StarDental DentalEZ Titan Blis-sonic K 264564 User manual

The Blis-sonic SW requires a separate installation of either a 2/3-line
swivel, StarDental Part #261550, or a 4-line swivel, StarDental Part #261547,
depending on the handpiece tubing conguration.
1. Grasp the swivel in front of the bearing assembly between the thumb and
index nger as shown.
2.Slidetubingcouplerbackuntil theendoftheISOconnectorisexposed.Hold
in hand as shown with ngers securing the tubing.
3. Align and insert swivel tubes into the ISO connector holes.
4.Whileapplyingslightpressuretohold ttingsintheconnectorholes, slidecouplerforwardandengagethreads.Continue threading
clockwise until hand tight against the swivel.
Note: The swivel tubing connection joint may loosen over time, resulting in water leaks and/or loss of power. Check for tightness
periodically.
Note: To disconnect scaler from swivel, grasp
scaler rmly with one hand and pull tubing
coupler withtheother.Donot pull on tubing.
Swivel remains with tubing as shown.
Scaler Swivel Tubing Coupler
Tubing
5. Using the universal wrench, StarDental Part #203048, hold the
swivel stationary while tightening the tubing coupler until snug.
Be sure to turn tubing coupler, not swivel. Do not overtighten.
6. Snap scaler onto swivel.
Swivel
Bearing
Assembly
Tubing
ISO Connector
Tubing
Coupler
Wrench Tubing
Coupler
Turn tubing coupler
while holding
swivel in place
Blis-sonic K 2/3-Line: StarDental®Part #264564
Blis-sonic K 4-Line: StarDental®Part #264565
Air Pressure: 40 psi (2.7 atmospheres)
at the scaler
Air Flow: 1.5 SCFM (42.0 SLM)
Water Flow: 25 cc/min. @ 15 psi
Length (with tip): 6.7 inches (17.0 cm)
Weight: 1.8 ounces (50.7 grams)
Specications (Nominal)
Blis-sonic SW: StarDental®Part #264563
(Swivel Sold Separately)
Air Pressure: 40 psi (2.7 atmospheres)
at the scaler
Air Flow: 1.5 SCFM (42.0 SLM)
Water Flow: 25 cc/min. @ 15 psi
Length (with tip): 6.8 inches (17.3 cm)
Weight: 1.9 ounces (54.6 grams)
Sterilizable up to 135°C
135°C
Contents
• Specications
•Blis-sonic SW & Blis-sonic K Installation
• Tip Installation
• Air Pressure and Water Flow Adjustments
• Tip Operation
• Cleaning & Sterilization
• Lubricating
• Troubleshooting
• Grip Replacement
• O-Rings & Rear Seal Replacement
• Rotor & Seal Rings Replacement
•Blis-sonic K Water Tube Gasket Replacement
• Replacement Part Numbers
• Warranty
LUBRICATING Blis-sonic SW and Blis-sonic K
Always lubricate scaler prior to each autoclave sterilization with DentaLube®for Sonic Scalers, StarDental Part #261818.
Lubricating
1. To remove Blis-sonic K Scaler from tubing: Unthread from the tubing coupler. Disconnect the
Blis-sonicSW from the swivel by grasping the scaler with one hand, pull coupler with other hand.
Do not pull on the tubing. When separated, the swivel remains with the tubing coupler.
2. Find drive air on Blis-sonic K or notched hole inside of Blis-sonic SW (see illustration).
3. Insert lubricator tip rmly into nozzle. Spray a short burst of DentaLube for Sonic Scalers into
the Blis-sonic Kdrive air tube or in a notched hole of the Blis-sonic SW.
4. Thread tubing coupler onto the Blis-sonic K or snap Blis-sonic SW onto swivel; attach tip.
Note: Never run scaler without the tip in place.
5. Run scaler for 20 seconds to circulate lubricant; remove the tip.
6. Steam sterilize, see Sterilization.
Aerosol Lubricant
and Conditioner
DentaLube forTitan Sonic Scalers
Contains a petroleum lubricant oil (CAS 64742-54-70),
Heptane (CAS 142-82-5), and CO
2
(CAS 124-38-9).
