
Operating Instructions--Type 7A13
Set the Type 7A13 front panel controls as follows:
COMPARISON VOLTAGE
(Vc) Polarity Either
+
or
-
Counter As
is
+
INPUT Mode GND
+
INPUT Connector Connected to battery posi-
tive terminal. Negative ter-
minal connected to ground
lug on front panel.
-
INPUT Mode GND
VARIABLE In (CAL)
STEP ATTEN BAL As
is
GAIN As
is
X10 BAL As
is
VAR BAL As
is
BW
5MHz
POSITION Midrange
1. Use the POSITION control to set the trace to DC-0 ref-
erence.
2. Place the
+
INPUT Mode switch to DC.
3. Multiply the number of divisions displayed by the
deflection factor (3 div
X
.5 V
=
1.5 V). The input voltage
amplitude
is
1.5 volts k0.3 Volts.
Note that if the input voltage were applied to the
-
IN-
PUT connector, with
+
INPUT Mode switch set to GND and
-
INPUT Mode switch set to DC, the 3 divisions deflection
would be downward, since the
-
INPUT causes an inverted
presentation.
Example
2
Differential Comparator Operation, DC measurement.
Measure the voltage output of a 1.5 volt battery.
Type 7A13 control settings are the same as for Example 1
above.
1. Set the COMPARISON VOLTAGE (Vc) Polarity switch to
+
and set the counter to 0150.
2. Set the VOLTS/DIV switch to 20 mV and pull the VARI-
ABLE knob out.
3. Using the POSITION control, set the trace to DC zero
reference (graticule center).
4. Simultaneously place the
+
INPUT Mode switch to DC
and the
-
INPUT Mode switch to Vc.
5. Using COMPARISON VOLTAGE (Vc) controls, reset the
trace to DC zero reference.
6.
Read the value of the input voltage on the COMPARI-
SON VOLTAGE (Vc) counter. It should be 1.5 volts
f
0.3
volts.
7.
Disconnect the battery from the Type 7A13.
Example
3
Single-Ended Operation-waveform measurement.
Measure the oscilloscope Calibrator
+
Volts waveform and
DC level.
Type 7A13 control settings are same as for examples 1 and
2 except for the
+
INPUT connector, which should be con-
nected to the Calibrator
+
VOLTS BNC connector on the
oscilloscope via coaxial cable.
1. Set the oscilloscope Calibrator control to 40 mV.
2. Set VOLTS/DIV Switch to 10mV position.
3. Set
+
INPUT Mode switch to AC and set the time base
plug-in controls for a stable display.
4. Compute AC waveform amplitude using the POSITION
control to establish a reference line. (4 divisions deflection
times 10mV/DIV yields 40 mV.)
5. Determine the DC level on which the waveform
is
rid-
ing, using the method shown in either example 1 or 2. Note
the DC level of the top or bottom of the waveform as de-
sired.
NOTE
For signals with a high
AC
to DC ratio, the AC and
DC signal components can be evaluated simultan-
eously by using DC coupling and single-ended
operation methods.
6.
Disconnect the calibrator signal.
Example
4
Differential Amplifier Operation. Measure the difference
between the oscilloscope Calibrator
+
Volts output and a
modified Calibrator
+
Volts output.
Type 7A13 control settings are the same as example 3
above with the following exception: The
-
INPUT Mode
switch should be set to DC.
1.
Set the VOLTS/DIV Switch to 20 mV position.
2. Connect a BNC T adapter to the Calibrator
+
Volts
output of the oscilloscope.
3. Build and connect a voltage divider to the BNC T as
illustrated in Fig. 2-2.
4. Connect the junction of the two resistors to the
-
IN-
PUT connector via coaxial cable.
5. Set the oscilloscope Calibrator
+
Volts output to 4 V.
6.
Simultaneously place the
+
INPUT and
-
INPUT Mode
switches to GND. Using the POSITION control, set the trace
to DC zero reference two divisions below the graticule hori-
zontal centerline.
7. Simultaneously set both the
+
and
-
INPUT Mode
switches to DC. Adjust the time-base plug-in triggering con-
trols if necessary to obtain a stable waveform.
8.
Measure the waveform amplitude. It will be equal to
the difference between the two input waveforms. (Approxi-
mately 40 mV; this
is
the voltage drop across the 1
k0
resis-
tor.)
