TA DTC-300 User manual

THERMAL CONDUCTIVITY METER
DTC-300
Getting Started Guide
Revision EIssued April 2020

Page 1 DTC-300 Getting Started Guide
Notice
The material contained in this manual is believed adequate for the intended use of the instrument. If the
instrument or procedures are used for purposes other than those specified herein, confirmation of their suit-
ability must be obtained from TA Instruments. Otherwise, TA Instruments does not guarantee any results
and assumes no obligation or liability. TA Instruments also reserves the right to revise this document and to
make changes without notice.
TA Instruments may have patents, patent applications, trademarks, copyrights, or other intellectual prop-
erty covering subject matter in this document. Except as expressly provided in written license agreement
from TA Instruments, the furnishing of this document does not give you any license to these patents,
trademarks, copyrights, or other intellectual property.
TA Instruments Operating Software, as well as Module, Data Analysis, and Utility Software and their asso-
ciated manuals, are proprietary and copyrighted by TA Instruments. Purchasers are granted a license to use
these software programs on the module and controller with which they were purchased. These programs
may not be duplicated by the purchaser without the prior written consent of TA Instruments. Each licensed
program shall remain the exclusive property of TA Instruments, and no rights or licenses are granted to the
purchaser other than as specified above.
TA Instruments can accept no liability for loss or damage, however caused, arising from the faulty or
incorrect use of its products.TA Instruments shall not be liable for any damages caused by interactions
between exogenous materials (e.g. chemicals) and parts of the instrument. This includes interactions of
gaseous, liquid or solid materials with parts of the instrument.
©2020 by
TA Instruments — Waters LLC
159 Lukens Drive
New Castle, DE 19720

DTC-300 Getting Started Guide Page 2
Introduction
Important: TA Instruments Manual Supplement
Please click the TA Manual Supplement link to access the following important information supplemental
to this Getting Started Guide:
• TA Instruments Trademarks
• TA Instruments Patents
• Other Trademarks
• TA Instruments End-User License Agreement
• TA Instruments Offices

Page 3 DTC-300 Getting Started Guide
Notes, Cautions, and Warnings
This manual uses NOTES, CAUTIONS, and WARNINGS to emphasize important and critical
instructions. In the body of the manual these may be found in the shaded box on the outside of the page.
NOTE: A NOTE highlights important information about equipment or procedures.
CAUTION: A CAUTION emphasizes a procedure that may damage equipment or cause loss of data
if not followed correctly.
MISE EN GARDE: UNE MISE EN GARDE met l'accent sur une procédure susceptible d'endom-
mager l'équipement ou de causer la perte des données si elle n'est pas correctement suivie.
Please heed the warning labels and take the necessary precautions when dealing with those parts of the
instrument. The DTC-300 Getting Started Guide contains cautions and warnings that must be
followed for your own safety.
Regulatory Compliance
Safety Standards
For Canada
CAN/CSA-C22.2 No. 61010-1 Safety requirements for electrical equipment for measurement, control, and
laboratory use, Part 1: General Requirements.
CAN/CSA-C22.2 No. 61010-2-010 Particular requirements for laboratory equipment for the heating of
materials.
For European Economic Area
(In accordance with Council Directive 2006/95/EC of 12 December 2006 on the harmonization of the laws
of Member States relating to electrical equipment designed for use within certain voltage limits.)
EN 61010-1:2012 Safety requirements for electrical equipment for measurement, control, and laboratory
use, Part 1: General Requirements + Amendments.
EN 61010-2-010:2003 Particular requirements for laboratory equipment for the heating of materials +
Amendments.
For United States
UL61010-1:2012 Electrical Equipment for Laboratory Use; Part 1: General Requirements.
UL61010A-2-010:2002 Particular requirements for laboratory equipment for the heating of materials +
Amendments.
A WARNING indicates a procedure that may be hazardous to the operator or to the environ-
ment if not followed correctly.
Un AVERTISSEMENT indique une procédure qui peut être dangereuse pour l'opérateur ou
l'environnement si elle n'est pas correctement suivie.