READ BACK PANEL CAREFULLY
.
CFC FREE
Toreorder use:
StarDental
®
Part #261818
StarDental® Instrument Solutions
™
DentaLube
for
Titan
®
Sonic Scalers
Blis-sonic K
Drive Air
Blis-sonic SW
Notched
Hole
Drive Air
Notched Hole
Use of lubricant other than StarDental “DentaLube for Titan Sonic Scaler” can result in
failure of the scaler.
BEFORE OPERATING THE HANDPIECE, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE THESE
INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Cleaning and Sterilization
866.DTE.INFO (383.4636)
www.DentalEZ.com
0344
064590 J Date of Issue 2019-10-01
StarDental®
1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
©2019 DentalEZ, Inc.
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
EC REP
StarDental® Instruments
Caution: Federal law restricts this device to
sale by or on the order of a dentist.
Titan® Blis-sonic™Scaler
Instruction Manual
Non-Sterile
Medical Device
Remember:
1) Use the last 1/16” of the tip.
2) Use the edges of the tip, not the ats.
3) Keep the tip moving in either of two
motions as shown.
Tip Operation
Edges of
Tip are
Effective
1/16” (1.5 mm)
Effective
Edge
Ineffective
Portion
Flats are
Not
Effective
Effective Scaling Motions Ineffective Motions
Operation of the Blis-sonic SW and K Ergonomic Sonic Scaler Tip is similar to that of an ultrasonic scaler. Follow these steps to
minimize the risk of damage to restorations.
Universal - StarDental Part #261668
• Use for all surfaces, medium and heavy calculus
Use tips for subgingival scaling and ne access. Select a tip according to the following applications.
ALWAYS USE WATER.
Note: Normal use results in tip wear. Replace tips annually. Use thetip lengthsindicatedto determinemorefrequent replacement.
Replace if shorter by .02” (1/2 mm) or more.
1. Always scale with tip at an angle to the restoration margin.
2. Use a “feather touch,” one to two ounces applied pressure.
3. Move tip in a crosshatching, circular or wave fashion (see illustration).
Sickle - StarDental Part #261667
• Recommended for sensitive patients
• Less energetic than the Universal Tip
• Less suitable for deep root scaling .970 in.
24.6 mm.
.950 in.
24 mm.
Perio - StarDental Part #261669
Perio Left - StarDental Part #263397
Perio Right- StarDental Part #263398
• Recommended for lingual supragingival scaling of
mandibular incisors and deep local pockets
• Provides the same energy as the Universal Tip
• Use to break up heavy calculus deposits .930 in. (23.6 mm.)
PERIO LEFT
PERIO RIGHT
PERIO
Note: Use softened and deionized water. Be sure lters operate efciently. Before installing scaler, thoroughly purge both the dental
unit air line and water line. Verify recommended ltration of 10 microns air and 25 microns water.
If delivery unit has oil/mist lubrication, detach and drain.
Connect the Blis-sonic K
by threadingtothe tubing
coupler.
Push the Blis-sonic SW
scaler onto swivel until it
snaps into place.
Blis-sonic SW & K Installation
Blis-sonic SW Scaler Swivel Tubing Coupler
Tubing Coupler
Blis-sonic K Scaler (4-line) 4-line Tubing connector
NOTICE
Do not exceed 40 psi. Patient sensitivity and tooth damage may result.
2. Mount 4/6-line pressure gauge, StarDental Part #262591, between the scaler and the air supply tubing for the
Blis-sonic K Ergonomic Scaler. Mount the pressure gauge between the swivel and air supply tubing for the
Blis-sonic SW Ergonomic Scaler. Snap scaler onto swivel.
3. Turn water OFF.
4. Fully depress foot control. Adjust air supply on
deliveryunituntil gaugeshows40psi,2.72kg/cm2.
1. Thread the selected tip. With the scaler operating, rotate tip a 1/4 turn with ngers.
Air Pressure and Water Flow Adjustments
To prevent damage to the scaler, do not tighten tips with the universal wrench. Use the wrench only to remove tips
as shown in Tip Installation.
