Sensor ThermCondSys 5500 User manual

MAIL ADDRESS PHONE E-MAIL
UL. ODLEWNIKÓW 18
PL-44-100 GLIWICE
POLAND +48-32 237-26-88 OFFICE@SENSOR-ELECTRONIC.PL
Edition 22.1
Thermal Conditions Measuring System
ThermCondSys 5500
O
Op
pe
er
ra
at
to
or
r’
’s
s
M
Ma
an
nu
ua
al
l

ThermCondSys 5500 2 –Operator’s Manual
1
1.
.
G
GE
EN
NE
ER
RA
AL
L
I
IN
NT
TR
RO
OD
DU
UC
CT
TI
IO
ON
N
The ThermCondSys 5500 is user friendly system for measuring, calculating and logging thermal comfort parameters
at four levels at the same time in moderate, hot and cold environments. It has been developed to meet the growing need
for indoor climate monitoring equipment such as:
-mobility (moving from place to place)
-easy fast configuration (connection)
-low power consumption and long logging time
-automatic sleeping mode and auto turning off
-wire-less communication between computer and devices.
ThermCondSys5500 offers simultaneous measurement following parameters depending on the system configuration
of measuring modules.
Modules
Measured results
Calculated thermal comfort results
SensoTCMod 5503
ta, tgc, tw,va,
WBGTin, WBGTout, SD, DR, Tu, tr, tadb, teq, twc
SensoTCMod 5506
ta, tgc, tw, va, RH
WBGTin, WBGTout, SD, DR, Tu, tr, to, teq, PMV, PPD, pa,
DEW, i, Ereq, Emax, Wreq, SWreq, twc, IREQmin, IREQneut,
ICLmin, ICLneut, DLEmin, DLEneut,
SensoTCMod 5507
ta, tgc, tw, va, RH, Pb
WBGTin, WBGTout, SD, DR, Tu, tr, to, teq, PMV, PPD, pa, x,
, DEW, i, Ereq, Emax, Wreq, SWreq, twc, IREQmin, IREQneut,
ICLmin, ICLneut, DLEmin, DLEneut,
tgc: corrected globe temperature
tnw: natural wet temperature
ta: air temperature
ts: supplementary temperature
va: air speed
RH: relative humidity
Pb: barometric pressure
WBGTin: wet bulb globe temperature
WBGTout: wet bulb globe temperature
in the presence of radiation
SD: standard deviation of air speed
DR: draught rating
Tu: turbulence intensity of air speed
tr: mean radiant temperature
tadb: adjusted dry-bulb temperature
to: operative temperature
teq: equivalent temperature
PMV: predicted mean vote
PPD: predicted percentage of dissatisfied
pa: partial water vapour pressure in the air
x: humidity ratio
air density
DEW: dew point temperature
i: enthalpy for humid air
Ereq:required evaporation rate
Emax: maximum evaporation rate
Wreq: required skin wettedness
SWreq: required sweat rate
twc : wind chill temperature
IREQ: required clothing insulation (minimal
and neutral)
ICL: required basic thermal insulation (minimal
and neutral) according to ISO 9920
DLE: duration limited exposure (minimal and
neutral)
Because the humidity RH and barometric pressure Pbare the same on the each level you can use measuring
configuration consisting of one SensoTCMod 5506 (SensoTCMod 5507) module and two or three SensoTCMod 5503
modules. ThermCondSys5000 is in compliant with following standards: ISO 7726, ISO 7730, ISO 13182, ISO 7243,
ISO 7933, ISO 9920, ISO11079 and ASHRAE 55.
2
2.
.
I
IN
NS
ST
TA
AL
LL
LA
AT
TI
IO
ON
N
O
OF
F
T
TH
HE
E
M
ME
EA
AS
SU
UR
RI
IN
NG
G
S
SY
YS
ST
TE
EM
M
Mount the measurement modules on a tripod.
Connect measurement modules with cables (an example: SensoCable 5601).
Connect one of the measurement module with SensoConnectUSB interface or SensoBee wire-less transmitter.
Connect the power supply adaptor to the SensoConnectUSB interface (SensoBee wire-less transmitter) or turn
on the power supply in the SensoBee transmitter (see Chapter 4C). The LED diode on the top of the
measurement modules should blink first for a few seconds and next light up continuously. If the LED diode
blinks permanently it means that the voltage of power supply is too small or batteries placed in the PowerBank
are used up.
Connect the SensoConnectUSB interface (SensoBeeUSB receiver) to the USB port in a computer. The LED
diode on the casing should light up continuously (for SensoConnectUSB) or blink (for SensoBeeUSB).