DTC-300 Getting Started Guide Page 4
Electromagnetic Compatibility Standards
For Australia and New Zealand
AS/NZS CISPR 11:2007 Industrial, Scientific and Medical (ISM) Equipment - Radio-frequency Distur-
bance Characteristics - Limits and Methods of Measurement.
For Canada
ICES-003 Issue 5, February 2004 Information Technology Equipment (ITE) - Interference-Causing Equip-
ment Standard - Limits and Methods of Measurements.
For the European Economic Area
(In accordance with Council Directive 2004/108/EC of 15 December 2004 on the approximation of the
laws of the Member States relating to electromagnetic compatibility.)
EN61326-1:2006 Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Require-
ments - Part 1: General Requirements.
Emissions: Meets Class A Requirements per CISPR 11.
Immunity: Per Table 1 - Basic Immunity Test Requirements.
For the United States
Code of Federal Regulations (CFR) Title 47 - Telecommunication, Chapter I - Federal Communications
Commission (FCC), Part 15 - Radio Frequency Devices, Subpart B - Unintentional Radiators for a Class A
Digital Device (FCC regulation pertaining to radio frequency emissions).
Supplier’s Declaration of Conformity
47 CFR § 2.1077 Compliance Information
Unique Identifier: 854030.901–.906
Responsible Party:
TA Instruments
159 Lukens Drive
New Castle, DE 19720
302-427-4000
www.tainstruments.com
FCC Compliance Statement:
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.

DTC-300 Getting Started Guide Page 5
Safety
Instrument Symbols
The following label is displayed on the instrument for your protection:
Please heed the warning labels and take the necessary precautions when dealing with these areas. This Get-
ting Started Guide contains cautions and warnings that must be followed for your own safety.
Warnings
Symbol Explanation
This symbol indicates that a hand crush hazard is present. Take care to keep hands
clear of the Test Stack during loading and unloading of samples.
Ce symbole indique la présence d'un risque de broyage de main. Prenez soin
d'éloigner vos mains de la cheminée d'essai pendant la mise en place et la
décharge des échantillons.
This symbol indicates that a hot surface may be present. Take care not to touch
this area or allow any material that may melt or burn to come in contact with this
hot surface.
Ce symbole indique la présence possible d'une surface chaude. Prenez soin de ne
pas toucher cette zone ou de laisser un matériau susceptible de fondre ou de brûler
entrer en contact avec cette surface chaude.
WARNING: The operator of this instrument is advised that if the equipment is used in a manner
not specified in this manual, the protection provided by the equipment may be impaired.
AVERTISSEMENT: L'utilisateur de cet instrument est prévenu qu'en cas d'utilisation contraire
aux indications du manuel, la protection offerte par l'équipement peut être altérée.
WARNING: Always unplug the instrument before performing any maintenance.
AVERTISSEMENT: Débranchez toujours l'instrument avant de procéder à la maintenance.
WARNING: No user serviceable parts are contained in the DTC-300. Maintenance not specified
in the DTC-300 Getting Started Guide and repairs must be performed by TA Instruments or other
qualified service personnel only.
AVERTISSEMENT: Aucune pièce du DTC-300 n'est réparable par l'utilisateur. Les travaux de
maintenance non spécifiés dans le guide de démarrage du DTC-300 et les réparations doivent
être effectués uniquement par TA Instruments ou tout autre personnel d'entretien qualifié.