5. After adjustment, uncouple the pressure gauge and install scaler directly to the tubing.
6. With scaler running and tip held vertically, adjust water ow until water atomizes at tip. When water
atomizes, the water ow needed is optimized (see illustration above).
Pressure Gauge
NOTICE
NOTICE
Hygiene
Strong
Do not sharpen or alter tips in any way.
Tip Installation
1. Thread the selected tip onto the scaler with the scaler operating until it stops turning.
Rotate slightly with ngers until tight.
E
C
C
S
T
A
Cover
DO NOT USE FOR ORAL SURGERY
2. When the procedure is complete and the scaler is no longer in use, remove the tip and
blow compressed air through the tip to remove water trapped inside the tip. Cap it with
the cover provided, StarDental Part #261358. This practice prevents injury to patient or
staff.
3. To change tips, always loosen or tighten with the scaler running. The vibration makes
changing tips easier.
To prevent damage to the scaler, do not tighten tips with the universal wrench. Use the universal wrench
ONLY to remove tips. Use the wrench as illustrated.
StarDental®
Instrument Solutions
StarDental®
Instrument Solutions
Water Tube
Insert Wire
Front
Blis-sonic K
Blis-sonic SW
CAUTION
CAUTION
CAUTION
Tip amplitude can exceed 200 µm (.0078 inches).
CAUTION Do not exceed 135°C
Do not use:
• dry heat sterilizer • cold sterilizers • ultrasonic cleaners • chemical disinfectors • immersion techniques
• surgical milk • surface wipes • hand soaps • detergents • surface sprays
• sterilizer bags containing disinfectants
Limits On Reprocessing: Repeated processing has minimal effect on this scaler. Endoflife isnormallydetermined bywear anddamagedueto use.
Instructions
Scaler, tip, grip, and wrench must be cleaned and sterilized before rst use and after each patient.
WARNING
Containment and
transportation: It is recommended that scalers are reprocessed as soon as is reasonably practical following use.
Preparation for cleaning: Wear puncture resistant gloves. After each patient, ush air and water by running scaler. Remove tip from
scaler. Lube scaler as described in lubrication.
1. Attach Blis-sonic Kto tubing by turning coupler clockwise. Attach Blis-sonic SW by snapping onto swivel.
2. Attach tip to scaler. Never run scaler without tip in place.
3. Run scaler 20 seconds to expel excess lube; remove tip, clean tip per instructions in Cleaning.
4. Detach scaler by reversing Step 1.
Cleaning - Automated: Washer-disinfectors are not recommended.
Cleaning - Manual: Clean all external surfaces using a small toothbrush with isopropyl alcohol.
Cleaning The Tips
WEEKLY - Blow compressed air through the tip to remove water
trapped inside the tip. Next, clean outside of tip with toothbrush
and isopropyl alcohol. Then run cleaning wire, StarDentalPart
#256719, through entire tip starting at opening in front.
Cleaning Blis-sonic SW or K Water Tube
On a regular basis or if a water tube blockage occurs,
remove the tip (and on the Blis-sonicSW also disconnect
from swivel).Push the cleaning wire through the internal
water tube using needle nose pliers at the front on both
the Blis-sonic SW and K as shown. Push small lengths of
3 to 4 mm at a time until the wire emerges at the opposite
end. Pull wire completely out.
Cleaning the Swivel
Note: To ensure peak performance and proper functioning of scaler,
clean swivel and mating components daily or as needed. Do not attempt
to disassemble swivel.
1. Disconnect swivel from scaler and tubing. Depress foot control;
verify that water exits tubing.
2. Connect swivel to tubing. Depress foot control; if almost no water
exits front of swivel, clean as follows.
3. Detach swivel from tubing and nd water tube.
4. Insert cleaning wire, StarDental Part #256719, into water tube.
When the wire will not insert, move wire back and forth to loosen
debris at blockage.
5. Remove wire and apply a blast of air into
water tube with syringe air.
Cleaning Mating Parts of the Swivel
To optimize operation of the swivel, disconnect
mechanism and swab all internal surfaces of
scaler with a cotton swab soaked in isopropyl alcohol until clean. Dry using syringe air.
