Sper scientific 800037 User manual

WBGT SD Card
Datalogger
800037
Instruction Manual

WBGT SD Card Datalogger 800037
Copyright ©2010 by Sper Scientic
ALL RIGHTS RESERVED
Printed in the USA
8281 E. Evans Rd. Suite #103,
Scottsdale, AZ 85260
Tel: (480) 948-4448 | Fax: (480) 967-8736
Web: www.sperscientic.com
The contents of this manual may not
be reproduced or transmitted in any
form or by any means electronic,
mechanical, or other means that do
not yet exist or may be developed,
including photocopying, recording,
or any information storage and
retrieval system without the express
permission from Sper Scientic.

TABLE OF CONTENTS
INTRODUCTION.......................................4
FEATURES................................................5
MATERIALS SUPPLIED............................6
POWER SUPPLY ......................................7
METER COMPONENTS ..........................8
KEYPAD ..................................................10
SETUP..................................................... 11
MEASUREMENT PROCEDURES ..........16
DATALOGGER ........................................20
BATTERY REPLACEMENT ...................23
TROUBLESHOOTING ............................24
PC CONNECTION ..................................25
APPENDICES .........................................29
OPTIONAL ACCESSORIES....................31
SPECIFICATIONS ..................................32
WARRANTY ...........................................36

INTRODUCTION
The Sper Scientic WBGT SD Card
Datalogger (Model 800037) is an intelligent
meter that reads wet bulb globe temperature,
wet bulb temperature, dew point temperature,
humidity and air temperature, and black globe
temperature.
You can use the meter almost anywhere to
determine appropriate exposure levels and
meet OSHA or other safety standards for high
temperatures to avoid heat exhaustion and
heat-related injuries during athletic events,
military exercises, and industrial activities.
This meter features a real time SD memory
card datalogger. Standard,portable SD
memory cards provide unlimited data storage
and upload pre-formatted data directly to
Excel, eliminating the need for cables or
software. Alternatively, manually record 99
readings for on-screen review or stream data
directly to a computer using the optional
software and cables. Each data set includes
the temperature measurement (wet bulb
globe temperature, wet bulb temperature,
dew point temperature, humidity and air
temperature or black globe temperature) and
the time and date.

FEATURES
● Reads indoor/outdoor wet bulb globe
temperature, wet bulb temperature,
dew point temperature, humidity and air
temperature, and black globe temperature
● SD memory card datalogger
● Manual datalogger option
● Highly accurate microcomputer circuit
● Highly accurate Pt sensor for globe
temperature measurement
● Fast response time for
humidity measurement
● Direct measurement of the radiation
eect with a 75 mm brass black ball
● Sensor protecting mechanism
● RS232 or USB PC connection
● Direct upload of data to Excel
● Internal clock and calendar
● WBGT alarm
● Touch-tone
● Tripod mounting screw
● Built-in tabletop stand
● Maximum and minimum
● Hold function
● Auto-power-o
● Low battery indicator
● Backlight

MATERIALS SUPPLIED
● Meter
● SD Card
● Brass Black Ball with Sensor
● 6 AA Batteries
● Instruction Manual
● Soft Carrying Case

POWER SUPPLY
This meter can be powered by six AA (1.5
V, UM3) batteries or an optional 9 Volt DC
adapter. See page 23 for battery replacement
instructions.
Plug the adaptor into the power port labeled
“DC 9V,” located on the side of the meter
(under the protective black cover).
Note…
When using the adapter, the meter will remain
permanently on and the POWER button will
be disabled.

SD Card
Socket
METER COMPONENTS
Bottom of Meter
Side of Meter
RS232 Output
RESET
DC 9V Adapter
Input Socket

METER COMPONENTS
Back of Meter
Black Ball
(Black
Globe
Temperature
Sensor)
Black Ball
Mounting
Screw
Battery
Compartment/
Cover
Humidity
Sensor/Air
Temperature
(TA) Sensor
Battery
Cover
Screws
Stand
Tripod
Mounting
Screw

POWER
(ESC,
Backlight)
SET (▼,
TIME
CHECK)
LCD
Display
HOLD
(Function,
Next)
REC/
ENTER
(Indoor/
Outdoor)
LOGGER ,
(▲,
SAMPLING
TIME
CHECK)
KEYPAD