DTC-300 Getting Started Guide Page 6
CAUTION: Do not power off the chiller during instrument operation. The top stack, guard furnace,
and bottom stack must be below 50°C. Failure to do so may result in damage to the instrument.
Electrical Safety
Supply voltage: 100 to 120VAC or 200 to 240VAC (Instrument Dependent, Reference the Voltage Rating
on the Back Panel)
Mains frequency: 50 to 60 Hz
Current Rating: 10A
Fuse type: 2X T 10A L 250V
Lifting the Instrument
WARNING: Do not raise the guard furnace until it is below 50°C. Failure to do so may result in
personal injury and/or damage to the instrument.
AVERTISSEMENT: Ne soulevez pas la fournaise de protection tant qu’elle n’est pas inférieure à
50 ° C. Dans le cas contraire, vous risquez de vous blesser et / ou d'endommager l'instrument.
DANGER: Because of the high voltages in this instrument, maintenance and repair of internal
parts must be performed by TA Instruments or other qualified service personnel only.
DANGER: À cause de la présence de tensions élevées dans cet instrument, la maintenance et la
réparation des pièces internes doivent être effectuées uniquement par TA Instruments ou tout
autre personnel d'entretien qualifié.
WARNING: Only use a mains power cable that meets the ratings displayed on the back on the
instrument being installed/reinstalled.
AVERTISSEMENT: Utilisez uniquement un câble du secteur électrique qui répond à la capacité
nominale affichée à l'arrière de l'instrument installé/réinstallé.
WARNING: Use two people to lift and/or carry the instrument. The instrument is too heavy for
one person to handle safely.
AVERTISSEMENT: Demandez à deux personnes de soulever et/ou de porter l'instrument.
L'instrument est trop lourd pour qu'une seule personne le manipule en toute sécurité.

DTC-300 Getting Started Guide Page 8
Table of Contents
Introduction.................................................................................................................................... 3
Important: TA Instruments Manual Supplement ...................................................................... 3
Notes, Cautions, and Warnings ................................................................................................. 4
Regulatory Compliance ............................................................................................................. 4
Safety Standards ................................................................................................................. 4
Electromagnetic Compatibility Standards .......................................................................... 5
Supplier’s Declaration of Conformity ....................................................................................... 5
Safety ......................................................................................................................................... 6
Instrument Symbols ............................................................................................................6
Warnings ............................................................................................................................. 6
Electrical Safety .................................................................................................................. 7
Lifting the Instrument .........................................................................................................7
Table of Contents........................................................................................................................... 8
Chapter 1: Introducing the DTC-300 ........................................................................................ 11
Overview ........................................................................................................................................ 11
DTC-300 System Components ......................................................................................................12
Electronic Enclosure ............................................................................................................... 12
DTC-300 Front Panel ....................................................................................................... 12
DTC-300 Back Panel ........................................................................................................ 14
Test Stack ................................................................................................................................ 15
Instrument Specifications .............................................................................................................. 16
Chapter 2: Installing the DTC-300 System ............................................................................... 18
Unpacking/Repacking .................................................................................................................... 18
Installing the System ...................................................................................................................... 18
Inspecting the System ............................................................................................................. 18
Choosing a Location ............................................................................................................... 19
In ....................................................................................................................................... 19
On ..................................................................................................................................... 19
Near .................................................................................................................................. 19
Away from ........................................................................................................................ 19
Moving the Instrument ............................................................................................................ 20
Connecting the Utilities ........................................................................................................... 20
Connecting the Power ....................................................................................................... 20
Connecting the Air Supply ............................................................................................... 21
Connecting the Drain Hose ............................................................................................... 21
Connecting the Coolant .................................................................................................... 22
Chapter 3: Operating the DTC-300 ........................................................................................... 23
Using the DTC-300 ........................................................................................................................ 23
Before You Begin ................................................................................................................... 23
Startup and Shutdown Procedures ................................................................................................. 23