Cleanareainthebearing/springassemblyatrearofswivelusingisopropylalcoholandatoothbrush.Dryusingsyringeair.
Cleaning Wire
Opening
Bearing/Spring Assembly
Internal Surfaces
Clean
Disinfection: Use steam sterilization exclusively for complete sterility as outlined below.
Drying: Dry the scaler for 30 minutes minimum using the sterilization dry cycle and allow the scaler to cool to room
temperature in bag prior to use. If moisture is present, increase your sterilizer dry cycle appropriately.
Lubrication & Maintenance: See Lubrication section of this manual. The scaler should be routinely maintained, checked for corrosion, worn
parts, etc. and should be replaced on an as needed basis.
Packaging: Single: A standard sterilization packaging material must be used. Ensure that the pack is large enough to contain
the scaler without stressing the seals.
In sets: scalers may be loaded into dedicated instrument trays or general purpose sterilization trays. Ensure
that cutting edges are protected and do not exceed the tray manufacturer’s limit per tray. Wrap the trays using
appropriate packaging materials and methods.
Wrapped Instrument
Sterilization: Gravity Type Autoclave: Pre-Vacuum Autoclave:
• Minimum of 15 minutes at 132° C • Minimum of 4 minutes at 132° C
• Minimum of 30 minutes at 121° C
Do not exceed 135° C
Storage: Store in bag after sterilization until ready to use.
Additional information: Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
Manufacturer contact: See front or back panels of this manual.
Inspection & function testing: See Tip Operation or Troubleshooting sections of this manual.
CAUTION
NOTICE It is the responsibility of the user to use only FDA compliant sterilizing equipment and accessories that are capable of
meeting the manufacturers recommendations below.
CAUTION
Indications For Use
The scaler is an air driven handpiece which runs at sonic frequency. It effectively removes calculus deposits and stains.
The scaler tips are designed for subgingival scaling and ne access. The Universal tip is for all surfaces and the removal of both heavy and
medium calculus. The Sickle tip is for use on sensitive patients and provides alternate access for difcult posterior areas. ThePerio tip is
for lingual supragingival scaling of mandibular incisors and deep local pockets.
NOTICE Any serious incident occurring from the use of this device should be reported to StarDental immediately. If incident occurred
in a EU Member State the Competent Authority must be notied.

Always insist on StarDental parts for best performance.
Replacement Part Numbers
Description Item Number
Troubleshooting
Problem: NO VIBRATION OR NOISE
Cause: No air
Correction: Check pressure at compressor and delivery unit.
Cause: Blocked air line
Correction: Check for air at tubing coupler.
Cause: Control unit not functioning
Correction: Check operation of handpieces and syringe. If needed, remove tubing from control block and check for blockage
of drive air tubing.
Problem: LOW VIBRATION LEVEL OR LOW POWER
Cause: Low air pressure
Correction: Check for full depression of foot control, unit pressure and compressor output. Check pressure at scaler with air
gauge 4/6-line, StarDental®Part #262591.
Cause: Tip loose
Correction: Finger tighten, do not tighten tip with wrench. Refer to Tip Installation.
Cause: Dirty rotor
Correction: Lubricate with DentaLube for Blis-sonic®Sonic Scalers, StarDental Part #261818.
Cause: Tip touches nose
Correction: Check all joints on nose, handle and torque reaction sleeve for tightness.
If OK, return scaler to factory for repair.
Cause: Rotor seal rings out of alignment
Correction: Remove scaler handle, relocate rotor seal rings in shallow grooves. Carefully thread
scaler handle on to prevent damage to seal rings.
Cause: Broken or worn tip
Correction: Refer to Tip Operation for tip lengths. Replace if shorter by .02” (1/2mm) or more.
Cause: Rotor broken
Correction: Remove scaler handle; replace rotor. Refer to Rotor & Seal Rings Replacement.
Cause: Broken rear suspension (very high air noise, low pitch)
Correction: Return scaler to dealer or factory for repair.
Cause: Defective swivel, O-rings and/or seal
Correction: Refer to O-Rings & Rear Seal Replacement.