ThermCondSys 5500 3 –Operator’s Manual
3
3.
.
E
EX
XA
AM
MP
PL
LE
ES
S
O
OF
F
C
CO
ON
NF
FI
IG
GU
UR
RA
AT
TI
IO
ON
N
A
A.
.
Multipoint wire configuration
B
B.
.
Multipoint wire-less configuration
SensoTCMod 5507
SensoTCMod 5503
SensoTCMod 5503 SensoData 5500 Bee
hand held data logger
SensoCable 5610 (length 10m)
optional wire communication
SensoData5500
IndoorClimateMeter
ESC
Off
Menu RecOn
Start
OK
electronic
max. 60m
SensoBee
wire-less transmitter
SensoBee
transmitt er
electronic
Power Supply
adapter
Power Bank
(optional)
SensoBee
transmitter
electronic
SensoBee
transmitt er
electronic
SensoBee USB
wire-less receiver or
SensoTCMod 5507
SensoTCMod 5503
SensoTCMod 5503 SensoData 5500
hand held data logger
SensoCable 5610 (length 10m)
SensoData5500
IndoorClimateMeter
ESC
Off
Menu RecOn
Start
OK
electronic
Power Supply
adapter
SensoBee
transmitter
electronic
SensoConnect USB
interface or

ThermCondSys 5500 4 –Operator’s Manual
4
4.
.
S
SP
PE
EC
CI
IF
FI
IC
CA
AT
TI
IO
ON
NS
S
A. SensoTCMod module
Measuring modules of the thermal comfort are made in the form of cylinders with dimensions of 100x48mm.
The SensoTCMod 5506 module includes following transducers: anemo, thermo, hygro. The SensoTCMod 5507
includes additionally barometric pressure sensor. The SensoTCMod 5503 module includes anemo and thermo
transducers, but doesn’t include baro and hygro transducer. Air temperature and relative humidity sensors are
placed on the top of the cylinder. Natural wet temperature, globe temperature and air speeds sensors are placed on
lateral horizontal supports. Air speed probe can be disconnected from the measuring module and may be
transported, stored or calibrated separately. The measuring modules can be mounted on a special folding tripod,
which enables mounting three or four units at the different heights.
Air speed probe includes the spherical omnidirectional air speed (1) and compensation temperature (2) sensors.
Both the sensors are vacuum covered with special aluminium coating that increases their resistance to contamination
and decreases the effect of thermal radiation on the accuracy of velocity measurement. The temperature sensor (2)
can be additionally protected against radiation by moving shielding tube (3) (it may be necessary only if the airflow
is measured very close to high temperature sources). Both sensors can be also protected against mechanical damage
using openwork basket (4) made with wire circles. If such a protection isn't necessary it is possible to removed them
from the shielding tube (3). Shielding tube (3) can be moved along the support of the probe and locked by the clamp
screw (5).
Technical data:
- diameter of the speed sensor: 2 mm
- measurement speed range: 0.05…5 m/s
- resolution: 0.001...0.01 m/s
- accuracy: ±0.02 m/s ±2% of readings
- automatic temperature compensation: < than 0.1%/K
- temperature compensation: -20OC…50OC
- upper frequency fup*): min. 1Hz, typ. 1.5 Hz
*)
The upper frequency is defined as the highest frequency up to which
the standard deviation ratio remains in the limits of 0.9 to 1.1 in
relation to the standard deviation of the frequency 0 Hz (see the
ASHRAE Transaction Vol.1,1998, paper No SF-98-20-2).