DTC-300 Getting Started Guide Page 9
Starting the DTC-300 System ................................................................................................. 23
Shutting Down the DTC-300 System ..................................................................................... 24
Running a DTC-300 Experiment ...................................................................................................24
Sample Requirements .............................................................................................................. 24
Low-Temperature Testing ....................................................................................................... 25
Loading the Sample .......................................................................................................... 25
Testing the Sample ........................................................................................................... 26
High-Temperature Testing ...................................................................................................... 27
Loading the Sample .......................................................................................................... 27
Testing the Sample ........................................................................................................... 28
Sub-Ambient Temperature Testing ......................................................................................... 28
Loading the Sample ......................................................................................................... 28
Testing the Sample ........................................................................................................... 29
Unknown Sample Testing ....................................................................................................... 29
Special Sample Testing ........................................................................................................... 31
Thin Film (Multiple) Sample Testing ............................................................................... 31
Non-Solid Sample Testing ................................................................................................32
Paste Cell Testing .............................................................................................................32
Polymer Melt Cell Testing ...............................................................................................33
Unloading the Sample ............................................................................................................. 34
Running a Conductivity Test ......................................................................................................... 35
Stopping a Test .............................................................................................................................. 39
Analyzing a Completed Test ......................................................................................................... 39
Reviewing the Results ............................................................................................................. 39
Reanalyzing the Test Results ..................................................................................................40
Manually Creating a Conductivity Test ............................................................................ 40
Chapter 4: Calibrating the DTC-300......................................................................................... 42
Calibration Procedures for Low Temperatures
(20°C to 100°C) ............................................................................................................................. 42
Calibration Procedures for High Temperatures
(100°C to 300°C) ........................................................................................................................... 43
Calibration Procedures for Sub-Ambient Temperatures
(-20°C to 20°C) .............................................................................................................................. 44
Calibration Procedure with the Paste Cell ..................................................................................... 45
Reanalyzing and Modifying a Calibration ..................................................................................... 45
Reanalyzing a Calibration ....................................................................................................... 45
Modifying a Calibration .......................................................................................................... 47
Adding a Reference .......................................................................................................... 47
Deleting a Reference ........................................................................................................ 48
Manually Creating a Calibration File ......................................................................................49
Chapter 5: Maintaining the DTC-300........................................................................................ 51
Cleaning the Test Stack Surface .................................................................................................... 51
Uninstalling and Installing the Stack Spacer ................................................................................. 51

DTC-300 Getting Started Guide Page 10
Uninstalling the Stack Spacer ................................................................................................. 51
Installing the Stack Spacer ...................................................................................................... 52
Uninstalling and Installing the Bottom Stack ................................................................................ 53
Uninstalling the Bottom Stack ................................................................................................ 53
Installing the Bottom Stack ..................................................................................................... 54
Replacement Parts ......................................................................................................................... 55

DTC-300 Getting Started Guide Page 11
Chapter 1:
Introducing the DTC-300
Overview
The Discovery Thermal Conductivity (DTC-300) instrument is used for measuring the thermal conductiv-
ity of materials in the range 0.1 to 30 W/(m.K). These include some metals and ceramics, polymers, com-
posites, glass, rubber, and carbon products. The DTC-300 can also be used to test thin samples like paper
products and polymeric films. Non-solids of high viscosity can be tested using a special container to hold
the material.
Standard thermal resistance range is 0.002 to 0.02 m2K/W. Optional ranges are 0.0005 to 0.01 and 0.01 to
0.05 m2K/W.
The DTC-300 is configured to test a 50.8 mm diameter sample. The instrument operates from -20 to
300°C–a suitable chiller/circulator is required to test from -20 to 20°C. Chiller setpoint temperatures must
be a minimum of 15°C below the desired sample temperature.
Input power options include 100–120 VAC or 200–240 VAC. Please check your instrument and make cer-
tain that you have received the option you ordered.
The DTC-300 system (shown below) is comprised of the test stack and electronic enclosure. The instru-
ment is PC-controlled and the operational software is also used to generate tabular and graphical data.
Figure 1 DTC-300 system.
Test Stack
Electronic Enclosure