Problem: UNABLE TO DETACH SWIVEL FROM SCALER
Cause: Lack of maintenance
Correction: Spray a short burst of DentaLube®for Sonic Scalers into swivel-tubing connector area; rotate swivel to circulate
lube. Disengage handpiece. Contact dealer if unable to disengage swivel. To prevent reoccurrence, refer to
Cleaning.
Problem: NO WATER
Cause: Water off
Correction: Check dental unit controls and inlet supply.
Cause: Tip blocked
Correction: Remove tip, check ow. If tip blocked, refer to Cleaning Tip.
Cause: Water tube blocked or swivel clogged
Correction: Clean scaler or swivel water tube. Refer to Cleaning Water Tube.
Cause: Worn swivel O-rings or seal
Correction: Replace O-rings and/or seal. Refer to O-Rings & Rear Seal Replacement.
Problem: WATER LEAKS AT REAR
Cause: Rear seal
Correction: Tighten tubing coupler. If leak persists, refer to O-Rings & Rear Seal Replacement.
Problem: WATER LEAKS BETWEEN TIP AND HANDLE
Cause: Tip shaft seat dirty
Correction: Clean shaft end recess with cotton swab soaked in isopropyl alcohol. Clean conical tip seat.
Problem: AIR LEAK INTO WATER STREAM
Cause: Rear seal leaks
Correction: Tighten tubing coupling. If leak persists, refer to O-Rings & Rear Seal Replacement.
Problem: WATER DOES NOT ATOMIZE
Cause: Front suspension O-rings
Correction: Replace front suspension O-Rings, refer to O-Rings & Rear Seal Replacement.
Problem: AUDIBLE IRREGULARITY, CHANGE IN PITCH OR NOISE
Cause: Tip loose
Correction: Finger tighten, do not tighten tip with wrench. Refer to Tip Installation.
Cause: Rotor broken
Correction: Remove handle assembly, replace rotor. Refer to Rotor & Seal Rings Replacement.
Cause: Rear suspension broken
Correction: Return to factory through authorized dealer for repair.
Note: If any problem persists, contact dealer for further assistance or contact StarDental Customer Service at 1-866-DTE-INFO.
1. Remove tip, see ‘Tip Installation’.
2. Remove grip, see ‘Grip Replacement - Removal of Grip’.
3. Unthread the front of the scaler nose to expose the front suspension O-rings.
4. Remove the two old O-rings and install the replacement O-rings, StarDental Part #256718.
5. Rethread scaler nose to original position.
O-Rings & Rear Seal Replacement
Blis-sonic K/SW Rear Seal & Swivel O-Rings
2/3-line for Blis-sonic SW and Blis-sonic K
Rear Seal
Blis-sonic K
2/3-line
Blis-sonic SW
2/3-line
Rear Seal
Swivel
1. Remove 2/3-line rear seal with an explorer.
2. Place new rear seal over drive air and water tubes.
3. Carefully push seal into place using a blunt instrument.
Swivel O-Rings: Periodically replace the swivel O-rings, StarDental
Part #259884.
Front Suspension O-Rings: After two years, front suspension O-rings
relax resulting in play between shaft and handle.
Seals wear with repeated use. Replace rear seals annually or as needed.
On either the 2/3-line swivel or Blis-sonic K 2/3-line scaler, replace the
2/3-linerearsealwithStarDentalPart#203103.Oneitherthe4-lineswivel
or Blis-sonic K 4-line scaler, replace seal with StarDental Part #202897.
If rotor fails, scaler emits a low pitch ‘burbling’ sound and power reduces
dramatically. Replace rotor and seals on Blis-sonic SW or Blis-sonic K
with Rotor Spares Kit, StarDental Part #256715.
1. Remove scaler tip.
2. Use clamp wrench provided to unthread handle from scaler torque
reaction sleeve exposing the shaft.
Handle precision rotor with extra care.
3. Slide off 2 pairs of seal rings and the old rotor.
4. Clean shaft with a 2 x 2 gauze.
5. Slide new rotor and new seal rings onto shaft; verify both pair of rings
are in both shallow grooves.