ThermCondSys 5500 5 –Operator’s Manual
Preparation for measurements:
1. Check if the openwork basket (4) is on the shielding tube (3). If not, put openwork basket (4) on the shielding
tube (3). Be particularly careful not to damage the speed sensor (1). To this purpose check if the shielding tube
(3) is on the end of the probe completely covering the speed sensor and the clamp screw (5) is properly locked
(it does not move). Put on the openwork basket (4) only then when the sensor (1) is covered by the shielding
tube (3).
2. Connect the air speed probe to measurement modules. Loosen the clamp screw (5) and move the shielding tube
(3) with the openwork basket (4) into the bottom uncovering the sensors (1) and (2).
3. Lock the clamp screw (5).
4. After the measurements move the shielding tube (3) back on the sensors (1) and (2).
5. The probe should be transported in the carrying case only with imposed openwork basket (4).
Air (ta), globe (tg) and natural wet (tnw) thermometers measure temperature using RTD sensors. Automatic
correction of thermal drift provides high measurement stability. The use of insulating elements in design of probes
(special wood and fiberglass) guarantees the appropriate thermal insulation and high measurement accuracy. In
order to improve the accuracy each sensor is individually calibrated.
Technical Data:
- type and diameter of thermometer: ta: Pt-100 cylindrical Ø 2.6 mm
tg: black globe ε0.95, Ø 70 mm, tickness 1.5 mm
tnw: Pt-100 cylindrical, length 30 mm, Ø 5 mm; support Ø 5 mm
- measurement range: ta: -20…60 OC
tg: -20…120 OC
tnw: 0…50 OC
- accuracy: ta: ±0.3 OC
tg: ±0.3 OC for range -20..50 OC i ±0.5 OC for range 50..120 OC
tnw: ±0.3 OC
- stabilization time of measurement: 15…20 min
Barometric pressure probe is located inside the cylinder of the measuring module. The probe utilizes specialized
piezoresistive micro-machined sensing element. Each probe is individually calibrated and thermally compensated.
The calibration coefficients are programmed into EEPROM memory.
Technical Data:
- measurement range: 500…1500 hPa
- accuracy: ±3 hPa
- response time: 2s
Relative humidity probe (optional) is located outside on the top of cylinder of the measuring module. The probe
includes a capacitive polymer sensing element for relative humidity and a bandgap temperature compensation
sensor. Each probe is individually calibrated in a precision humidity chamber with a chilled mirror hygrometer as
reference.
Technical Data:
- measurement range: 0…100 % RH
- accuracy: typ. ±1,5% in range 10…90% RH,max ±3%
- long term stability: <0.25% RH/rok
- response time: <4s
B. SensoBee wire-less transmitter
The SensoBee transmitter contains an RF module for
wire-less data transmission. The RF module operates
under ZigBee (802.15.4) protocol accepted by ETSI/EC
(Europe), FCC (U.S.A) and IC (Canada). The
transmitter should be powered by an external power
supply adapter or by PowerBank. RF switch should be
in the position RF and RF led should flash.
Measurement modules must be connected to the socket
B. In the case of the absence or errors in the wire-less

ThermCondSys 5500 6 –Operator’s Manual
transmission, the socket A enables optional wired connection of the transmitter with wire-less receiver. To do this,
move RF switch to the Off position, the RF led is not lit. The transmitter is equipped with a latch for mounting it in a
DIN rail.
Technical Data:
indoor (urban) range: up to 60m
outdoor line-of-sight range: up to 100m
transmit power output: 100mW (20dB)
operating frequency: 2.4GHz
RF data rate: 250 kbps
power supply: 5…9 VDC/1A
Remark:
Pay attention that the power supply adapter should have the voltage polarity on the plug compatible with
power socket in the casing (+ out).
C. SensoBeeUSB wire-less receiver
The SensoBeeUSB wire-less receiver contains the
RF module for wire-less data transmission. It enables
wire-less data transmission between transducers (via
SensoBee) and USB port of computer. The RF module
operates under ZigBee (802.15.4) protocol accepted by
ETSI/EC (Europe), FCC (U.S.A) and IC (Canada). The
receiver is being powered from the USB port of
computer (RF led flashes when connected to USB port),
therefore it doesn't require the use of external power
supply adapter.
The SensoBeeUSB can also be used for wired data
transmission. In this case, you should:
-connect to the RJ45 socket one end of the
connecting cable and the other end depending on the configuration, either directly to one of the measurement
module or SensoBee wire-less transmitter
-RF switch in the SensoBee transmitter change from RF position to Off (disable RF module).
Power socket placed on the side of the casing is used to connect the power supply adapter in case of direct power
connection to the measurement modules without using the SensoBee wire-less transmitter.
D. SensoConnect USB interface
The SensoConnect USB interface enables data
transmission to a computer through the USB port. It is being
powered from the USB port of computer, therefore doesn't
require external power supply adapter. The power socket
located on the side of the casing allows the user to connect an
external power supply adapter used to power the measurement
modules. The USB led indicates the connection interface to the
computer.