DTC-300 Getting Started Guide Page 12
DTC-300 System Components
A DTC-300 system has two major components: DTC-300 electronic enclosure and the test stack.
Electronic Enclosure
The DTC-300 electronic enclosure contains the system electronics used to read and display the output of
the temperature sensors.
DTC-300 Front Panel
The heater control for each of the heaters is sensed and automatically adjusted by the software, and can
also be adjusted manually using the PID temperature controllers located on the front of the electronic
enclosure. The various parameter settings are preset at the factory and should not be adjusted or manipu-
lated without first consulting TA. These parameters are re-initialized at the beginning of each test to ensure
proper starting conditions. An independent over-temperature monitor is employed as a safety for the
device. The fault indicator on the front panel lights when an over-temperature condition is detected. Pres-
sure applied to the sample can be adjusted using the air pressure regulator and monitored on the built-in
pressure gauge. The movement of the test stack is controlled with the UP/DOWN, FREE/MOVE, and
SPEED switches.
Refer to Figure 2 below for the locations of knobs, switches, and indicators on the DTC-300 front panel
and refer to Table 1 for a description of their functions.
Figure 2 DTC-300 front panel.
HEATER POWER switch
Heater temperature display
(TOP, GUARD, BOTTOM)
SPEED/INCREASE knob
MOVE/FREE switch
UP/DOWN switch
Air pressure gauge
Air pressure regulator knob
FAULT indicator

DTC-300 Getting Started Guide Page 13
Table 1: DTC-300 Front Panel
Indicator Function
HEATER POWER switch Powers the heaters ON and OFF
FAULT indicator Indicates (lights) when an over-temperature condition is
detected
Heater temperature display: TOP, GUARD,
BOTTOM
Displays current temperature and setpoint for each heater
SPEED/INCREASE knob Controls the speed at which the top stack moves up or down
MOVE/FREE switch Activates the top stack pneumatics
UP/DOWN switch Moves the top stack up/down
Air pressure gauge Air pressure gauge for monitoring air pressure applied to the
top stack
Air pressure regulator knob Air pressure regulator allows you to adjust the air pressure to
the top stack

DTC-300 Getting Started Guide Page 14
DTC-300 Back Panel
The DTC-300 back panel has five ports. Figure 3 below shows the rear connections and Table 2 provides a
description of the function of each port.
Figure 3 DTC-300 rear panel connections.
Table 2: DTC-300 Back Panel
Port Function
USB Provides communications to the Temperature Measurement
System
RS232 Provides communication to the PID Heater Control System
DRAIN Removes condensing water (when needed)
AIR IN Provides the air or inert gas used to operate the DTC-300
pneumatics
MAIN POWER port Provides power to the instrument
MAIN POWER switch Powers the DTC-300 ON (I) / OFF (0)
MAIN POWER port
USB
RS232
AIR IN
DRAIN
MAIN POWER switch

DTC-300 Getting Started Guide Page 15
Test Stack
The sample to be tested is held under a reproducible compressive load between two polished metal sur-
faces (the “test stack”), each controlled at a different temperature. The lower contact surface is part of a
calibrated heat flux transducer. As heat flows from the upper surface through the sample to the lower sur-
face, a uniaxial temperature gradient is established in the stack. By measuring the temperature difference
across the sample along with the output from the heat flux transducer, thermal conductivity of the sample
can be determined when the sample thickness is known.
Figure 4 Test stack on the DTC-300.
add image

