dis-
play mode selection
di5 Select to display In-
stantaneous Flow or
Accumulated Flow
Mode.
[ d5p] Flow sensor dis-
play mode selection
in5
[F-91] [ eco] Power-Save
mode selection
eco Select if turn on
power-save mode to
reduce power con-
sumption
[ eco] Power-Save
mode selection
no
[F-92] [ inp] External input se-
lection
inp Select for Accumu-
lated flow rate zero
clear, Auto-Shift or
Auto-Shift zero.
[ inp] External input se-
lection
[F-93][ ] Modbus RTU set-
ting
Set ID number, baud
rate and transmission
format.
[ id] ID number setting 1
[ rat] Baud rate setting 96 (9600 Bd)
[ for] Transmission for-
mat setting
n81
[ tra] Communications
protocol setting
rtv
[F-94] [fine] Fine adjustment
Setting
fine The displayed value
can be adjusted
slightly.
[fine] Fine adjustment
Setting
off
[F-95] [fovt] Forced output
function
fovt To turn the analog
ON / OFF forcibly.

Product Description
EN-US · 30.30.01.02533 · 00 · 11/20 13 / 42
[ovt1] Forced output
function
off
[ovt2] Forced output
function
off
[F-99] [re5t] Reset to the de-
fault setting
re5t Return to the factory
default setting.
[ r5t] Reset to the de-
fault setting
off
3.5.2Measurement
Item Explanation
Pressure display Display pressure value.
Flow display Display instantaneous flow rate.
Accumulated flow rate display Display accumulated flow rate.
Pressure zero setting The displayed pressure value can be adjusted to "0".
Instantaneous Flow rate zero set-
ting
The displayed instantaneous flow rate value can be adjusted to
"0".
Accumulated flow rate zero clear The accumulated flow rate can be set to "0".
Peak value display The maximum pressure or instantaneous flow can be detected
when the power is supplied for a period.
Bottom value display The minimum pressure or instantaneous flow can be detected
when the power is supplied for a period.
Key lock/unlock mode To prevent errors occurring due to unintentional changes of the
set values.

Technical Data
14 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
4Technical Data
4.1General Parameters
Parameter Unit FS-5 FS-10 FS-50 FS-100 FS-200
Fluid Dry air, N2, Non-corrosive / Non-flammable gas
Sensor
Ele-
ment
Flow Measured flow
rate range
0 - 5 L/
min
0 - 10 L/
min
0 - 50 L/
min
0 - 100 L/
min
0 - 200 L/
min
Flow Direction Unidirection
Pressure Rated Pressure
Range
-90 ~ 800 kPa
Display 4 digital * 4 digital, 7 segment LCD display (Red / Green / Or-
ange)
Instant
Flow
Rate
Display Range 0 - 5.00 L/
min
0 -10.00
L/min
0 - 50.0 L/
min
0 - 100 L/
min
0 - 200 L/
min
Mini-
mum
Setting
Scale
LPM 0.01 L/min 0.1 L/min 1 L/min
CFM 1) 0.1 ft3/min 1 ft3/min
Accu-
mu-
lated
Flow
Display Range 999999.99 L 9999999.9 L 99999999 L
Minimum Setting
Scale 1)
0.01 L 0.1 L 1 L
0.1 ft31 ft3
Pressure
Display
Display Range -100 - 1000 kPa
Mini-
mum
Setting
Scale
kPa 1
kfg/cm20.01
bar 0.01
psi 0.1
Accu-
racy
Flow Guaranteed Range 2 - 100 % F.S.
Indicator Accuracy ± 3% F.S. ± 1 digit 2)
Analog Output Ac-
curacy
± 5% F.S. 2)
Repeatability ± 1% F.S. ± 1 digit 3)
Linearity ±3% F.S. 3)
Temp. Characteris-
tic
± 2% F.S. (15 - 35°C) ; ± 5% F.S. (0 - 15°C · 35 - 50°C) (compare
with 3))
Pressure Character-
istic
± 5% F.S. ± 1 digit 4)
Pressure Guaranteed Range 0 - 100 % F.S.
Indicator Accuracy ± 2% F.S. ± 1 digit 5)
Analog Output Ac-
curacy
± 2.5% F.S. 5)
Repeatability ± 0.2% F.S. ± 1 digit 5)
Linearity ±1% F.S. 5)
Temp. Characteris-
tic
± 2% F.S. (compare with 5))
Switch
Out-
put
2NPN: open collector 2 outputs
Max. Load Current: 125 mA
Max. Supply Voltage: 28 V DC
Voltage Drop: ≤ 1.5 V