Scans by ArtekMedia © 2007
Operating Instructions-Type 7A13
Set the Type 7A
13
front panel controls as follows:
COMPARISON
VOLTAGE
(V
c)
Polarity
Counter
+
INPUT
Mode
+
INPUT
Connector
-
INPUT
Mode
VOLTS/DIV
VARIABLE
STEP
A
HEN
BAL
GAIN
XlO
BAL
VAR
BAL
BW
POSITION
Either + or -
As
is
GND
Connected to battery posi-
tive terminal. Negative ter-
minal connected to ground
lug
on front panel.
GND
.5V
In
(CAL)
As
is
As
IS
As
is
As
is
5MHz
Midrange
1.
Use
the POSITION control to set the trace to
DC-O
ref-
erence.
2.
Place the +
INPUT
Mode switch to
DC.
3.
Multiply the number of divisions displayed by the
deflection factor
(3
div X
.5
V =
1.5
V).
The
input voltage
amplitude
is
1.5
volts
+0.3
Volts.
Note that
if
the input voltage were applied to the -
IN-
PUT
connector, with +
INPUT
Mode switch set to GND and
-
INPUT
Mode switch set to
DC,
the 3 divisions deflection
would be downward, since the -
INPUT
causes an inverted
presentation.
Example 2
Differential Comparator Operation,
DC
measurement.
Measure the voltage output of a
1.5
volt battery.
Type 7A
13
control settings are the same as for Example 1
above.
l.
Set the
COMPARISON
VOLTAGE
(Vc)
Polarity switch to
+and set the counter to 0150.
2.
Set the
VOLTS/DIV
switch to
20
mV
and
pull
the
VARI-
ABLE
knob out.
3.
Using
the POSITION control, set the trace to
DC
zero
reference (graticule center).
4.
Simultaneously place the +
INPUT
Mode switch to
DC
and the -
INPUT
Mode switch to
Vc.
5.
Using
COMPARISON
VOLTAGE
(Vc)
controls, reset the
trace to
DC
zero reference.
6.
Read the value of the input voltage on the
COMPARI-
SON
VOLTAGE
(Vc)
counter.
It
should be l.5 volts
+0.3
volts.
7.
Disconnect the battery
from
the Type 7A
13.
2-4
Example 3
Single-Ended
Operation-waveform
measurement.
Measure the oscilloscope Calibrator +
Volts
waveform and
DC
level.
Type
7A13
control settings are same as for examples 1 and
2 except for the +
INPUT
connector, which should be con-
nected to the Calibrator +
VOLTS
BNC
connector
on
the
oscilloscope via coaxial cable.
1.
Set the oscilloscope Calibrator control to
40
mV.
2.
Set
VOLTS/DIV
Switch to
10
mV
position.
3.
Set +
INPUT
Mode switch to
AC
and set the time base
plug-in controls for a stable display.
4.
Compute
AC
waveform amplitude using the
POSITION
control to establish a reference line.
(4
divisions deflection
times
10
mV
/DIV yields
40
mV.)
5.
Determine the
DC
level
on
which the waveform
is
rid-
ing, using the method shown
in
either example 1 or
2.
Note
the
DC
level of the top or bottom of the waveform as de-
sired.
NOTE
For signals with a high
AC
to
DC
ratio,
the
AC
and
DC
signal
components
can
be
evaluated
simultan-
eously
by
using
DC
coupling
and
single-ended
operation
methods.
6.
Disconnect the calibrator signal.
Example 4
Differential Amplifier Operation. Measure the difference
between the oscilloscope Calibrator +
Volts
output and a
modified Calibrator + Volts output.
Type
7A13
control settings are the same as example 3
above with the following exception:
The
-
INPUT
Mode
switch should be set to
DC.
1.
Set the
VOLTS/DIV
Switch to
20
mV
position.
2.
Connect a
BNC
T
adapter
to the Calibrator +
Volts
output of the oscilloscope.
3.
Build
and connect a voltage divider to the
BNC
T as
illustrated
in
Fig.
2-2.
4.
Connect the junction of the two resistors to the -
IN-
PUT
connector via coaxial cable.
5.
Set the oscilloscope Calibrator +
Volts
output to 4
V.
6.
Simultaneously place the +
INPUT
and -
INPUT
Mode
switches to GND. Using the POSITION control, set the trace
to
DC
zero reference two divisions below the graticule hori-
zontal centerline.
7.
Simultaneously set both the + and -
INPUT
Mode
switches to
DC.
Adjust the time-base plug-in triggering con-
trols
if
necessary to obtain a stable waveform.
8.
Measure the waveform amplitude.
It
will
be equal to
the difference between the two input waveforms. (Approxi-
mately
40
mV;
this
is
the voltage drop across the 1 kn
resis-
tor.)
®