DTC-300 Getting Started Guide Page 16
Instrument Specifications
The table below contains the technical specifications for the DTC-300.
Table 3: DTC-300 Technical Specifications
Item/Area Specifications
Dimensions Depth 71 cm (28 in)
Width 64 cm (25 in)
Height 94 cm (37 in)
Weight 44.5 kg (98 lbs)
Powerab
a. This instrument is supplied with a power cord approximately 2.4 m (7.75 ft) long. The 220V instrument
power cord comes supplied with a USA NEMA 6-15P style plug; the 110V instrument power cord comes
supplied with a USA NEMA 5–15P style plug. The TA Representative installing the instrument will pro-
vide the appropriate power cable for the geographical location where the instrument is being installed.
Ensure that the mains assigned do not also supply power to noise generating equipment nearby, such as
motors, welders, transformers, etc.
100–120V (for 110V instrument) or 200–240 VAC (for 220V instru-
ment)
50/60 Hz
10A max current supply
Supply voltages lower than indicated limit may result in a degrada-
tion of performance.
Fuse Type 2X T 10A L 250V
Operating Environmental Conditions Temperature: 15 to 35°C
Relative humidity: 5 to 80% (non-condensing)
Maximum Altitude 2000 m (6560 ft)
Locate instrument in a ventilated space (hood, etc.) if noxious gases
or vapors are generated during the heating of samples.
The degree of protection for this instrument according to EN 60529
is IP20.
Purge Gas System Requirements Nitrogen Purge Gasc
1.5 psi (0.1 bar) nitrogen gas furnace flush is required for operating
at temperatures below 20°C.
Pneumatic System Requirements Compressed Air/Inert Gasd
Maximum inlet pressure to the instrument: 60 psi (4.1 bar)
Typical operating pressure: 40 psi (2.8 bar). The flow rate is less
than 1 L/day. Compressed inert gas or air from a gas cylinder may
be used. A proper gas cylinder pressure regulator/reducing regulator
is required.
Cooling System Requirements Chiller/Circulator or Cooling Water and Draine
Permissible coolant temperatures are between -40 and 20°C,
depending upon testing considerations. Excessively cold water may
result in “sweating” and possible corrosion of cooled metal surfaces.
A nitrogen furnace cavity “flush” is required for sub-ambient opera-
tion. Warm water may not permit required cooling and stack tem-
perature gradients during testing.

DTC-300 Getting Started Guide Page 17
b. An independent GROUND is required on the power connection where the instrument is being installed.
Improper grounding may cause severe damage for which supplier will not accept responsibility.
c. The instrument is supplied with a 1.22m (4 ft) length of 3.175 mm (.125 in) tube and a 3.175 mm (.125
in) to 6.35 mm (.25 in) expander fitting to connect the purge gas to the instrument. The customer is
required to make all hose connections. A flow meter is required for control of the furnace chamber
“flush” gas (not supplied).
d. The customer is required to supply either compressed air, dried and filtered, to 10 microns, or laboratory
grade inert gas (i.e.N2from a high pressure cylinder) for operating the pneumatic system.
e. Operating below the minimum chiller coolant level will result in erratic operation. A chiller/circulator is
recommended for this instrument. Consult the chiller/circulator manual for specifications concerning
appropriate fluid types. The instrument is supplied with insulated hoses to attach to a chiller/circulator.
Table 4: DTC-300 Performance Specifications
Item/Area Specifications
Test Method ASTM E1530
Thermal Conductivity Range 0.1 – 40W/mK
Thermal Resistance Range Low-Range: 0.0005 – 0.010 m2K/W
Mid-Range: 0.002 – 0.020 m2K/W
Hi-Range: 0.01 – 0.05 m2K/W
Accuracy ± 3%
Reproducibility ± 2%

DTC-300 Getting Started Guide Page 18
Chapter 2:
Installing the DTC-300 System
Unpacking/Repacking
You may wish to retain all of the shipping hardware, the plywood, and boxes from the instrument in the
event you wish to repack and ship your instrument.
Installing the System
Before shipment, the DTC-300 is inspected both electrically and mechanically so that it is ready for opera-
tion upon proper installation. Installation involves the following procedures:
• Inspecting the system for shipping damage and missing parts
• Choosing a location for instrument installation
• Installing the instrument
• Connecting the utilities
It is strongly recommended that you have your DTC-300 system installed by a TA Instruments Service
Representative; call for an installation appointment when you receive your instrument.
CAUTION: To avoid mistakes, read this entire chapter before you begin installation.
MISE EN GARDE: Pour éviter de commettre des erreurs, lisez tout le chapitre avant de commencer
l'installation.
Inspecting the System
When you receive your DTC-300 system, look over the instrument and shipping container carefully for
signs of shipping damage, and check the parts received against the enclosed shipping list.
• If the instrument is damaged, notify the carrier and TA Instruments immediately.
• If the instrument is intact but parts are missing, contact TA Instruments.