ThermCondSys 5500 7 –Operator’s Manual
5
5.
.
A
AC
CC
CE
ES
SS
SO
OR
RI
IE
ES
S
A. PowerBank
PowerBank is a high performance external battery pack with latch for DIN rail. The latch allows for easy
installation of the device on a tripod along with other parts of the measuring system. Includes a charging cable (USB
to micro USB) and spiral power cable (USB-plug DC) with a length of 30-100 cm.
Technical Data:
Power port: 2 x USB DC 5V/1A
Real capacity*): 7000mA or 1400mA
Minimum supply current: 135mA
Charging port: micro USB 5V/2.1A max
Charging time: >15 h
Long time efficiency: 300 cycles of full charge and discharge cycles will leave 85% of nominal
capacity)
*) Capacity due to reduced efficiency when converting from battery voltage 3.6 V to output voltage 5V (theoretical
capacity of the battery is 10000mAh or 20000mAh)
Remarks:
1. Turn off the PowerBank by disconnecting the power cable when you
are finished working. Otherwise, this may lead to rapid discharge of
the battery.
2. Charge and discharge the device once every 3 months, if it is not used
for a longer period of time.
3. Due to the minimal supply current, PowerBank may automatically turn
off after a few seconds in the case of smaller receivers power supply
current consumption (for example: a single SensoAnemo transducer).
Proper operation is achieved when are powered at least two
SensoAnemo transducers and SensoBee wire-less transmitter.
4. See picture next to how to connect the PowereBank with SensoBee
wire-less transmitter.

ThermCondSys 5500 8 –Operator’s Manual
B. Folds-out tripod
Measurement system can be equipped with a fold-out tripod adapted to mount several measuring transducers
with accessories. The tripod consists of the following parts presented in the photos below.
The way of assembling of the tripod is shown in
the photos below. Attention should be paid to the
pipes D1 ... D3 turn in the order corresponding to the
incisions, ie. D1 pipe (single) to the D2 (dual), and D2
to D3 pipe (triple). D1 pipe should be inserted to the
base of the tripod and locked by the knob. Then place
the grips for fastening the measuring transducers and
DIN grips for the accessories and lock them using the
knobs.
C. Grip for the modules
The grip enables attaching on the tripod the
measuring module at the any height from the floor.
D. DIN rail-mounted grip
The DIN rail-mounted grip enables attaching the
devices equipped with rail clamps, so as: SensoBee
wire-less transmitter, SensoBox adapter, SensoHygBar
transducer or SensoData 5500 datalogger.