Technical Data
EN-US · 30.30.01.02533 · 00 · 11/20 15 / 42
Parameter Unit FS-5 FS-10 FS-50 FS-100 FS-200
2PNP: open collector 2 outputs
Max. Load Current: 125 mA
Max. Supply Voltage: 24 V DC
Voltage Drop: ≤ 1.5 V
Response Time Flow 800 ms (50 ms, 80 ms, 120 ms, 200 ms, 400 ms, 1500 ms se-
lectable)
Pres-
sure
2.5 ms (25 ms, 100 ms, 250 ms, 500 ms, 1000 ms, 1500 ms se-
lectable)
Output Mode Flow Hysteresis Mode, Window Comparator Mode, Accumulated
Output, Accumulated Pulse Output
Pres-
sure
One Point Set Mode, Hysteresis Mode, Window Comparator
Mode
Hysteresis Adjustable
Output Short Circuit Proterct-
lon
Yes
Accumulated Pulse Output 1) 0.05 L/
Pulse
0.1 L/
Pulse
0.5 L/
Pulse
1 L/Pulse 2 L/Pulse
0.2 ft3/
Pulse
0.4 ft3/
Pulse
2 ft3/Pulse 4 ft3/Pulse 7 ft3/Pulse
Ana-
log
Out-
put
Voltage Output Voltage Output Range: 1 - 5 V 6)
Output Impedance: 1 kΩ
Current Output Voltage Output Range: 4 - 20 mA 6)
Load Impedance: ≤ 300 Ω
Response Time Pressure: ≤ 50 ms ; Flow: ≤ 100 ms
External Input Non-voltage input, < 0.4 V , ≥ 30 ms
Communication interface RS-485 7)
Power Power Supply Voltage 12 - 24V DC ±10% · Ripple (P-P) ≤ 10%
Current Consumption ≤ 50 mA
Envi-
ron-
ment
Withstand Pressure 1000 kPa
Enclosure IP40
Working Fluid Temp. 0 - 50°C (No condensation or freezing)
Ambient Temp. Range Operation: 0 - 50°C ; Storage: -10 - 60°C (No condensation or
freezing)
Ambient Humidity Range Operation / Storage: 35 - 85 % R.H. (No condensation)
Insulation Resistance > 50 MΩ ( 500 V DC , between case and lead wire )
Withstand Voltage 1000 V AC 1 -min ( between case and lead wire )
Vibration Total amplitude 1.5 mm or 10 G, 10 Hz - 55 Hz - 10 Hz scan
for 1 minute, 2 hours each direction of X, Y and Z
Shock 100 m/s2 (10 G), 3 times each in direction of X, Y and Z
EMC IEC 61000-6-2, IEC 61000-6-4
Lead Wire Ø4 Oil-resistance cable - 26 AWG (0.15 mm2) - 6 cores
Port
Size
G1/8 Internal Thread — —
G1/8 Internal Thread — — —
Weight (with 2 Meter Lead Wire) Approx. 107 g (G1/8-IG) ; Approx. 110.5 g (G1/4-IG)
Accumulated Pulse Output 1)
NOTE:

Technical Data
16 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
1) CFM (ft3/min * 10-2) and ft3 *10-2.5) Outlet flow rate = 0 L/min, 25 °C.
2) CONDITION : Inlet Pressure: 300 kPa , Outlet
Pressure: 1 atmospheric pressure, 25 °C.
6) PWM output, corresponding to pressure sensor 0
- 1000 kPa.
3) CONDITION: Outlet Pressure: 1 atmospheric pres-
sure, 25 °C.
7) This function only available for Output Specifica-
tion -02 and -04.
4) -90 - 800 kPa, Outlet Pressure: 1 atmospheric
pressure, 25 °C.
4.2Dimensions
Lk SW1 L1 X1 d1 d2 Y1 B H H1
2.000 17 84,2 28 2,5 3 8 17 50 52
H2 H3 H4 L L2 L3 G1 G2 X2
3 9,5 36,7 55 18,5 18,6 G1/4"-IG G1/4"-IG 18
All data in mm.

