Filter Rates are:
Filter Rate(RD 0ff) Rate(RD on) Notes
0 19msec 26msec 25dB rejection @ 60Hz
1 19msec 26msec 25dB rejection @ 50Hz
2 30msec 31msec
3 50msec 51msec 80dB rejection @ 60Hz
61msec 62msec 65dB reject @ 50 & 60Hz
5 62msec 63msec 80dB rejection @ 50Hz
6 105msec 106msec 67dB reject @ 50 & 60Hz
7 21msec 22msec
8 8 2msec 8 3msec
9 1.7sec 1.7sec
The best setting depends upon the application. For example, to capture fast peaks 0 or 1 would be
best, a fast bag filler might be best set to filter , but a tank content indicator would be best on filter 9.
Use the 5 6 keys to change the filter value. When correct press PROG.
3.3 Least significant digit count
Press the PROG button then press 5 or 6 until lsd is displayed. Press the 4 key to
indicate the current setting. The value n indicates the least significant digit increments. This can be
1's, 2's, 5's or 10's. This feature is used where the accuracy or resolution of the load
cells makes the display appear unstable and does not justify counting the least
significant digits in 1's.
Use the 5 6 keys to change the value. When correct press PROG.
3. Hold mode
Press the PROG button then press 5 or 6 until HOLD is displayed. Press the 4 key to
indicate state of peak hold mode, on or off. When hold mode is on, the instrument
will measure peaks and troughs of the measured input. The display will show the c
current peak value. Holding in the 5 key will show the trough (minimum) value.
Pressing the RESET key will clear the current values.
Use the 5 6 keys to toggle between the states. When correct press PROG.
3.5 Setpoint relays operating mode
Press the PROG button then press 5 or 6 until RELAY is displayed. Press the 4
key to indicate state of relay operating mode, latch or free. In latching mode, once a
relay is energised, it will stay on until reset. In free mode the relay will be
energised only whilst the setpoint value is exceeded. If both setpoint relays are latched,
press RESET to clear relay one then press RESET again to clear relay two. If only one relay
is latched, pressing RESET will clear it.
Use the 5 6 keys to toggle between the states. When correct press PROG.
RS232 option
.1 Baud rate
Press the PROG button then press 5 or 6 until BAUD is displayed.
Press the 4 key to indicate the current value. Baud rates available are: 115200,
38 00, 9600 and 2 00. Use the 5 6keys until the required value is shown. Press PROG to save
and exit.
.1a RS 85 address
If the instrument has been supplied for use with an RS 85 adaptor, such as the SY 85,
it requires a device address to be identifiable on the RS 85 network. The current
address is shown. To be distinguished as an address, the MSB of the transmitted byte is 1. If the
address was set as 001, the device would be accessed by transmitting $81. Use the 5 6 & 4 keys
to adjust the value.
lsd
lsd n
Hold
On
OFF
rELAY
LAtCH
FrEE
AUd
dAnnn