SETUP MODE
The advanced Setup Mode allows you to
customize the following meter preferences
and defaults:
● Real Time Clock
● WBGT Alarm
● Decimal Type
● Auto Power O
● Touch-Tone
● Temperature Units
● Sampling Time
● SD Memory Card Format
Note…
The setup functions can be performed under
any parameter but not while utilizing the
datalogger function. Once selections are
saved, the meter will default to the selected
preferences.
1. Press POWER to turn the meter on.
2. Press SET for 2 seconds or longer to enter
Setup Mode.
3. Press NEXT to cycle through the
setup functions.
Note…
Press ESC to exit Setup Mode. The meter will
return to Normal Mode.

SETUP MODE
Real Time Clock
1. Enter the clock function from Setup Mode
(see page 11). “dAtE” appears on the LCD.
2. Press ENTER. The year will ash on
the LCD.
3. Press ▲ or ▼ to adjust the value. Press
ENTER to save the value.
4. Repeat Step 3 to adjust the month, date,
hour, minute and second.
Note…
This procedure adjusts the meter’s internal
clock. The internal clock will function when
the meter is turned o but only when the
batteries have adequate power (not with low
battery power.)
WBGT Alarm
1. Enter the WBGT alarm function from
Setup Mode (see page 11). “AL” appears
on the LCD.
2. Press ▲ or ▼ to adjust the WBGT alarm
value. Press ENTER to save the value.
Note…
The alarm will sound when the WBGT
measurement exceeds the set WBGT
alarm value.

SETUP MODE
Decimal Type
Although the decimal is commonly expressed
as the “.” symbol (i.e., 20.6 or 1000.53), some
(European) countries use a “,” symbol to
represent the decimal (i.e., 20,6 or 1000,53).
The meter defaults to the period symbol.
To adjust:
1. Enter the decimal type function from
Setup Mode (see page 11). “dEC” appears
on the LCD.
2. Press ▲ or ▼ to select Basic (.) or Euro (,).
Press ENTER to save the selection.
Auto Power O
The meter automatically turns o after 10
minutes of inactivity, however this function
can be disabled.
1. Enter the auto power o function from
Setup Mode (see page 11). “PoFF” appears
on the LCD.
2. Press ▲ or ▼ to select yes (auto power o
enabled) or no (auto power o disabled).
Press ENTER to save the selection.
Touch-Tone
1. Enter the touch-tone function from Setup
Mode (see page 11). “bEEP” appears on
the LCD.

SETUP MODE
2. Press ▲ or ▼ to select yes (touch-tone
enabled) or no (touch-tone disabled). Press
ENTER to save the selection.
Temperature Units
1. Enter the temperature units function from
Setup Mode (see page 11). “t-CF” appears
on the LCD.
2. Press ▲ or ▼ to select C (degrees
Celcius) or F (degrees Farhenheit).
Press ENTER to save the selection.
Sampling Time
The sampling time (seconds) is the time
allotted between successive measurements.
To adjust the sampling time:
1. Enter the sampling time function from
Setup Mode (see page 11). “SP-t”
appears on the LCD.
2. Press ▲ or ▼ to adjust the value (0, 1, 2,
5, 10, 30, 60, 120, 300, 600, 1800, 3600).
Press ENTER to save the value.
SD Memory Card Format
Enabling this function will format the SD
memory card to work specically with your
meter. Formatting the SD card will erase any
previous memory on the card. New SD cards
should be formatted to work with your meter.

SETUP MODE
1. Enter the SD memory card format function
from Setup Mode (see page 11). “Sd F”
appears on the LCD.
2. Press ▲ or ▼ to select yes (format the SD
memory card) or no (do not format the SD
memory card). Press ENTER to conrm
your selection. If selecting yes, “yES Enter”
will appear on the LCD, press ENTER again
and the meter will format the SD card.