Installation
EN-US · 30.30.01.02533 · 00 · 11/20 17 / 42
5Installation
5.1Installation Instructions
• Do not use near a surge voltage generated area. Solenoid lifters, high frequency Induction furnaces
and motors, etc. can generate high surge voltages, if using near the sensor will cause the internal
circuit components to deteriorate and cause damages.
• Sensors can not withstand lightning strikes. The product is CE compliant, but can not resist surge
voltage of lightning strikes, take measures to avoid lightening strikes in the system.
• Do not use in an environment where sensors could be splashed by water or oil. Enclosure rating is
IP40, please avoid water or oil splashed environment to prevent adversely effects.
• Do not use in an environment subject to large temperature cycling. Internal components of the sen-
sor will be affected adversely by large heating/cooling cycles other than ordinary changes in temper-
ature.
• Do not mount the product in locations where it is exposed to radiant heat. This could result in dam-
age and/or malfunction.
• Ensure the flow direction of the fluid. Follow the flow direction indicator for installation and piping.
• Flush out all dirt and dust by air blow before connect the piping to the sensor.
• When installation, do not drop, hit or apply excessive shock (100m/s2). Internal damage can cause
malfunction even if the housing appears to be undamaged.
• Do not install multiple products in close proximity. The heat generated from each product could
cause the temperature to rise and change the characteristics of product or deterioration of the plas-
tic parts. Please position the products 10mm apart from each other.
• Hold the sensor body when installing. The tensile strength of the cable is 24.5N and apply excessive
pulling force can cause damage to the sensor.
NOTE
The pressure or flow rate exceeded the specified range
The sensing element may be damaged
4Check the regulator and flow adjustment valve before introducing the fluid.
The sensing element cannot measure properly if foreign matter adheres to it.
On the inlet side, be sure to install an air filter below the filtration level of 10µm.
Recommended Equipments and Installation
Inlet Outlet
FS
Flow Sensor
Pressure Sensor
Air Source Air Dryer Filter Regulator Micro Mist
Separator
(Oil Mist Filter)
When measuring the pressure of the inlet side, install a throttle valve or solenoid valve on the
outlet side.

Installation
18 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
When measuring the pressure of the outlet side, install a throttle valve or solenoid valve on
the inlet side.
5.2Mounting Bracket / Optional Parts
Do not mount the sensor in a place that will be used as a foothold. The product may damage if sit or step
on it accidentally.
The LCD display may be difficult to see at certain angles.
The sensor can be installed horizontally or vertically, but the flow rates may change because of the instal-
lation way of the product or piping.
The tightening torque for screws should be under 0.5 ±0.1Nm.
Horizontal mounting (by Through-Hole)
4Mounting screws
Bracket (no. 10.06.04.00011) mounting
1. Fix the bracket to the flow sensor. Self-Tap-
ping-Screw: P Type, size M3.0, length L =
6mm.

Installation
EN-US · 30.30.01.02533 · 00 · 11/20 19 / 42
2. Mounting screws.
5.3Piping
Install the pipe by following the arrow indication that shows the air flow direction on the product.
Flow direction
Use straight piping 8cm or longer to connect the Piping Port (Inlet side). If straight piping is not installed,
the accuracy may vary by ± 2% F.S..
Straight Piping: The pipe is without bending and the cross sectional areas of the pipe keeps
the same.
Piping Port
(Inlet side)
straight piping
8 cm or longer
Piping Port
(Outlet side)

Installation
20 / 42 EN-US · 30.30.01.02533 · 00 · 11/20
Blow the air to flush out the foreign matters, dust and
etc. before installing the pipe. Uncleaned air may cause
malfunction or damage to the product.
When mounting the fitting, a wrench should be used
on the metal part.
Using on other parts of the product with a wrench may
damage the product.
If the tightening torque is exceeded, the product can
be broken.
If the tightening torque is insufficient, the fitting may
become loose and cause air leakage.
After installation completed, turn on the gas and
power supply for proper operation and leaking test to
confirme whether the installation is correct.
Caution
5.4Electrical Connection
NOTE
Incorrect power supply
Destruction of the integrated electronics
4Operate the product using a power supply unit with protected extra-low voltage
(PELV).
4The system must incorporate safe electrical cut-off of the power supply in compliance
with EN60204.
4Do not connect or disconnect the connector under tension and/or when voltage is ap-
plied.
NOTE
Connect with the power turned on
Damage to the electronics and/or malfunction
4Switch off the power supply before connecting cables!
The folw sensor is supplied with a 6-wire connection cable with open cable ends.
Integrate the folw sensor into your application in accordance with the electrical circuit diagram. The fol-
lowing points must be taken into account for this:
• Check wire color and terminal number when wiring. Incorrect wiring can cause permanent damages
to the sensor, check wire color and terminal number against the manual before wiring.
• Avoid repeatedly bending or stretching the lead wire. It can cause damage to the sheath, or break-
age of the wire.
• Confirm wiring insulation. Please avoid poor insulations (and interference from another circuit, poor
insulation between terminals, etc.) it can lead to over current being applied to the product, causing
damage.
• Do not route wires and cables together with power or high voltage cables. The product may mal-
function due to interference or noise and surge voltage from power and high voltage cables.
• Do not short-circuit the load. When the load is short-circuited, an error will be displayed. But excess
current may cause damage to the sensor.
This manual suits for next models
19
Table of contents
Other Schmalz Accessories manuals
Popular Accessories manuals by other brands

Gira
Gira System 106 installation instructions

Scott
Scott WORKS FILM SYSTEM 50 quick start guide

Dytran Instruments
Dytran Instruments 3312A2T operating guide

SMC Networks
SMC Networks PFM5 Series Installation & maintenance manual

Waterpik
Waterpik Sidekick WF-04 quick start guide

Vetek
Vetek DGTPK-F FORMULA user manual