6. Thread scaler handle assembly onto torque reaction sleeve. Take care to not catch seal rings with handle assembly.
7. Firmly tighten handle assembly to shaft.
8. Lubricate scaler and reconnect to handpiece tubing.
9. If power is inadequate after replacing rotor, reverse the rotor on the shaft as shown.
Never remove scaler nose nor handle with tip assembled to the scaler, this could damage the shaft assembly.
Rotor & Seal Rings Replacement
Replace the washer gasket or o-ring on the Blis-sonic K watertube assemblyannuallywiththegasketprovidedintheStarDentalSpares
Kit #256888 for the 2/3-line, or the o-ring provided in the StarDental Spares Kit #264592 for the 4-line.
Water Tube Gasket Replacement
1. Remove the water tube assembly. Remove rear seal with an old dental instrument
such as an explorer. Gently tug on the air tube with needle nose pliers to remove
water tube assembly. Retaining ring pops out when removing assembly. Take care not
to lose retaining ring.
2. Slide old gasket off water tube.
3. Slide on new gasket and seat in place.
4. Insert watertubeinto handle, rotate slowly between thumb and forenger to locate internal water tube assembly. Push
lightly to fully seat tube.
Note: Reinstall retaining ring. Then assemble rear seal over tubes and retaining ring.
Use needle nose pliers carefully to prevent distortion of the air tube when
removing water tube assembly.
Universal Tip (Plastic Cover Included)..................................................................................................................................................................... 261668
Sickle Tip (Plastic Cover Included)..............................................................................................................................................................................261667
Perio Tip (Plastic Cover Included) ...............................................................................................................................................................................261669
Perio Left Tip (Plastic Cover Included)...................................................................................................................................................................... 263397
Perio Right Tip (Plastic Cover Included)................................................................................................................................................................... 263398
Tip, Plastic Cover ................................................................................................................................................................................................................261358
TITAN®Blis-sonic SW Titan Scaler (4 Line)............................................................................................................................................................264563
TITAN®Blis-sonic KTitan Scaler (2/3 Line) ...........................................................................................................................................................264564
TITAN®Blis-sonic KTitan Scaler (4 Line)................................................................................................................................................................264565
Cleaning Wire .008 Gauge (Package of 2) ................................................................................................................................................................256719
Retaining Ring..................................................................................................................................................................................................................... 261349
O-Rings - Swivel (2 large, 2 small)...............................................................................................................................................................................259884
O-Rings - Front Suspension (Package of 2)..............................................................................................................................................................256718
Rotor Spares Kit (Rotor, 4 Seal Rings).........................................................................................................................................................................256715
Rear Seal, 2/3-line (Package of 3) ................................................................................................................................................................................203103
Rear Seal, 4-line (Package of 3).................................................................................................................................................................................... 202897
Water Tube Assembly, Gasket, Retaining Ring, 2/3-line TITAN®Blis-sonic K..........................................................................................256454
Water Tube Assembly, 4-line TITAN®Blis-sonic SW...........................................................................................................................................260503
DentaLube for Sonic Scalers (8.0 .oz. Spray Can)................................................................................................................................................261818
Air Pressure Gauge (ISO, 4/6-line)...............................................................................................................................................................................262591
Universal Wrench .............................................................................................................................................................................................................203048
Rubber Grip Replacements (1 each: Aqua, Blue, Lavender and Pink)..........................................................................................................264571
Rubber Grip Replacement, Four Aqua...................................................................................................................................................................... 264555
Rubber Grip Replacement, Four Blue .......................................................................................................................................................................264556
Rubber Grip Replacement, Four Lavender..............................................................................................................................................................264557
Rubber Grip Replacement, Four Pink........................................................................................................................................................................264558
For new products, manuals and technical information, call
1-866-DTE-INFO or visit our web site @ www.dentalez.com.
Rear Seal
on the 4-line
Blis-sonic K
Rear Seal on
the 4-line Swivel
O-Ring
Rear Seal
Unthread
Water Tube Assembly
Water Tube
Water
Manifold
4-Line
Grip Replacement
Removal of grip: Installation of grip:
1. With thumb at rear of grip, push down
and forward.