ThermCondSys 5500 9 –Operator’s Manual
6
6.
.
S
SO
OF
FT
TW
WA
AR
RE
E
A
AN
ND
D
D
DR
RI
IV
VE
ER
R
I
IN
NS
ST
TA
AL
LA
AT
TI
IO
ON
N
1. Insert the CD-ROM in your CD drive, choose setup.exe in folder ThermCondSys_Inst and run.
2. Follow the on-screen instructions.
3. Locate the USB port and connect the device to the computer. The driver for new device will be installed
automatically after finishing the software installation. In the case of trouble it is possible to install the driver into
the following way:
Option 1
select SensoUSB DriverInstaller icon from SENSOR Electronic folder located in Program Menu Folder or
click it on the Desktop Panel (the SensoUSB DriverInstaller.exe file is also in SensoUSB_driver folder on the
CD-ROM.
locate the USB port and plug in the device (SensoConnectUSB or SensoBeeUSB).
Windows automatically identifies the new device and select the appropriate driver.
Option 2
locate the USB port and plug in the device (SensoConnectUSB or SensoBeeUSB).
Windows will detect the new device and will initiate the Found New Hardware Wizard prompting you to
install the software for the detected USB device.
select the “Specify a location” and choose the correct path. Enter E:\SensoUSB_driver\. The “E” is the Disk
that Package CD put in. Click “OK” to start the search.
once Windows finds the correct driver, click “Next” to install the driver.
4. To verify whether the device is properly installed and to determine the COM port assignment for the device
follow the steps:
open the “Control Panel”, and then double click on the “System” icon.
once the System Properties window is displayed select the “Device Manager” button.
double-click on Ports (COM &LPT).
if the hardware is correctly installed, you should now see the USB device listing and the assigned COM port.
5. If you need to assign the COM port name to another Port Number:
double-click on the USB device to view the properties
once the properties window opens click on the Port Settings tab and select the advanced button
when the Advanced Settings dialog box appears on the screen, click on the COM Port Number drop down the
box to check what other port number are available
try to reassign COM port to a unused port number
click OK when finished.
7
7.
.
D
DE
ES
SC
CR
RI
IP
PT
TI
IO
ON
N
O
OF
F
T
TH
HE
E
T
TH
HE
ER
RM
MC
CO
ON
ND
DS
SY
YS
S
5
55
50
00
0
P
PR
RO
OG
GR
RA
AM
M
Features and Requirements
The ThermCondSys 5500 program provides:
- automatic scanning of the configuration of the measuring system
- calculating thermal comfort parameters
- on-line viewing and logging all data in a PC computer
- identification of the low battery level for each module
- automatic identification of the faulty probe and over range
- ECO mode that reduces power consumption
Program requirements:
- operating system: Microsoft Windows
- minimal monitor resolution: 1024x760 pixels
- size of text and other elements in Windows: default 100%

ThermCondSys 5500 10 –Operator’s Manual
The program contains the Context Help window which displays a description when the user moves the cursor
over the property. The user can enable or disable the Context Help by selecting ‘Show Context Help’ from the
menu item ‘Help…’ on the menu bar on the front panel.
Updating the program and downloading the newest version of the instruction are available from the menu bar.
The program communicates only at the speed 115 200 BPS. It is very important not to start up the program
before SensoConnectUSB interface or SensoBeeUSB receiver are connected to the USB port and the power
supply is turned on.
The DataSocket connection automatically convert measurement data into a stream of bytes that is sent across the
network to another application.
The measurement data can be transfer across the network by TCP/IP protocol.
S
St
ta
ar
rt
ti
in
ng
g
a
an
nd
d
C
Co
on
nf
fi
ig
gu
ur
ra
at
ti
io
on
n
In the case of select automatic mode the program is scanning all ports of the computer in order to find connected
devices. In the manual mode the user have to choose the number of COM port. After establishing contact the
program is reading the information of all devices.
All connected devices are specified on the ‘Configuration’ window which appears after establishing contact. The
user can deactivate (set as OFF) or activate (set as ON) the devices to be used. The program automatically will
adjust the device as OFF if the calibration date is expired. However the user can change it and to activate him. If
the voltage of power supply is too small or batteries are discharge the indicator of power supply is shining red
and the transducer is automatically deactivate.
F
Fr
ro
on
nt
t
P
Pa
an
ne
el
l
Running Data
Results are displayed every 2 seconds with the chosen averaging time. Averaging Time determines the time
of calculating an average of the measurement results. Averaging is continuous, i.e. the oldest result is being
replaced with the youngest result in the cycle of averaging. Averaging Time is independent from Interval
Time which is a parameter of logging.
Logged Data
Only logged results are displayed according to logging parameters: Interval Time and Logging Time.
Buttons Selector of Results
It is possible to select results to displaying and logging by pressing the appropriate buttons above a
multicolumn listbox.
Visible results depending on buttons selector
Buttons
selector
Thermal comfort results
Input
parameters
Q-ty
none
ta, tgc, va, tr, to, PMV, PPD, RH, Pb
W, M, Icl, Ar/Adu
10
Hot
Environment
tnw, WBGTin, WBGTout, Ereq, Emax, SWreq, Wreq
7
Cold
Environment
twc, IREQmin, IREQneut, ICLmin, ICLneut, DLEmin, DLEneut
w, p
7
tadb, teq
tadb, teq
2
pa, x, ro,
Dew, i
pa, x, ro, Dew, i
5

