troubleshooting guide
The ozone meter cannot be turned on.
The meter cannot be charged.
No response to the test value of the gas
on the meter.
The concentration value of the instrument in
clean air is unstable.
The battery is exhausted. Restart the meter
after charging. If it cannot be started, contact
SAT.
Measure if the charger output voltage is
4.75 - 5.5v. If the output voltage is abnormal,
replace the charger.
- The oxygen content of the introduced gas
is too low: <5%.
- The life of the sensor has ended.
Place the instrument in a clean place
without interfering gas, and the instrument
value will fall to the normal value.
problem solution
the high alarm value or the low alarm value to be modified. The user can press the “OK” key to
operate the value, change the cursor to select a value through the keys “◄” y “►”, add or sub-
tract the value through “▲” and “▼”,
press the “OK” key to confirm when the value is set, move the highlight to the storage by pres-
sing the “◄” and “►” keys, and press it again to save.
At this point, the system indicates that the save is successful and the voice mentions “Save suc-
cessful”. Press “▼” to return to the menu.
There are three detector alarm modes: sound, vibration and
light. When our company’s gas detector leaves the factory,
the alarm modes are activated by default and can be adjusted
by the user according to actual needs.
System configuration:
User can move the cursor to the “system setting” icon and
press “OK” key to enter the system setting selection interface.
Year, month, day and time setting: You can set the time with “◄” and “►”, to finish press OK. Press
the “▼” key during setting to return to the menu.
Communication settings: you can set the direction and speed of transmission
of the instrument. When importing data to the computer,
Password setting: Default is “1111”. If you change it and forget it, you can only contact the company
staff to send it back to the factory for recovery.
Air pump settings: Set the air pump to close
automatically at a fixed time, with a fixed maximum time of 999 seconds. If not required
time, change the time to 000 seconds and the air pump will be normally open.