MEASUREMENT PROCEDURES
Turning the Unit On/O
1. Press POWER to turn the meter on.
2. Press and hold POWER for 2 seconds to
turn the meter o.
Selecting the Function
1. Press and hold FUNC to cycle through the
options listed below. Release FUNC when
you reach the desired function.
● Wet bulb globe temperature
“_bgt” appears on the LCD.
● Wet bulb temperature
“_b” appears on the LCD.
● Dew point temperature
“dP” appears on the LCD.
● Humidity and air temperature
“tA” appear on the LCD.
● Black globe temperature
“tg” appears on the LCD.
Note…
The meter defaults to °C. To change the
temperature unit, see page 14.
Wet Bulb Globe Temperature
1. Press POWER to turn the meter on.
2. Press and hold FUNC to cycle through

MEASUREMENT PROCEDURES
the options until “_bgt” appears on the LCD.
Release the FUNC button. The WBGT value
appears on the upper display and “_bgt in”
or “_bgt out” appears on the lower display to
indicate the setting for indoor or outdoor use.
Note…
Use the “in” setting for indoor use or outdoor
with no sun. Use the “out” setting for outdoor
use with full sun. To change the setting, press
and hold In/Outdoor and release when the
desired selection appears on the LCD.
3. To enable/disable the WBGT audible alarm,
see page 12.
Wet Bulb Temperature
1. Press POWER to turn the meter on.
2. Press and hold FUNC to cycle through
the options until “_b” appears on the LCD.
Release the FUNC button. The wet
bulb temperature value appears on the
upper display.
Dew Point Temperature
1. Press POWER to turn the meter on.
2. Press and hold FUNC to cycle through
the options until “dP” appears on the LCD.
Release the FUNC button. The dew
point temperature value appears on the
upper display.

MEASUREMENT PROCEDURES
Humidity and Air Temperature
1. Press POWER to turn the meter on.
2. Press and hold FUNC to cycle through the
options until “tA” appears on the LCD. Release
the FUNC button. The humidity value appears
on the upper display and the air temperature
value appears on the lower display.
Black Globe Temperature
1. Press POWER to turn the meter on.
2. Press and hold FUNC to cycle through the
options until “tg” appears on the LCD. Release
the FUNC button. The black globe temperature
value appears on the upper display.
Hold Function
1. When measuring any parameter, press
HOLD to freeze the reading on the display.
“HOLD” will appear on the LCD. Press
HOLD again to release the hold function.
Maximum and Minimum
To record maximum and minimum readings:
1. When measuring any parameter, press
REC to begin recording the maximum and
minimum values. “REC” appears on the LCD.

MEASUREMENT PROCEDURES
2. Press REC. The maximum value and “REC
MAX” will appear on the LCD.
3. Press REC. The minimum value and “REC
MIN” will appear on the LCD.
4. To delete the maximum or minimum value,
press HOLD. “REC” appears on the LCD
and the meter will begin recording the
maximum and minimum values again.
5. To exit the min/max function, press and
hold REC for 2 seconds or longer. The
meter will return to Normal Mode.
Backlight
1. The backlight turns on automatically when
the meter is turned on.
2. Press to turn the backlight o or on.
View Real Time Clock
1. Press TIME CHECK. The time information
(Year, Month/Date, Hour/Minute) will
appear on the lower display.
View Sampling Time
1. Press SAMPLING CHECK. The sampling
time (in seconds) will appear on the lower
display.
Note…
The time information and sampling time can
be viewed in Normal Mode.

Preparing the Datalogger
1. Insert the SD card into the SD card socket
on the bottom of the meter, ensuring that
the front of the SD card faces the back of
the meter.
2. Format the SD card as needed
(see page 14).
3. Set the real time clock if using the meter for
the rst time (see page 12).
4. Set the decimal type if using the meter for
the rst time (see page 13.)
Auto Datalogging
1. Set the sampling time to ≥ 1 second. Refer
to page 14.
2. Press REC. “REC will appear on
the LCD.
3. Press LOGGER. “REC” will ash on the
LCD and the alarm will sound while the
measurement data and time information
are saved to memory.
4. To pause datalogging, press LOGGER.
The meter will temporarily stop recording
and “REC” will stop ashing on the
LCD. Press LOGGER again to resume
datalogging. “REC” will ash on the LCD.
5. To nish datalogging, while the datalogger
is paused, press REC for 2 seconds or
longer. “REC” will