ThermCondSys 5500 11 –Operator’s Manual
L
Lo
og
gg
gi
in
ng
g
P
Pr
ro
oc
ce
ed
du
ur
re
e
In order to begin the logging procedure one should press the ‘Start’ button. A ‘Select file for data logging’
window appears and the user should choose the name of a file for data logging. Next, there appears a new
window ‘Comment’ in which the user can enter any comment which will be placed at the beginning of a logging
file. After pressing “OK” key the logging procedure starts. The start time of the logging and remaining time to
the end of logging are shown on appropriate indicators. During the logging procedure the logging indicator is
blinking. Some buttons on the front panel are disabled in order not to disturb the logging process. Interrupting
the logging in every moment is possible, after pressing the ‘Stop’ button.
Interval Time determines an interval between logged data. It can be chosen in the range from 2s to 6h.
Logging Time determines how long a data will be logged. It can be chosen in the range from 10s to 48h. It is
possible to choose unlimited time as well.
The logged data are stored in the text file. This file can be easily imported to MS Excel. The data logging and
statistical analyzing procedure used is made clear in the following figure.
An example of settings with Interval Time of 1 min, Averaging Time of 3 min and Logging Time of 6 min.
E
Ec
co
on
no
om
my
y
M
Mo
od
de
e
If the Economy Mode switch is ON, devices can be switch into the state of asleep when reading results isn't
necessary. Asleep is only possible during the logged. Economy Mode reduces the consumption of energy what
influences increasing the working time on batteries.
The Economy Mode switch is active when the Interval Time is bigger than 3 min and the Averaging Time is
smaller than the half of the Interval Time.
A
Au
ut
to
om
ma
at
ti
ic
c
E
En
nd
d
When AutoEnd switch is ON, the program automatically will be closed and all devices will be switched into the
state of asleep after finishing the logging procedure.
D
Da
at
ta
aS
So
oc
ck
ke
et
t
C
Co
on
nn
ne
ec
ct
ti
io
on
n
When you use this protocol, the program communicates with the DataSocket Server. You must provide a named
tag for the data, which is appended to the URL. The data connection uses the named tag to address a particular
data item on a DataSocket Server. To use this protocol, you must run a DataSocket Server.
In order to enable or disable the DataSocket connection, select ‘DataSocket’ from the menu item ‘Data
Communication…’ on the menu bar on front panel.
You publish (write) data by specifying a URL. You should enter 'localhost' on the publisher computer and IP
address of the publisher on the subscriber computer.
The example of Data Socket Reader is available in the Program Menu Folder (SENSOR Electronic\Examples).
min.
1
0
2
3
4
5
6
A v e r a g i n g T i me
A v e r a g i n g T i me
A v e r a g i n g T i me
A v e r a g i n g T i me
Interval Time
Interval Time
Interval Time
Interval Time
Interval Time
Interval Time
Logging Time

ThermCondSys 5500 12 –Operator’s Manual
T
TC
CP
P/
/I
IP
P
C
Co
on
nn
ne
ec
ct
ti
io
on
n
The built-in TCP/IP connection enables to communicate between different computers by the networks. TCP/IP
routes data from server (ThermCondSys 5500 program) to client.
The default TCP port number is 5000. In order to change the TCP port number select ‘TCP connection’from
the menu item ‘Data Communication…’ on the menu bar on front panel.
The frequency of data transfer for TCP connection is specified by the Interval Time.
The data transferred by TCP server are separated by TAB delimiters. The data string can be column headers or
the results depending on setting the TCP connection. The part of string with column headers is ended by
CR carriage return element. On the end of result string is EOL
Example data string with results:
23.1 TAB 23.2 TAB 22.9 EOL
Example data string with column headers:
ta[degC] TAB tnw[degC] TAB tg[degC] CR 23.1 TAB 23.2 TAB 22.9 EOL
The example of TCP Data Client is available in the Program Menu Folder (SENSOR Electronic\Examples).
P
Ph
hy
ys
si
ic
ca
al
l
m
mo
od
de
el
ls
s
a
an
nd
d
p
pr
ro
oc
ce
ed
du
ur
re
e
Standard deviation of velocity [m/s]
2
2vvSD
, where
n
v
v
n
i
i
1
,
n
v
v
n
i
i
1
2
2
Turbulence Intensity [%]
%100
a
v
SD
Tu
, where SD –standard deviation
Draught Rate (The percentage of people dissatisfied due to draught) [%]
14.33705,034 62,0 SDvtDR aa
, for ta>34 or va<0.05 DR=0
Partial Water Vapour Pressure in the air [hPa]
a
a
tB
At
aRHp
10061078.0
for ta>0A=7.5; B=237.3; for ta0A=9.5; B=265.5
Humidity ratio [g/kg]
ab
apP p
x
98.612
Dewpoint Temperature [OC]
1121,09.0112
100
8 aa tt
RH
Dew
Enthalpy for humid air [kJ/kg]
250184.1005.1 aa txti