DTC-300 Getting Started Guide Page 19
Choosing a Location
Because of the sensitivity of DTC-300 experiments, it is important to choose a location for the instrument
using the following guidelines. The DTC-300 system should be:
In
• A temperature and humidity controlled area.
• A clean, vibration-free environment.
• An area with ample working and ventilation space.
On
• A stable, non-flammable work surface.
Near
• A power outlet for the specific voltage of the instrument.
• Compressed lab air and purge gas supplies with suitable regulators and filtering where needed.
• Cooling water/coolant (stable chiller/circulator preferred).
Away from
• Dusty environments.
• Exposure to direct sunlight.
• Direct air drafts (fans, room air ducts).
• Poorly ventilated areas.
• Electrically noisy areas or areas prone to mechanical vibrations.
NOTE: Do not place equipment against walls or cabinets that might impede air flow. Leave at least 7.5 cm
(3 in) clearance between the back of the instrument and any objects.
WARNING: For safety, position the equipment in a manner that allows access to
the power cord for emergency disconnection.
AVERTISSEMENT: Par mesure de sécurité, placez l'équipement de sorte qu'il per-
mette d'accéder facilement au cordon d'alimentation en cas de débranchement
d'urgence.
WARNING: Protect power and communications cable paths. Do not create
tripping hazards by laying the cables across access ways.
AVERTISSEMENT: Protégez les chemins de câble électriques et de câbles de
télécommunication. Ne créez pas de risques de déclenchement en posant des
câbles sur les voies d'accès.

DTC-300 Getting Started Guide Page 20
Moving the Instrument
Lifting the instrument should always be done with two people. With each person standing on either side
of the instrument, place hands between the feet on the bottom of the instrument and lift upwards, keeping
the instrument in an upright position.
Figure 5 Lifting the instrument.
Connecting the Utilities
Connecting the Power
Plug the power cord from the MAIN POWER port on the back of the DTC-300 to an electrical outlet
rated in accordance with the power input information on the back of the instrument.
CAUTION: Whenever plugging or unplugging power cords, handle them by the plugs, not by the
cords.
MISE EN GARDE: Chaque fois que vous branchez ou débranchez les cordons d'alimentation,
tenez-les par les fiches et non par les cordons.
WARNING: Use two people to lift and/or carry the instrument. The instrument is
too heavy for one person to handle safely.
AVERTISSEMENT: Demandez à deux personnes de soulever et/ou de porter
l'instrument. L'instrument est trop lourd pour qu'une seule personne le manipule
en toute sécurité.
WARNING: Only use a mains power cable that meets the ratings displayed on the
back of the instrument being installed/reinstalled.
AVERTISSEMENT: Utilisez uniquement un câble du secteur électrique qui répond
à la capacité nominale affichée à l'arrière de l'instrument installé/réinstallé.
Table of contents
Other TA Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Varian
Varian FRG-700 series instruction manual

Allen-Bradley
Allen-Bradley 809S-E2A manual

PCB Piezotronics
PCB Piezotronics 3503A1160KG/-0002 Installation and operating manual

Banner
Banner SureCross Series manual

dymax
dymax ACCU-CAL 160 user guide

MBA Instruments
MBA Instruments MLA 900 operating instructions

Growatt
Growatt ShineWiFi-S Quick Installation Guideline

Saunier Duval
Saunier Duval MiLink Plus installation instructions

AEMC instruments
AEMC instruments 502 manual

Symmons
Symmons Evolution EVOUM1KIT installation guide

Klein Tools
Klein Tools ET140 instruction manual

Assured Automation
Assured Automation WM-NLD Series Installation, operation & maintenance manual