Operating Instructions --7A29
DETAILED OPERATING
INFORMATION
PROBES
Probes recommended for use with the 7A29 are the
P6056 (10X) or P6057 (100X) Both probes are
compatible with 50 ohm systems, and will allow good
frequency response These probes also contain trace
IDENTIFY and readout encoding functions
If the P6201 Probe is used, aone megohm input
impedance may be achieved (with some loss in
bandwidth). Consult your local Tektronix Field Engineer
or Representative for further information
VERTICAL GAIN CHECK AND ADJUSTMENT
To check the gain of the 7A29. set the VOLTS/DIV
switch to 01Vand connect a04V. 1kHz signal from
the oscilloscope mainframe calibrator to the 7A29 INPUT
connector
NOTE
Oscilloscope calibrators typically have an
output impedance of either 50 Oor 450 Cl.
The 50- ohm input impedance of the 7A29
forms avoltage divider with the calibrator
output impedance, reducing the voltage to the
7A29 by aprecise factor of 2(for 50- ohm
calibrators) or 10 (for 450-ohm Calibrators).
Thus the 4 V calibrator output of the 7104
and 7904 must be selected for an output
voltage of 04volts.
The vertical deflection should be exactly four divisions. If
not. adjust the front-panel GAIN adjustment for exactly
four vertical divisions.
INPUT COUPLING
The AC-OFF-DC switch allows achoice of input coupling
methods The type of display desired and the applied
signal will determine the coupling to use.
The DC coupling position can be used for most
applications. For signals with frequencies below about 1
kHz the DC coupling position is necessary to avoid signal
attenuation
In the AC coupling position the dc component of the
Signal is blocked by a2microfarad capacitor in the input
circuit. The AC coupling position provides the best display
of signals with adc component much larger than the ac
component.
In the OFF position of the input coupling switch, the
signal path of the amplifier is interrupted so that there is
no displayed signal However, asignal connected to the
INPUT connector will be loaded by approximately 390
kilohms dc, or by 195 kilohms ac. At extremely high
frequencies a small portion of the input signal will still be
displayed. For example, the signal attenuation in the OFF
position is about 32 dB at 1gigahertz.
The OVERLOAD protection feature of the 7A29 activates
when asignal capable of causing damage (to 7A29
internal circuitry) is applied to the INPUT connector To
reset the 7A29. turn the AC-OFF-DC switch to the OFF
position and remove the cause of overloading The 7A29
is then ready for normal operation
The 7A29 is capable of withstanding short term
overloads in excess of its continuous rating of 10 Vrms.
Thus, it is possible to apply 50 Vpeak (the same as 50 V
rms for asingle pulse) for 20 ms, or a20 Vrms signal for
125 ms. The 7A29 integrates the rms input power and
disconnects the signal before damage can occur.
Generally speaking, the 7A29 will protect itself from
excessive (damaging) signals; however, there are
exceptions. Continuous wave or pulsed radio frequency
signals above 100 MHz are not sensed by the input
protection circuitry and may cause damage to the input
attenuator if in excess of the 10 Vrms. 50 Vpeak, rating.
The relay contacts which disconnect the 7A29 from
excessive voltages may themselves be damaged by
arcing. If the contacts open with asignal greater than 50
Vapplied to the INPUT, the contacts may be permanently
damaged.
The ac coupling capacitor used in the AC position of the
AC-OFF-DC switch is rated at 100 Vdc Since the input
protection circuitry is located after the ac coupling
capacitor, application of up to 100 Vdc to the INPUT is
possible in the AC coupled mode. However, in order to
avoid afast-rise pulse being sent to the 7A29 from a
signal in excess of 50 Vdc. turn the AC-OFF DC switch
to OFF before connecting such asignal to the 7A29
INPUT In the OFF position, the ac coupling capacitor is
safely precharged to the average dc voltage
Figure 1-2 in section 1shows the maximum rms
amplitude which may safely be applied to the 7A29
INPUT Figure 1-2 also corresponds approximately to the
rms voltage the 7A29 will accept without automatically
disconnecting itself Acomplex pulse, whose rms value is
equal to the amount shown in Figure 1-2. is safe if it is
also less than 50 Vpeak.
For example, a30 Vrms sinusoid may be applied for up
to 55 ms. but a36 Vrms sinusoid may not be applied
because the peak voltage exceeds 50 VThe 7A29 INPUT
may be connected to a50 Vpower supply for up to 20 ms
because the rms and peak values of the resulting pulse
are both 50 V
2-4