ThermCondSys 5500 13 –Operator’s Manual
Air Density [kg/m3]
273622.062.4 1
a
b
txPx
Mean Radiant Temperature [OC]
273
1038.5
273 4
1
8
4
ag
cggr
tt
htt
, where
25,0
6,0
675.2
755.17
max_
ag
m
cg tt
v
ofh
where tgis globe temperature for globe diameter of 7.5cm
WBGT-Index (Wet-Bulb-Globe Temperature) [OC]
gcnwin ttWBGT 3.07.0
agcnwout tttWBGT 1.02.07.0
where tgc is corrected globe temperature corresponds to globe with a diameter 15 cm
Operative Temperature [OC]
rc
rrac
ohh thth
t
, where
25,0
6.0
38,2
17.812.55.3
max_
ask
arararar
ctt
vforvorvforv
ofh
38 )273
2
(109996.21
skr
rtt
ARh
7.0
)5815.58(0052.0
min_ M
ofvv aar
claarask IMpvttt 553.0074432.0254.0571.0045.0093.030(
where M [Met] –metabolic rate
Icl [Clo] - clothing isolation
AR [-] - body area fraction exposed
Required Evaporation Rate [W/m2]
RCONECWME resresreq )(15.58
, where
)35(08141.0 ares tMC
)619.5(005995.1 ares pME
)( askclcttFhCON
)( rskclrttFhR

ThermCondSys 5500 14 –Operator’s Manual
rc
cl
clrc
cl
hh I
Ihh
F
30535.01 1
1
155.0)(1
1
Maximal Evaporation Rate [W/m2]
)(
max askepphE
, where
pclce Fhh 7.16
rc
cl
clc
pcl
hh I
Ih
F
30535.01 1
1
155.022.21
1
3.237
27.17
exp6105.0
sk
sk
sk tt
p
Required Skin Wettedness [-],
Required Sweat Rate [W/m2]
max
E
E
Wreq
req
2
5.01 req
req
req W
E
SW
if
0
req
E
then
0
req
W
,
0
req
SW
if
0
max E
then
2
req
W
,
reqreq ESW 2
if
1
req
W
then
reqreq ESW 2
Adjusted Dry-BulbTemperature [OC]
raadb tAAtt 1
, where A=0.5 for
2.0
a
v
m/s
A=0.6 for
6.02.0 a
v
m/s
A=0.7 for
6.0
a
v
m/s
Equivalent Temperature [OC]
teq=tofor va<0,1m/s
aclaraeq tIvttt 5,361/75,024,045,055,0
for va0,1m/s
Wind Chill Temperature [OC]
16.0
10
16.0
10 3965.037.116215.012.13 vtvtt aawc
, gdzie
]/[4.5]/[
10 smvhkmvm

ThermCondSys 5500 15 –Operator’s Manual
8
8.
.
G
GU
UA
AR
RA
AN
NT
TE
EE
E
A
AN
ND
D
R
RE
EP
PA
AI
IR
RS
S
Manufacturer guarantees the correct operation of the devices. The guarantee period is 24 months, beginning
from the date of sale. All defects due to faulty material or manufacturing will be repaired. What is under guarantee is
repair and replacement of detective parts. Damages of the devices owing to wrong transport or use will not be
recovered.
Table of contents
Popular Measuring Instrument manuals by other brands

THORLABS
THORLABS Redstone OSA305 user guide

FLIR
FLIR MR77 user manual

Tektronix
Tektronix PQA600C Software upgrade instructions

Keysight Technologies
Keysight Technologies MY53280000 Installation notes

PCB Piezotronics
PCB Piezotronics IMI SENSORS TO602D11 Installation and operating manual

Codonics
Codonics Virtua 83C Series Technical brief