2. Continue pushing until grip clears nose.
PLEASE NOTE: DO NOT PULL OR PEEL TO REMOVE
Swivel O-Rings
Front Suspension O-Rings
Scaler Nose Handle
Torque Reaction Sleeve
Rear of Grip
Front Recess
Wall
Rear Recess Wall
Grip Recess Area
Front of Grip
1. Push grip onto nose at front of grip.
2. Continue pushing until grip ts in
nose grip recess area. Grip should
rest against front and rear recess
walls.
3. Grip must not protrude past front
of handpiece (scaler nose).
Water
Tube
Plate
Retaining
Ring
Gasket
2/3-line
Water Tube
Assembly
Water
Manifold
Nose with Grip Handle Blis-sonic SW
Rotor
Shallow
Groove
1st pair
of Seal Rings
(Slid Out Of Shallow Groove)
2nd pair of
Seal Rings in
Shallow Groove
Shaft
Blis-sonic Scaler
Clamp Wrench
Rotor
2 pairs
of Seal Rings
Torque Reaction
Sleeve
Torque
Reaction
Sleeve
Blis-sonic K
2/3 Line:
1. Remove rear seal. Then, using a 7/16” open end wrench, or an adjustable wrench,
unthread manifold assembly (water tube assembly).
2. Slide old o-ring off of water tube.
3. Slide on new o-ring and seat in place.
4. Insert water tube into handle, rotate slowly between thumb and forenger to locate
internal water tube assembly. Thread manifold assembly (water tube assembly) into
torque reaction sleeve. Tighten using 7/16” open end wrench, or adjustable wrench.
Note: Reinstall rear seal.
4 Line:
Important Notice About
Returning Product To The DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain the proper RETURN AUTHORIZATION NUMBER from the
Customer Service Department. The RA Number must appear clearly marked on both the outside and inside of the returned product carton.
When calling for a return authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
• Product Name, Model Number or Part Number
• Serial Number of Product
• Dealer Return Purchase Order Number
• Reason For Return
• Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with this procedure will help us expedite
processing of any returns.
Never remove scaler nose nor handle with tip assembled to the scaler, this could damage the shaft assembly.
NOTICE
NOTICE
NOTICE
NOTICE
Limited Warranty
O-Rings, Rotor and Tips specifically excluded from this warranty.
DentalEZ®warrants the Blis-sonic™ SW and K Ergonomic Sonic Scalers to be free of defects in material and workmanship,
under normal usage, under the following terms:
StarDental® Products: Warranty Period:
Blis-sonic™ SW and K Ergonomic Sonic Scalers 1 year from date of purchase*
Please note the following additional terms of our warranty and return policy:
• Warranty covers manufacturing defects only and does not cover defects resulting from abuse, accidents, misuse, improper handling,
cleaning, care or maintenance or the failure to follow or observe our operating/use, maintenance and/or installation instructions.
Damage resulting from the use of chemicals, cleaners, disinfectants or sterilization is not covered under this warranty nor is normal
wear and tear. Failure to use authorized parts or an authorized repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective part(s) or products at our sole discretion. All other liabilities,
in particular liability for damages, including, without limitation, special, exemplary, consequential or incidental damages
are excluded
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE,
REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY
OTHER WARRANTY.
WARRANTY REPAIRS: Parts repaired or replaced on a product that is in warranty will be warranted for the duration of
that product's original warranty.
NON-WARRANTY REPAIRS:The warranty on parts either repaired or replaced on an out-of-warranty product will cover
the repaired part only and will be for the time frame of a new parts warranty period.
PRODUCT RETURN: Opened products or product returns more than a year old cannot be returned for credit.There
will be a 15% ($25.00 minimum) restocking charge on all items authorized for return.
*Provided conditions defined in the Instruction Manual are met.
WARRANTY REGISTRATION
PLEASE VISIT THE WEBSITE BELOW TO REGISTER YOUR STARDENTAL PRODUCT. THIS WILL
ENSURE PROPER REGISTRATION AND PROMPT AND ACCURATE SERVICE SHOULD ANY OF YOUR
PURCHASES NEED EVALUATION OR REPAIR DURING THE WARRANTY PERIOD.
StarDental® Products: www.dentalez.com/warranty